Found 1 results

510(k) Data Aggregation

    K Number
    K252313

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-01-14

    (173 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The gloves was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978, "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs".

    The following drugs have been tested with these gloves:

    Chemotherapy Drug & ConcentrationMinimum Breakthrough Detection Time in Minutes, 0.01 μg/cm2/min:
    Bendamustine HCl, 5 mg/ml (5,000 ppm)>240 min
    Bleomycin Sulfate, 15 mg/ml (15,000 ppm)>240 min
    Busulfan, 6 mg/ml (6,000 ppm)>240 min
    Carboplatin, 10 mg/ml (10,000 ppm)>240 min
    Carmustine 3.3 mg/ml (3,300 ppm)10.0 min
    Chloroquine, 50 mg/ml (50,000 ppm)>240 min
    Cisplatin, 1mg/ml (1000 ppm)>240 min
    Cyclophosphamide 20mg/ml (20,000 ppm)>240 min
    Cytarabine HCl (Cytosine), 100 mg/ml (100,000 ppm)>240 min
    Dacarbazine (DTIC), 10 mg/ml (10,000 ppm)>240 min
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240 min
    Epirubicin HCl, 2 mg/ml (2,000 ppm)>240 min
    Etoposide, 20 mg/ml (20,000 ppm)>240 min
    Fludarabine, 25 mg/ml (25,000 ppm)>240 min
    Fluorouracil (5 Flu), 50mg/ml (50,000ppm)>240 min
    Gemcitabine HCl, 38 mg/ml (38,000 ppm)>240 min
    Idarubicin HCl, 1 mg/ml (1,000 ppm)>240 min
    Ifosfamide (IFEX), 50 mg/ml (50,000 ppm)80.0 min
    Mechlorethamine HCl, 1 mg/ml (1,000 ppm)>240 min
    Melphalan HCl, 5 mg/ml (5,000 ppm)>240 min
    Methotrexate, 25mg/ml (25,000ppm)>240 min
    Mitomycin C, 0.5mg/ml (500 ppm)>240 min
    Mitoxantrone HCl (2.0 mg/mL) (2000 ppm)>240 min
    Oxaliplatin, 5 mg/ml (5,000 ppm)>240 min
    Paclitaxel, 6mg/ml (6,000ppm)>240 min
    Rituximab, 10 mg/ml (10,000 ppm)>240 min
    Thiotepa, l0mg/ml (10,000ppm)20.0 min
    Trisenox (Arsenic Trioxide), 1 mg/ml (1,000 ppm)>240 min
    Vincristine sulfate, 1mg/ml (1000 ppm)>240 min
    Vinorelbine, 10 mg/ml (10,00 ppm)>240 min
    Opioid DrugMinimum Breakthrough Detection Time in Minutes, 0.01 μg/cm2/min:
    Fentanyl Citrate Injection, 100mcg/2mg>240 min

    *Please note that the following drugs have low permeation times:
    Carmustine: 10.0 minutes, Thiotepa: 20.0 minutes
    Warning: Do not use with Carmustine & Thiotepa.
    Caution: Testing showed breakthrough time of 80.0 minutes with Ifosfamide IFEX (50.0 mg/mL).

    Device Description

    Not Found

    AI/ML Overview

    N/A