510(k) Data Aggregation

    K Number
    K200648

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-17

    (158 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EKOS PE Endovascular Device with CU 4.0 is indicated for the:

    • Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
    • Infusion of solutions into the pulmonary arteries.
    • Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
    Device Description

    The EKOS PE Endovascular System consists of an EKOS PE Endovascular Device and CU4.0 (Control Unit 4.0 and Connector Interface Cables). The EKOS PE Endovascular Device consists of a single-use, disposable infusion catheter with removable ultrasound core. The infusion catheter contains multiple side holes distributed over the length of the treatment zone. The ultrasound core contains up to 20 ultrasound elements, evenly spaced over the treatment zone. Thermal sensors in the treatment zone monitor catheter temperature. The Control System generates and controls the delivery of radiofrequency energy to the ultrasound core while monitoring and controlling the temperature of the treatment zone.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for a medical device called the EKOS PE Endovascular Device with Control Unit 4.0. This document describes the device, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

    However, the information provided does not contain acceptance criteria or a study proving the device meets specific performance metrics in the way your request describes (e.g., using a test set with ground truth established by experts, MRMC studies, or standalone algorithm performance).

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (EkoSonic Endovascular Device with CU4.0) through a comparison of technological characteristics and performance testing against internal specifications and external standards (like ISO and IEC).

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and the reported device performance (in the context of clinical efficacy or diagnostic accuracy, as implied by "device performance" in your prompt). The table provided in the document (Table 1) lists "Product Specification" and "Purpose" with "Pass" results, which are more akin to design verification tests rather than clinical performance metrics.
    • Sample size used for the test set and the data provenance
    • Number of experts used to establish the ground truth
    • Adjudication method
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    • The type of ground truth used
    • The sample size for the training set
    • How the ground truth for the training set was established

    The device described is an "Embolectomy Catheter" used for ultrasound-facilitated infusion of fluids (including thrombolytics) for the treatment of pulmonary embolism and in peripheral vasculature. It is a physical medical device, not a diagnostic algorithm or AI system that would typically require the detailed clinical validation studies you're asking about.

    What the document does provide regarding performance:

    The document includes a "Performance Data" section (Section VII) which states: "Testing has confirmed that the EKOS PE Endovascular Device functions as intended and is substantially equivalent to the predicate device."

    It then presents Table 1: Testing Overview Supporting the EKOS PE Endovascular System. This table lists various engineering and safety-related tests, their purpose, and the outcome ("Pass" for both initial and accelerated aged conditions). These are verification and validation tests to ensure the device meets its design specifications and complies with relevant standards.

    Table 1: Testing Overview Supporting the EKOS PE Endovascular System (from the document)

    Product SpecificationPurposeEKOS PE Endovascular Device (Subject Device) T = 0Accelerated Aged (T=3 Years)
    Physical DimensionsVerified dimensional compatibility and stability with the MSD and DDC.PassPass
    Tip ShapeVerified tip conformity to ISO 10555-1 requirements.PassPass
    Kink ResistanceVerified device can be used within the target vasculature.PassPass
    Tensile StrengthVerified device conformity to ISO 10555-1 requirements.PassPass
    Electrical SafetyVerified device conformity IEC 60601-1 and IEC 60601-1-2 requirements.PassPass
    Acoustic CharacteristicsVerified device met specified acoustic design requirements.PassPass
    RadiopacityVerified device can be viewed under standard imaging techniques.PassPass
    Pressure ResistanceVerified device lumens conformity to ISO 10555-1.PassPass
    Temperature SensingVerified the device temperature sensors met specified design and functional requirements.PassPass
    Functional LifeVerified device met specified functional life requirements.PassPass
    BiocompatibilityVerified the device met biocompatibility requirements per ISO 10993-1.PassPass
    SterilizationVerified the device met sterilization requirements per ISO 11135.PassPass
    Shelf LifeVerified the device met functional requirements after a 3-year shelf life.N/A.Pass
    System Integration with CU4.0Verified the device integrates with the control unit, specifically, its acoustic protocol and enabled temperature safety features.PassN/A.

    The "acceptance criteria" here are generally that the device "Passed" these various engineering and safety tests, demonstrating it functions as intended and is safe for its intended use, aligning with the standards and the predicate device's characteristics. There's no further detail on the methodology of these "Pass" assessments (e.g., specific quantitative thresholds beyond meeting standard requirements).

    Ask a Question

    Ask a specific question about this device

    K Number
    K111705

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-03

    (47 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EkoSonic Endovascular System is intended for the infusion of solutions into the pulmonary arteries.

    Device Description

    The EkoSonic Endovascular System is an infusion Device designed to deliver fluids via a multi sidehole catheter. The fluid is dispersed via multiple ultrasound transducers distributed linearly along the length of an MicroSonic Device which is placed into the center lumen of the catheter. This device is intended to deliver physician-specified agents or fluids into the peripheral vasculature.

    AI/ML Overview

    The provided documents are FDA 510(k) letters related to the EKOS EkoSonic Endovascular System. These documents confirm the regulatory classification and substantial equivalence of the device but do not contain information about specific acceptance criteria or the study data proving the device meets those criteria, especially not in the context of an AI/ML-driven device or study.

    The description you've requested (acceptance criteria, test set details, expert qualifications, adjudication, MRMC study, standalone performance, ground truth, training set details) is typically found in detailed clinical study reports or 510(k) summaries for AI/ML devices, which this document is not. This document is a regulatory approval letter for a medical device (a catheter system), not an AI algorithm.

    Therefore, I cannot fulfill your request using only the provided text. The information is not present in these FDA 510(k) general correspondence letters.