510(k) Data Aggregation

    K Number
    K232336

    Validate with FDA (Live)

    Device Name
    ES Series V2
    Manufacturer
    Date Cleared
    2023-09-26

    (53 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound tissue brightness parameters such as ultrasound beam attenuation and backscattering coefficient, and estimates of speed of sound, in internal structures of the body. The device is intended to be used by trained healthcare professionals, in a healthcare environment.

    The device is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

    In particular, the device is intended to provide:

    • Linear distance measurements of anatomical structures,

    • Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen.

    • Estimates of ultrasound tissue brightness parameters in the liver at selected ultrasound frequencies,

    • Measurement of brightness ratio between structures and in particular between the liver and the kidnev.

    • Estimates of speed of sound in the liver.

    The shear wave speed measurements, ultrasound tissue brightness parameters, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.

    Device Description

    E-Scopics' ES Series V2 is an ultraportable ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging and quantitative imaging studies. ES Series V2 consists of a Software App (running on a consumer off-the-shelf Selected Host) and an accessory curved array probe. The system produces images and quantifications, which are displayed on the monitor of the Selected Host. ES Series V2 is operated from the Selected Host multi-touch screen. ES Series V2 also allows the user to perform measurements, to capture images, and to generate printable reports.

    ES Series V2 is designed to perform non-invasive measurements of liver/spleen shear wave speed and to estimate tissue stiffness. The e.C5-1 probe is equipped with a mechanical vibrator to produce shear waves in tissue. When used on the spleen applications, shear waves travel through the skin and intercostal space into the abdominal organ. 2D ultrasound is used to track the shear wave and measure its speed. The system then provides an estimate of tissue stiffness. In addition, ES Series V2 is designed to measure several quantitative parameters from B Mode imaging: ultrasound tissue brightness parameters such as ultrasound attenuation and backscattering coefficient, speed of sound, and compute ultrasound brightness ratio between 2 regions in the image. The results of ES Series V2 quantitative imaging modalities are displayed on the selected host monitor.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the ES Series V2 device, based on the provided text:

    Acceptance Criteria and Device Performance

    The device's performance was evaluated against precision (within subject coefficient of variation, wCV) and bias (difference from ground truth) on calibrated phantoms. The stated acceptance criteria are reflected in the reported performance.

    ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (ES Series V2) - wCV (Precision)Reported Device Performance (ES Series V2) - Bias
    B-Mode ImagingAxial Resolution: < 1 mmLateral Resolution: < 2mm
    Contrast Resolution: distinguishes targets with +/- 3 dB contrast ratio
    Speed of Sound (SOS)Bias: (-0.7%) - (0.7%)wCV: [0.1%, 0.4%] (CI 95% [0.1%, 2.5%])Bias: [-3.2%, -1.2%] (CI 95% [-4.0%, -0.2%])
    Precision: (0.05%) - (0.3%)
    Shear Wave Speed (SWS) / Liver Stiffness (2D TE)Bias (kPa): (-12.5%) - (28.5%)wCV (SWS): [0.8%, 3.2%] (CI 95% [0.4%, 20%])Bias (SWS): [-15%, 2.2%] (CI 95% [-20%, 9%])
    Bias (m/s): (-6.4%) - (13.2%)wCV (Young's modulus/stiffness): [1.2%, 6.3%] (CI 95% [0.6%, 40%])Bias (Young's modulus/stiffness): [-27%, 5.6%] (CI 95% [-35%, 18%])
    Precision (kPa): (0.8%) - (2.7%)
    Precision (m/s): (0.4%) - (1.4%)
    Elasticity ImagingwCV (SWS): [0.1 %, 0.9 %] (CI 95% [0.1%, 5.3%])Bias (SWS): [3.8 %, 11.2 %] (CI 95% [3.0%, 13.6%])
    wCV (Young's moduli): [0.3 %, 1.7 %] (CI 95% [0.1%, 10.6%])Bias (Young's moduli): [7.5 %, 23.7 %] (CI 95% [5.9%, 28.9%])
    Average axial resolution: 4.3 mmAverage lateral resolution: 4.0 mm
    Ultrasound Attenuation Coefficient (ATT)Bias (dB/m @ 3.5 MHz): (-11.6%) - (1.4%)wCV: [1.2%, 2.0%] (CI 95% [0.6%, 12.6%])Bias: [-26%, 6.1%] (CI 95% [-28%, 16%])
    Precision (dB/m @ 3.5 MHz): (1.5%) - (4.5%)
    Ultrasound Backscatter Coefficient (BSC)(Predicate Device used Hepatpreanal Index (HRI), not direct BSC)wCV: [0.8%, 1.7%] (CI 95% [0.3%, 10.8%])Bias: [-13%, -1.1%] (CI 95% [-18%, 0.8%])

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The test set consisted of a "sample of calibrated phantoms." The exact number of phantoms is not specified.
      • The data provenance is in-vitro performance assessment. The document does not specify the country of origin for the phantoms or if the data was retrospective or prospective, as it's an in-vitro study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable for this study. The ground truth for the phantoms was established by their inherent calibrated properties, not by expert consensus.
    3. Adjudication method for the test set:

      • Not applicable for this study, as the ground truth was based on calibrated phantoms.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical testing is necessary to demonstrate substantial equivalence between ES Series V2 and its Predicate Device." This study focused on in-vitro technical performance and comparison to a predicate device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance testing described is a standalone (algorithm only) assessment. The device's quantitative modes (Speed of Sound, Shear Wave Speed/Stiffness, Ultrasound Attenuation, Ultrasound Backscatter Coefficient) were evaluated algorithmically against known ground truth values from calibrated phantoms.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth used was calibrated phantoms with known reference values.
    7. The sample size for the training set:

      • The document does not explicitly state the sample size for the training set. However, given that this is a 510(k) submission primarily focused on demonstrating substantial equivalence to a predicate device through performance testing, the details of the training set used for the underlying algorithms are not typically provided in this summary. The focus is on the performance of the final, developed algorithms.
    8. How the ground truth for the training set was established:

      • The document does not explicitly describe how the ground truth for the training set was established. This information is typically proprietary to the developer's internal development process and not required for a 510(k) summary demonstrating equivalence via performance testing.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213102

    Validate with FDA (Live)

    Device Name
    ES1 System
    Manufacturer
    Date Cleared
    2022-01-11

    (109 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ES1 Ultrasound Diagnostic System is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound beam attenuation and estimates of speed of sound, in internal structures of the body. This device is intended to be used by trained healthcare professionals, in a healthcare environment.

    The ES1 Ultrasound Diagnostic System is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

    In particular, the ES1 ultrasound diagnostic system is intended to provide:

    • Linear distance measurements of anatomical structures,

    • Measurement of shear wave speed at selected shear wave frequencies, and estimates in the liver and the spleen,

    • Estimates of ultrasound attenuation coefficient in the liver at 3.5 MHz.

    • Measurement of brightness ratio between structures and in particular between the liver and the kidney,

    • Estimates of speed of sound in the liver.

    The shear wave speed measurements, ultrasound beam attenuation, elastic properties estimates ratio may be used as an aid to the diagnosis, montoring and clinical management of adult and pediatric patients with liver disease.

    Device Description

    E-Scopics' ES1 System is an ultraportable ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging and quantitative imaging studies. The ES1 System consists of a Software App (running on a consumer off-the-shelf Selected Host) and an accessory curved array probe. The system produces images and quantifications, which are displayed on the monitor of the Selected Host. The ES1 System is operated from the Selected Host multi-touch screen. The ES1 System also allows the user to perform measurements, to capture images, and to generate printable reports.

    The ES1 System is designed to perform non-invasive measurements of liver/spleen shear wave speed and to estimate tissue stiffness. The e.C5-1 probe is equipped with a mechanical vibrator to produce shear waves in tissue. When used on the liver or the spleen applications, shear waves travel through the skin and intercostal space into the abdominal organ. 2D ultrasound is used to track the shear wave and measure its speed. The system then provides an estimate of tissue stiffness. In addition, the ES1 System is designed to measure several quantitative parameters from B Mode imaging: ultrasound attenuation, speed of sound, and compute ultrasound brightness ratio between 2 regions in the image. The results of ES1 quantitative imaging modalities are displayed on the Selected Host monitor.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ES1 System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/MeasurementAcceptance Criteria (Implicit from "Performance Test Summary")Reported Device Performance (In-vitro)
    B-Mode ImagingDistinguish targets with contrast ±3 dBDistinguishes targets with a contrast ratio of +/- 3 dB
    B-Mode ImagingAxial Resolution1 mm (constant over whole image depth)
    B-Mode ImagingLateral Resolution1 mm to 4 mm (depending on depth)
    Speed of Sound EstimationBiasRanges in [-0.7 %, 0.7 %]
    Speed of Sound EstimationPrecision (wCV)Ranges in [0.05 %, 0.3 %]
    Shear Wave Speed (SWS) / Stiffness (kPa) Estimation (2D TE)Bias (SWS)Ranges in [-8.6 %, 5.7 %]
    Shear Wave Speed (SWS) / Stiffness (kPa) Estimation (2D TE)Bias (Young's modulus/stiffness - kPa)Ranges in [-15.7 %, 11.7 %]
    Shear Wave Speed (SWS) / Stiffness (kPa) Estimation (2D TE)Precision (wCV) (SWS)Ranges in [0.1 %, 2.9 %]
    Shear Wave Speed (SWS) / Stiffness (kPa) Estimation (2D TE)Precision (wCV) (Young's modulus/stiffness - kPa)Ranges in [0.2 %, 6.0 %]
    Shear Wave Speed (SWS) Correlation with Predicate DeviceHigh Pearson correlation coefficient0.997
    Ultrasound Attenuation Coefficient EstimationBiasRanges in [-22.2 %, 1.4 %]
    Ultrasound Attenuation Coefficient EstimationPrecisionRanges in [1.5 %, 4.5 %]
    Ultrasound Attenuation Correlation with Predicate DeviceHigh Pearson correlation coefficient0.999
    Acoustic Output (MI, TI, ISPTA3)Below FDA Guidance limits for non-ophthalmic useAlways below the limits

    2. Sample Size for the Test Set and Data Provenance

    The document explicitly states that the tests were conducted "on a sample of calibrated phantoms." It does not specify the exact number of phantoms used.

    • Data Provenance: The tests were "in-vitro experiments" on phantoms and "on a few subjects in a clinical setting" for B-Mode and Speed of Sound. This indicates a mix of controlled lab settings and limited prospective clinical observation for certain aspects. The country of origin for the data is not specified, but the manufacturing company is E-Scopics S.A.S., France.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not mention the use of experts to establish ground truth for the test set. The ground truth was established by "calibrated phantoms with known reference values."

    4. Adjudication Method for the Test Set

    Not applicable, as ground truth was established by calibrated phantoms, not human expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document focuses on the technical performance of the device against phantoms and comparison with a predicate device, not on human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the Loop) Performance Study

    Yes, the performance data presented is for the standalone algorithm as applied to phantoms. The "Performance Data Discussion" section describes "accuracy and precision of the device ultrasound modalities...tested on calibrated phantoms" and "in-vitro experiments." While "use of the ES1 System on a few subjects in a clinical setting was also demonstrated to be feasible and safe," the quantitative performance metrics (bias, precision, correlation) are derived from the in-vitro phantom studies.

    7. Type of Ground Truth Used

    The ground truth used for the quantitative measurements (Speed of Sound, Shear Wave Speed/Stiffness, Ultrasound Attenuation) was derived from known reference values of calibrated phantoms. For B-Mode, it involved "distinguish[ing] targets having a contrast ratio of +/- 3 dB as compared to the phantom background," and assessing resolution (axial and lateral).

    8. Sample Size for the Training Set

    The document does not provide any information about the sample size for a training set. This typically implies that the device's algorithms were developed without a distinct "training set" in the machine learning sense, or that the training data details were not part of this 510(k) summary. The development may have relied on established physics-based algorithms.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no information regarding a training set is provided.