Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K182008

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-12-14

    (140 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Phasix™ ST Mesh is indicated for use in the reinforcement of soft tissue, where weakness involving soft tissue repair, such as for the repair of hernias.

    The Echo 2™ Positioning System is intended to facilitate the delivery and positioning of the mest during laparoscopic hernia repair.

    Device Description

    The Phasix™ ST Mesh with Echo 2™ Positioning System is a sterile single use mesh preattached to the Echo 2™ Positioning System. The subject device is a combination of the FDA cleared Phasix™ ST Mesh (K173143) and the Echo 2™ Positioning System that is cleared by the FDA in the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294). The Echo 2™ Positioning System is removed following the mesh placement and fixation. As with the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294), many sizes of the subject device include an introducer tool accessory.

    The subject device is intended for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias and will utilize a positioning system to facilitate the delivery and positioning of the mesh during laparoscopic hernia repair.

    Procedures involving "laparoscopic hernia repair" also pertain to those which are conducted under laparoscopic visualization through a trocar, such as robotic assisted laparoscopic ventral hernia repair (LVHR).

    The Phasix™ ST Mesh in the subject device is a fully resorbable mesh with a resorbable hydrogel coating and is identical to the secondary predicate device, Phasix™ ST Mesh (K173143). It is co-knitted using Poly-4-hydroxybutyrate (P4HB) and Polyglycolic Acid (PGA) fibers to form two sided mesh. The PGA side of the mesh is coated with a bioresorbable, chemically modified sodium hyaluronate (HA), Carboxymethylcellulose (CMC) and Polyethylene Glycol (PEG) based hydrogel. The fascial P4HB side of the mesh allows for a prompt fibroblastic response through its interstices, allowing for complete tissue ingrowth, similar to P4HB mesh alone. The visceral PGA side of the mesh coated with a bioresorbable hydrogel separates the mesh from underlying tissues and organ surfaces to help minimize tissue attachment to the mesh. Shortly after hydration, the biopolymer coating becomes a hydrated gel that is resorbed from the site in less than 30 days.

    Preclinical implantation studies indicate that resorption of the P4HB fibers is minimal throughout the 12 week expected healing period and up to 26 weeks post implantation. Significant degradation of the mesh fibers is observed in preclinical studies within 12 to 18 months, indicating loss in mechanical integrity and strength of the mesh. While fiber segments were observed at 18 months, they continued to degrade.

    This Phasix™ ST Mesh (K173143) is preattached to the Echo 2™ positioning system on the hydrogel side of the mesh using nylon connectors. The Echo 2™ Positioning System is composed of nitinol wire, encased in a sealed nylon frame. A Polyethylene Terephthalate (PET) center hoisting suture is attached to the nylon frame. The Echo 2™ positioning system utilized in the subject device is identical in materials and construction to the Ventralight™ ST Mesh with Echo 2™ Positioning System (K170294).Once inserted into the abdomen, the positioning system facilitates both deployment and positioning of the mesh. After initial fixation is complete, the Echo 2™ Positioning System is manually detached from the mesh and then removed from the body through a trocar or skin incision.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Phasix™ ST Mesh with Echo 2™ Positioning System." This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than presenting a de novo study with strict acceptance criteria and a definitive study to prove the device meets those criteria from scratch.

    Therefore, the information you're requesting regarding a detailed breakdown of acceptance criteria met by a specific study is not explicitly provided in the typical format of a clinical trial or performance study report. Instead, the document discusses performance data in the context of demonstrating similarity and equivalence to predicate devices.

    Here's an attempt to answer your questions based on the available information, with explanations for what is not present:


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a formal table with specific numerical acceptance criteria and corresponding reported device performance values. Instead, it states that "All the samples of the subject device successfully met the established acceptance criteria and performed as intended and similar to the predicate devices" for the product testing specified.

    The types of tests performed and their general outcomes are summarized below:

    Type of Performance TestAcceptance Criteria (Implicit)Reported Device Performance
    Biocompatibility TestingConformity to ISO 10993-1 and established biocompatibility of known materials. No adverse interactions between components.Successfully passed Cytotoxicity, Pyrogenicity, and Chemical Characterization tests. Demonstrated to be biocompatible with no adverse interactions between the Phasix™ ST Mesh and the Echo 2™ Positioning System.
    Mechanical TestingMesh integrity (gel disruption), connector-to-mesh attachment strength, and ball burst performance comparable to predicate devices.Passed all test requirements; test results demonstrated substantial equivalence to predicate devices.
    Simulated Use Evaluations (Bench Top Simulator)Satisfactory performance in delivery, support, access/fixation, and removal.Successfully met established acceptance criteria and performed as intended and similar to predicate devices.
    Performance Testing (Porcine Model)Satisfactory performance in delivery, support, access/fixation, and removal.Successfully met established acceptance criteria and performed as intended and similar to predicate devices.

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Sizes: The document does not specify the exact sample sizes (e.g., number of meshes, animals, or test replicates) used for each individual test or evaluation. It broadly refers to "all the samples of the subject device."
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of a 510(k) submission and the focus on bench and animal testing, these are typically prospective laboratory or animal studies, not human clinical trials with geographical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable to the type of tests described. The evaluations are primarily engineering, chemical, and biological performance tests (biocompatibility, mechanical, simulated use, animal model performance), not diagnostic or interpretative tasks requiring human expert "ground truth" in the way a medical imaging AI would. The "ground truth" for these tests is based on objective measurements and predefined specifications / predicate device performance.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or expert reviews of complex data (e.g., medical image interpretation) where there might be disagreement among reviewers. The tests described are laboratory and animal-based, with results determined by objective measurements and passing/failing criteria without such an adjudication process.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. An MRMC study is designed to evaluate the performance of diagnostic devices, often medical imaging AI, and how it impacts expert human reader performance. The Phasix™ ST Mesh with Echo 2™ Positioning System is a surgical mesh with an associated positioning system, not a diagnostic device or AI software. There is no AI component in this device, nor is it intended for interpretation by human readers.


    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm or software. It is a physical medical device (mesh and positioning system) used by a surgeon.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the tests performed can be characterized as follows:

    • Biocompatibility: Established scientific standards (ISO 10993-1) and objective laboratory measurements for cytotoxicity, pyrogenicity, and chemical characterization.
    • Mechanical Testing: Objective measurements against predefined engineering specifications and performance profiles of the predicate devices.
    • Simulated Use Evaluations (Bench Top Simulator) & Performance Testing (Porcine Model): Observation and objective assessment of the device's functional performance (delivery, support, access/fixation, removal) against expected results and the demonstrated performance of predicate devices.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set" for an algorithm. All the data described relates to verification and validation testing of the physical medical device.


    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI algorithm, this question is not relevant to the described device.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VENTRALIGHT™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias.

    The ECHO 2™ Positioning System is intended to facilitate the delivery and positioning of soft tissue prostheses during laparoscopic hernia repair.

    Device Description

    The Subject Device, VENTRALIGHT™ ST Mesh with ECHO 2TM Positioning System is provided sterile for single use and consists of an implantable surgical mesh (VENTRALIGHT™ ST Mesh) with a pre-attached mesh positioning system (ECHO 2™ Positioning System). The Subject Device is designed for the reconstruction of soft tissue deficiencies during laparoscopic ventral hernia repair.

    AI/ML Overview

    The provided text describes the regulatory clearance for a medical device, the VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System, and its substantial equivalence to a predicate device. However, it does not detail specific acceptance criteria and the comprehensive study results in the format requested for an AI/algorithm-based device.

    This document pertains to a surgical mesh and a positioning system, which are physical medical devices, not an AI or algorithm-based system. Therefore, many of the requested sections related to AI/algorithm performance metrics (like sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of device and are not present in the provided text.

    Based on the provided information, here's what can be extracted and what cannot:

    1. Table of acceptance criteria and the reported device performance:

    The document states that "All samples tested met the established acceptance criteria" for product testing and cadaveric studies, and "The VENTRALIGHT™ ST Mesh with ECHO 2™ Positioning System passed all the test requirements and demonstrated substantial equivalence to the test results of the Predicate Device." However, the specific numerical acceptance criteria or the raw performance data are not provided in this summary. The general categories of testing and the conclusion of meeting criteria are mentioned.

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilitySuccessfully performed tests: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogenicity. Met established acceptance criteria.
    Product Testing (Simulated use)Passed all test requirements; demonstrated substantial equivalence to Predicate Device. Met established acceptance criteria.
    Product Testing (Mechanical Mesh)Passed all test requirements; demonstrated substantial equivalence to Predicate Device. Met established acceptance criteria.
    Product Testing (Mechanical Positioning System)Passed all test requirements; demonstrated substantial equivalence to Predicate Device. Met established acceptance criteria.
    Cadaveric Study (Safety and Feasibility)Demonstrated safe delivery and support of mesh construct without interfering with fixation or frame removal; performed similarly to Predicate Device. Met established acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: The document states "All samples tested met the established acceptance criteria" for product testing and cadaveric studies, but it does not specify the numerical sample sizes used for these tests.
    • Data Provenance: Not specified.
    • Retrospective/Prospective: Not specified, although cadaveric studies would inherently be prospective in their execution given the study design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable as this isn't an AI/algorithm device requiring expert ground truth for classification. The "ground truth" for this physical device is based on objective measurements and observations during mechanical and cadaveric testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. Clinical studies were not performed, and this is not an AI-assisted device for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For biocompatibility, the ground truth is established by adherence to ISO 10993-1 standards and successful performance in standardized biological tests. For product testing (mechanical and simulated use) and cadaveric studies, the "ground truth" or acceptance criteria are based on pre-defined engineering specifications, functional requirements, and safety/feasibility observations relative to the predicate device.

    8. The sample size for the training set:

    Not applicable. There is no AI/algorithm training set for this physical device.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142808

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-03-09

    (161 days)

    Product Code
    Regulation Number
    878.4750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CapSure ™ Permanent Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

    Device Description

    The CapSure™ Permanent Fixation System is a sterile single use device that delivers either 15 or 30 permanent fasteners via a straight shaft. The CAPSURE™ Permanent Fixation System is 37 cm in length. The fasteners are designed with a 316L Stainless steel helical coil and polyetheretherketone (PEEK) cap on the proximal end to support mesh or tissue.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for the CapSure™ Permanent Fixation System, a medical device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityConforms to FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process." Specific tests to include Cytotoxicity, Sensitization, Irritation, Pyrogen Testing (for device); Cytotoxicity, Maximization Sensitization Study, Intracutaneous Study, Acute Systemic Toxicity Study, Pyrogenicity, Hemolysis, Complement Activation Assay, Intramuscular Implant (12 weeks), Rabbit Femoral Bone Implant (12 & 26 weeks), Bacterial Reverse Mutation (Ames) Assay, In Vitro Mouse Lymphoma Mutation Assay, Mouse Peripheral Blood Micronucleus Test, Subacute (14-Day) Intraperitoneal Toxicity Study Mice, Subchronic (14-Day) Intravenous Toxicity Study Mice, Subchronic (13-Week) Toxicity Study in Rats, In Vivo Neurotoxicity, Non-Volatile Residue, Residue on Ignition, Turbidity, UV Absorption (for implant materials). Implantable fastener material conforms to ASTM F138 (316L Stainless Steel) and ASTM F2026 (PEEK).All samples tested met the acceptance criteria.
    Mechanical TestingNot explicitly stated in quantitative terms but implied to demonstrate substantial equivalence to the predicate device (ProTack™). Specific tests include Trigger Force, Mesh Compatibility, Deployment Reliability, MR Compatibility, Burst Testing.CapSure™ passed all the test requirements and showed substantial equivalence to the results of the predicate device - ProTack™.
    Animal StudiesNot explicitly stated in quantitative terms but implied to demonstrate substantial equivalence to the predicate device. Specific study: Porcine implantation strength study.CapSure™ passed all the test requirements and showed substantial equivalence to the results of the predicate device - ProTack™.

    2. Sample Size for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for the mechanical and animal studies. For biocompatibility, it refers to "All samples tested," but the specific number is not provided. The data provenance is from internal company testing conducted by Davol Incorporated. It is retrospective in the sense that it's reported after the tests were performed to support the 510(k) submission. No information about the country of origin of the data is specified beyond "the company."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    This information is not applicable as the studies described are performance and safety tests (biocompatibility, mechanical, animal studies) for a medical device that performs a mechanical function, not an AI or diagnostic device that requires expert-established ground truth from images or clinical data.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as point 3.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a surgical fixation system, not an AI-powered diagnostic tool, so MRMC studies involving human readers and AI assistance are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a physical surgical tool; there is no "algorithm only" performance to be evaluated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the tests performed can be characterized as follows:

    • Biocompatibility: Established by adherence to ISO 10993 standards and the expected biological responses (e.g., absence of cytotoxicity, irritation, sensitization).
    • Mechanical Testing: Established by engineering specifications, physical measurements, and comparison to the performance of the legally marketed predicate device (ProTack™).
    • Animal Studies: Established by anatomical and physiological assessment in the animal model, and comparison to the performance of the predicate device.

    8. The Sample Size for the Training Set:

    This is not applicable. The device is a physical medical device, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    This is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112791

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-02-24

    (151 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Site-Scrub IPA Device is intended for use on injection ports and female luer hubs as a disinfecting cleaner.

    Device Description

    The subject Site-Scrub IPA Device is a handheld, sterile, single use device containing isopropyl alcohol (IPA) 70% (v/v) solution. The Site-Scrub IPA Device, by design, provides active mechanical friction and IPA to clean and disinfect surfaces, such as injection ports and female luer hubs, in accordance with 2011 CDC guidelines that recommend for needleless intravascular catheter systems that contamination be minimized by "scrubbing the access port with an appropriate antiseptic (including 70% alcohol) and accessing the port only with sterile devices." The foam design allows for application of the IPA on internal and external surfaces. The Site-Scrub IPA Device has been specifically designed to disinfect female luer hubs and injection ports and has undergone in vitro testing to validate its effectiveness for this purpose. The results of in vitro antimicrobial efficacy testing show the Site-Scrub IPA Device is effective, when used as intended, for significantly reducing (> 4 Log10 or 99.99% reduction) microbial load of the following microbes known to be associated with catheter line-associated blood stream infections (CLABSI) which include: 1. Candida albicans, 2. Candida parapsilosis, 3. Escherichia coli, 4. Pseudomonas aeruginosa, 5. Staphylococcus aureus (MRSA), and 6. Staphylococcus epidermidis (MRSE).

    AI/ML Overview

    The information provided describes the Site-Scrub IPA Device, a handheld, sterile, single-use device containing 70% (v/v) isopropyl alcohol (IPA) solution, designed for disinfecting injection ports and female luer hubs.

    Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from Device Description)Reported Device Performance (from Device Description)
    Significantly reducing (> 4 Log10 or 99.99% reduction) microbial load of specified microbes associated with CLABSI when used as intended.In vitro antimicrobial efficacy testing demonstrated effective reduction (> 4 Log10 or 99.99% reduction) of microbial load.

    Specific Microbes Mentioned:

    • Candida albicans
    • Candida parapsilosis
    • Escherichia coli
    • Pseudomonas aeruginosa
    • Staphylococcus aureus (MRSA)
    • Staphylococcus epidermidis (MRSE)

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "in vitro testing" to validate effectiveness. However, it does not specify the sample size used for this testing.

    Regarding data provenance, the study appears to be in-vitro (laboratory-based) rather than involving human clinical data (retrospective or prospective). The country of origin for the data is not explicitly stated, but the submission is to the US FDA, implying testing was conducted to meet US regulatory requirements, likely within the US or a country with comparable standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth. Since the testing is described as "in vitro antimicrobial efficacy testing," the "ground truth" would be established by standard microbiological methods and controls, rather than expert human interpretation of medical images or clinical outcomes.

    4. Adjudication Method for the Test Set

    This information is not applicable as the described study is "in vitro antimicrobial efficacy testing" and does not involve human readers or complex interpretations that would require an adjudication method like 2+1 or 3+1. The results are based on objective measurement of microbial reduction.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human readers interpret medical cases, and the AI's impact on their performance is evaluated. The Site-Scrub IPA Device is a disinfectant, not an AI-powered diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Since the device is a physical disinfectant and not an algorithm or AI system, the concept of "standalone performance" for an algorithm is not applicable. The device's performance is inherently "standalone" in the sense that it functions physically to disinfect.

    7. The Type of Ground Truth Used

    The ground truth used for this study is based on in-vitro microbiological cultures and quantitative reduction of microbial load. This is determined by standard laboratory techniques to count colony-forming units (CFUs) or other measures of microbial viability before and after disinfection, compared against controls.

    8. The Sample Size for the Training Set

    This information is not applicable. The Site-Scrub IPA Device is a physical medical device, not an AI/machine learning algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1