Found 6 results

510(k) Data Aggregation

    K Number
    K250739

    Validate with FDA (Live)

    Date Cleared
    2025-11-07

    (241 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Primeprint Aligner is a light curing resin intended for the treatment of tooth malocclusion in patients with developed permanent dentition. Primeprint Aligner repositions teeth by way of continuous gentle force.

    Intended place of use: Orthodontic/dental practices/clinics, dental laboratories

    Device Description

    Primeprint Direct Aligner is a light-curing material for the production of aligners using the Primeprint Solution (Primeprint 385nm DLP 3D printer and Primeprint Post Processing Unit). The material is intended for the treatment of tooth malocclusion in patients with developed permanent detention. Primeprint Direct Aligner repositions teeth by way of continuous gentle force.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K200580

    Validate with FDA (Live)

    Device Name
    FotoDent denture
    Date Cleared
    2020-11-19

    (259 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FotoDent® denture is a light curing resin intended for manufacturing of full and partial removable dentures.

    Device Description

    FotoDent® denture is a light curable resin for 3D printing of full and partial dentures.

    AI/ML Overview

    The provided submission describes FotoDent® denture, a light-curing resin for 3D printing of full and partial dentures. This is a Class II device (21 CFR 872.3760).

    1. Acceptance Criteria and Reported Device Performance

    The device was tested against the applicable requirements of ISO 20795-1:2013 (Dentistry – Base polymers – Part 1: Denture base polymers) and ISO 7491:2000 (Dentistry – Dental materials – Determination of colour stability). The submission states that FotoDent® denture "met the applicable requirements" of these standards.

    A direct comparison of physical properties between FotoDent® denture and the predicate device (DENTCA Denture Base II) is provided, which implicitly sets the acceptance criteria based on demonstrating "similar physical properties" to the predicate.

    AttributeAcceptance Criteria (Predicate)FotoDent® denture Performance (Reported)
    Indications
    Manufacturing of full denturesYesYes
    Manufacturing of partial denturesYesYes
    Physical Property
    Type 4 (light-activated) acrylic resinYes (per ISO 20795-1)Yes (per ISO 20795-1)
    Before Curing (liquid state)
    Viscosity1000 < X < 2000 cps400 < X < 600 cps
    Density1.05 < X < 1.2 g/cm³1.05 – 1.15 g/cm³
    After Curing (solid state)
    Density1.15 < X < 1.25 g/cm³1.05 – 1.15 g/cm³
    Flexural Strength> 65 MPa> 80 MPa
    Flexural Modulus> 2000 MPa> 2000 MPa
    Material TypeResin basedResin based
    Chemical CharacterizationMethacrylate/acrylate resins with photo-initiators, pigments and additivesMethacrylate/acrylate resins with photo-initiators, pigments and additives
    Shelf Life2 years2 years
    Storage15-25°C, Do not expose to direct sunlight18-28°C, Do not expose to direct sunlight
    Physical ConfigurationSupplied in liquid formSupplied in liquid form
    Shades4 shades of pink2 shades of pink
    Application3D Printing3D Printing
    Fabrication TypeAutomated 3D printing of resin in multiple layers, each layer light-cured before adding next layerAutomated 3D printing of resin in multiple layers, each layer light-cured before adding next layer
    Polymerization (Curing Method)Light-curing resinLight-curing resin
    Post Curinglight-curing unitlight-curing unit
    Teeth AssemblyBondingBonding
    SterileNoNo
    Single UseNoNo
    Environment of UseDental LaboratoryDental Laboratory

    Note: The submission clarifies that "The differences in physical properties between FotoDent denture and the predicate device does not impact safety and effectiveness, as the finished clinical product is a custom-fitted denture base regardless of the material variation." This indicates that while some properties differ, they are deemed within an acceptable range for the intended use and do not negatively affect clinical outcomes compared to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The submission mentions "Bench test results" and "Non-Clinical Performance Testing" and "Biocompatibility" testing.

    • Sample Size: The document does not specify the sample sizes used for these tests.
    • Data Provenance: The document does not specify the country of origin of the data or whether the tests were retrospective or prospective. Given these are bench tests and biocompatibility evaluations, they are inherently prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The tests performed are non-clinical (bench tests and biocompatibility), and therefore do not involve human expert ground truth establishment in the way clinical studies with image interpretation or diagnosis would.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the tests are non-clinical, and thus an adjudication method for human interpretation is irrelevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states under "Clinical Performance Data": "Not applicable. No human clinical testing was performed to support the substantial equivalence of FotoDent® denture."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. FotoDent® denture is a physical material (resin) for 3D printing, not an algorithm or a software device that would have standalone performance.

    7. The Type of Ground Truth Used

    For the non-clinical performance and biocompatibility testing, the "ground truth" is established by the specifications and standard methodologies outlined in the referenced ISO standards (ISO 20795-1:2013, ISO 7491:2000, ISO 10993-1:2009, ISO 7405:2008). These standards define the acceptable range or characteristics for the material properties.

    8. The Sample Size for the Training Set

    Not applicable. As FotoDent® denture is a physical material, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190571

    Validate with FDA (Live)

    Date Cleared
    2019-11-07

    (246 days)

    Product Code
    Regulation Number
    872.3660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StoneBite® and StoneBite® scan are addition-curing silicones for bite registrations.

    StoneBite® is indicated for:

    • bite registrations
    • I key for intraoral registration

    StoneBite® scan is indicated for:

    • i bite registrations
    • key for intraoral registration
    • . optical collection of data in CAD/CAM/CIM systems
    Device Description

    StoneBite and StoneBite scan are addition-curing silicones for bite registration, used in dental applications to produce an accurate reproduction of the patient's occlusal record.

    StoneBite and StoneBite scan are vinyl polysiloxane based materials.

    StoneBite and StoneBite scan are medium-body elastomeric bite registration materials for dental applications to produce an accurate reproduction of the patient's occlusal record. StoneBite and StoneBite scan are two-component (base and catalyst) additioncuring materials.

    AI/ML Overview

    The document describes the non-clinical performance testing for dental impression materials, specifically StoneBite and StoneBite scan. This is not medical AI device and therefore does not contain any of information requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171562

    Validate with FDA (Live)

    Date Cleared
    2018-04-13

    (318 days)

    Product Code
    Regulation Number
    872.3660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dynax® Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.

    Device Description

    Dynax® Dental Impression Materials are designed for dental applications to define and reproduce the structure of a patient's teeth and gums for producing crowns, bridges, occlusal and dental implant. Base and catalyst components are mixed in equal ratio 1:1, placed into an impression tray and inserted into the patient's mouth. The material will conform to the patient's dentition and when set will produce a reproduction of the patient's teeth and occlusion.

    AI/ML Overview

    The acceptance criteria and study proving device performance are detailed below based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Predicate Device K053427 Fresh Bold)Reported Device Performance (Dynax® clear, putty, heavy body, light, mono)
    Working / Processing time90 - 120 sec90 sec
    Setting time / Time in the mouth120 - 140 sec90 sec
    Hardness42 - 70 Shore A46 - 70 Shore A
    Working humidity50 %50 %
    Dimensional accuracy99.9 % - 99.2 %99.9 % - 99.2 %
    Stability (linear dimensional change)< 0.1 % typical< 0.2 % typical
    ConsistencyType 0 - type 3 DIN EN ISO 4823Type 0 - type 3 DIN EN ISO 4823

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set. It mentions "a series of in-house tests have been conducted to verify the intended signals are accurate and can maintain performance over its useful life" and that the device meets its product specification.

    The data provenance is not explicitly mentioned as country of origin, but the manufacturer is Dreve Dentamid GmbH, located in Unna/Germany. The testing appears to be "in-house."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The testing described focuses on physical and chemical properties of the impression material, not on interpretations by experts.

    4. Adjudication Method for the Test Set

    Adjudication methods (e.g., 2+1, 3+1) are not applicable or mentioned, as the testing relates to measurable physical properties rather than subjective interpretations by multiple experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document presents a comparison of physical and chemical properties of the device with a predicate device. It does not describe an MRMC comparative effectiveness study involving human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. The device is a dental impression material, not an algorithm or AI. Performance is evaluated based on its physical properties.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is established by standardized physical and chemical property measurements, benchmarked against the acceptable ranges defined by the predicate device and relevant international standards (e.g., ISO 4823). These are objective measurements rather than expert consensus, pathology, or outcomes data in a medical diagnostic sense.

    8. The Sample Size for the Training Set

    This information is not applicable and is not provided. The development and testing of this dental impression material do not involve a "training set" in the context of machine learning or AI models.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this type of device. The ground truth for the product's performance is derived from its adherence to documented specifications and international standards for dental impression materials.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162408

    Validate with FDA (Live)

    Date Cleared
    2018-04-05

    (584 days)

    Product Code
    Regulation Number
    872.3310
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Varnish for final coating for resin temporaries to produce a smooth and shiny surface.

    Dental sealing lacquer for coating resin temporaries

    Device Description

    The light-curing one-component dental sealing lacquers NanoVarnish, Plaquit and Lightpaint on Surface are designed for coating resin parts of complete and partial dental prosthesis and for provisional crowns and bridges. The lacquer is applied to a dental restoration and cured with the help of a dental light-curing unit.

    AI/ML Overview

    This document is a 510(k) premarket notification for a dental sealing lacquer. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of AI-powered device performance. Therefore, many of the requested criteria regarding AI acceptance, human readers, and ground truth for AI model training cannot be extracted from this document.

    Here's what can be extracted based on the provided text:

    Acceptance Criteria and Device Performance

    For this type of device (a dental material), "acceptance criteria" are based on demonstrating comparable physical and chemical properties to a legally marketed predicate device, rather than accuracy metrics for an AI system. The study performed is a comparative study of the new device against a predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Predicate: VITA ENAMIC® Glaze K123761)Reported Device Performance (NanoVarnish, Plaquit, Lightpaint on Surface)
    Intended UseSealing lacquer for characterization of dental restorationsDental sealing lacquer for coating resin temporaries
    Indications for UseFinal coating for shade customization/characterization, protection from abrasion wearVarnish for final coating for resin temporaries to produce a smooth and shiny surface
    DesignSupplied in glass bottlesSupplied in glass bottles
    Chemical DescriptionOne component UV-light curing (meth)acrylic lacquer radical polymerizationOne component UV-light curing (meth)acrylic lacquer radical polymerization
    Appearance (visual)Colorless, liquidColorless, liquid
    Viscosity at 23°C≤ 0.15≤ 0.15 - 0.02
    Pencil hardness (acc. to Wolff Wilborn)2 H2 H
    Working / Processing time< 1 min.< 1 min.
    Curing timeapprox. 3 min.approx. 3 - 5 min.
    TemperatureRoom temperatureRoom temperature
    Shelf life2 years2 years
    Conformance to consensus standardsEN ISO 13485: 2012 + AC:2012, ISO 7405: 2008, ISO 14971: 2007, ISO 10993-1:2009EN ISO 13485: 2012 + AC:2012, ISO 7405: 2008, ISO 14971: 2007, ISO 10993-1: 2009

    Explanation of Differences (as stated in the document): The document explicitly states that "The difference of the Intended Use and the Indications for Use of the Dreve dental sealing lacquers (Plaquit, NanoVarnish and Lightpaint on Surface) and the predicate are due to the availability of the predicate K123761 VITA ENAMIC® Glaze in a set of devices." It clarifies that the Glaze material within the predicate's set is used "for coating purpose only" and provides "protection of the stains from abrasion wear." The document concludes that "The technological characteristics of the subject device and the predicate are virtually identical and both devices have generically the same intended use."


    Information NOT available in the document (due to the nature of the device and the submission):

    This document describes a conventional medical device (a dental varnish), not an AI/ML-powered device. Therefore, the following information, which is relevant for AI acceptance criteria, is not present and cannot be extracted:

    1. Sample sizes for test set and data provenance: Not applicable. The "study" here is a set of comparable physical and chemical tests, not a clinical trial with patient data or an AI model's test set.
    2. Number of experts and their qualifications for ground truth: Not applicable for a physical material. Ground truth in this context refers to standard measurements/tests.
    3. Adjudication method for the test set: Not applicable.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not applicable. This is for AI-assisted human performance studies.
    5. Standalone (algorithm only without human-in-the loop performance): Not applicable as there is no algorithm.
    6. Type of ground truth used: For a physical material, the "ground truth" is established through standardized laboratory testing and material characterization against defined specifications and recognized consensus standards (e.g., ISO standards for dental materials). The document states: "Testing has confirmed these devices meet their product specification."
    7. Sample size for the training set: Not applicable. This document is not about an AI model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171729

    Validate with FDA (Live)

    Device Name
    Fixtemp C&B
    Date Cleared
    2017-12-13

    (184 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fixtemp® C&B is a resin based material used to fabricate temporary crowns and bridges.

    Device Description

    Fixtemp® C&B is a two-component automatically mixable composite on the basis of multifunctional (meth)acrylates provided in standard double cartridges. The material is dispensed and mixed by cartridge/mixing tip combination and it is used for the manufacturing of temporary crowns and bridges for use until the permanent restoration is fabricated. It can be also used for inlays, onlays and veneers and is available in tooth-colors A1, A2, A3, A3.5, B1, Bleach X and D2. This is a prescription only material. The labeling and working instructions are designed for health care professionals.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Fixtemp® C&B

    The Fixtemp® C&B device is a resin-based material used to fabricate temporary crowns and bridges. Its performance was evaluated against a set of non-clinical criteria to establish substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (or Reference Standard)Reported Device Performance
    Working / Processing time≥ 45 sec≥ 45 sec
    Curing time in mouth (flexible phase)≥ 2 - 3 min≥ 2 - 3 min
    Setting time≥ 6 min≥ 6 min
    Hardness (1h)≥ 75 Shore D≥ 75 Shore D
    Hardness (24h)≥ 80 Shore D≥ 80 Shore D
    Compressive strength≥ 200 MPa (according to DIN EN ISO 4049:2010)≥ 200 MPa (according to DIN EN ISO 4049:2010)
    Flexural strength≥ 60 MPa (according to DIN EN ISO 4049:2010)≥ 60 MPa (according to DIN EN ISO 4049:2010)
    E-modulus> 1500 (according to DIN EN ISO 4049:2010)> 1500 (according to DIN EN ISO 4049:2010)
    Water absorptionPass (according to DIN EN ISO 4049:2010)Pass (according to DIN EN ISO 4049:2010)
    Radiopacity≥ 1.00 mm aluminium/mm material (according to DIN EN ISO 4049:2010)≥ 1.00 mm aluminium/mm material (according to DIN EN ISO 4049:2010)
    CytotoxicityNo biological reactivity (Grade 0) based on ISO 10993-5No biological reactivity (grade 0) of the cells exposed to the test article. (Test report: "Toxikon Europe Final GLP Report: 17-01389-G1")
    Carcinogenic potentialNo evidence for causing cancer within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for causing cancer within the time period of use in the patients mouth for up to 29 days.
    Mutagenic potentialNo evidence for mutagenic reactions within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for mutagenic reactions within the time period of use in the patients mouth for up to 29 days.
    Irritating potentialNo evidence for irritation within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for irritation within the time period of use in the patients mouth for up to 29 days.
    Sensitizing potentialNo evidence for sensitization within the time period of use (up to 29 days)Different evaluations and assessments have been made for cured Fixtemp® C&B and they have shown that there is no evidence for sensitization within the time period of use in the patients mouth for up to 29 days.
    Shelf life2 yearsThe results of stability tests at 23 °C/37 °C justify to assume a shelf life of 2 years for Fixtemp® C&B.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual test (e.g., number of specimens for compressive strength). It generally refers to "a series of in-house tests" and adherence to specific ISO standards (e.g., DIN EN ISO 4049:2010).

    For the cytotoxicity testing, the test was conducted by the independent laboratory Toxikon Europe. The provenance of the data is indicated as "Toxikon Europe Final GLP Report: 17-01389-G1". The country of origin for the manufacturing company (Dreve Dentamid GmbH) is Germany. It is implied that these are prospective tests performed on the Fixtemp® C&B material.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The performance evaluation relies on standardized tests (e.g., ISO standards) and laboratory findings rather than expert consensus on a test set in the typical sense of diagnostic imaging or clinical studies. The cytotoxicity test was conducted by an "independent laboratory," but the qualifications of the individuals establishing the ground truth (i.e., interpreting the cytotoxicity results) are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable. The reported tests are objective measurements based on standardized procedures and laboratory analysis, not subjective interpretations requiring adjudication by multiple experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data. Fixtemp® C&B is a material for fabricating dental prosthetics, and its performance is evaluated through material properties and biocompatibility, not reader interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. Fixtemp® C&B is a dental material, not an algorithm or software. Its performance is inherent to its physical and chemical properties.

    7. The Type of Ground Truth Used

    The ground truth for the performance tests consists of:

    • Standardized measurement directly from testing: This applies to properties like working time, curing time, setting time, hardness, compressive strength, flexural strength, E-modulus, water absorption, and radiopacity, where the "ground truth" is a measured physical or chemical property that meets or exceeds a predefined numerical threshold from an established standard (e.g., DIN EN ISO 4049:2010).
    • Laboratory findings: For cytotoxicity, the ground truth is established by the results of the ISO 10993-5 test conducted by an independent laboratory, indicating the absence of biological reactivity.
    • Evaluations and assessments: For carcinogenic, mutagenic, and irritating/sensitizing potentials, the ground truth is based on "different evaluations and assessments" indicating no evidence of these effects within the specified time frame. While the specifics of these evaluations are not detailed, they likely rely on established toxicological principles and potentially other in-vitro or in-vivo studies not explicitly listed.

    8. The Sample Size for the Training Set

    Not applicable. Fixtemp® C&B is a physical material, not an AI model or software that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.