510(k) Data Aggregation

    K Number
    K253186

    Validate with FDA (Live)

    Date Cleared
    2026-02-24

    (151 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoSphere Nano™ Monitor when used with a compatible non-invasive finger cuff is indicated for adult patients (≥ 18 years of age) in whom cardiac function parameters need to be evaluated as part of a patient's assessment. The HemoSphere Nano™ Monitor and compatible finger cuffs non-invasively measure blood pressure and associated hemodynamic parameters.

    Device Description

    The HemoSphere Nano Monitor is a hand-held monitoring device that measures the arterial pressure waveform collected from a connected non-invasive finger cuff and displays blood pressure and derived hemodynamic parameters. These parameters are continuously displayed for up to a period of 10 minutes, serving as a series of spot-check or point-in-time measurements. As such, the device does not feature any physiological alarms. It is compatible for use with the single use Acumen IQ Plus finger cuff (AIQCA2; cleared via K243781).

    The HemoSphere Nano Monitor utilizes the same principle of operation, algorithms, and mechanism for non-invasive monitoring hemodynamic parameters as the primary predicate, HemoSphere Advanced Monitoring Platform (K243781). When compared to the primary predicate, the subject device introduces a new hardware configuration with a smaller form factor. This new hardware configuration is powered by a rechargeable battery and features a smaller display to enable the product to be hand-held. Contrary to the predicate which incorporates a modular approach, the subject integrates all existing non-invasive technology (i.e., the ClearSight Module and Pressure Controller) into the HemoSphere Nano Monitor body itself to provide a unified and compact solution for non-invasive hemodynamic monitoring.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242909

    Validate with FDA (Live)

    Date Cleared
    2025-03-21

    (178 days)

    Product Code
    Regulation Number
    870.2850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloTrac sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. They are intended to be used in adult patients.

    The FloTrac Jr sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. The FloTrac Jr sensor is indicated for use in pediatric patients ≥ 12 years of age.

    The Acumen IQ sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output. They are intended to be used in adult patients.

    The VolumeView sensor is indicated for use in intravascular pressure monitoring. It is also indicated for use with the Edwards arterial pressure based cardiac output monitoring devices or hardware to measure cardiac output.

    Device Description

    The FloTrac, Acumen IQ, and VolumeView sensors are constructed from two disposable pressure transducers that convert a physiological signal (or mechanical pressure) to an electrical signal that is transmitted through the cable to the patient monitor. The sensors have a straight, flow-through design in which fluid is passed directly across the pressure sensor. The sensors are comprised of a pressure sensitive silicon chip with two electrodes for excitation voltage and two electrodes for signal output. A polycarbonate housing with an integral stopcock at one end, and an integral flush device at the other end, encloses the sensors.

    AI/ML Overview

    The provided FDA 510(k) summary for the FloTrac, FloTrac Jr, Acumen IQ, and VolumeView sensors does not contain detailed information about specific acceptance criteria and the study that proves the device meets those criteria in the typical format requested for an AI/ML-based medical device.

    This document describes a substantial equivalence determination for extravascular blood pressure transducers (sensors) manufactured by Edwards Lifesciences, LLC. The core of the submission is that the subject devices are identical to the predicate devices in terms of intended use, indications for use, and technological characteristics, EXCEPT for changed pressure tubing and IV set component materials.

    Therefore, the "study" described here is primarily focused on demonstrating that these material changes do not introduce new safety or effectiveness concerns, rather than validating an AI/ML algorithm's diagnostic performance against established ground truth.

    Here's a breakdown based on the provided text, addressing your questions where possible, and noting where the information is not applicable or not present:


    Overview of the Device and Study's Focus:

    The devices in question are FloTrac, FloTrac Jr, Acumen IQ, and VolumeView sensors, which are intravascular pressure monitoring devices that also work with Edwards' arterial pressure-based cardiac output monitoring hardware. The 510(k) submission (K242909) is for modifications to these existing devices, specifically changes to the pressure tubing and IV set component materials. The premise of the submission is that these material changes do not alter the fundamental performance or safety in a way that would require new clinical performance studies typical for an AI/ML device.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document states:

    • "All testing met the existing predetermined acceptance criteria."
    • "Based on the performance testing and the technological characteristics, the FloTrac sensors, Acumen IQ sensors, and VolumeView sensors meet the established performance criteria and are substantially equivalent to the predicate."

    However, the specific quantitative acceptance criteria (e.g., accuracy +/- X mmHg, drift < Y%, etc.) and the reported numerical performance data from these tests are not provided in this summary document. The summary indicates that "Performance verification was performed in accordance with Edwards' current design requirements," and "shelf-life and sterilization validations have been performed to existing specifications." These are standard engineering and manufacturing performance metrics, not AI/ML performance metrics.

    Acceptance Criteria (Quantitative examples are not provided in this document)Reported Device Performance (General Statement)
    Implicit Criteria: Performance equivalent to predicate devices despite material changes; Biocompatibility, Pressure Accuracy, Shelf-life, Sterilization."All testing met the existing predetermined acceptance criteria."
    "The differences do not raise any new concerns of safety and effectiveness.""The provided risk analysis and performance verification testing demonstrates the subject devices are substantially equivalent to the predicates."

    2. Sample Size Used for the Test Set and Data Provenance

    The document refers to "Performance verification" and "Biocompatibility testing." These generally involve lab-based testing (e.g., mechanical stress tests, pressure accuracy checks on a test bench, chemical compatibility tests) rather than patient data test sets in the context of an AI/ML algorithm.

    • Sample Size for Test Set: Not specified. The nature of the "test set" here refers to physical product samples for engineering verification and manufacturing quality, not a dataset of patient measurements for AI validation.
    • Data Provenance: The studies mentioned (biocompatibility, performance verification, shelf-life, sterilization) are likely conducted in a controlled lab/manufacturing environment, not directly on patient data. Therefore, questions of "country of origin" or "retrospective/prospective" data collection are not applicable in the context of this type of submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Given this is not an AI/ML submission related to diagnostic imaging or interpretation, the concept of "experts establishing ground truth" in the clinical sense is not applicable. The ground truth for this device's performance would be established through:

    • Engineering specifications and testing standards for pressure transducers.
    • Biocompatibility standards (e.g., ISO 10993-1).
    • Sterilization validation protocols.
    • Comparison to the performance of the predicate device.

    4. Adjudication Method for the Test Set

    Not Applicable. Adjudication methods like "2+1" or "3+1" are used in clinical studies, particularly for diagnostic interpretations where multiple readers assess cases. This submission is about the physical device's material change and its impact on manufacturing and engineering performance, not clinical interpretation.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is designed to assess the impact of an AI algorithm on human reader performance for tasks like diagnosis or detection. Since this submission is for material changes to a physical sensor, and there is no AI component described, an MRMC study was not performed and is not relevant.


    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No. This question pertains to AI/ML algorithms. The device described (pressure sensor) does not have a standalone AI algorithm that would produce a diagnostic output. Its function is to convert a physiological signal to an electrical signal for a monitor.


    7. The Type of Ground Truth Used

    The "ground truth" for this submission is based on:

    • Engineering specifications and validated testing methods: For pressure accuracy, electrical signal output fidelity, and physical integrity.
    • International standards (e.g., ISO 10993-1): For biocompatibility.
    • Defined shelf-life and sterilization parameters: Within established industry and regulatory norms.
    • Comparison to the performance characteristics of the legally marketed predicate devices.

    This is not ground truth derived from expert consensus, pathology, or outcomes data in the clinical sense, as it's not a diagnostic AI device.


    8. The Sample Size for the Training Set

    Not Applicable. Training sets are relevant for machine learning algorithms. This submission does not describe an AI/ML component or the training of such a component.


    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As there's no mention of an AI/ML training set, the establishment of ground truth for it is irrelevant to this 510(k) summary.


    In summary: This FDA 510(k) summary provides a high-level overview of a submission for material changes to existing medical devices (intravascular pressure sensors). The document focuses on demonstrating substantial equivalence by confirming that these material changes do not negatively impact the device's established performance, biocompatibility, shelf-life, or sterilization. It does not contain the detailed information typically associated with the validation of an AI/ML diagnostic algorithm, such as specific performance metrics, test set sizes for clinical data, expert panel ground truth adjudication, or MRMC study results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240078

    Validate with FDA (Live)

    Date Cleared
    2024-08-28

    (230 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fogarty thru-lumen embolectomy catheter is indicated for adults requiring the removal of fresh, soft emboli and thrombi from vessels in the arteries of the peripheral vasculature. Additionally, it may also be used for temporary occlusion of blood vessels, infusion of fluids, and blood sampling in general surgical procedures.

    Device Description

    The Fogarty Thru-Lumen Embolectomy Catheter is a double lumen catheter with a latex balloon at the distal end. One lumen is used for balloon inflation and is accessed through the gate valve at the proximal end. The thru-lumen is used for infusion of fluids, sampling of blood or guidewire passage through the distal tip. A removable stainless steel stylet(s) is provided.

    AI/ML Overview

    It appears there might be a misunderstanding of the provided text. The document is an FDA 510(k) clearance letter for a medical device: the Fogarty Thru-Lumen Embolectomy Catheter.

    This document is not about an AI/software-as-medical-device (SaMD) or a device that relies on complex algorithms requiring a study to prove its performance against acceptance criteria like sensitivity, specificity, or AUC based on a test set, ground truth, or expert readers.

    Instead, this device is a physical catheter used to remove emboli and thrombi. The "acceptance criteria" and "study" proving its performance in this context relate to engineering and biocompatibility testing, not to AI model performance or interpretation of medical images.

    Here's why the questions about AI/software-related study methodologies are not applicable to the provided document:

    • No mention of AI, algorithms, or software: The document describes a physical medical device. There is no indication that it incorporates AI, software, or image analysis to function.
    • "Study that proves the device meets the acceptance criteria" refers to bench testing: The document explicitly states under "Device Testing": "Bench testing was performed in accordance with Edwards' current design requirements. In addition, shelf-life, packaging, and sterilization validations have been performed to existing specifications. All testing met the existing predetermined acceptance criteria." This refers to mechanical, material, and sterilization performance, not diagnostic accuracy or AI performance.
    • Comparison to a predicate device: The basis for clearance is "substantial equivalence" to a predicate device (Fogarty Thru-Lumen Embolectomy Catheter cleared in K892410). The comparative analysis states, "The subject device is identical to the predicate device in terms of the intended use, indications for use, and technological characteristics (including design, material, chemical composition, principle of operation)." This reinforces that it's a physical device, not an AI system that would typically undergo MRMC studies or performance validation against a ground truth dataset of patient images.

    Therefore, I cannot provide the requested information (table of acceptance criteria, sample size, expert qualifications, adjudication, MRMC, standalone performance, ground truth types, training set details) because the provided FDA document pertains to a physical medical device and does not involve AI or software development with the type of validation methodology you are asking about.

    To summarize, the document is about the FDA clearance of a conventional medical device, not an AI-powered one. The "acceptance criteria" and "study" refer to engineering and biological safety tests, not the performance metrics of an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233984

    Validate with FDA (Live)

    Date Cleared
    2024-08-02

    (228 days)

    Product Code
    Regulation Number
    870.5600
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Edwards Lifesciences Acumen Assisted Fluid Management (AFM) Software Feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM Software Feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM Software Feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the Assisted Fluid Management suggestions.

    Device Description

    The Acumen AFM Software Feature (core AFM algorithm + AFM Graphical User Interface) was originally granted in De Novo, DEN190029, on November 13, 2020, to inform clinicians about a patient's fluid responsiveness. The performance of the AFM Software Feature in predicting a patient's fluid responsiveness is measured using response rate and is calculated by reporting the percentage of followed AFM recommendations ("Fluid Bolus Suggested" and "Test Bolus Suggested" prompts) that have the desired change in stroke volume (SV), divided by the total number of AFM recommendations.

    With this submission, Edwards is seeking clearance for the AFM Prompt Reclassifier algorithm (AFM PR algorithm) to the Acumen AFM Software Feature. The AFM Prompt Reclassifier algorithm is intended to be used in conjunction with the core AFM algorithm to re-assess the fluid bolus recommendations provided by the core alqorithm. It analyzes the patient's current hemodynamics for either confirming (corroborating) the original prompt or reclassifying the prompts (i.e., reclassify a "Test Bolus Suggested" prompt to a "Fluid Bolus Suggested" prompt or vice versa). In doing so, it acts as a secondary check for the fluid bolus prompts such that a greater number of the "Fluid Bolus Suggested" prompts lead to the desired change in stroke volume. Through refined prompt adjustments informed by real-time hemodynamic data, the AFM PR algorithm aims to improve patient responsiveness, thereby optimizing the impact of the AFM Software Feature on patient hemodynamics.

    AI/ML Overview

    The FDA 510(k) summary for the Acumen Assisted Fluid Management (AFM) Software Feature describes the acceptance criteria and the study conducted to demonstrate the device meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated as a numerical target, but the overall goal was to demonstrate an improvement in the response rate for "Fluid Bolus Suggested" prompts due to the addition of the AFM Prompt Reclassifier (AFM PR) algorithm. This improvement should confirm that a greater number of these prompts lead to desired changes in the patient's stroke volume.The study "demonstrated an improvement in response rate for 'Fluid Bolus Suggested' prompts, thus demonstrating that the AFM PR algorithm met the predefined acceptance criteria." The results showed that the differences in fluid bolus suggestions introduced by the AFM PR algorithm do not raise any safety and effectiveness concerns.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The algorithm performance was analyzed on an archived dataset consisting of 1229 data points from 307 patients.
    • Data Provenance: The data came from the US IDE Study, G170204. The study involved 9 independent U.S. sites. The data is retrospective as it was an "archived dataset."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not provided in the document. The study uses physiological response (change in stroke volume) as the outcome measure, implying a physiological ground truth rather than expert interpretation of images or other data.

    4. Adjudication Method for the Test Set

    • This information is not provided in the document. Given the nature of the ground truth (physiological response), a traditional adjudication method for subjective assessments might not be directly applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • A MRMC comparative effectiveness study was not explicitly mentioned or described.
    • The study focuses on the algorithm's performance in improving the response rate of its suggestions, rather than comparing human reader performance with and without AI assistance. The device offers "suggestions" and the "decision to administer a fluid bolus is made by the clinician."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone algorithm performance study was done. The document states, "Algorithm performance was analyzed on the archived dataset... The validation was to demonstrate the impact on the response rate of the AFM Software Feature's fluid bolus prompts due to the addition of AFM Prompt Reclassifier algorithm." This implies evaluating the algorithm's predictions against the measured physiological outcomes.

    7. The Type of Ground Truth Used

    • The ground truth used is physiological response/outcomes data. Specifically, the "desired change in stroke volume (SV)" following AFM recommendations was used to measure the "response rate."

    8. The Sample Size for the Training Set

    • The document does not explicitly state the sample size for the training set. It only mentions the "archived dataset from the US IDE Study, G170204" used for algorithm performance analysis/validation.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not explicitly describe how the ground truth for the training set was established. It focuses on the validation of the AFM Prompt Reclassifier algorithm using an existing dataset. Given that the core AFM algorithm was granted in De Novo DEN190029, the ground truth for its original training would likely have involved similar physiological outcome data from clinical studies where fluid administration decisions were made and subsequent stroke volume changes were observed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233819

    Validate with FDA (Live)

    Date Cleared
    2024-05-22

    (173 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fogarty Venous Thrombectomy Catheters are indicated for use in the peripheral venous vasculature of adult patients for the removal of emboli and thrombi, and may be used for temporary vessel occlusion.

    Device Description

    Fogarty venous thrombectomy catheters are specifically designed for the performance of venous thrombectomy in adult patients. Fogarty venous thrombectomy catheters have a long, soft tip which is intended to facilitate advancement past the venous valves while minimizing trauma. The flexibility required for the venous procedure is supplied by the spring-wound body. The spring is covered with a braiding for additional strength. The gate valve is designed for single-handed operation and to minimize the possibility of leakage. The gate valve has an arrow to indicate the "open" and "closed" positions. There are three available catheter sizes (6 French (2.0 mm), 8 French (2.7 mm), and 8/10 French (2.7/3.3 mm)). Each may be quickly identified by the color-coded body. The size and filling capacity are printed on each catheter.

    AI/ML Overview

    This document is a 510(k) summary for the Fogarty Venous Thrombectomy Catheter. It outlines the device description, indications for use, and a comparison to a predicate device. Importantly, it focuses on demonstrating substantial equivalence to a preamendment predicate device through bench testing, biocompatibility testing, shelf-life, packaging, and sterilization validations. It does not present a study proving the device meets acceptance criteria related to AI/software performance or human-in-the-loop improvements, nor does it discuss ground truth establishment, expert consensus, or MRMC studies.

    Therefore, I cannot extract the information required to answer your prompt because the provided text pertains to a traditional medical device (catheter) and does not contain information about AI, software, or the types of studies you are asking about (e.g., MRMC studies, standalone algorithm performance).

    The document explicitly states: "Bench testing inclusive of design verification, packaging, sterilization, and biocompatibility testing all demonstrate that the subject devices are substantially equivalent to the predicate devices." This highlights that the device's performance is demonstrated through physical and biological testing, not through AI/software evaluation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230612

    Validate with FDA (Live)

    Date Cleared
    2023-11-17

    (256 days)

    Product Code
    Regulation Number
    870.2700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Edwards Algorithm for Measurement of Blood Hemoglobin is indicated for continuously monitoring changes to hemoglobin concentration in the circulating blood of adults ≥ 40 kg receiving advanced hemodynamic monitoring using HemoSphere ForeSight Oximeter Cable and non-invasive ForeSight sensors (large) in cerebral locations.

    The Edwards Algorithm for Measurement of Blood Hemoglobin is intended for use as an adjunct monitor of relative and total hemoglobin concentration of blood in individuals at risk for reduced-flow or no-flow ischemic states in surgical and ICU settings.

    Device Description

    The Edwards Algorithm for Measurement of Blood Hemoglobin is intended for continuously and non-invasively monitoring the relative and total hemoglobin values in the blood of patients requiring advanced hemodynamic monitoring in a critical care environment. The outputs of the algorithm include the relative changes in total hemoglobin in blood ( $\Delta$ tHb) and total hemoglobin in blood (tHb) parameters and are derived from the relative change in concentration of total tissue hemoglobin ( $\Delta$ ctHb parameter) measured by the ForeSight Oximeter Cable on the HemoSphere Advanced Monitoring Platform (K213682, cleared June 22, 2022).

    The subject algorithm provides relative blood hemoglobin ( $\Delta$ tHb; measured in g/dL of blood) values continuously as a change over time from 0 g/dL. It can also be calibrated using an optional input of reference blood hemoglobin measurements such as ones obtained in vitro from a blood gas analyzer. When calibrated, it provides the value of total blood hemoglobin (tHb).

    Additionally, the algorithm also provides three secondary output flags:
    o DoNotCalibrate Flag: This flag is intended to indicate when a calibration should not be performed.
    o Recalibrate Flag: This flag is intended to indicate when a new calibration is recommended.
    o Unstable Flag: This flag is intended to indicate when the input signal ( $\Delta$ ctHb) is unstable.

    AI/ML Overview

    The provided text describes the Edwards Algorithm for Measurement of Blood Hemoglobin, an algorithm intended for continuously monitoring changes to hemoglobin concentration. Here's a breakdown of the acceptance criteria and the study proving its performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Target)Reported Device Performance
    Accuracy (Bias)Bias close to 0 g/dLBias close to 0 g/dL
    Precision (Standard Deviation)Precision less than 1 g/dLPrecision less than 1 g/dL
    Overall Error (RMSE/ARMS)1 g/dLMet 1 g/dL

    Note: The document states "The results demonstrated that the subject device subject device met the acceptance criteria of 1g/dL with a bias close to 0 and precision less than 1g/dL". While it mentions meeting the criteria, the exact numerical values for bias and precision are not explicitly provided in the text beyond "close to 0" and "less than 1g/dL". The 1g/dL criteria appears to be for RMSE/ARMS.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 83 patients
    • Data Provenance:
      • Country of Origin: Data was collected from 5 sites across the US and EU.
        • Amsterdam, The Netherlands (European Union) - 27 patients (32.53%)
        • Santander, Spain (European Union) - 8 patients (9.64%)
        • Greenville, North Carolina, USA - 18 patients (21.69%)
        • Sacramento, California, USA - 11 patients (13.25%)
        • Chicago, Illinois, USA - 19 patients (22.89%)
      • Retrospective or Prospective: Retrospective analyses were performed on data already collected, independent of the device development.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts used or their specific qualifications for establishing the ground truth. However, it indicates that the device's performance was compared against a "laboratory co-oximeter," which implies a gold standard measurement method rather than expert consensus on images.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by laboratory co-oximeter measurements, not through expert adjudication of human interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study focused on the algorithm's direct performance against a laboratory co-oximeter. The text states, "No clinical trial was performed in support of the subject 510(k) submission."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The performance data was assessed by comparing the device's output to laboratory co-oximeter measurements, without human interaction with the algorithm's output during the measurement process. The algorithm's outputs are numerical values for hemoglobin concentration.

    7. The Type of Ground Truth Used

    The ground truth used was laboratory co-oximeter measurements of total hemoglobin values in blood. This is considered a highly accurate and objective reference method for hemoglobin quantification.

    8. The Sample Size for the Training Set

    The document does not provide information regarding the sample size used for the training set. The descriptions provided are solely for the retrospective analysis performed for validation (the test set).

    9. How the Ground Truth for the Training Set was Established

    The document does not provide information regarding how the ground truth for the training set was established, as details about the training phase are not included in the provided text.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoSphere Alta monitor when used with the HemoSphere Alta Swan-Ganz patient cable and Edwards Swan-Ganz catheters is indicated for use in adult and pediatic critical care patients requiring of cardiac output (continuous [CO] and intermittent [CO]) and derived hemodynamic parameters in a hospital environment. Pulmonary artery blood temperature monitoring is used to compute continuous and intermittent CO with thermodilution technologies. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.

    The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. The GHI algorithm provides the risk of a global hypoperfusion event (defined as SvO2 ≤ 60% for at least 1 minute) occurring in the next 10-15 minutes. The GHI algorithm is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring with the Swan-Ganz catheter. The GHI algorithm is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions.

    HemoSphere Alta monitor with HemoSphere Oximetry Cable

    The HemoSphere Alta monitor when used with the HemoSphere oximetry cable and Edwards oximetry catheters is indicated for use in adult and pediatric crtical care patients requring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Alta Monitor with HemoSphere Pressure Cable

    The HemoSphere Alta monitor when used with the HemoSphere pressure cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring of hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac sensor, Acumen IQ sensor, and TruWave DPT indications for use statements for information on target patient populations specific to the sensor/transducer being used.

    The Edwards Acumen Hypotension Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or non-surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.

    When used in combination with the HemoSphere Pressure Cable connected to a compatible Swan-Ganz catheter, the Edward Lifesciences Right Ventricular Pressure (RVP) algorithm provides the clinician with physiological insight into the hemodynamic status of the right ventricle of the heart. The RVP algorithm is indicated for critically ill patients over 18 years of age receiving advanced hemodynamic monitoring in the operating room (OR) and intensive care unit (ICU). The RVP algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Right Ventricular Pressure (RVP) parameters.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Alta Monitor with ForeSight Oximeter Cable

    The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states. The ForeSight oximeter cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight oximeter cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere Alta monitor.

    · When used with large sensors, the ForeSight oximeter cable is indicated for use on adults and transitional adolescents ≥40 kg.

    · When used with medium sensors, the ForeSight oximeter cable is indicated for use on pediatric subjects ≥3 kg.

    · When used with small sensors, the ForeSight oximeter cable is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Alta Monitor with ClearSight Technology

    The HemoSphere Alta when used with the pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with comorbidities for which hemodynamic optimization is desired and invasive measurements are difficult.

    The HemoSphere Alta monitor and compatible Edwards finger cuffs non-invasively measures blood pressure and associated hemodynamic parameters. Refer to the ClearSight finger cuff and Acumen IQ finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.

    The Edwards Acumen Hypotension Index (HP)) feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Acumen Hypotension Prediction Index (HPI) parameter.

    Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.

    HemoSphere Alta Monitor with Acumen Assisted Fluid Management Feature and Acumen IQ Sensor

    The Acumen assisted fluid management (AFM) software feature provides the clinician with physiological insight into a patient's estimated response to fluid therapy and the associated hemodynamics. The Acumen AFM software feature is intended for use in surgical patients ≥18 years of age, that require advanced hemodynamic monitoring. The Acumen AFM software feature offers suggestions regarding the patient's physiological condition and estimated response to fluid therapy. Acumen AFM fluid administration suggestions are offered to the clinician; the decision to administer a fluid bolus is made by the clinician, based upon review of the patient's hemodynamics. No therapeutic decisions should be made based solely on the assisted fluid management suggestions.

    The HemoSphere Alta Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting.

    The HemoSphere Alta Advanced Monitoring Platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, FloTrac sensors, Acumen IQ sensors, TruWave DPTs, ForeSight /ForeSight Jr sensors, and ClearSight/Acumen IQ finger cuffs.

    Device Description

    The HemoSphere Alta™ Advanced Monitoring Platform is Edwards' next-Device generation platform that provides a means to interact with and visualize Description: hemodynamic and volumetric data on a screen. The HemoSphere Alta™ Monitoring Platform provides an improved user interface utilizing the existing Edwards technologies and algorithms commercially available in the HemoSphere Advanced Monitoring Platform.

    AI/ML Overview

    This FDA 510(k) summary for the Edwards Lifesciences HemoSphere Alta Advanced Monitoring Platform (K232294) primarily focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and non-clinical performance validation. It explicitly states that "No new clinical testing was performed in support of the subject 510(k)." As such, the document does not provide specific acceptance criteria for AI/algorithm performance or details of a study proving the device meets such criteria through clinical data.

    Instead, the submission emphasizes the device's functional and safety aspects, along with the integration of existing, previously cleared technologies and algorithms into a new hardware and software platform with an improved user interface.

    Therefore, many of the requested sections below cannot be fully answered based on the provided text, as the focus was on non-clinical verification and substantial equivalence rather than new clinical performance studies for AI/algorithm features.


    1. Table of Acceptance Criteria and Reported Device Performance

    As per the provided document, specific acceptance criteria and detailed device performance metrics for individual AI/algorithm features (like HPI, GHI, AFM, RVP) are not detailed as part of a new clinical study for this 510(k) submission. The submission states, "No new clinical testing was performed in support of the subject 510(k)." The "Performance Data" section primarily discusses non-clinical verification.

    The document states:

    • "Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications."
    • "Measured and derived parameters were tested using a bench simulation."
    • "System integration and mechanical testing was successfully conducted to verify the safety and effectiveness of the device. All tests passed."
    • "Software verification testing were conducted... All tests passed."

    This indicates that internal performance specifications were met, but these specifications themselves are not provided, nor is the performance against them quantified in this public summary.

    2. Sample Size Used for the Test Set and Data Provenance

    No test set for clinical performance of AI/algorithm features is described, as "No new clinical testing was performed." The device leverages existing, previously cleared algorithms.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable, as no new clinical test set for AI/algorithm performance is described. The AI/algorithm features leverage ground truth established in prior clearances for the predicate devices.

    4. Adjudication Method for the Test Set

    Not applicable, as no new clinical test set for AI/algorithm performance is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study is mentioned, as "No new clinical testing was performed." The submission focuses on the HemoSphere Alta platform being a new generation integrating existing Edwards technologies and algorithms with an improved user interface and hardware.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The document does not describe new standalone performance studies for the AI/algorithm features. The AI/algorithm features (HPI, GHI, AFM, RVP) themselves were likely evaluated in standalone fashion during their original predicate clearances (e.g., K231038 for GHI). This 510(k) integrates these existing algorithms into a new platform.

    7. Type of Ground Truth Used

    The type of ground truth for the AI/algorithm features (HPI, GHI, AFM, RVP) would have been established during their original clearances. For this 510(k) submission, this information is not provided. Typically, hemodynamic algorithms like HPI or GHI rely on physiological measurements (e.g., direct arterial pressure, SvO2 from Swan-Ganz catheter, outcomes data related to hypotension or hypoperfusion events) as ground truth.

    8. Sample Size for the Training Set

    No details regarding training set sample sizes for the AI/algorithm features are provided in this 510(k) summary, as it covers the integration of existing algorithms. The training data would have been described in the original 510(k) submissions for those predicate algorithms (e.g., for Acumen HPI feature, Global Hypoperfusion Index, Right Ventricular Pressure algorithm, Acumen Assisted Fluid Management).

    9. How the Ground Truth for the Training Set Was Established

    As with the training set size, the method for establishing ground truth for the training set of the AI/algorithm features is not detailed in this 510(k) summary because it pertains to existing algorithms. This would have been covered in their individual predicate 510(k) submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230919

    Validate with FDA (Live)

    Date Cleared
    2023-10-24

    (204 days)

    Product Code
    Regulation Number
    870.1130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acumen™ IQ finger cuff is indicated for patients over 18 years of age to noninvasively measure blood pressure and associated hemodynamic parameters when used with the HemoSphere advanced monitoring platform.

    Device Description

    The Acumen IQ finger cuff is intended to noninvasively measure blood pressure and use the information to derive hemodynamic parameters when connected to an Edwards' HemoSphere Advanced Monitoring platform.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Acumen IQ finger cuff, which is a modified version of a previously cleared device (K190130). The modification involves expanding the finger circumference size range and removing a finger sizing aid. The submission states that the device is identical to the predicate device in terms of design, intended use, and technology.

    The document mentions that usability testing was conducted in accordance with ANSI/AAMI HE75 and that clinical performance testing was completed to ensure the finger cuffs perform as intended, and that the Acumen IQ finger cuff has successfully passed all testing. However, the actual acceptance criteria and device performance results from these studies are not explicitly provided in the given text. The text only broadly states that the device "successfully passed all testing."

    Without the specific details of the clinical performance testing, a comprehensive table of acceptance criteria and reported device performance cannot be generated. Similarly, concrete information regarding sample sizes, data provenance, ground truth establishment, MRMC studies, or standalone performance is not available in the provided document.

    Therefore,Based on the provided text, a complete answer to your request is not possible. The document states that clinical performance testing was completed and successfully passed, but it does not provide the specific acceptance criteria or the numerical results of this performance testing.

    Here's a breakdown of what can be inferred and what is missing:

    1. Table of Acceptance Criteria and the Reported Device Performance: This information is not provided in the document. The text only states that "clinical performance testing was completed to ensure the finger cuffs perform as intended" and that the device "has successfully passed all testing." No specific criteria (e.g., accuracy, precision) or numerical performance values are given.

    2. Sample size used for the test set and the data provenance: This information is not provided in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided in the document.

    4. Adjudication method for the test set: This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This information is not provided in the document. The device is a "noninvasive blood pressure measurement system," not an AI diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The device itself is a measurement device. Its "performance" would inherently be standalone in its measurement capability once the cuff is applied. However, specific methodology (e.g., comparison to invasive measurements) and results are not provided.

    7. The type of ground truth used: Given that the device non-invasively measures blood pressure, the ground truth for clinical performance testing would typically be established using a reference method for blood pressure measurement, such as invasive arterial line measurements or a highly accurate non-invasive oscillometric device that meets specific standards (e.g., ISO or BHS protocols). However, this is not explicitly stated in the document.

    8. The sample size for the training set: This information is not provided in the document. (Note: For a device like a blood pressure monitor, there might not be a "training set" in the same sense as an AI algorithm that learns from data. It's more about calibration and validation against ground truth.)

    9. How the ground truth for the training set was established: This information is not provided in the document.

    In summary, the provided FDA clearance letter and 510(k) summary confirm that usability and clinical performance testing were conducted and passed, but they do not detail the specific acceptance criteria, performance results, or methodologies of those studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230579

    Validate with FDA (Live)

    Date Cleared
    2023-08-18

    (169 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used in combination with a Swan-Ganz catheter connected to a pressure transducer, the Edwards Lifesciences Smart Wedge algorithm measures and provides pulmonary artery occlusion pressure and assesses the quality of the pulmonary artery occlusion pressurement. The Smart Wedge algorithm is indicated for use in critical care patients over 18 years of age receiving advanced hemodynamic monitoring. The Smart Wedge algorithm is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Smart Wedge algorithm parameters.

    Device Description

    The Smart Wedge algorithm is designed to provide the value at end-expiration of the pulmonary artery occlusion pressure (PAOP) signal, also called pulmonary wedge pressure, pulmonary capillary wedge pressure (PCWP), or pulmonary artery wedge pressure (PAWP), and to assess the quality of the pulmonary artery occlusion pressure measurement.

    The Smart Wedge algorithm is intended to be used with a Swan-Ganz pulmonary artery catheter connected to a pressure cable and pressure transducer.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance

    The device performance is reported for two main aspects: PAOP (Pulmonary Artery Occlusion Pressure) Identification and PAOP Measurement. The acceptance criteria can be inferred from the reported performance results and the comparison to the predicate device, especially the statement: "Results for the Smart Wedge algorithm met or exceeded predicate device performance." While explicit numerical acceptance criteria aren't listed as "targets," the provided performance values serve as the acceptable outcomes.

    Smart Wedge Algorithm ParameterImplicit Acceptance Criteria (Target/Goal)Reported Device Performance (Mean with 95% CI)
    PAOP IdentificationHigh Sensitivity (close to 100%)Sensitivity: 100% [100, 100]
    (based on 225 PAP waveforms)High Specificity (close to 100%)Specificity: 96% [92, 100]
    High Positive Predictive Value (PPV) (close to 100%)PPV: 95% [89, 99]
    High Negative Predictive Value (NPV) (close to 100%)NPV: 100% [100, 100]
    PAOP MeasurementLow Mean Absolute Error (MAE) (e.g., < 4 mmHg)MAE: 1.1 mmHg [0.8, 1.5]
    (based on 110 PAOP measurements)Low Bias (close to 0 mmHg)Bias: 0.4 mmHg [0.1, 0.7]
    Low Standard Deviation (Std)Std: 1.7 mmHg [1.4, 2.0]
    High Correlation (r) (close to 1.0)Correlation (r): 0.98

    Note: The document explicitly states for PAOP Measurement: "PAOP within mean absolute error < 4 mmHg accuracy." This serves as a clear numerical acceptance criterion for MAE.

    2. Sample Sizes and Data Provenance

    • Test Set (PAOP Identification): 225 PAP waveforms from 129 patients.
    • Test Set (PAOP Measurement): 110 PAOP measurements from 59 patients.
    • Data Provenance: Retrospectively collected from ICU and OR patients. The country of origin is not specified, but given the Edwards Lifesciences headquarters in Irvine, California, it's likely primarily US-based or multi-site.

    3. Number of Experts and Qualifications

    • Number of Experts: Three experienced healthcare providers (HCPs) were used to establish the ground truth.
    • Qualifications: Described as "experienced healthcare providers (HCPs)." Specific qualifications (e.g., "radiologist with 10 years of experience") are not detailed, but the term "experienced" suggests domain expertise relevant to pulmonary artery occlusion pressure waveforms.

    4. Adjudication Method for the Test Set

    • For PAOP Identification: "Mode of three HCP annotations." This means the most frequent annotation among the three experts was
      taken as the ground truth.
    • For PAOP Measurement: "Average PAOP measurement of three HCPs." This implies the numerical average of the three experts' measurements was used as the ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • The document explicitly states: "No clinical trial was performed in support of the subject 510(k)."
    • Therefore, an MRMC comparative effectiveness study, which typically involves human readers improving with AI vs. without AI assistance, was not conducted. The study focuses on the algorithm's performance against expert consensus.

    6. Standalone (Algorithm Only) Performance

    • Yes, the reported study describes the standalone performance of the Smart Wedge algorithm. The tables ("Performance Results of PAOP Identification" and "Performance Results of PAOP Measurements") present the algorithm's capabilities (Sensitivity, Specificity, MAE, Bias, Correlation) measured against the established expert consensus ground truth. There's no mention of a human-in-the-loop component in the reported performance metrics.

    7. Type of Ground Truth Used

    • The ground truth used was expert consensus.
      • For PAOP Identification: "Mode of three HCP annotations."
      • For PAOP Measurement: "Average PAOP measurement of three HCPs."

    8. Sample Size for the Training Set

    • The document does not explicitly state the sample size used for the training set. It mentions the verification was performed using "waveforms retrospectively collected from ICU and OR patients", but the specific number for training versus testing is not provided. The provided numbers (225 waveforms/129 patients for identification, 110 measurements/59 patients for measurement) are for the test set.

    9. How Ground Truth for Training Set Was Established

    • Similar to the training set sample size, the document does not explicitly detail how the ground truth for the training set was established. Typically, for machine learning models, the training data also requires labeled ground truth, often established similarly to the test set (e.g., expert annotation or other reliable sources). However, this specific 510(k) summary focuses on the verification and validation of the algorithm's performance and the ground truth establishment for the test data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231038

    Validate with FDA (Live)

    Date Cleared
    2023-07-26

    (105 days)

    Product Code
    Regulation Number
    870.2210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. The GHI algorithm provides the risk of a global hypoperfusion event (defined as SvO2 ≤ 60% for at least 1 minute) occurring in the next 10-15 minutes.

    The GHI algorithm is intended for use in surgical patients receiving advanced hemodynamic monitoring with the Swan-Ganz catheter.

    The GHI algorithm is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the GHI algorithm predictions.

    Device Description

    The Global Hypoperfusion Index (GHI) parameter provides the clinician with physiological insight into a patient's likelihood of a global hypoperfusion event on average 10-15 minutes before mixed venous oxygen saturation (SvO2) reaches 60%. The GHI feature is intended for use in surgical or nonsurgical patients. The product predictions are adjunctive for reference only and no therapeutic decisions should be made based solely on the GHI parameter.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a comprehensive study report with all the requested information. It primarily presents the FDA's 510(k) clearance letter and a summary of the device, its indications for use, and a comparison to predicate devices, stating that performance testing was executed and that no clinical trial was performed for the 510(k) submission.

    However, based on the available information, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions that the GHI algorithm provides the risk of a global hypoperfusion event (defined as SvO2 ≤ 60% for at least 1 minute) occurring in the next 10-15 minutes and alerts the clinician on average 10-15 minutes before SvO2 reaches 60%. It also states that the GHI algorithm provides an index from 0 to 100 where the higher the value, the increased likelihood that a global hypoperfusion event will occur.

    While specific numerical acceptance criteria (e.g., minimum sensitivity, specificity, or AUC) and their corresponding achieved performance values are not explicitly stated in the provided text, the overall conclusion is that the algorithm "has successfully passed functional and performance testing" and "meets the predetermined design and performance specifications." This implies that internal acceptance criteria were met, even if they are not detailed here.

    Example (Hypothetical, as not provided in text):

    MetricAcceptance Criteria (Hypothetical)Reported Device Performance (Implied as "met")
    Time to AlertAverage 10-15 minutes before eventAchieved average 10-15 minutes before event
    Ability to Identify RiskGHI 0-100, higher = increased riskGHI provides increased likelihood with higher values
    Overall PerformanceMeets predetermined specificationsMet predetermined specifications

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The text states, "Prospective analyses of retrospective clinical data from multiple independent datasets, comprised of data from a diverse set of patients over the age of 18 years undergoing surgical procedures with invasive monitoring, were analyzed to verify the safety and performance of the subject device." However, the exact sample size (number of patients or data points) for the test set is not specified.
    • Data Provenance:
      • Country of Origin: Not specified in the provided text.
      • Retrospective or Prospective: "Prospective analyses of retrospective clinical data" implies that existing (retrospective) data was collected and then analyzed in a forward-looking (prospective) manner for the purpose of the study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The text does not provide any information regarding the number of experts, their qualifications, or their involvement in establishing ground truth for the test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    The text does not provide any information regarding an adjudication method for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    The text explicitly states: "No clinical trial was performed in support of the subject 510(k)." This indicates that an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted for this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance evaluation was conducted. The text states:

    • "Algorithm performance was tested using clinical data."
    • "The algorithm was tested at the algorithm level to ensure the safety of the device. All tests passed."
    • "Prospective analyses of retrospective clinical data... were analyzed to verify the safety and performance of the subject device."

    This confirms that the algorithm's performance was assessed independently.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for a "global hypoperfusion event" is explicitly defined in the Indications for Use as: "SvO2 ≤ 60% for at least 1 minute." This is an objective physiological measurement (outcomes data) rather than expert consensus or pathology.

    8. The sample size for the training set:

    While the text mentions that "patient waveforms were collected in support of the development and validation of the GHI algorithm," the sample size for the training set is not specified.

    9. How the ground truth for the training set was established:

    Given that the ground truth for the device's output is based on SvO2 measurements, it is highly probable that the ground truth for the training set was established using the same objective physiological measurement: SvO2 ≤ 60% for at least 1 minute. The text implies that clinical data (patient waveforms) were used for both development and validation.