510(k) Data Aggregation

    K Number
    K251203

    Validate with FDA (Live)

    Date Cleared
    2025-12-03

    (229 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AVIEW Lung Nodule CAD is a Computer-Aided Detection (CAD) software designed to assist radiologists in the detection of pulmonary nodules (with diameter 3-20 mm) during the review of CT examinations of the chest for asymptomatic populations. AVIEW Lung Nodule CAD provides adjunctive information to alert the radiologists to regions of interest with suspected lung nodules that may otherwise be overlooked. AVIEW Lung Nodule CAD may be used as a second reader after the radiologist has completed their initial read. The algorithm has been validated using non-contrast CT images, the majority of which were acquired on Siemens SOMATOM CT series scanners; therefore, limiting device use to use with Siemens SOMATOM CT series is recommended.

    Device Description

    The AVIEW Lung Nodule CAD is a software product that detects nodules in the lung. The lung nodule detection model was trained by Deep Convolution Neural Network (CNN) based algorithm from the chest CT image. Automatic detection of lung nodules of 3 to 20mm in diameter CT images. By complying with DICOM standards, this product can be linked with Picture Archiving and Communication System (PACS) and provides a separate user interface to provide functions such as analyzing, identifying, storing, and transmitting quantified values related to lung nodules. The CAD's results could be displayed after the user's first read, and the user could select or de-select the mark provided by the CAD. The device's performance was validated with SIEMENS' SOMATOM series manufacturing. The device is intended to be used with a cleared AVIEW platform.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for AVIEW Lung Nodule CAD (K251203) does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and the study proving device performance.

    Specifically, the document states: "This Medical device is not new; therefore, a clinical study was not considered necessary prior to release. Additionally, there was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing."

    This indicates that clinical performance (e.g., accuracy against a medical ground truth) was not a primary focus of this particular submission, but rather a demonstration of equivalence to a predicate device and non-clinical testing (software verification/validation, cybersecurity, OTS testing).

    Therefore, based solely on the provided text, I cannot provide details on specific acceptance criteria related to a clinical performance study (like sensitivity, specificity, or FROC scores) or the specifics of a study that proves the device meets those criteria from a clinical perspective.

    However, I can extract the available information.


    Acceptance Criteria and Study for AVIEW Lung Nodule CAD (K251203)

    Based on the provided FDA 510(k) Clearance Letter, the primary "acceptance criteria" for this submission appear to be related to demonstrating substantial equivalence to a predicate device and adherence to non-clinical software and cybersecurity standards, rather than establishing new clinical performance metrics. The document explicitly states that "a clinical study was not considered necessary" and "no clinical testing required."

    1. Table of Acceptance Criteria and Reported Device Performance

    As specific clinical performance metrics (e.g., sensitivity, specificity, FROC analysis) and their associated acceptance criteria are not detailed in this 510(k) summary, I can only infer the "performance" in terms of equivalence and successful non-clinical testing.

    Acceptance Criterion (Inferred from Submission)Reported Device Performance (Summary from Submission)
    Substantial Equivalence to Predicate DeviceDevice has the "same purpose and operating principle and has same functions" as the predicate. "Differences between the prior device and the proposed device are not significant because they do not cause new or potential safety risks... and do not raise questions about safety or effectiveness."
    Software Verification & Validation"Results of the software verification and validation tests concluded that the proposed device is substantially equivalent to the predicates device." Unit Test, System Test, and Regression Test were conducted.
    Cybersecurity CompliancePenetration Test was conducted to comply with cybersecurity requirements.
    Off-The-Shelf (OTS) Software ComplianceOTS Test Report was conducted to comply with OTS requirements.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify a sample size for a clinical test set or the provenance of any data used for clinical validation, as it states clinical testing was "not considered necessary" for this submission to establish substantial equivalence. The predicate device (K221592) presumably had a clinical validation, but those details are not in this document.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document, as a clinical study involving expert ground truth establishment was not a requirement for this specific submission.

    4. Adjudication Method for the Test Set

    This information is not provided in the document, as a clinical study involving adjudication was not a requirement for this specific submission.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not performed for this submission. The document explicitly states that "clinical testing was not required."

    6. Standalone (Algorithm Only) Performance Study

    A standalone performance study with clinical metrics (e.g., sensitivity, specificity, FROC curves) was not detailed or required for this particular submission for substantial equivalence. The submission focuses on non-clinical software validation and equivalence.

    7. Type of Ground Truth Used

    Based on the information provided, a clinical ground truth (e.g., expert consensus, pathology, outcomes data) was not established or utilized within the scope of this specific 510(k) submission, as clinical testing was not required. The "ground truth" for the non-clinical tests would relate to expected software behavior and security standards.

    8. Sample Size for the Training Set

    The document states, "The lung nodule detection model was trained by Deep Convolution Neural Network (CNN) based algorithm from the chest CT image." However, the sample size for this training set is not provided in the given text.

    9. How the Ground Truth for the Training Set Was Established

    The document mentions that the model was trained using "chest CT image" data; however, the method by which the ground truth for this training set was established (e.g., expert annotations, pathology, outcomes) is not detailed in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221592

    Validate with FDA (Live)

    Date Cleared
    2023-02-24

    (267 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AVIEW Lung Nodule CAD is a Computer-Aided Detection (CAD) software designed to assist radiologists in the detection of pulmonary nodules (with diameter 3-20 mm) during the review of CT examinations of the chest for asymptomatic populations. AVIEW Lung Nodule CAD provides adjunctive information to alert the radiologists to regions of interest with suspected lung nodules that may otherwise be overlooked. AVIEW Lung Nodule CAD may be used as a second reader after the radiologist has completed their initial read. The algorithm has been validated using non-contrast CT images, the majority of which were acquired on Siemens SOMATOM CT series scanners; therefore, limiting device use to use with Siemens SOMATOM CT series is recommended.

    Device Description

    The AVIEW Lung Nodule CAD is a software product that detects nodules in the lung. The lung nodule detection model was trained by Deep Convolution Network (CNN) based algorithm from the chest CT image. Automatic detection of lung nodules of 3 to 20mm in chest CT images. By complying with DICOM standards, this product can be linked with the Picture Archiving and Communication System (PACS) and provides a separate user interface to provide functions such as analyzing, identifying, storing, and transmitting quantified values related to lung nodules. The CAD's results could be displayed after the user's first read, and the user could select or de-select the mark provided by the CAD. The device's performance was validated with SIEMENS’ SOMATOM series manufacturing. The device is intended to be used with a cleared AVIEW platform.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the AVIEW Lung Nodule CAD, as derived from the provided document:

    Acceptance Criteria and Reported Device Performance

    Criteria (Standalone Performance)Acceptance CriteriaReported Device Performance
    Sensitivity (patient level)> 0.80.907 (0.846-0.95)
    Sensitivity (nodule level)> 0.8Not explicitly stated as separate from patient level, but overall sensitivity is 0.907.
    Specificity> 0.60.704 (0.622-0.778)
    ROC AUC> 0.80.961 (0.939-0.983)
    Sensitivity at FP/scan < 2> 0.80.889 (0.849-0.93) at FP/scan=0.504

    Study Details

    1. Acceptance Criteria and Reported Device Performance (as above)

    2. Sample size used for the test set and data provenance:

    • Test Set Size: 282 cases (140 cases with nodule data and 142 cases without nodule data) for the standalone study.
    • Data Provenance:
      * Geographically distinct US clinical sites.
      * All datasets were built with images from the U.S.
      * Anonymized medical data was purchased.
      * Included both incidental and screening populations.
      * For the Multi-Reader Multi-Case (MRMC) study, the dataset consisted of 151 Chest CTs (103 negative controls and 48 cases with one or more lung nodules).

    3. Number of experts used to establish the ground truth for the test set and their qualifications:

    • Number of Experts: Three (for both the MRMC study and likely for the standalone ground truth, given the consistency in expert involvement).
    • Qualifications: Dedicated chest radiologists with at least ten years of experience.

    4. Adjudication method for the test set:

    • Not explicitly stated for the "standalone study" ground truth establishment.
    • For the MRMC study, the three dedicated chest radiologists "determined the ground truth" in a blinded fashion. This implies a consensus or majority vote, but the exact method (e.g., 2+1, 3+1) is not specified. It does state "All lung nodules were segmented in 3D" which implies detailed individual expert review before ground truth finalization.

    5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study:

    • Yes, an MRMC study was performed.
    • Effect size of human readers improving with AI vs. without AI assistance:
      * AUC: The point estimate difference was 0.19 (from 0.73 unassisted to 0.92 aided).
      * Sensitivity: The point estimate difference was 0.23 (from 0.68 unassisted to 0.91 aided).
      * FP/scan: The point estimate difference was 0.24 (from 0.48 unassisted to 0.28 aided), indicating a reduction in false positives.
    • Reading Time: "Reading time was decreased when AVIEW Lung Nodule CAD aided radiologists."

    6. Standalone (algorithm only without human-in-the-loop performance) study:

    • Yes, a standalone study was performed.
    • The acceptance criteria and reported performance for this study are detailed in the table above.

    7. Type of ground truth used:

    • Expert consensus by three dedicated chest radiologists with at least ten years of experience. For the standalone study, it is directly compared against "ground truth," which is established by these experts. For the MRMC study, the experts "determined the ground truth." The phrase "All lung nodules were segmented in 3D" suggests a thorough and detailed ground truth establishment.

    8. Sample size for the training set:

    • Not explicitly stated in the provided text. The document mentions the lung nodule detection model was "trained by Deep Convolution Network (CNN) based algorithm from the chest CT image," but does not provide details on the training set size.

    9. How the ground truth for the training set was established:

    • Not explicitly stated in the provided text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K214036

    Validate with FDA (Live)

    Device Name
    AVIEW
    Date Cleared
    2022-12-23

    (365 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AVIEW provides CT values for pulmonary tissue from CT thoracic and cardiac datasets. This software can be used to support the physician providing quantitative analysis of CT images by image segmentation of sub-structures in the lung, lobe, airways, fissures completeness, cardiac, density evaluation, and reporting tools. AVIEW is also used to store, transfer, inquire and display CT data set on-premises and as a cloud environment to allow users to connect by various environments such as mobile devices and Chrome browsers. Converts the sharp kernel for quantitative analysis of segmenting low attenuation areas of the lung. Characterizing nodules in the lung in a single study or over the time course of several thoracic studies. Characterizations include type, location of the nodule, and measurements such as size (major axis, minor axis), estimated effective diameter from the volume of the nodule, the volume of the nodule, Mean HU(the average value of the CT pixel inside the nodule in HU), Minimum HU, Max HU, mass (mass calculated from the CT pixel value), and volumetric measures(Solid major, length of the longest diameter measure in 3D for a solid portion of the nodule, Solid 2nd Major: The size of the solid part, measured in sections perpendicular to the Major axis of the solid portion of the nodule), VDT (Volume doubling time), and Lung-RADS (classification proposed to aid with findings.) ). The system automatically performs the measurement, allowing lung nodules and measurements to be displayed and, integrate with FDA certified Mevis CAD (Computer aided detection) (K043617). It also provides the Agatston score, and mass score by the whole and each artery by segmenting four main arteries (right coronary artery, left main coronary, left anterior descending, and left circumflex artery). Based on the calcium score provides CAC risk based on age and gender. The device is indicated for adult patients only.

    Device Description

    The AVIEW is a software product that can be installed on a PC. It shows images taken with the interface from various storage devices using DICOM 3.0, the digital image and communication standard in medicine. It also offers functions such as reading, manipulation, analyzing, post-processing, saving, and sending images by using software tools. And is intended for use as a quantitative analysis of CT scanning. It provides the following features such as segmentation of lung, lobe, airway, fissure completeness, semi-automatic nodule management, maximal plane measures and volumetric measures, automatic nodule detection by integration with 3rd party CAD. It also provides the Brocks model, which calculates the malignancy score based on numerical or Boolean inputs. Follow-up support with automated nodule matching and automatically categorize Lung-RADS score, which is a quality assurance tool designed to standardize lung cancer screening CT reporting and management recommendations that are based on type, size, size, size, size, size, size, size, size, size, size change, and other findings that are reported. It also provides a calcium score by automatically analyzing coronary arteries from the segmented arteries

    AI/ML Overview

    The provided document does not contain specific acceptance criteria and detailed study results for the AVIEW device that would allow for the construction of the requested table and comprehensive answer. The document primarily focuses on demonstrating substantial equivalence to a predicate device and briefly mentions software verification and validation activities.

    However, I can extract the information that is present and highlight what is missing.

    Here's an analysis based on the provided text, indicating where information is present and where it is absent:


    Acceptance Criteria and Device Performance (Partial)

    The document mentions "pre-determined Pass/Fail criteria" for software verification and validation, but it does not explicitly list these criteria or the numerical results for them. It broadly states that the device "passed all of the tests."

    Table of Acceptance Criteria and Reported Device Performance

    Feature/MetricAcceptance CriterionReported Device Performance
    General Software PerformancePassed all tests based on pre-determined Pass/Fail criteriaPassed all tests
    Unit TestSuccessful functional, performance, and algorithm analysis for image processing algorithm componentsTests conducted using Google C++ Unit Test Framework
    System Test (Defects)No 'Major' or 'Moderate' defects foundNo 'Major' or 'Moderate' defects found (implies 'Passed' for this criterion)
    Kernel Conversion (LAA result reliability)LAA result on kernel-converted sharp image should have higher reliability with soft kernel than LAA results on sharp kernel image not applying Kernel Conversion.Test conducted on 96 total images (53 US, 43 Korean). (Result stated as 'A', indicating this was a test conducted but no specific performance metric is given for how much higher the reliability was).
    Fissure CompletenessCompared to radiologists' assessmentEvaluated using Bland-Altman plots; Kappa and ICC reported. (Specific numerical results are not provided).

    Detailed Breakdown of Study Information:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria: Not explicitly stated with numerical targets. The document mentions "pre-determined Pass/Fail criteria" for software verification and validation and "Success standard of System Test is not finding 'Major', 'Moderate' defect." For kernel conversion, the criterion is stated qualitatively (higher reliability). For fissure completeness, it's about comparison to radiologists.
      • Reported Device Performance:
        • General: "passed all of the tests."
        • System Test: "Success standard... is not finding 'Major', 'Moderate' defect."
        • Kernel Conversion: "The LAA result on kernel converted sharp image should have higher reliability with the soft kernel than LAA results on sharp kernel image that is not Kernel Conversion applied." (This is more of a hypothesis or objective rather than a quantitative result here).
        • Fissure Completeness: "The performance was evaluated using Bland Altman plots to assess the fissure completeness performance compared to radiologists. Kappa and ICC were also reported." (Specific numerical values for Kappa/ICC are not provided).
    2. Sample sizes used for the test set and the data provenance:

      • Kernel Conversion: 96 total images (53 U.S. population and 43 Korean).
      • Fissure Completeness: 129 subjects from TCIA (The Cancer Imaging Archive) LIDC database.
      • Data Provenance: U.S. and Korean populations for Kernel Conversion, TCIA LIDC database for Fissure Completeness. The document does not specify if these were retrospective or prospective studies. Given they are from archives/databases, they are most likely retrospective.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not specified in the provided text. For Fissure Completeness, it states "compared to radiologists," but the number and qualifications of these radiologists are not detailed.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not specified in the provided text.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not specified. The document mentions "compared to radiologists" for fissure completeness, but it does not detail an MRMC study comparing human readers with and without AI assistance for measuring an effect size of improvement.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the performance tests described (e.g., Nodule Matching, LAA Comparative Experiment, Semi-automatic Nodule Segmentation, Fissure Completeness, CAC Performance Evaluation) appear to be standalone evaluations of the algorithm's output against a reference (ground truth or expert assessment), without requiring human interaction during the measurement process by the device itself.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Fissure Completeness, the ground truth appears to be expert assessment/consensus from radiologists implied by "compared to radiologists."
      • For other performance tests like "Nodule Matching," "LAA Comparative Experiment," "Semi-automatic Nodule Segmentation," "Brock Model Calculation," etc., the specific type of ground truth is not explicitly stated. It's likely derived from expert annotations or established clinical metrics but is not detailed.
    8. The sample size for the training set:

      • Not specified in the provided text. The document refers to "pre-trained deep learning models" for the predicate device, but gives no information on the training data for the current device.
    9. How the ground truth for the training set was established:

      • Not specified in the provided text.

    Summary of Missing Information:

    The document serves as an FDA 510(k) summary, aiming to demonstrate substantial equivalence to a predicate device rather than providing a detailed clinical study report. Therefore, specific quantitative performance metrics, detailed study designs (e.g., number and qualifications of readers, adjudication methods for ground truth, specifics of MRMC studies), and training set details are not included.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220408

    Validate with FDA (Live)

    Device Name
    AVIEW RT ACS
    Date Cleared
    2022-11-10

    (269 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AVIEW RT ACS provides deep-learning-based auto-segmented organs and generates contours in RT-DICOM format from CT images which could be used as an initial contour for the clinicians to approve and edit by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.

    • a. Deep learning contouring from four body parts (Head & Neck, Breast, Abdomen, and Pelvis)
    • b. Generates RT-DICOM structure of contoured organs
    • c. Rule-based auto pre-processing
      Receive/Send/Export medical images and DICOM data
      Note that the Breast (Both right and left lung, Heart) were validated with non-contrast CT. Head & Neck (Both right and left Eyes, Brain and Mandible), Abdomen (Both right and Liver), and Pelvis (Both right and left Femur and Bladder) were validated with Contrast CT only.
    Device Description

    The AVIEW RT ACS provides deep-learning-based auto-segmented organs and generates contours in RT-DICOM format from CT images. This software could be used by the radiation oncology department planning, or other professions where a segmented mask of organs is needed.

    • Deep learning contouring: it can automatically contour the organ-at-risk (OARs) from four body parts (Head ● & Neck, Breast, Abdomen, and Pelvis)
    • . Generates RT-DICOM structure of contoured organs
    • . Rule-based auto pre-processing
      Receive/Send/Export medical images and DICOM data
    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The general acceptance criterion for the AVIEW RT ACS device appears to be comparable performance to a predicate device (MIM-MRT Dosimetry) in terms of segmentation accuracy, as measured by Dice Similarity Coefficient (DSC) and 95% Hausdorff Distance (HD). While explicit numerical acceptance thresholds are not stated in the provided text (e.g., "DSC must be greater than X"), the study is structured as a comparative effectiveness study. The expectation is that the AVIEW RT ACS performance should be at least equivalent to, if not better than, the predicate device.

    The study's tables (Tables 1-30) consistently show the AVIEW RT ACS achieving higher average DSC values (closer to 1, indicating better overlap) and generally lower average 95% HD values (closer to 0, indicating less maximum distance between contours), across various organs, demographic groups, and scanner parameters, compared to the predicate device.

    Table of Acceptance Criteria and Reported Device Performance:

    Metric / Organ (Examples)Acceptance Criterion (Implicit)AVIEW RT ACS Performance (Mean ± SD, [95% CI])Predicate Device Performance (Mean ± SD, [95% CI])Difference (AVIEW - Predicate)Meets Criteria?
    Overall DSCShould be comparable to or better than predicate device.(See tables below for individual organ results)(See tables below for individual organ results)Mostly positiveYes
    Overall 95% HD (mm)Should be comparable to or better than predicate device (i.e., lower HD).(See tables below for individual organ results)(See tables below for individual organ results)Mostly negative (indicating better AVIEW)Yes
    Brain DSCComparable to or better than predicate.0.97 ± 0.01 (0.97, 0.98)0.96 ± 0.01 (0.96, 0.96)0.01Yes
    Brain 95% HD (mm)Comparable to or better than predicate (lower HD).6.92 ± 20.46 (-1.1, 14.94)4.61 ± 2.17 (3.76, 5.46)2.31Mixed (Higher HD for AVIEW, but wide CI)
    Heart DSCComparable to or better than predicate.0.94 ± 0.03 (0.93, 0.95)0.78 ± 1.20 (0.70, 8.56)0.16Yes (Significantly better)
    Heart 95% HD (mm)Comparable to or better than predicate (lower HD).6.19 ± 4.21 (4.73, 7.65)18.90 ± 5.09 (17.14, 20.67)-12.71Yes (Significantly better)
    Liver DSCComparable to or better than predicate.0.96 ± 0.01 (0.96, 0.97)0.87 ± 0.06 (0.85, 0.90)0.09Yes
    Liver 95% HD (mm)Comparable to or better than predicate (lower HD).7.17 ± 12.07 (2.54, 11.81)24.62 ± 15.16 (18.79, 30.44)-17.44Yes (Significantly better)
    Bladder DSCComparable to or better than predicate.0.88 ± 0.14 (0.84, 0.93)0.52 ± 0.26 (0.44, 0.60)0.36Yes (Significantly better)
    Bladder 95% HD (mm)Comparable to or better than predicate (lower HD).10.55 ± 20.56 (3.74, 17.36)30.48 ± 22.76 (22.94, 38.02)-19.93Yes (Significantly better)

    Note: The tables throughout the document provide specific performance metrics for individual organs and sub-groups (race, vendors, slice thickness, kernel types). The general conclusion from these tables is that the AVIEW RT ACS consistently performs as well as or better than the predicate device across most metrics and categories.

    Study Details for Acceptance Criteria Proof:

    1. Sample Size Used for the Test Set: 120 cases.

      • Data Provenance: The dataset included cases from both South Korea and the USA. It was constructed with various ethnicities (White, Black, Asian, Hispanic, Latino, African, American, etc.), and from four major vendors (GE, Siemens, Toshiba, and Philips).
      • Retrospective/Prospective: Not explicitly stated, but the mention of a data set constructed for validation suggests a retrospective collection.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

      • Number of Experts: 3 radiation oncology physicians.
      • Qualifications: All were trained by "The Korean Society for Radiation Oncology," board-certified by the "Ministry of Health and Welfare," with a range of 9-21 years of experience in radiotherapy. The experts included attending assistant professors (n=2) and professors (n=1) from three institutions.
    3. Adjudication Method for the Test Set:

      • The method was a sequential editing process:
        1. One expert manually delineated the organs.
        2. The segmentation results from the first expert were then sequentially edited by the other two experts.
        3. The first expert made corrections.
        4. The result was then received by another expert who finalized the gold standard.
      • This can be considered a form of sequential consensus or collaborative review rather than a strict N+1 or M+N+1 method.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • Yes, a comparative effectiveness study was done. The study directly compares the AVIEW RT ACS against a predicate device (MIM-MRT Dosimetry).
      • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance:
        • The study does not measure the improvement of human readers with AI assistance. Instead, it evaluates the standalone performance of the AI device against the standalone performance of a predicate AI device, both compared to expert-generated ground truth. The "human readers" (the three experts) were used solely to create the ground truth, not to evaluate their performance with or without AI assistance.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done. The study compares the auto-segmentation results of the AVIEW RT ACS directly to the expert-derived "gold standard" and also compares it to the auto-segmentation of the predicate device. This is purely an algorithm-only evaluation.
    6. The Type of Ground Truth Used:

      • Expert Consensus. The ground truth was established by three radiation oncology physicians through a sequential delineation and editing process to create a "robust gold standard."
    7. The Sample Size for the Training Set:

      • Not specified within the provided text. The document refers only to the validation/test set.
    8. How the Ground Truth for the Training Set Was Established:

      • Not specified within the provided text. Since the training set size and characteristics are not mentioned, neither is the method for establishing its ground truth.