Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K251955

    Validate with FDA (Live)

    Device Name
    Onflex™ Mesh
    Date Cleared
    2025-07-24

    (29 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OnFlex™ Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, such as in the repair of inguinal hernias.

    Device Description

    The OnFlex™ Mesh is a self-expanding, non-absorbable, sterile device made from monofilament polypropylene mesh and has a lightweight large pore design. This construction allows a prompt fibroblastic response through the interstices of the mesh as observed in a preclinical model, which may not correlate to performance in humans. The OnFlex™ Mesh has an anatomical shape design to cover potential defect areas. The OnFlex™ Mesh also contains a pocket on the larger medial apex of the mesh to facilitate insertion and positioning of the device.

    The OnFlex™ Mesh contains SorbaFlex™ Memory Technology comprised of an absorbable PDO monofilament which forms an interrupted ring. SorbaFlex™ Memory Technology provides memory and stability to the device, facilitating ease of initial insertion and proper placement of the device. The PDO monofilament is folded and welded onto itself at the ends. The PDO monofilament fully degrades by means of hydrolysis in vivo in 6 – 8 months. The PDO monofilament is dyed violet by adding D & C Violet No. 2. The interrupted ring is placed within a mesh tube constructed from a knitted polypropylene monofilament. The purpose of the mesh tube is to contain the interrupted recoil ring during the degradation process. The interrupted ring, contained in the mesh tube, is sewn between two layers of mesh with polytetrafluoroethylene (PTFE) monofilament.

    The OnFlex™ Mesh has a blue limit line at the lateral portion of the mesh, composed of polypropylene monofilament dyed with Phthalocyaninato(2-) copper colorant, to provide visual feedback for where the device can be tailored. The OnFlex™ Mesh can be tailored at the opening of the interrupted ring or outside of the blue limit line.

    The OnFlex™ Mesh is offered in two sizes: medium (0115410) and large (0115411). The OnFlex™ Mesh is considered a tissue contacting permanent implant.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the OnFlex™ Mesh (K251955) indicates that this is a Special 510(k) submitted to notify the FDA of changes to the device labeling only. The device itself remains identical to its predicate, OnFlex™ Mesh (K142711), cleared in 2015.

    Therefore, the submission explicitly states: "No non-clinical or clinical testing was provided in support of this Special 510(k)."

    This means that the document does not contain any information regarding acceptance criteria or a study proving that the device meets those criteria, as no new testing was performed for this specific submission. The substantial equivalence is based on the device's identity to its predicate and the conclusion that the labeling changes do not affect safety or effectiveness.

    Because no new studies were conducted or presented in this document for the K251955 submission, I cannot provide the requested information. The document focuses solely on demonstrating that the labeling updates do not alter the substantial equivalence to the original predicate device (K142711).

    Ask a Question

    Ask a specific question about this device

    K Number
    K251557

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-12

    (22 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bard® Mesh is indicated to reinforce soft tissue where weakness exists, i.e., repair of hernias and chest wall defects.

    Bard® Mesh Pre-shaped is indicated for the repair of inguinal hernia defects.

    Device Description

    The Bard® Mesh is a nonabsorbable, sterile mesh designed for the reinforcement of soft tissue where weakness exists, i.e., repair of hernias and chest wall defects. The Bard® Mesh Pre-Shaped is a nonabsorbable, pre-trimmed sterile mesh designed for reinforcement of inguinal hernias. The Bard® Mesh is offered in several sizes of rectangular flat sheets and in several pre-shaped configurations, known as Bard® Mesh Pre-Shaped. Bard® Mesh product family products consist of the standard, rectangular or square sizes which can be trimmed as needed to provide surgeons with more freedom to customize the prosthesis prior to implantation. The Bard® Mesh portfolio currently ranges from 1" x 4" to 10" x 14" rectangle mesh. The Bard® Mesh is comprised of knitted polypropylene monofilaments in a square or rectangular flat sheet with selvage edges. The Bard Mesh Pre-Shaped is comprised of knitted polypropylene monofilaments in a pre-shaped configuration with a pre-cut into a shape commonly used by surgeons to protect the spermatic cord or similar structure as part of an inguinal hernia repair. All clinical, biological, and technical characteristics of the Bard® Mesh and Bard® Mesh Pre-Shaped are identical except that Bard® Mesh Pre-Shaped is only indicated for inguinal hernias due to the pre-cut/pre-shaped nature of the product.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for a medical device called Bard® Mesh. However, this document does not contain the information typically found in a study demonstrating how an AI/algorithm-driven device meets acceptance criteria.

    The Bard® Mesh is a surgical mesh, a physical medical device used to reinforce soft tissue, such as in hernia repair. It is a Class II device, and the clearance is based on its substantial equivalence to a preamendment predicate device (meaning it was marketed before May 28, 1976).

    The document specifically mentions:

    • "As there are no new or increased risks identified, no verification or validation activities are required for these labeling changes. Design verification testing was conducted to support the performance of the pre-shaped configuration." This indicates that for this type of physical device and the nature of the changes (labeling and a pre-shaped configuration), extensive clinical studies or complex performance evaluations are not required in the same way they would be for a novel diagnostic or AI-powered device.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment, which are typical for studies validating AI or diagnostic algorithms, is not present in this 510(k) letter. This clearance is based on the device's established technological characteristics and historical use, demonstrating its equivalence to a long-marketed product, rather than proving performance against specific acceptance criteria through a new study as would be done for an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250098

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-09

    (85 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bard® Soft Mesh is indicated for the repair of ventral, incisional, and inguinal hernias.
    Bard® Soft Mesh Pre-Shaped is indicated for the repair of inguinal hernias.

    Device Description

    The Bard® Soft Mesh is a nonabsorbable, sterile mesh designed for reinforcement of soft tissue deficiencies in adults with ventral, incisional, and inguinal hernias. The Bard® Soft Mesh Pre-Shaped is a nonabsorbable, sterile mesh designed for reinforcement of soft tissue deficiencies in adults with inguinal hernias. The knit construction allows the mesh to be stretched in both directions, in order to accommodate and reinforce tissue defects. The Bard® Soft Mesh is comprised of knitted polypropylene monofilaments in a rectangular flat sheet ranging from 2" x 4" to 12" x 12" with smooth radiused corners. The Bard® Soft Mesh Pre-Shaped is comprised of knitted polypropylene monofilaments in a pre-shaped configuration pre-cut into a shape commonly used by surgeons to protect the spermatic cord or similar structure as part of an inguinal hernia repair.

    AI/ML Overview

    This 510(k) clearance letter pertains to a surgical mesh (Bard Soft Mesh), not an AI/software device. Therefore, the detailed questions about acceptance criteria, study design referencing AI performance, expert consensus, MRMC studies, training and test sets, and ground truth establishment are not applicable to this submission.

    The 510(k) is for changes to the labeling of an existing device, with no changes to the device itself. The primary purpose of the submission is to align the labeling with European medical device regulations and to document historical product changes (like adding a new size) that were previously handled via internal documentation.

    Here's a breakdown of the relevant information provided in the document:

    1. Acceptance Criteria and Device Performance:

    Since this 510(k) is not for a new device or a functional change, but rather for labeling updates and documentation of previous minor changes, the concept of "acceptance criteria" for a new device performance study (like for an AI model) does not apply. The acceptance criteria here are implicitly related to demonstrating that the labeling changes and historical product modifications do not impact the safety or effectiveness of the device and maintain substantial equivalence to the predicate device.

    The document states: "Through the risk analysis conducted, it was determined that the labeling changes described in this 510(k) do not result in any new or increased risks. As there are no new or increased risks identified, no verification or validation activities are required for these labeling changes."

    "Non-clinical testing including mesh thickness, mesh knit construction, mesh pore size, mesh density, tensile strength, device stiffness, ball burst (strength), tear resistance, device stiffness, and suture pull out were considered in accordance with Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh issued on March 2, 1999. The non-clinical testing supported those historical changes taken under documentation to file, including dimensional line extension, do not impact the safety or effectiveness of the subject device nor do they impact the substantial equivalence between the subject and predicate devices."

    Therefore, the "performance" shown is that the device, with its updated labeling and previously documented size extension, continues to meet the physical and mechanical properties expected of a surgical mesh and is substantially equivalent to the predicate.

    Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):

    Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device PerformanceStudy Type
    Risk AssessmentNo new or increased risks from labeling changes or historical modifications.Determined that labeling changes and historical modifications do not result in any new or increased risks.Risk Analysis
    Physical PropertiesMaintenance of established physical properties (mesh thickness, knit construction, pore size, density).Non-clinical testing supported that historical changes (e.g., dimensional line extension) do not impact safety or effectiveness.Non-Clinical Testing
    Mechanical PropertiesMaintenance of established mechanical properties (tensile strength, device stiffness, ball burst (strength), tear resistance, suture pull out).Non-clinical testing supported that historical changes (e.g., dimensional line extension) do not impact safety or effectiveness.Non-Clinical Testing
    Substantial EquivalenceSubject device remains substantially equivalent to the predicate device."The subject device and predicate device have the same intended use and technological characteristics... The subject device... is substantially equivalent to the predicate device."Comparison Assessment

    2. Sample Size for Test Set and Data Provenance:

    • Sample Size for Test Set: Not applicable. This submission is based on non-clinical testing of physical device properties and a comparison to a predicate, not clinical studies with a "test set" of patient data for an AI model. The "test set" refers to the specific mesh samples used for physical and mechanical property testing. The document states "Non-clinical testing including mesh thickness, mesh knit construction, mesh pore size, mesh density, tensile strength, device stiffness, ball burst (strength), tear resistance, device stiffness, and suture pull out were considered," implying relevant samples of the mesh were tested, but specific sample numbers are not provided in this summary.
    • Data Provenance: Not applicable in the context of patient data. The "data" pertains to the characteristics of the physical mesh device.

    3. Number of Experts used to establish ground truth:

    • Not applicable. There is no "ground truth" establishment by experts in the context of an AI model's performance on patient data. The non-clinical testing results establish the performance of the physical device.

    4. Adjudication Method:

    • Not applicable. No expert adjudication process for an AI model's output is relevant to this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not applicable. This is not an AI-assisted diagnostic device study.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. This is not an algorithm.

    7. Type of Ground Truth Used:

    • Not applicable in the context of an AI model. The "ground truth" for this device's performance is established by objective non-clinical engineering and material property testing methods as per relevant standards (e.g., those referenced in the 1999 FDA guidance for surgical mesh).

    8. Sample Size for Training Set:

    • Not applicable. No AI model is being trained here.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241250

    Validate with FDA (Live)

    Device Name
    3DMax Light Mesh
    Date Cleared
    2024-06-14

    (42 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K233402

    Validate with FDA (Live)

    Device Name
    Bard 3DMax Mesh
    Date Cleared
    2023-11-01

    (28 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bard® 3DMax™ Mesh is indicated for use in the reinforcement of soft tissue where weakness exists, in the repair of inguinal hernias.

    Device Description

    The Bard® 3DMax™ Mesh is anatomically designed to fit the inquinal anatomy. The device is curve shaped and preformed with sealed edges that allow for easier positioning of the device than a traditional flat sheet of mesh in a laparoscopic inguinal hernia repair. The device is constructed of knitted polypropylene monofilaments 7.5 millimeters in diameter. The knit construction allows the mesh to be stretched in both directions in order to accommodate and reinforce tissue defects. The device contains an orientation marker in an M shape with an arrow to facilitate mesh positioning and placement.

    AI/ML Overview

    This submission, K233402, describes a Special 510(k) for the Bard 3DMax Mesh, which focuses solely on labeling changes to align with European medical device regulations and other devices in the Davol Inc. portfolio. The submission explicitly states:

    "No non-clinical or clinical testing was provided in support of this Special 510(k)." and "The purpose of this Special 510(k) is to notify FDA of the changes to the Bard® 3DMax™ Mesh labeling. There are no changes to the device itself."

    Therefore, based on the provided document, there are no acceptance criteria or studies associated with this specific submission to describe regarding device performance, as the device itself (materials, design, manufacturing, etc.) has not changed from its predicate, K081010.

    To provide the requested information, one would need access to the original 510(k) submission (K081010) that established the substantial equivalence of the device itself, rather than just its labeling.

    Since the prompt asks for information from the provided input, and the input explicitly states no testing was done, I cannot fulfill the request for acceptance criteria and study details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191287

    Validate with FDA (Live)

    Date Cleared
    2019-07-16

    (64 days)

    Product Code
    Regulation Number
    878.4750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

    Device Description

    The OptiFix™ AT Absorbable Fixation System with Articulating Technology, herein after referred to as "OptiFix™ AT", is a sterile (via gamma) single use device that is comprised of a deployment component and an absorbable fastener component. The device is available in either 15 or 30 preloaded fasteners. The tip of the shaft can be articulated, and the handle of the device can be rotated 360 degrees to facilitate access for fixation during surgery. The fasteners are designed with retention features and are manufactured from Poly (L-lactide and glycolide) and are dyed with D&C Violet No. 2.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state specific quantifiable acceptance criteria (e.g., minimum tensile strength, maximum deployment force) or numerical performance metrics for the OptiFix™ AT device. Instead, it uses qualitative statements about the device meeting established requirements.

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilitySuccessfully completed testing "in accordance with the FDA Guidance" and "International Standard ISO 10993-1."
    Product Testing"Passed all the test requirements and demonstrated that the proposed device design meets product specifications and intended uses. All samples tested met the established acceptance criteria."
    Package Integrity"Completed... in accordance with the following FDA consensus standards: ISO 11607-1+A1:2014" and "ISO 11607-2:2006+A1:2014."

    2. Sample Sizes and Data Provenance

    The document does not specify the sample sizes used for the product testing (design verification and validation).

    • Test Set Sample Size: Not specified.
    • Data Provenance: Not explicitly stated, but given the nature of pre-market submissions in the US, the testing would generally occur in the US or in adherence to US/international standards. It is retrospective in the sense that the testing was completed prior to the submission.

    3. Number of Experts and Qualifications for Ground Truth Establishment

    This information is not applicable as the device is a physical medical stapling system, not an AI/software device that requires expert-established ground truth for a test set. The performance is assessed through engineering and biological testing.

    4. Adjudication Method for the Test Set

    This information is not applicable as the device is a physical medical stapling system. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or for setting ground truth in imaging/AI studies, which were not conducted for this submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical studies were not performed for the submission of this device nor were clinical studies performed for the predicate device." MRMC studies are typically for evaluating diagnostic devices where human readers interpret results.

    6. Standalone (Algorithm Only) Performance Study

    No, a standalone performance study was not done. This device is a physical medical staple system and does not involve an algorithm.

    7. Type of Ground Truth Used

    The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. For this physical device, the "ground truth" for verifying performance relies on:

    • Engineering Specifications: The device's design, materials, and functional performance are compared against predetermined engineering specifications and safety standards.
    • Biocompatibility Standards: Performance against established international standards (ISO 10993).
    • Packaging Standards: Performance against established international standards (ISO 11607).

    8. Sample Size for the Training Set

    This information is not applicable. This is a physical medical device, not an AI algorithm, so there is no training set in the AI sense. The manufacturing process and quality control would involve internal testing of components and assemblies that could be analogous to "training" in a broader engineering sense, but not for an "AI training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as point 8. The "ground truth" for the design and manufacturing of such a device is established through engineering design principles, material science, applicable regulatory standards, and risk assessments.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101851

    Validate with FDA (Live)

    Date Cleared
    2010-07-15

    (14 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ventralight ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.

    Device Description

    The Ventralight ST Mesh is a low profile, sterile, single use device indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. This device is co-knitted using polypropylene (PP) and polyglycolic acid (PGA) fibers to result in a two-sided mesh with a PP surface and a PGA surface. The mesh is coated on the PGA surface with a bioresorbable, chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC) and polyethylene glycol (PEG) based hydrogel. The fascial side of the mesh allows a prompt fibroblastic response through the interstices of the mesh, encouraging complete tissue ingrowth, similar to polypropylene mesh alone. The visceral side of the mesh is a bioresorbable coating, separating the mesh from underlying tissue and organ surfaces to minimize tissue attachment to the mesh. Shortly after placement, the biopolymer coating becomes a hydrated gel that is resorbed from the site in less than 30 days.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Ventralight ST Mesh, a surgical mesh. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria based on quantitative performance metrics like sensitivity, specificity, or F-score often associated with AI/ML-based devices.

    Therefore, many of the requested elements for an AI/ML device study, such as sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable in this context.

    Here's an analysis based on the information provided, highlighting the aspects that are relevant to this type of medical device submission:


    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of traditional surgical mesh device, "acceptance criteria" are generally tied to demonstrating substantial equivalence to a legally marketed predicate device, focusing on similar safety and effectiveness. The performance is assessed through non-clinical testing.

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Properties Comparability: Device should maintain appropriate mechanical characteristics for its intended use and be comparable to the predicate device.Bench Testing Results:- PGA pullout strength testing completed.- Burst strength testing completed.- Tear strength testing completed.- Suture retention strength testing completed.- Dry bond strength testing completed.- Hydrogel disruption testing after laparoscopic deployment completed.Conclusion: "Results of testing demonstrate that the Ventralight ST Mesh design meets product specifications and intended uses."
    Biocompatibility: Device materials should be biocompatible and comparable to the predicate device.Biocompatibility Testing: Completed (similar to predicate device, as there were no material changes).
    Tissue Interaction (reduced attachment, ingrowth): The device should demonstrate similar tissue ingrowth and minimal tissue attachment as the predicate device.Animal Study Results (4-week post-implantation porcine model, comparative): Evaluated:- Tissue attachment- Mesh contracture- Tissue in-growth- Host inflammatory and fibrotic responseConclusion: Results indicated similar performance to the predicate device ("performance testing demonstrates that these differences do not adversely affect the safety and effectiveness of the proposed device.").
    Sterilization: Device sterilization method should be effective and comparable to the predicate.Sterilization: Stated as "the same as the predicate device."
    Packaging: Device packaging should maintain integrity and sterility.Packaging: Stated as "the same as the predicate device." Tested seal strength and package qualification.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated with specific numbers, but the document mentions a "4 week post-implantation study in a porcine model." This implies a limited number of animal subjects. Bench testing also does not specify the number of samples tested for each metric, but these are typically standardized tests performed on multiple samples.
    • Data Provenance: The animal study was conducted in a "porcine model," indicating an animal study rather than human clinical data. Bench testing is laboratory-based. This is prospective testing of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This concept is not applicable here because "ground truth" in this context would refer to objective measurements from bench tests or histological evaluations from the animal study. These would typically be performed and interpreted by laboratory technicians or veterinary pathologists, not "experts" in the sense of clinical reviewers establishing diagnostic truth for AI.

    4. Adjudication method for the test set

    • Not applicable. Adjudication methods are used when multiple human readers interpret an image or clinical case and their interpretations need to be reconciled to establish a ground truth. For bench tests and animal studies, results are typically quantitative measurements or pathological findings with established protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic devices, not for a surgical mesh.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical surgical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the animal study: Pathology/Histology (e.g., evaluating tissue attachment, ingrowth, inflammatory response) and observational data from the porcine model.
    • For bench testing: Objective physical measurements (e.g., strength values, disruption levels) using laboratory equipment and established test methods.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device, so there is no training set in the computational sense. The "training" for the device's development would be through design iterations, material science knowledge, and manufacturing processes.

    9. How the ground truth for the training set was established

    • Not applicable. Since there is no training set in the AI/ML context, there is no "ground truth" to establish for a training set. The device design and materials are based on established engineering principles and prior knowledge from the predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092032

    Validate with FDA (Live)

    Date Cleared
    2009-12-08

    (155 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bard The PerFix™ Light Plug is indicated for reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as groin hernia defects.

    Device Description

    The Proposed Product, Bard® PERFIX™ Light Plug is a pre-formed three-dimensional (cone shape) device constructed of a fluted outer layer and multiple inner layers of mesh attached at the tip. The device has a large pore design and is constructed of knitted polypropylene monofilaments. The pores in the mesh allow for tissue ingrowth. A separate flat, pre-shaped onlay patch (2.3" x 5.4") will be packaged with each device. The Bard® PERFIX™ Light Plug will be offered in a range of sizes with the largest being the extra large PerFix™ Plug 1.5" x 2.0" (4.1cm x 5.0cm). The extra large size consists of three (3) preformed medium shells held in place, side by side, inside a pre-formed large shell. The small, medium and large sizes consist of an outer cone with 2 pieces of mesh (petals) to help the plug maintain its fluted form. The inner petals and cones are sewn together at the tip with a single polypropylene monofilament thread.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Bard® PERFIX™ Light Plug:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/AttributeAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Physical PropertiesThicknessSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Pore SizeSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    DensitySubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    StiffnessSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Mechanical PropertiesTensile StrengthSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Percent ElongationSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Suture Pullout StrengthSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Burst StrengthSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Tear ResistanceSubstantially equivalent to predicates"Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing the PERFIX™ Light Plug is substantially equivalent to the referenced Predicates." (General statement, no specific values reported)
    Material PropertiesBiocompatibilityAcceptable results obtained"Biocompatibility testing was completed on the polypropylene material used in the Proposed PERFIX™ Light Plug and has received acceptable results."
    Overall EquivalenceSubstantial EquivalenceDemonstrated through laboratory testing"The results show that the Proposed Product is substantially equivalent to the currently marketed Predicate Products. Therefore, based on laboratory testing and biocompatibility data, the Proposed Product, PERFIX™ Light Plug, is safe and effective for its intended use."

    Study Description:

    The study conducted to demonstrate the safety and effectiveness and to meet the acceptance criteria of the Bard® PERFIX™ Light Plug was a series of laboratory bench tests and biocompatibility testing. This was a comparative study against legally marketed predicate devices: Bard® Soft Mesh, Bard® PERFIX™ Plug, Bard® Mesh Dart, and Ethicon Mersilene Mesh.

    The study aimed to show substantial equivalence of the new device to these predicates, as required for a 510(k) premarket notification. The guidance followed was "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh" (March 2, 1999).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample sizes for each of the laboratory bench tests (e.g., number of mesh samples tested for tensile strength).
    • Data Provenance: The data is from laboratory bench testing, implying it was conducted in a controlled lab environment. The document does not explicitly state the country of origin, but given it's a 510(k) submission to the FDA, it's presumed the testing was conducted in a manner acceptable for US regulatory purposes, likely either domestically or by a certified international lab. The study is retrospective in the sense that it's comparing a new device against existing, predicate devices using established testing methods.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" for this device, which is a physical surgical mesh, is established through objective, quantifiable laboratory measurements of its physical, mechanical, and biological properties, not by human expert assessment of images or clinical outcomes.

    4. Adjudication Method for the Test Set

    This is not applicable. The data presented relies on objective laboratory measurements rather than subjective expert consensus or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This type of study is typically relevant for interpretative devices (e.g., imaging AI) where human readers evaluate cases with and without AI assistance. The Bard® PERFIX™ Light Plug is a physical surgical implant, so an MRMC study is not relevant.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    A standalone performance study was performed, but it's not an "algorithm only" study. It's a "device only" study focusing on the intrinsic material and design characteristics of the surgical mesh without human interaction during the performance testing itself. The laboratory bench tests (thickness, pore size, stiffness, tensile strength, etc.) and biocompatibility tests represent the standalone performance of the device's physical and biological properties.

    7. Type of Ground Truth Used

    The ground truth used was based on objective, quantifiable laboratory measurements of the physical, mechanical, and material properties (biocompatibility) of the device and its predicates. These measurements are considered the "true" performance characteristics against which the new device was compared to demonstrate substantial equivalence.

    8. Sample Size for the Training Set

    This information is not applicable. The Bard® PERFIX™ Light Plug is a physical medical device, not a machine learning algorithm or AI model that requires a "training set" in the computational sense. The design and manufacturing process for the device would have been developed through R&D and engineering processes, not through machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as #8. There is no "training set" in the context of this device. The "ground truth" for the device's design and manufacturing would come from engineering specifications, material science principles, and established medical device standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081010

    Validate with FDA (Live)

    Device Name
    BARD 3DMAX MESH
    Date Cleared
    2008-10-07

    (181 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bard 3DMAX Mesh is a sterile, single use device indicated to reinforce soft tissue where weakness exists e.g. for repair of hernias and chest wall defects.

    Device Description

    The modified Bard 3DMAX Mesh is anatomically designed to fit the inguinal anatomy. The device is curve shaped and preformed with sealed edges that allow for easier positioning of the device than a traditional flat sheet of mesh in a laparoscopic inguinal hernia repair. The device is constructed of knitted polypropylene monofilaments 7.5 millimeters in diameter. The knit construction allows the mesh to be stretched in both directions in order to accommodate and reinforce tissue defects. The device contains an orientation marker in an M shape with an arrow to facilitate mesh positioning and placement.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Bard 3DMAX Mesh. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific acceptance criteria through a clinical study with detailed performance metrics.

    Therefore, much of the requested information (such as specific performance metrics, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) is not applicable to this type of regulatory submission and is not present in the provided text.

    The "study" referenced in the document is a comparison to predicate devices, focusing on technological characteristics, performance, and intended use, rather than a clinical trial with acceptance criteria for a new AI or diagnostic device.

    Here's a breakdown of what can be extracted based on the provided text, and what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria or specific quantitative performance metrics like sensitivity, specificity, or accuracy. The "performance" assessment focuses on demonstrating substantial equivalence to predicate devices.

    Acceptance Criteria (Not Explicitly Stated)Reported Device Performance
    Biocompatibility: Device materials must not cause adverse biological reactions."Biocompatibility testing was performed on the modified 3DMAX device. Testing performed to date indicates that changes made to the 3DMAX device do not impact its biocompatibility profile."
    Technological Characteristics: Must be similar to predicate devices."laboratory testing was performed to compare the modified 3DMAX Mesh device to the currently marketed 3DMAX Mesh and Bard Mesh devices. The results show that the modified device is substantially equivalent to the currently marketed predicate devices." "The modified 3DMAX Device differs from the currently marketed 3DMAX device in medial marker material only. The main components of the medial marker, polypropylene and Phthalocyaninato(2-) copper dye will remain the same. However, these items will be purchased from a different vendor and extruded to form a dyed polypropylene monofilament."
    Intended Use: Must be the same as predicate devices."The intended use for the modified device is exactly the same as the predicate devices, 3DMAX Mesh, Bard Mesh and Usher's Marlex Tubular Mesh."
    Safety and Effectiveness: Must be safe and effective for its intended use."Therefore, based on laboratory testing and biocompatibility data, the modified 3DMAX device is safe and effective for its intended use."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The "testing" mentioned refers to laboratory and biocompatibility tests, not a clinical test set with human subjects or data provenance in the context of AI.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The assessment relies on laboratory and biocompatibility testing, not expert-adjudicated ground truth data for a diagnostic algorithm.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. The Bard 3DMAX Mesh is a surgical mesh, not an AI-powered diagnostic tool, so MRMC studies are irrelevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" equivalent for this device is based on biocompatibility testing standards and laboratory tests comparing material properties and performance characteristics to the existing predicate devices.

    8. The sample size for the training set

    This information is not applicable and not provided. There is no "training set" in the context of a physical surgical mesh.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061754

    Validate with FDA (Live)

    Date Cleared
    2006-10-23

    (124 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bard Composix L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects.

    Device Description

    The Composix® L/P Mesh, will consist of a layer of polypropylene mesh and a layer of expanded polytetrafloroethylene (ePTFE) stitched together using Polytetrafluroethylene (PTFE) monofilament thread sewn in a lock stitch formation. The polypropylene layer is knitted from 0.0048 inch polypropylene monofilament, which is identical to the currently marketed Large Pore Soft™ Mesh. The thickness of the ePTFE layer is identical to the currently marketed Composix® E/X Mesh. As with the currently marketed Composix E/X, the peripheral edge of the polypropylene mesh will be heat sealed to the ePTFE layer.

    The layers of the Composix L/P Mesh will be sewn together using rows of interspaced stitching, which will allow the surgeon to tailor the product. Using the stitch pattern as a guide, the surgeon can tailor the product. For user convenience to facilitate the deployment of the larger sized Composix L/P products, the larger sizes of the Composix L/P products will be packaged with an Introducer Tool. This Introducer Tool is similar to the rolling tines/plunger assembly and T-handle of the PrecisionPass™ Delivery Device. The Introducer Tool consists of a handle with a thin stainless steel rod. The stainless steel rod is split into two pieces along its length and has ridges around the perimeter to facilitate use. Similar to the T-handle provided with the PrecisionPass Device, the Introducer Tool's T-cap will be used to secure the open end of the metal rolling tines during the rolling process, then removed prior to delivering the device through the trocar.

    AI/ML Overview

    This document is a 510(k) Summary for the Bard Composix® L/P Mesh, a surgical mesh. The submission focuses on demonstrating substantial equivalence to predicate devices, rather than establishing de novo acceptance criteria through a clinical study with detailed performance metrics. Therefore, many of the requested elements for a typical AI/software device study are not directly applicable or available in this type of submission.

    Here's an analysis of the provided text in relation to your questions:

    1. A table of acceptance criteria and the reported device performance

    This
    submission does not define specific quantitative "acceptance criteria" in the way an AI/software device would, nor does it report numerical "device performance" in terms of accuracy, sensitivity, or specificity. The acceptance criteria in a 510(k) for a medical device like a surgical mesh are typically based on demonstrating substantial equivalence in materials, design, manufacturing, intended use, and performance characteristics to a legally marketed predicate device.

    The "performance" reported primarily revolves around demonstrating that the new device is as safe and effective as the predicate devices.

    Acceptance Criteria (Inferred from Substantial Equivalence)Reported Device Performance (as stated in the submission)
    Intended Use Equivalence: Same indications for use as predicate devices."Bard Composix® L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects." (Same as predicate)
    Technological Characteristics Equivalence: Similar physical attributes, performance characteristics, and materials as predicate meshes."The Proposed Product has similar physical attributes, performance characteristics, and materials as the Predicates Bard Composix® E/X and Bard Large Pore Soft™ Mesh."
    Manufacturing Process Equivalence: Similar manufacturing methods."Furthermore, it has similar manufacturing methods as the Predicate Bard Large Pore Soft™ Mesh and Predicate Bard Composix E/X Mesh."
    Material Equivalence: Materials used are safe and similar to those in predicate devices.The device consists of "polypropylene mesh and a layer of expanded polytetrafloroethylene (ePTFE)" similar to predicate components. "Biocompatibility testing completed on the product manufactured with the same materials and similar processes... demonstrate the device is non-toxic and non-sensitizing to biological tissues consistent with its intended use."
    Performance (Bench Test) Equivalence: Laboratory testing demonstrates material chosen and design utilized are substantially equivalent."Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing Bard Composix® L/P Mesh is substantially equivalent to the referenced Predicate Products."
    Safety and Effectiveness: Biocompatibility and bench testing support safety and effectiveness for intended use."Biocompatibility and bench testing provided in this submission support the safety and effectiveness of Bard Composix® L/P Mesh for its intended use."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The submission does not describe a "test set" in the context of evaluating an AI/software device. The performance data is based on "Biocompatibility and bench testing." These types of tests involve material samples and controlled laboratory experiments, not patient data in the typical sense of a "test set" for diagnostic performance. Therefore, sample sizes for patient data, country of origin, and retrospective/prospective nature are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is a physical surgical mesh, not an AI/software device that requires expert-established ground truth on patient data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is a physical surgical mesh, not an AI/software device that requires expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical surgical mesh, not an AI or diagnostic device that would involve human readers or MRMC studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical surgical mesh, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for a physical device like a surgical mesh is established through a combination of:

    • Material specifications and standards: Confirming the physical and chemical properties of the materials meet established safety and performance benchmarks.
    • Biocompatibility testing: Demonstrating the material does not cause adverse biological reactions.
    • Bench testing: Laboratory tests to evaluate mechanical properties (e.g., tensile strength, tear resistance) and other functional aspects relevant to its use as a surgical mesh.
    • Predicate device history: Relying on the established safety and effectiveness profile of the legally marketed predicate devices, combined with evidence that the new device is sufficiently similar.

    The document states: "Biocompatibility and bench testing provided in this submission support the safety and effectiveness of Bard Composix® L/P Mesh for its intended use." and "Laboratory testing discussed in this submission demonstrate that the material chosen and the design utilized in manufacturing Bard Composix® L/P Mesh is substantially equivalent to the referenced Predicate Products."

    8. The sample size for the training set

    Not applicable. This is a physical surgical mesh, not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical surgical mesh, not an AI/software device that requires a training set with ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2