510(k) Data Aggregation

    K Number
    K241887

    Validate with FDA (Live)

    Date Cleared
    2024-07-25

    (27 days)

    Product Code
    Regulation Number
    876.1520
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K233964

    Validate with FDA (Live)

    Date Cleared
    2024-01-12

    (28 days)

    Product Code
    Regulation Number
    876.1520
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GI Genius™ system is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius™ computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

    Device Description

    GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.

    GI Genius is composed of software (namely, ColonPRO™ 4.0) and hardware (namely, GI Genius™ Module 100 and 200).

    GI Genius™ Module 100 and 200 are compatible with Video Processors featuring SDI (SMPTE 259M) or HD-SDI (SMPTE 292M) output ports and endoscopic display monitors featuring SDI (SMPTE 259M) or HD-SDI (SMPTE 292M) input ports. GI Genius™ Module 200 is also compatible with Video Processors featuring the 4K UHD standard.

    The GI Genius system is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.

    During live video streaming of the endoscopic video image, GI Genius generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.

    The operating principle of the subject device is identical to that of the predicate device, this being a computerassisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems and the accessories given by the footswitch and the USB K-switch.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the GI Genius™ Module 100, GI Genius™ Module 200, and ColonPRO™ 4.0, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" for each performance metric, but it does present a comparison table that shows the performance of the Subject Device (ColonPRO™ 4.0) against its Predicate Device (GI Genius™ System 100 and 200). The implication is that the subject device's performance, being "improved" or "same" compared to the already cleared predicate, meets the necessary equivalence for clearance.

    CharacteristicAcceptance Criteria (Implied: at least as good as predicate)Reported Device Performance (Subject Device - ColonPRO™ 4.0)Comparison to Predicate (Performance of Predicate)
    Lesion-based sensitivity≥ 86.5%88.07 %Improved (86.5 %)
    Frame-level True Positive≥ 269,223277,738Improved (269,223)
    Frame-level True NegativeFor 150 videos/338 polyps: ≥ 5,239,1285,248,406Improved (5,239,128)
    Frame-level False PositiveFor 150 videos/338 polyps: ≤ 104,66995,391Improved (104,669)
    Frame-level False NegativeFor 150 videos/338 polyps: ≤ 192,567184,052Improved (192,567)
    True positive rate per frameMean: ≥ 58.30 %, % of polyps: 100 %Mean: 60.14 %, % of polyps: 100 %Improved (Mean: 58.30 %, % of polyps: 100 %)
    False positive rate per frameMean: ≤ 1.96 %Mean: 1.79 %Improved (Mean: 1.96 %)
    Frame-Based TPr/FPr ROC curve, AUC≥ 0.7960.826Improved (0.796)
    False positive clusters per patient< 500 ms: ≤ 106, > 500 ms: ≤ 11< 500 ms: 97, > 500 ms: 10Improved (< 500 ms: 106, > 500 ms: 11)
    Video delay, signal in to signal outAs per predicate (1.52 µs for Module 100, 0.74 µs for Module 200)1.52 µs (Module 100), 0.74 µs (Module 200)Same

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document explicitly states the frame-level performance was assessed using 150 videos / 338 polyps. It doesn't specify if this refers to the number of patients or individual lesions.
    • Data Provenance: The document does not provide information on the country of origin of the data or whether it was retrospective or prospective. It only mentions that "the baseline clinical validation for the subject device was conducted and reviewed in DEN200055 and is still applicable." Since this is a Special 510(k) for a software update (version 4.0.0 replacing 3.0.2), the primary performance data seems to derive from the re-training of the neural network rather than a new clinical study. The "Non-clinical testing" section mentions that "Tests according to the Standalone Performance Testing Protocol v2.0, submitted as part of the K231143 predicate device submission, have been repeated for the applicable parts of the subject device." This suggests the test set for this submission is the same as, or comparable to, that used for the predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts used or their qualifications for establishing the ground truth on the test set.

    4. Adjudication Method for the Test Set

    The document does not specify the adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention or present an MRMC comparative effectiveness study where human readers improve with AI vs. without AI assistance. The device is described as a "computer-assisted reading tool," suggesting it's intended to work alongside an endoscopist, but no study on human performance improvement with the device is provided in this submission or summary. It refers to the "baseline clinical validation" for the predicate device, but the details of that validation are not present here.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a standalone performance study was done. The performance metrics listed in the table (Lesion-based sensitivity, Frame-level True Positive/Negative/False Positive/Negative, True positive rate per frame, False positive rate per frame, Frame-Based TPr/FPr ROC curve, AUC, False positive clusters per patient) all refer to the algorithm's performance without direct human-in-the-loop interaction for the purpose of these specific measurements. The "Non-clinical testing" section explicitly states: "Tests according to the Standalone Performance Testing Protocol v2.0, submitted as part of the K231143 predicate device submission, have been repeated for the applicable parts of the subject device."

    7. Type of Ground Truth Used

    The document implies the ground truth for polyps and lesions was used to evaluate detection performance. However, it does not explicitly state the method for establishing this ground truth (e.g., expert consensus, pathology, outcome data). Likely, for lesion detection in endoscopic videos, ground truth would typically be established by expert endoscopist review, potentially confirmed by pathology for detected lesions.

    8. Sample Size for the Training Set

    The document does not specify the sample size for the training set. It only mentions "retraining of the neural network" as the source of improved detection performance for ColonPRO™ 4.0.

    9. How the Ground Truth for the Training Set Was Established

    The document does not specify how the ground truth for the training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231143

    Validate with FDA (Live)

    Date Cleared
    2023-05-19

    (28 days)

    Product Code
    Regulation Number
    876.1520
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

    Device Description

    GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.

    GI Genius is compatible with Video Processors featuring SDI (SMPTE 259M) or HD-SDI (SMPTE 292M) output ports and endoscopic display monitor featuring SDI (SMPTE 259M) or HD-SDI (SMPTE 292M) input ports.

    GI Genius is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.

    During live video streaming of the endoscopic video image, GI Genius generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.

    The operating principle of the subject device is identical to that of the predicate device, this being a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the GI Genius System (System 100 and System 200) meets these criteria, specifically in the context of its substantial equivalence to a previously cleared device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The relevant performance metrics are presented as a comparison between the subject device (GI Genius System 100/200) and its predicate device. Since the claim is substantial equivalence, the performance of the subject device is shown to be at least as good as the predicate.

    CharacteristicAcceptance Criteria (Predicate Performance)Reported Device Performance (Subject Device)Comparison
    Lesion-based sensitivity86.5 %86.5 %Same
    Frame Level Performance (150 videos / 338 polyps)
    True positive269,223269,223Same
    True negative5,239,1285,239,128Same
    False positive104,669104,669Same
    False negative192,567192,567Same
    True positive rate per frame (Mean)58.30 %58.30 %Same
    True positive rate per frame (% of polyps)100 %100 %Same
    False positive rate per frame (Mean)1.96 %1.96 %Same
    Frame-based TPr/FPr ROC curve, AOC0.7960.796Same
    False positive clusters per patient (< 500 ms)40 less than baseline40 less than baselineSame
    False positive clusters per patient (> 500 ms)1 more than baseline1 more than baselineSame
    Video delay, signal in to signal out (GI Genius System 200)1.52 µs (Predicate)0.74 µsImproved
    Video delay, signal in to signal out (GI Genius System 100)1.52 µs (Predicate)1.52 µsSame

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The performance data is based on 150 videos containing 338 polyps.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, it references the "baseline clinical validation for the subject device was conducted and reviewed in DEN200055," suggesting the source of this test data comes from a prior, perhaps larger, study. The testing for the current submission involved repeatable performance tests and non-inferiority testing on "42 pre-recorded procedures."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    The document does not provide details on the number or qualifications of experts used to establish the ground truth for the test set. It mentions "per-frame assessment," which implies a detailed, possibly manual, review process.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (e.g., 2+1, 3+1).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study to assess how human readers improve with AI vs. without AI assistance. The study focuses on the device's standalone performance and its non-inferiority (or equivalence) to a predicate device, rather than human-AI collaboration.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The document states:

    • "Tests according to the Standalone Performance Testing Protocol v2.0, submitted as part of the K211951 predicate device submission, have been repeated for the applicable parts of the subject device."
    • The performance metrics provided in the table (True Positive, True Negative, False Positive, False Negative, ROC curve AOC, per-frame rates, and false positive clusters) are indicative of standalone algorithm performance.
    • "Non-inferiority of performance of GI Genius with the Olympus CV-1500 EVIS X1 UHD video processor has been established by means of a per-frame assessment on 42 pre-recorded procedures." This further supports standalone assessment.

    7. The Type of Ground Truth Used

    The ground truth appears to be based on expert consensus or a similar form of expert review, as evidenced by the "per-frame assessment" and lesion-level evaluation related to polyps and adenomas, which are typically confirmed by medical professionals. Pathology or outcomes data are not explicitly mentioned as the primary ground truth.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size for the training set. It mentions that "GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images." The performance evaluation focuses on the test set, not the training set details.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide details on how the ground truth for the training set was established. It only states that the device "has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps."

    Ask a Question

    Ask a specific question about this device

    K Number
    K211951

    Validate with FDA (Live)

    Device Name
    GI Genius
    Date Cleared
    2021-07-23

    (30 days)

    Product Code
    Regulation Number
    876.1520
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GI Genius System is a computer-assisted reading tool designed to aid endoscopists in detecting colonic mucosal lesions (such as polyps and adenomas) in real time during standard white-light endoscopy examinations of patients undergoing screening and surveillance endoscopic mucosal evaluations. The GI Genius computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. This device is not intended to replace clinical decision making.

    Device Description

    GI Genius is an artificial intelligence-based device that has been trained to process colonoscopy images containing regions consistent with colorectal lesions like polyps, including those with flat (non-polypoid) morphology.

    GI Genius is compatible with the following Video Processors: Fujifilm VP-7000, Olympus CV-180 EXERA II, Olympus CV-190 EXERA III, Fujifilm VP-4450HD, and Pentax EPK-i7000 Video Processor.

    GI Genius is connected between the video processor and the endoscopic display monitor. When first switched on, the endoscopic field of view is clearly identified by four corner markers, and a blinking green square indicator appears on the connected endoscopic display monitor to state that the system is ready to function.

    During live video streaming of the endoscopic video image, GI Genius generates a video output on the endoscopic display monitor that contains the original live video together with superimposed green square markers that will appear when a polyp or other lesion of interest is detected, accompanied by a short sound. These markers will not be visible when no lesion detection occurs.

    The operating principle of the subject device is identical to that of the predicate device, this being a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device includes hardware to support interfacing with video endoscopy systems.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the comparison to the predicate device (GI Genius v1.0.2). The goal of the study for GI Genius v2.0.0 was to demonstrate non-inferiority or improved performance compared to the predicate.

    CharacteristicAcceptance Criteria (Predicate Device Performance)Reported Device Performance (GI Genius v2.0.0)Result (vs. Predicate)
    Lesion-based sensitivity82.0 %86.5 %Improved
    Frame-level True Positive228,929269,223Improved
    Frame-level True Negative5,235,6825,239,128Improved
    Frame-level False Positive108,115104,669Improved
    Frame-level False Negative232,861192,567Improved
    True Positive Rate per frame Mean49.57 %58.30 %Improved
    True Positive Rate per frame % of polyps99.7 %100 %Improved
    False Positive Rate per frame Mean2.02 %1.96 %Improved
    Frame-Based TPr/FPr ROC curve, AUC0.7230.796Improved
    False positive clusters < 500 msBaseline40 less than baselineImproved
    False positive clusters > 500 msBaseline1 more than baselineSimilar
    Video delay, signal in to signal out1.52 μs1.52 μsSame

    2. Sample Size Used for the Test Set and Data Provenance

    The non-clinical testing for standalone performance used 150 videos containing 338 polyps.

    The data provenance is not explicitly stated regarding country of origin or whether it was retrospective or prospective. However, given that it's a re-training of a neural network with "additional procedure videos," it suggests a retrospective dataset of previously recorded endoscopy procedures.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given text.

    4. Adjudication Method for the Test Set

    This information is not provided in the given text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned for the GI Genius v2.0.0. The evaluation focuses on standalone performance relative to its predicate.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone performance test was done. The text states: "Standalone Performance Testing has been carried out to assess the performance of the subject device in accordance with the same test protocol as that used for the predicate device, the results of which demonstrate substantial equivalence to the predicate device."

    7. The Type of Ground Truth Used

    The ground truth used for performance evaluation appears to be based on the identified "polyps and adenomas" within the procedure videos. While not explicitly stated as "expert consensus" or "pathology," the context of "colonic mucosal lesions (such as polyps and adenomas)" generally implies a clinical or pathological reference standard.

    8. The Sample Size for the Training Set

    The document mentions that the neural network was re-trained with "additional procedure videos" and "improved data augmentation," but it does not specify the sample size of the training set (number of videos or lesions).

    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. However, given that the device aids in "detecting colonic mucosal lesions (such as polyps and adenomas)," it implies that these lesions were identified and annotated within the training videos, likely by expert endoscopists or pathologists.