510(k) Data Aggregation

    K Number
    K243841

    Validate with FDA (Live)

    Date Cleared
    2025-08-01

    (231 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sparta Infusion Set for Insulin is indicated for the subcutaneous infusion of insulin, administered by an external pump. The infusion set is indicated for use with adult and pediatric users weighing greater than 10 kg. The infusion set is indicated for single-use.

    Device Description

    The Sparta Infusion Set is a sterile, single-use, subcutaneous infusion set that establishes a sealed fluid path from an external portable infusion pump to the patient. Fluid is delivered to the patient through a 6 mm, 90°, soft cannula inserted in the subcutaneous tissue. The cannula is inserted using the preloaded, single-use mechanical Inserter. The infusion set has a 23-inch tubing length that terminates in a Luer connecter.

    AI/ML Overview

    The provided FDA 510(k) summary for the Sparta Infusion Set for Insulin (K243841) does not contain information about acceptance criteria and a study proving a software-driven device meets those criteria. The device described is an infusion set, which is a physical medical device, not an AI/software device. The document explicitly states:

    "No clinical testing is required to support the Subject Device's indications for use or a substantial equivalence determination."

    And regarding human factors, it says:

    "Human Factors validation testing was performed... Testing demonstrated that use of the Sparta Infusion Set is safe and effective for its intended users, uses, and use environments." This is typical for physical devices, not an AI output study.

    Therefore, I cannot provide the specific details about acceptance criteria, study design, sample sizes, expert involvement, or MRMC studies that you requested, as these are typically applicable to AI/software-driven diagnostic or imaging devices.

    The document primarily focuses on:

    • Biocompatibility Testing: To ensure the materials used are safe for contact with the body.
    • Bench Performance Testing: To verify physical performance characteristics like leak resistance, insertion performance, needle safety, occlusion performance, etc.
    • Sterilization and Shelf Life: To ensure the device remains sterile and functional over time.
    • Human Factors: To ensure the device can be used safely and effectively by its intended users.

    These are standard types of studies for a physical medical device like an infusion set, not for a software or AI product.

    If you have a document describing an AI/software medical device, I would be happy to analyze it against your criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250930

    Validate with FDA (Live)

    Device Name
    twiist system
    Date Cleared
    2025-04-02

    (5 days)

    Product Code
    Regulation Number
    880.5730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The twiist system is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 6 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The twiist system is intended for single patient, home use and requires a prescription.

    The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

    Device Description

    The twiist system described herein contains a modification of the labeling of the previously cleared DEKA ACE Pump System (K233952) to add the ConvaTec contact™ detach 23" subcutaneous infusion set and the ConvaTec inset™ 23" and 32" subcutaneous infusions sets as compatible for use with the twiist system.

    The twiist system is a wearable alternate controller enabled (ACE) insulin infusion pump intended to subcutaneously deliver insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The pump is able to reliably and securely communicate with compatible, digitally connected devices to receive, confirm, and execute commands. The pump is intended for single patient, ambulatory use and requires a prescription. The twiist system is intended for the management of diabetes mellitus in persons six years of age and greater.

    The twiist system, consistent with the predicate K233952, consists of the following durable and disposable components:

    1. Pump: A durable pump that incorporates fluid delivery algorithms and interfaces to a cassette, external wireless user interface, and iCGM. The pump is powered by a rechargeable lithium ion battery.
    2. Cassette: A single-use pumping cassette that combines microfluidic valves, a pump chamber, insulin reservoir, and Acoustic Volume Sensing (AVS) measurement chamber. The cassette interfaces to a pump and off-the-shelf infusion sets.
    3. Sequel twiist App: An iOS mobile application that serves as the primary user interface for the system. The twiist app can be downloaded onto the user's personal iPhone.
    AI/ML Overview

    The provided FDA Clearance Letter, specifically the 510(k) Summary, details the technical characteristics and testing for the "twiist system" and its substantial equivalence to the "DEKA ACE Pump System" (K233952).

    This submission (K250930) is a Special 510(k), meaning it focuses on a modification rather than a completely new device. The primary modification here is the addition of compatibility with new infusion sets (ConvaTec contact™ detach 23" and ConvaTec inset™ 23" and 32") to an already cleared device. This type of submission relies heavily on demonstrating that the performance with the new components is equivalent to the established performance of the predicate device under the original clearance.

    Therefore, the study described here is primarily focused on bench testing to demonstrate compatibility and similar performance with the new infusion sets, using established methodologies and acceptance criteria from the previous clearance (K233952).

    Here's the breakdown of the acceptance criteria and the study as described in the provided document:


    Acceptance Criteria and Device Performance for the twiist system (K250930)

    This 510(k) submission for the twiist system (K250930) is a Special 510(k), specifically addressing the addition of new compatible infusion sets. As such, the acceptance criteria and reported performance for key functional aspects of the pump itself (e.g., basal and bolus accuracy, occlusion detection) are identical to those established and validated for the predicate device, DEKA ACE Pump System (K233952). The study conducted for K250930 primarily focused on demonstrating that these established performance characteristics are maintained when using the newly added infusion sets.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since the core pump performance characteristics are "Unchanged from K233952" and "Same" as the predicate device, the acceptance criteria and reported performance values listed in the "Technological Characteristic Comparison" table reflect the established performance of the twiist system (and its predicate) for various parameters. The specific study for K250930 demonstrated that the device continues to meet these previously set criteria with the new infusion sets.

    CharacteristicAcceptance Criteria (from K233952)Reported Device Performance (with new infusion sets, as per K250930)
    Insulin Basal Rate Delivery Range0 units/hour - 30 units/hour0 units/hour - 30 units/hour (Same as predicate)
    Insulin Bolus Delivery RangeProgrammable from 0.05 - 25.00 Units in 0.01 Unit increments.Programmable from 0.05 - 25.00 Units in 0.01 Unit increments (Same as predicate)
    Basal AccuracyUnchanged from K233952 (Specific criteria not detailed in this summary, but implied met)Unchanged from K233952 (Specific performance not detailed, but implied met through testing with new sets)
    Bolus AccuracyUnchanged from K233952 (Specific criteria not detailed in this summary, but implied met)Unchanged from K233952 (Specific performance not detailed, but implied met through testing with new sets)
    Bolus Volume after Occlusion ReleaseNo more than 0.74 units.No more than 0.74 units. (Same as predicate)
    Time to occlusion alarm10 min (Bolus); 3 hours (Basal, 1 U/h); 6 hours (Basal, 0.1 U/hr)10 min (Bolus); 3 hours (Basal, 1 U/h); 6 hours (Basal, 0.1 U/hr) (Same as predicate)

    The supporting language states: "Performance testing with the added ConvaTec infusion sets demonstrates that subject device performance is equivalent to the predicate device. The methods and acceptance criteria used in this testing are well established in the previous clearance (K233952)."


    Study Details for K250930

    2. Sample size used for the test set and the data provenance

    The document states: "Bench performance testing was conducted to qualify the additional infusion sets for use with the subject device, and to establish substantial equivalence to the predicate device in terms of safety and effectiveness."

    • Test Set Sample Size: The exact sample size (number of devices, number of tests run per device/infusion set combination) for the bench testing is not specified in this 510(k) summary. However, it's implied that sufficient testing was performed to "demonstrate" performance equivalence.
    • Data Provenance: The data is from bench testing, which typically occurs in a controlled laboratory environment. The country of origin of the data is not explicitly stated, but given DEKA Research & Development Corp. is based in Manchester, New Hampshire, USA, it's highly probable the testing was conducted in the USA. The data is prospective in the sense that it was generated specifically for this 510(k) submission to qualify the new infusion sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this type of device and study. The ground truth for infusion pumps involves precise physical measurements of fluid delivery, rather than subjective expert interpretations (like in radiology). The "ground truth" for the test set is established by calibrated laboratory equipment and measurement techniques compliant with relevant standards (e.g., ISO 13485:2016).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies where subjective human interpretation (e.g., image reading) requires consensus building. For bench testing of an infusion pump, the measurements are objective.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. MRMC studies are specific to diagnostic devices (often AI-powered) that assist human interpretation of medical images or data. The twiist system is an insulin infusion pump, a therapeutic device, not a diagnostic one.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in a sense. The "pump" itself (the microprocessor-controlled mechanism with AVS feedback) delivers insulin "standalone" based on its algorithms. The bench testing performed to verify its accuracy and safety with the new infusion sets is a form of standalone performance evaluation in a controlled environment, demonstrating the device's inherent functional capabilities without human intervention during the actual delivery process.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for this device's performance testing (and for the predicate) is objective physical measurement based on highly accurate and calibrated laboratory equipment designed for fluid delivery assessment. This would involve measuring the actual volume of insulin delivered, flow rates, pressure changes, and time to alarm, compared to the programmed parameters.

    8. The sample size for the training set

    This question is not applicable. Infusion pumps like the twiist system are engineered medical devices that operate based on deterministic algorithms and physical principles, not machine learning or AI models that require "training sets" in the typical sense.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as #8. The device's "behavior" is designed and verified rather than "learned" from a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250357

    Validate with FDA (Live)

    Date Cleared
    2025-03-11

    (32 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RemunityPRO™ Pump for Remodulin® (treprostinil) Injection (the RemunityPRO System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in patients, ages 17 and older.

    Device Description

    The RemunityPRO™ Pump for Remodulin® (treprostinil) Injection is a wearable infusion pump designed to deliver Remodulin® (treprostinil) subcutaneously for the treatment of pulmonary arterial hypertension (PAH). Remodulin® was FDA-approved under NDA 021272. The system consists of several components: a wearable pump assembly, a remote interface, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the system to the patient. The RemunityPRO™ Pump for Remodulin® (treprostinil) Injection is prescription use only. The pump assembly is composed of a durable pump and a disposable, single-use cassette with a userfilled drug reservoir. The pump infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The subject system utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy. The subject device has updated branding and also modifies the predicate device to be used with an additional infusion set.

    AI/ML Overview

    The provided text describes a 510(k) summary for the RemunityPRO™ Pump for Remodulin® (treprostinil) Injection. This is a submission to demonstrate substantial equivalence to a legally marketed predicate device, rather than a study designed to establish acceptance criteria for an entirely new device or an AI algorithm. Therefore, many of the requested categories related to medical AI studies (like expert involvement, MRMC studies, or training sets) are not applicable.

    The core of this submission is to show that the minor changes to an existing device (the RemunityPRO™ Pump compared to the Remunity® 2.0 Pump) do not raise new questions of safety or effectiveness. The primary change highlighted is the addition of compatibility with an additional infusion set.

    Here's an attempt to fill in the requested information based on the provided text, noting where information is not applicable to this type of regulatory submission:

    Acceptance Criteria and Device Performance

    The provided document details a comparison between the subject device (RemunityPRO™ Pump) and its predicate (Remunity® 2.0 Pump). The "acceptance criteria" can be inferred from the performance characteristics of the predicate device, which the subject device is shown to match or be equivalent to.

    Acceptance Criteria (Based on Predicate Performance)Reported Device Performance (Subject Device)
    Device Classification: Class II Infusion Pump, 21 CFR 880.5725, Product Code: FRNNo change (Equivalent)
    Indications for Use: Continuous subcutaneous delivery of Remodulin (treprostinil) Injection for patients, ages 17 and older.No change (Equivalent)
    Prescription Use or OTC: PrescriptionNo change (Equivalent)
    Intended Patient Population: Age 17 years and olderNo change (Equivalent)
    Patient Environment: On-body wearable ambulatory pumpNo change (Equivalent)
    Delivery Method: Microprocessor controlled micro-dosing pump mechanism supplemented with acoustic volume sensor (AVS) feedback for monitoring delivery accuracyNo change (Equivalent)
    Delivery type: Subcutaneous infusionNo change (Equivalent)
    Dimensions: 6 cm x 6 cm x 2 cm (2.4 in x 2.4 in x 0.4 in)No change (Equivalent)
    Weight: 50 g (1.76 oz.)No change (Equivalent)
    Basal Delivery Rate Range: 8 µL/hr – 225 µL/hr, with increments of 1 µL/hrNo change (Equivalent)
    Basal Accuracy: ±6%±6% (No change/Equivalent)
    Bolus Volume after Occlusion Release: < 8 µL at all rates< 8 µL at all rates (No change/Equivalent)
    Time to Occlusion Alarm: Maximum time to occlusion alarm: • ≥ 100 µL/hr: 12 minutes • < 100 µL/hr: 8 hoursMaximum time to occlusion alarm: • ≥ 100 µL/hr: 12 minutes • < 100 µL/hr: 8 hours (No change/Equivalent)
    Material Biocompatibility: Fluid contacting: Polycarbonate (PC), COP, Bromobutyl, SEBS, Polyurethane. Patient contacting: PC, Acrylic, Polyurethane, ABS, Aluminum.No change (Equivalent)
    Cassette Shelf Life: 2 years2 years (No change/Equivalent)
    Service Life: 3 years3 years (No change/Equivalent)
    Applicable Safety Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-24, ISO 11137-1, ISO 10993-1, ISO 14971, IEC 60601-4-2, ANSI C63.27, AIM 7351731No change (Equivalent)
    Power Source: Rechargeable Lithium Ion BatteryNo change (Equivalent)
    Pump Storage Conditions: Temperature: -25 °C to 70 °C (-13 °F to 158 °F), Non-condensing humidity: up to 90%, Pressure: 50 kPa to 106 kPaNo change (Equivalent)
    Operating Conditions: Temperature: 5 °C to 40 °C (41 °F to 104 °F), Non-condensing humidity: up to 90%, Pressure: 70 kPa to 106 kPaNo change (Equivalent)
    Remote Interface System User Feedback: Touchscreen Device, Visual, audible, vibratoryNo change (Equivalent)
    Battery Operating Time: 72 hours72 hours (No change/Equivalent)
    Infusion Set Compatibility: Medtronic Quick-set, Medtronic Silhouette, Smiths Medical Cleo 90 Infusion SetsEquivalent - The subject device adds compatibility with the Neria Guard Infusion Set. Performance testing with Neria Guard demonstrated equivalent performance to the predicate device with established methods and acceptance criteria.
    Cassette: Unity 2.0 Cassettes, 3 mL, user-filledNo change (Equivalent)
    Priming Method: Primed by Pump after Cassette attachment to the PumpNo change (Equivalent)

    Study Details

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • The document states: "Bench performance testing was conducted in order to qualify the additional infusion set for use with the subject device, and to establish substantial equivalence to the predicate device in terms of safety and effectiveness."
      • Specific sample sizes for these bench tests are not provided in the summary.
      • Data provenance is not specified, but it would typically be internal laboratory testing by DEKA Research & Development Corp.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not Applicable. This is a bench performance study for a physical medical device (infusion pump), not a study involving interpretation of medical data by experts to establish ground truth for an AI algorithm. The performance metrics are objective and measurable (e.g., flow rate accuracy, occlusion alarm time).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. As this is a bench performance study of a physical device, there is no "adjudication" in the sense of resolving discrepancies between human readers or AI outputs. The tests involve direct measurement against established physical standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI-assisted diagnostic or interpretative device; it's an infusion pump. Therefore, an MRMC study is not relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. While the pump contains a microprocessor and control algorithms, the "standalone" context typically refers to the performance of an AI algorithm in isolation for diagnostic tasks. This device is a complete system with mechanical, electronic, and software components, and its performance is evaluated as a whole in bench testing.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" for this type of device performance would be established physical and electrical measurement standards, regulatory standards (e.g., IEC 60601-2-24 for infusion pumps), and the known performance characteristics of the predicate device. These are objective engineering and regulatory requirements, not medical "ground truth" derived from patient data or expert consensus.
    7. The sample size for the training set:

      • Not Applicable. This device, while containing software, is not an AI/ML-driven device that undergoes a "training" phase in the context of machine learning. Its operation is based on programmed control logic and physical components.
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no AI/ML training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241736

    Validate with FDA (Live)

    Date Cleared
    2025-01-16

    (213 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Remunity® 2.0 Pump for Remodulin® (treprostinil) Injection (the Remunity 2.0 System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in patients, ages 17 and older.

    Device Description

    The Remunity 2.0 System is a wearable infusion pump designed to deliver Remodulin® (treprostinil) subcutaneously for the treatment of pulmonary arterial hypertension (PAH). Remodulin® was FDA-approved under NDA 021272. The system consists of several components: a wearable pump assembly, a remote interface, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the system to the patient. The Remunity 2.0 System is prescription use only.

    The pump assembly is composed of a durable pump and a disposable, single-use cassette with a user-filled drug reservoir. The pump infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The subject system utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.

    AI/ML Overview

    The provided text describes the Remunity 2.0 Pump for Remodulin® (treprostinil) Injection, an infusion pump, and its substantial equivalence to a predicate device. However, it does not contain information about the acceptance criteria and study that proves a device meets those criteria in the context of an AI/ML medical device.

    The document is a 510(k) summary for an infusion pump, which is a hardware device for drug delivery. The "Performance Data" section lists various functional areas evaluated for device performance (e.g., delivery accuracy, occlusion performance, durability) but does not provide specific acceptance criteria values or detailed results for these evaluations. It also mentions software, cybersecurity, human factors, electrical safety, EMC, and alarms as evaluated areas, but again, without specific acceptance criteria or study outcomes.

    Therefore, I cannot populate the requested table and answer questions 2 through 9 based on the provided text. The information requested (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, type of ground truth, training set size and ground truth establishment) is typically associated with studies for AI/ML-driven diagnostic or prognostic devices, which is not the type of device described here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243354

    Validate with FDA (Live)

    Device Name
    Remunity System
    Date Cleared
    2024-11-27

    (29 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Remunity® Pump for Remodulin® (treprosimil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old).

    Device Description

    The Remunity Subcutaneous Delivery System for Remodulin® (Remunity System) is a wearable infusion pump designed to deliver Remodulin® for the treatment of pulmonary arterial hypertension (PAH). It is intended for continuous subcutaneous delivery of FDA-approved Remodulin® (treprostinil) (hereinafter referred to as 'Remodulin®' or 'Remodulin® (treprostinil)'), NDA 021272. The Remunity System consists of several components: a wearable pump assembly, a remote interface, a filling and priming aid, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the Remunity System to the patient.

    The pump assembly is composed of a reusable pump and a disposable single-use cassette with a pharmacy-filled or user-filled drug reservoir, which infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The Remunity System utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.

    The device is prescription use only.

    The modified device can be used with an additional infusion set.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Remunity System," which is an infusion pump. The submission is for a modified version of an already cleared device, and the primary modification discussed is the addition of a compatible infusion set.

    Here's an analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative manner for specific performance characteristics of the new infusion set. Instead, it states that "Performance testing with the Neria Guard demonstrates that subject device performance is equivalent to the predicate device. The methods and acceptance criteria used in this testing are well established in previous clearances."

    However, the table comparing the predicate device and the subject device does list several performance characteristics and their values, which implicitly serve as the performance requirements that the modified device must meet to be considered equivalent. Since the subject device is stated to have "No change" in all these characteristics except the infusion set, it implies that the modified device performs identically to the predicate device for these parameters.

    Acceptance Criteria (Implicit from Predicate Device Performance)

    CharacteristicAcceptance Criteria (Predicate Device Performance)Reported Device Performance (Subject Device)
    Basal Delivery Rate Range16 µL/hr – 225 µL/hr, increments of 1 µL/hrNo change (implies 16 µL/hr – 225 µL/hr, increments of 1 µL/hr)
    Basal Accuracy±6%No change (implies ±6%)
    Bolus Volume after Occlusion Release< 40 µL at all ratesNo change (implies < 40 µL at all rates)
    Time to Occlusion Alarm (≥ 100 µL/hr)Maximum 12 minutesNo change (implies Maximum 12 minutes)
    Time to Occlusion Alarm (< 100 µL/hr)Maximum 8 hoursNo change (implies Maximum 8 hours)
    Pump Service Life3 yearsNo change (implies 3 years)
    Ingress ProtectionIP58 when connected to the CassetteNo change (implies IP58)
    Battery Operating Time72 hoursNo change (implies 72 hours)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses "bench performance testing." It does not specify the sample size for this testing. It also does not mention the country of origin of the data or whether the data was retrospective or prospective. Given that it is bench testing, it would inherently be prospective data collected in a laboratory setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The testing described is "bench performance testing" which typically does not involve human experts establishing "ground truth" in the same way clinical or diagnostic studies do. The "ground truth" in this context would be physical measurements of pump performance against engineering specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided and is not applicable for bench performance testing of an infusion pump. Adjudication methods are typically used in clinical studies, especially those involving human interpretation of medical images or outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. The device in question is an infusion pump, which is a hardware device for delivering medication, not an AI-powered diagnostic tool requiring human reader interpretation. Therefore, questions about human readers and AI assistance are not relevant to this device and its clearance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device is an infusion pump; therefore, the concept of "standalone algorithm performance" as typically applied to AI/ML devices is not directly applicable. The device's performance, as measured in bench testing, is inherently its "standalone" performance without "human-in-the-loop" assistance in terms of its core function (drug delivery accuracy, occlusion detection, etc.).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench performance testing of the infusion pump, the "ground truth" would be engineering specifications and metrology measurements. This isn't based on expert consensus, pathology, or outcomes data. It's about measuring physical phenomena (e.g., fluid volume delivered, time to alarm) against predefined design requirements.

    8. The sample size for the training set

    This information is not provided and is not applicable. This device is an infusion pump, which is a physical medical device, not a machine learning or AI algorithm that requires a "training set" of data in the conventional sense. Its function is based on mechanical and electronic engineering principles.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable as explained in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242693

    Validate with FDA (Live)

    Date Cleared
    2024-10-07

    (28 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DEKA Infusion System is indicated for continuous delivery of general maintenance fluids and supported antibiotics to adult patients through a peripheral intravenous (IV) site under the supervision of a qualified clinician.

    The DEKA Administration Set is indicated for gravity administration of general maintenance fluids and supported antibiotics approved for peripheral intravenous (IV) infusion to adult patients under the supervision of a qualified clinician. The DEKA Administration Set is indicated for single patient use only. The DEKA Administration Set is not indicated for administration of high-alert medications, blood, and blood products.

    The DEKA Administration Set is compatible with the DEKA Infusion Controller or can be used as a stand-alone administration set for performing manual gravity infusions.

    Device Description

    The subject device is a modified version of the predicate DEKA Infusion System. Updates have been made to the predicate device software to increase the maximum programmable flow rate (from 300 to 600 mL/h), update the labeling to reflect the increased maximum programmable flow rate and update the performance of tilt angle detection. The administration set is unchanged.

    The DEKA Infusion System, comprised of the DEKA Infusion Controller and DEKA Administration Set, is a gravity-based infusion controller. The Infusion Controller uses a camera to view drop formation in an Administration Set Drip Chamber in order to measure flow rate. During an infusion, the Infusion Controller can adjust the Flow Control Valve (FCV) as necessary against the DEKA Administration Set to maintain the desired flow rate without user intervention.

    The DEKA Administration Set is a single-use, non-DEHP intravascular administration set provided sterile. It is compatible with the DEKA Infusion Controller or can be used as a standalone administration set for performing manual gravity infusions.

    AI/ML Overview

    The provided text is a 510(k) summary for the DEKA Infusion System and DEKA Administration Set. It primarily focuses on demonstrating substantial equivalence to a predicate device after a modification, specifically an increased maximum programmable flow rate and an update to tilt angle detection.

    However, the document does not contain the detailed information required to answer your specific questions about acceptance criteria, the study proving the device meets the criteria, sample sizes, expert involvement, ground truth, or MRMC studies for an AI/ML-based device.

    This document is for a mechanical/electrical infusion system that uses a camera for drop formation detection and flow rate measurement, not an AI/ML diagnostic or prognostic device that would involve a training set, test set, expert adjudication, or MRMC studies in the way your questions imply. The "Vision System Performance" mentioned is likely related to the camera's ability to accurately detect drops, not a complex AI model requiring extensive human-in-the-loop validation or ground truth established by multiple medical experts interpreting images.

    Therefore, I cannot extract the requested information from this document. The questions you've asked are typical for the validation of Artificial Intelligence/Machine Learning (AI/ML) medical devices, particularly those involving image interpretation or complex diagnostic algorithms. This 510(k) summary does not describe such a device or its corresponding validation studies.

    The closest aspects discussed are:

    • Acceptance Criteria for Flow Rate Accuracy: Mentioned as "± 10% nominally" and "± 25% under environmental limits of normal use." The document states "All tests met acceptance criteria."
    • Study Proving Device Meets Criteria: "Formal performance testing was conducted on all areas of the subject device that are impacted by its changes with respect to the predicate device."
    • Performance Data: "Flow Rate Accuracy," "Vision System Performance," "Stopping Flow," and "Device Handling Alarms" were tested.

    But none of these provide the level of detail (sample size, expert ground truth, adjudication, MRMC) specific to AI/ML device validation that your questions are designed to elicit.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240920

    Validate with FDA (Live)

    Date Cleared
    2024-07-12

    (99 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoCare Bicarbonate Concentrate Set (BCS) is indicated for use in the preparation of dialysate in a 3-stream proportioning hemodialysis machine with a compatible mechanical interface, such as the HemoCare Hemodialysis System. The BCS is intended for extracorporeal bicarbonate hemodialysis for chronic renal failure, according to a physician's prescription.

    Device Description

    The HemoCare Bicarbonate Concentrate Set (BCS) is the Subject Device of this submission. It consists of powdered sodium bicarbonate in a single use cartridge with an integrated tubing assembly for interfacing with hemodialysis systems with compatible three-stream proportioning connector ports such as the HemoCare Hemodialysis System. The HemoCare System uses the bicarbonate concentrate in 3-stream, 45x proportioning to generate dialysate for bicarbonate hemodialysis therapy.

    AI/ML Overview

    This document is a 510(k) summary for the HemoCare Bicarbonate Concentrate Set (BCS). It outlines the device, its intended use, comparison to a predicate device, and performance data. However, it does not describe a study involving an AI/Machine Learning (ML) device.

    Therefore, I cannot provide information on acceptance criteria, reported device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details as they pertain to an AI/ML device.

    The document focuses on the HemoCare Bicarbonate Concentrate Set, which is a medical accessory for hemodialysis, and its performance data relates to biocompatibility and bench testing to verify specified performance, not the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233557

    Validate with FDA (Live)

    Date Cleared
    2024-07-12

    (249 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HemoCare® Hemodialysis System is indicated for hemodialysis treatment, including short daily and nocturnal hemodialysis, of renal failure patients. The HemoCare Hemodialysis System is indicated for use in chronic dialysis facilities, self-care dialysis facilities, and the home setting. All treatments must be prescribed by a physician and administered by a trained operator. Treatments must be performed with the assistance or observation of an individual who has been trained and is considered competent in the use of the device. The HemoCare Hemodialysis System is not indicated for solo hemodialysis in the home setting.

    Device Description

    The HemoCare® Hemodialysis System is a self-contained, software-controlled device that provides hemodialysis treatments, including short daily and nocturnal hemodialysis, for patients with renal failure. The device is intended for use by a single patient in chronic dialysis facilities, self-care dialysis facilities, and the home setting. The HemoCare® Hemodialysis System is composed of the following components:

    • The HemoCare® Treatment Device is a hemodialysis delivery system. When provided a source of water for dialysis, it produces infusion grade dialysate using liquid acid and powder bicarbonate concentrates. The Treatment Device provides secure, 2-way communication with the cloud for transmitting patient registration, physician-prescribed dialysis treatments, patient monitoring, and device status information.
    • The HemoCare® Water Device is a water purification system that produces water for dialysis through distillation of EPA drinking water. The Water Device interfaces with the Treatment Device.
    • The Blood Treatment Set connects to the patient access site and interfaces with the Treatment Device to provide extracorporeal blood flow through a dialyzer for dialysis treatment. The Blood Treatment Set is terminally sterilized and includes displacement blood pumps, a heparin delivery system, and an access disconnect sensor. The Blood Treatment set can be used for up to 16 treatments.
    AI/ML Overview

    The provided text describes the HemoCare® Hemodialysis System, a medical device. However, it does not contain the specific information required to complete the detailed table about acceptance criteria and device performance results as requested, nor does it detail a study proving the device meets acceptance criteria in the context of an AI/ML algorithm.

    The document is a 510(k) summary for a hemodialysis system, focusing on its substantial equivalence to predicate devices. It outlines performance data categories like Biocompatibility, Electrical Safety, Software Verification and Validation, Bench Performance Testing, Sterilization and Shelf Life, Human Factors, and Clinical Testing.

    Importantly, the document does not describe an AI/ML algorithm that would have "acceptance criteria," "test sets," "ground truth," "expert readers," or "MRMC studies" in the context typically associated with an AI/ML medical device submission. The "Software Verification and Validation Testing" section states that "Testing results demonstrate that the device meets all software performance requirements," but it does not specify what those requirements are, nor does it indicate these requirements relate to an AI/ML diagnostic or predictive function.

    Therefore, I cannot fulfill the request as the type of information sought (related to AI/ML algorithm performance) is not present in the provided text. The document is about a physical medical device (hemodialysis system) and its regulatory clearance as substantially equivalent to existing devices, not about a software-as-medical-device (SaMD) or AI-enabled medical device performance study with the detailed metrics requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240256

    Validate with FDA (Live)

    Device Name
    Remunity System
    Date Cleared
    2024-06-12

    (133 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Remunity® Pump for Remodulin® (treprosimil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old).

    Device Description

    The Remunity Subcutaneous Delivery System for Remodulin® (Remunity System) is a wearable infusion pump designed to deliver Remodulin® for the treatment of pulmonary arterial hypertension (PAH). It is intended for continuous subcutaneous delivery of FDA-approved Remodulin® (treprostinil) (hereinafter referred to as 'Remodulin®' or 'Remodulin® (treprostinil)'), NDA 021272. The Remunity System consists of several components: a wearable pump assembly, a remote interface, a filling and priming aid, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the Remunity System to the patient.

    The pump assembly is composed of a reusable pump and a disposable single-use cassette with a pharmacy-filled or user-filled drug reservoir, which infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The Remunity System utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.

    The device is prescription use only.

    The modified device utilizes a touchscreen remote controller.

    AI/ML Overview

    The provided text describes the Remunity System, an infusion pump. However, the document is a 510(k) summary for a substantial equivalence determination, not a study report proving a device meets specific acceptance criteria.

    The 510(k) process primarily demonstrates that a new device is "substantially equivalent" to a legally marketed predicate device, meaning it has the same intended use and similar technological characteristics, or that any differences do not raise new questions of safety or effectiveness. It does not typically involve the presentation of detailed clinical study results or performance data in the format of acceptance criteria and proven performance as might be found in a clinical trial report or a comprehensive validation study.

    Specifically, the document states:

    • "No animal study or clinical trial data was obtained in support of this premarket submission."
    • "Performance testing was conducted in order to verify and validate the subject device remote interface for its intended use, and to establish substantial equivalence to the respective predicate device remote interface in terms of safety and effectiveness." This indicates that performance testing was done, but the details of acceptance criteria and reported numerical performance values are not provided within this summary.
    • The only "performance data" mentioned are the types of tests conducted (e.g., Touchscreen Remote Reliability, Functionality, Wireless Communication, Software and Cybersecurity, Human Factors, Electrical Safety, EMC, Alarms), not the quantitative results of those tests.

    Therefore, Based solely on the provided FDA 510(k) summary (K240256), I cannot extract the specific information requested regarding acceptance criteria and a study proving the device meets those criteria, as detailed as your prompt requires. The document attests that appropriate testing was performed and that the modified device (with a touchscreen remote controller) is substantially equivalent to its predicate, but it does not present the raw or summarized performance data against explicit acceptance criteria.

    The information you are asking for, such as sample size for test sets, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, standalone performance, and training set details, are typically found in more detailed study reports (e.g., clinical study reports, validation reports) that might be referenced in a 510(k) submission but are not fully elaborated in the summary itself.

    In summary, the provided text does not contain the detailed information required to fill out the table and answer the specific questions about acceptance criteria and performance study methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241178

    Validate with FDA (Live)

    Date Cleared
    2024-05-24

    (25 days)

    Product Code
    Regulation Number
    880.5730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages six and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.

    The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

    Device Description

    The DEKA ACE Pump System described herein contains a modification of the labeling of the previously cleared DEKA ACE Pump System (K233952) to add Novolog U-100 (insulin aspart) as a compatible insulin. The device is a wearable alternate controller enabled (ACE) insulin infusion pump intended to subcutaneously deliver insulin. at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The pump is able to reliably and securely communicate with compatible, digitally connected devices to receive, confirm, and execute commands. The pump is intended for single patient, ambulatory use and requires a prescription. The DEKA ACE Pump System is intended for the management of diabetes mellitus in persons six years of age and greater.

    The pump was previously cleared and is indicated for use with Humalog U-100 insulin. This Special 510(k) utilizes the same methodology and acceptance criteria used to obtain the clearance and indication for Humalog U-100 to add the indication for Novolog U-100.

    The DEKA ACE Pump System, consistent with the predicate K233952, consists of the following durable and disposable components:

    1. Pump: A durable pump that incorporates fluid delivery algorithms and interfaces to a cassette, external wireless user interface, and iCGM. The pump is powered by a rechargeable lithium ion battery.
    2. Cassette: A single-use pumping cassette that combines microfluidic valves, a pump chamber, insulin reservoir, and Acoustic Volume Sensing (AVS) measurement chamber. The cassette interfaces to a pump and off-the-shelf infusion sets.
    3. DEKA Loop App: An iOS mobile application that serves as the primary user interface for the system. The DEKA Loop app can be downloaded onto the user's personal iPhone.

    Also consistent with the predicate, the DEKA ACE Pump system includes the following electronic interfaces:

    • Dexcom G6 iCGM: The DEKA ACE Pump System is compatible with Dexcom G6 iCGMs. Communication between the Dexcom G6 and DEKA ACE Pump is unchanged from the predicate device. The BLE central role on the ACE Pump radio processor connects directly to the Dexcom iCGM transmitter using the sensor's medical slot. The ACE Pump communicates with the iCGM transmitter at a regular interval to provide iCGM sensor data to the iAGC.
    • Halo Cloud: Halo Cloud is a digital platform that connects the DEKA ACE Pump System to a variety of cloud-related services. These services include:
    • Patient onboarding and device pairing via secure key transfer
    • Prescription setting downloads to the ACE Pump
    • Event log uploads from the ACE pump to the cloud
    • Remote (OTA) software updates

    No changes to the hardware or software of the system from that of the predicate are necessary to add the Novolog U-100 compatibility claim.

    AI/ML Overview

    The provided text describes a 510(k) submission for the DEKA ACE Pump System, specifically seeking to add Novolog U-100 (insulin aspart) as a compatible insulin. The core of this submission is to demonstrate substantial equivalence to a previously cleared version of the same pump (K233952), which was cleared for use with Humalog U-100 insulin.

    Based on the provided text, the device in question is an insulin pump, not an AI/Software as a Medical Device (SaMD) for image analysis or diagnosis. Therefore, many of the typical acceptance criteria and study aspects related to AI/SaMD (like MRMC studies, expert ground truth establishment for diagnostic images, sample sizes for training/test sets for AI models, etc.) are not applicable to this type of medical device submission.

    The "acceptance criteria" in this context refer to the demonstration that the modified device (with Novolog U-100 compatibility) is substantially equivalent to the predicate device (with Humalog U-100 compatibility) in terms of safety and effectiveness, and that it meets all applicable Special Controls for Alternate Controller Enabled Infusion Pumps (21 CFR 880.5730, Product Code QFG).

    Here's an attempt to answer the questions based on the provided document, highlighting what is and isn't applicable:

    1. A table of acceptance criteria and the reported device performance

      The document doesn't present a table of quantitative acceptance criteria and reported numerical performance for a new study, but rather asserts that the same methodology and acceptance criteria used for Humalog U-100 were applied to Novolog U-100, implying the new performance matches the previously accepted thresholds. The comparison focuses on demonstrating equivalence across various characteristics.

      Implicit Acceptance Criteria (based on equivalence to predicate and Special Controls):

      • Indications for Use: No change.
      • Prescription Use: Yes.
      • Intended Population: Persons with Diabetes Mellitus ages 6 and above.
      • Environment of Use: Professional healthcare facilities and home healthcare environments.
      • Patient Environment: On-body wearable ambulatory pump.
      • Delivery Method: Microprocessor controlled Micro-dosing pump mechanism supplemented with acoustic volume sensor (AVS) feedback for monitoring delivery accuracy.
      • Insulin Basal Rate Delivery Range: 0 units/hour - 30 units/hour.
      • Insulin Bolus Delivery Range: Programmable from 0.05 - 25.00 Units in 0.01 Unit increments.
      • Basal Accuracy: Unchanged from K213536 (predicate's reference).
      • Bolus Accuracy: Unchanged from K213536 (predicate's reference).
      • Bolus Volume after Occlusion Release: No more than 0.74 units.
      • Time to occlusion alarm: 10 min (Bolus); 3 hours (Basal, 1 U/h); 6 hours (Basal, 0.1 U/hr).
      • Material Biocompatibility: Compliant with ISO-10993.
      • Cartridge/Cassette Shelf Life: 1 year.
      • Ingress Protection: IP28.
      • Applicable Safety Standards: Compliance with listed IEC and ISO standards (e.g., IEC 60601-1, ISO 14971).
      • Power Source: Rechargeable Lithium Ion Battery.
      • Pump Storage Conditions: Temperatures of -25 °C to 70 °C, Non-condensing humidity 15% to 90%.
      • Operating Conditions: Temperatures of 5 °C to 40 °C, Non-condensing humidity of 15% to 90%.
      • System User Feedback: Visual, audible, and vibratory.
      • Battery Operating Time: 72 hours.
      • New "Acceptance": The addition of Novolog U-100 compatibility does not impact the safety and effectiveness, and testing with Novolog U-100 is equivalent to that performed with Humalog U-100.

      Reported Device Performance (based on assertion of equivalence):
      For all the criteria listed above, the "Subject Device" (DEKA ACE Pump with Novolog U-100 compatibility) is stated to have the exact same characteristics and performance as the "Predicate Device" (DEKA ACE Pump with Humalog U-100 compatibility), or that its performance is equivalent to the predicate's established performance for the relevant aspects (e.g., Novolog U-100 testing being equivalent to Humalog U-100 testing). The key is that "No new or modified risks" are introduced.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      The document states: "No clinical data was obtained in support of this premarket submission."
      It also states that "Performance testing was performed in order to establish substantial equivalence for the compatibility claim of Novolog U-100 in comparison to the previously cleared compatibility claim for Humalog U-100 in terms of both safety and effectiveness..."

      This heavily implies in vitro or bench testing rather than human subject testing. The "test set" in this context would likely refer to the number of tests/replicates performed with Novolog U-100 to demonstrate equivalent performance to Humalog U-100, not a dataset for an AI model. This information (specific sample sizes for bench testing, e.g., number of pump cycles, number of insulin lots tested) is not provided in this summary document. Data provenance for such non-clinical testing is typically internal lab data, not human patient data with country of origin.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      Not applicable. This is not an AI/SaMD for diagnostic image analysis. Ground truth would be established through engineering specifications, validated test methods, and measurement standards for pump performance, not expert consensus on medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      Not applicable. This is not an AI/SaMD for diagnostic image analysis.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      Not applicable. This is not an AI/SaMD for diagnostic image analysis. The device is an insulin pump.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      Not applicable. The device has algorithms (fluid delivery, bolus calculator) but it's a physical pump with user interaction, not a standalone diagnostic algorithm being evaluated. Its performance is inherent to the pump's mechanical and electronic design working with insulin.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      The ground truth for the performance of the insulin pump (e.g., basal/bolus accuracy, occlusion detection, battery life) would be established through engineering specifications, standardized test methods, and quantifiable measurements (e.g., precision volumetric measurements, pressure sensor readings, time-to-event logging). For this submission, the "ground truth" for Novolog U-100 compatibility is that its chemical and physical properties allow the pump to perform identically or equivalently to how it performs with Humalog U-100, which has already been established as safe and effective.

    8. The sample size for the training set

      Not applicable. This device does not use machine learning that requires a "training set" in the conventional AI sense for diagnostic tasks. The "training" of the pump would be its design, development, and validation testing based on engineering principles and regulatory standards.

    9. How the ground truth for the training set was established

      Not applicable. As above, there isn't a "training set" in the AI sense. The "ground truth" for the pump's design and function comes from established medical device standards, diabetes management requirements, and rigorous engineering testing.