Search Results
Found 83 results
510(k) Data Aggregation
(65 days)
The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures.
The Y-Knotless™ Flex Anchors may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor systems thereby stabilize the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.
The Y-Knotless™ Flex Anchors may be used in the following orthopedic procedures:
- Shoulder Labrum
- MPFL Reconstruction
- Foot and Ankle
- Biceps
- Meniscal Root Repair
The Y-Knotless™ Flex Anchors are a non-absorbable, all-suture, knotless implant, that are supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10-6. The implant anchors are composed of a white flat braided ultra-high molecular weight polyethylene (UHMWPE) suture anchor, a tapered #2 Hi-Fi® repair suture either white/black or green/white/black, and a #2-0 Hi-Fi® shuttle suture, blue and black. The implants are pre-loaded on a disposable driver with sutures cleated to the handle. The anchors require a 1.8 mm pre-drilled bone hole created using Class I, exempt instrumentation.
N/A
Ask a specific question about this device
(30 days)
The BioBrace® Extra-Articular Ligament Augmentation Kit is intended to reattach soft tissue to bone and reinforce soft tissue where weakness exists in extra-articular knee ligament surgical procedures.
The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.
The BioBrace® Reinforced BioInductive Implant is intended for reinforcement of soft tissues that are repaired by suture or suture anchors during knee repair surgery including reinforcement of patellar ligament, medial collateral ligament, lateral collateral ligament or other knee extra-articular ligaments. It is not intended to replace normal body structures or provide the full mechanical strength to the repair. Sutures used to repair the tear, and sutures or bone anchors used to attach the tissue to bone, provide mechanical strength for the tendon repair.
The BioBrace® Extra-Articular Ligament Augmentation Kit is a single-use augmentation procedure kit for use in general surgical procedures for BioBrace® Reinforced BioInductive Implant ("BioBrace® implant") reinforcement of soft tissue where weakness exists, specifically in the extra-articular knee ligaments.
The kit is comprised of bioabsorbable implants and instrumentation. The kit includes one (1) tapered BioBrace Implant 5x250mm, which is threaded through the anchor eyelet.
Kit components are assembled in a single PETG tray and retainer then sealed in a Tyvek pouch. The BioBrace® Extra-Articular Ligament Augmentation Kit is supplied sterile (ETO).
Identical to K240090
Argo Knotless® GENESYS™ Anchor: The Non-Self-Punching Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10-6. The anchor configuration requires a pre-prepared bone hole. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK material. Each size features a single use driver, a threaded anchor, a PEEK suture eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.
Identical to K244025
Argo Knotless® GENESYS™ Anchor: The Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10-6. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK Optima material. The device features a single use driver, a threaded anchor, a PEEK-Optima eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures or one (1) bioresorbable reinforced implant. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.
Identical to K242187
BioBrace® is intended for use in surgical procedures for reinforcement of soft tissue where weakness exists.
BioBrace® is also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. BioBrace® is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the repair.
N/A
Ask a specific question about this device
(30 days)
The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures.
The CONMED Argo Knotless® Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.
The CONMED Argo Knotless® Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10⁻⁶. The threaded anchor and the suture eyelet are manufactured of PolyEtherEtherKetone (PEEK) material. Each size features a single use driver, a threaded anchor, suture eyelet, a 1.0mm UHMWPE Hi-Fi™, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. A maximum of six (6) Hi-Fi #2 suture, or 1.3mm Hi-Fi ribbon suture, or (2) two limbs of Hi-Fi tape, or one (1) bioresorbable reinforced implant can be threaded through the eyelet. The repair sutures are then tensioned and cleated on the driver cleat, if deemed necessary by the surgeon. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.
The provided FDA 510(k) clearance letter pertains to the CONMED Argo Knotless® Anchor, a medical device for reattaching soft tissue to bone. This document is a clearance letter, not a detailed study report. Therefore, it does not contain the level of detail typically found in a clinical study for AI/software devices. In particular, it does not include any information about AI or software, MRMC studies, or standalone algorithm performance. The acceptance criteria and performance data described relate to the mechanical and material properties of the physical medical device.
Here's an analysis of the provided information based on your questions:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state quantitative acceptance criteria for each test. Instead, it indicates that "Testing has been completed to demonstrate that CONMED Argo Knotless® Anchor performs as intended and is substantially equivalent to the predicate device." The performance data is summarized by the types of tests conducted.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Mechanical Performance | |
| Insertion success | Testing completed successfully |
| Ultimate Pull Displacement Resistance | Testing completed successfully |
| Cyclic Loading performance | Testing completed successfully |
| Material/Biocompatibility | |
| Biocompatibility | In accordance with ISO 10993-1 and FDA# G95-1 |
| Pyrogen (Bacterial Endotoxin) | Testing completed successfully |
| Sterilization & Packaging | |
| ETO Sterilization to SAL of 10⁻⁶ | Device sterilized via Ethylene Oxide Sterilization - SAL (10⁻⁶) |
| Shelf-life validation | Testing completed successfully |
| Packaging integrity (Transportation) | Testing completed successfully |
| Packaging and Labeling User Validation | Testing completed successfully |
| User Interaction | |
| User Validation | Testing completed successfully |
2. Sample size used for the test set and the data provenance
The document does not specify specific sample sizes for any of the performance tests (e.g., insertion, pull displacement, cyclic loading).
The data provenance is from CONMED Corporation, the manufacturer, as part of their 510(k) submission. These are typically retrospective in the sense that the device has been manufactured and tested, and the results are then compiled for the submission. The tests are likely performed in a controlled laboratory or testing facility setting and do not involve human patient data in the context of device performance testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is generally applicable to AI/software performance where ground truth is established by human experts. For this physical device, the "ground truth" for mechanical properties is established through engineering standards and validated test methods. Therefore, experts would be engineers and technicians familiar with these standards and methods, but the document does not specify their number or qualifications beyond their implied role in executing the listed tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies to resolve discrepancies in expert interpretations (e.g., radiology reads). For physical device performance testing, the results are typically quantitative and objective, measured against established engineering specifications, rather than subject to expert interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical medical implant, not an AI or software device. Therefore, MRMC studies and "human readers improving with AI assistance" are not relevant to this 510(k) submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. As noted above, this clearance is for a physical medical device, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the performance of this physical device is based on engineering specifications, industry standards (e.g., ISO 10993-1 for biocompatibility), and validated test methods. These tests establish whether the mechanical properties (insertion, pull displacement, cyclic loading) and material properties (biocompatibility, pyrogenicity) meet predetermined safety and performance benchmarks for orthopedic implants.
8. The sample size for the training set
Not applicable. This device is a physical medical device, not a machine learning model. Therefore, there is no "training set."
9. How the ground truth for the training set was established
Not applicable. As there is no "training set" for a physical medical device, this question is not relevant.
Ask a specific question about this device
(54 days)
The BioBrace® RC Delivery System is intended for use in rotator cuff surgical procedures for reinforcement of soft tissue where weakness exists. The BioBrace® Implant is intended for reinforcement of soft tissues that are repaired by suture or suture anchors, during rotator cuff tendon repair surgery. The BioBrace® Implant is not intended to replace normal body structures or provide the full mechanical strength to support the rotator cuff. Sutures used to repair the tear, and sutures or bone anchors used to attach the tissue to bone, provide mechanical strength for the tendon repair.
The Disposable Inserter is intended to facilitate the delivery and positioning of the implant during soft tissue repairs.
The BioBrace® RC Delivery System is comprised of a BioBrace® Implant pre-stitched with HI-FI® suture, a threader assembly that facilitates passing sutures through the BioBrace® Implant, and an inserter to facilitate placement of the BioBrace® Implant into the subacromial space.
The previously cleared BioBrace® Implant is a bioresorbable, reinforced Implant composed of a highly porous collagen sponge made from insoluble bovine tendon type-1 collagen and reinforced with poly-L-lactic-acid (PLLA) multifilament yarn. The BioBrace® Implant is 80% porous with a median pore diameter of 19 µm. BioBrace® implants are approximately 3 mm thick and provided in two rectangular sizes of 23 x 25 mm and 35 x 25 mm.
The BioBrace® RC Delivery System is single-use and supplied sterile (ETO).
The provided text is a 510(k) clearance letter for a medical device called the BioBrace® RC Delivery System. This type of document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed clinical study for novel acceptance criteria.
Therefore, the document does not contain the information requested regarding detailed acceptance criteria for diagnostic performance, a specific study proving the device meets those criteria, sample sizes for test/training sets, data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone algorithm performance, or how ground truth was established in the context of diagnostic accuracy.
The document describes performance testing that focused on demonstrating the functional equivalence and safety of the new delivery system and integrated implant compared to the previously cleared implant. This testing includes:
1. Table of Acceptance Criteria and Reported Device Performance (based on provided text):
Since the document does not present quantitative diagnostic performance metrics or acceptance criteria for such, the table below reflects the types of performance aspects mentioned for this delivery system and implant combination. All "reported device performance" are stated as having been successfully completed and demonstrating equivalence/effectiveness/safety.
| Acceptance Criterion Type (Inferred) | Reported Device Performance |
|---|---|
| Functional Equivalence | Demonstrated performance effectiveness and safety similar to the predicate device. |
| Biocompatibility | Demonstrated compliance (in accordance with ISO 10993-1 for the predicate). |
| Packaging Integrity | Testing completed successfully. |
| Packaging and Labeling User Validation | Completed successfully. |
| Performance Testing (of the delivery system aspects) | Completed successfully. |
| Shelf-life | 12 months (for the proposed device). |
| Sterilization Efficacy | Ethylene Oxide Sterilization achieving a SAL (10-6) (same as predicate). |
| Transportation Integrity | Testing completed successfully. |
| User Validation (of the delivery system) | Completed successfully. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not specify sample sizes for any test sets in the context of diagnostic performance. The studies mentioned appear to be engineering and biological performance tests rather than clinical diagnostic studies. Data provenance (country of origin, retrospective/prospective) is also not provided.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not applicable as the document describes performance testing for a surgical mesh delivery system and implant, not a diagnostic device requiring expert-established ground truth for classification.
4. Adjudication Method for the Test Set:
This information is not applicable for the reasons stated above.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
An MRMC study was not conducted (or at least not reported in this document), as this is not a diagnostic imaging device intended for interpretation by human readers.
6. Standalone (Algorithm Only) Performance:
This information is not applicable as the device is a physical surgical mesh delivery system, not an algorithm or AI software within a diagnostic context.
7. Type of Ground Truth Used:
This information is not applicable as the described tests relate to material properties, sterility, packaging, and functional delivery of a surgical implant, not diagnostic accuracy. For the biocompatibility claims, "ground truth" would be established by the ISO 10993-1 standards themselves.
8. Sample Size for the Training Set:
This information is not applicable as the device is not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable for the reasons stated above.
In summary: The provided 510(k) clearance letter details the regulatory submission for a surgical implant delivery system. The "performance data" section refers to a range of engineering, biological, and functional tests demonstrating that the physical device meets safety and effectiveness requirements and is substantially equivalent to a predicate. It does not contain the kind of diagnostic performance study information that would typically involve acceptance criteria for accuracy, ground truth establishment, or human reader performance for AI software.
Ask a specific question about this device
(32 days)
The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.
The Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK Optima material. The device features a single use driver, a threaded anchor, a PEEK-Optima eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures or one (1) bioresorbable reinforced implant. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.
The provided text is a 510(k) Summary for the Argo Knotless® GENESYS™ Anchor, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and study results in the format typically found in clinical trial reports for software devices.
Based on the information provided, here's what can be extracted and what is not present:
Missing Information:
The document does not describe acceptance criteria in the typical statistical sense (e.g., specific thresholds for sensitivity, specificity, or other performance metrics). It also does not describe a study that proves the device meets these types of acceptance criteria. The performance data mentioned refers to mechanical testing and user validation, not a clinical study to establish analytical or clinical performance of an AI/software device.
Therefore, the following points cannot be answered from the provided text:
- A table of acceptance criteria and the reported device performance: Not provided in terms of statistical performance metrics.
- Sample sized used for the test set and the data provenance: Not applicable in the context of AI/software performance. The "test set" here refers to mechanical testing or cadaveric validation.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is a physical bone anchor, not an AI/software product that assists human readers.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used: Not applicable for an AI/software performance study. For the mechanical and user validation tests, the "ground truth" would be the physical properties or user feedback.
- The sample size for the training set: Not applicable as this is not an AI/software device.
- How the ground truth for the training set was established: Not applicable.
What is present in the document about performance and validation:
The document briefly mentions "performance data" in Section VII, but this refers to mechanical and user validation for a physical medical device.
1. Acceptance Criteria and Reported Device Performance (As implied for a physical device, not statistical AI performance):
While not explicitly stated as "acceptance criteria," the document indicates that the device's performance was evaluated to demonstrate substantial equivalence to its predicate. The implied acceptance criterion is that the new device performs as intended and similarly to the predicate device in terms of mechanical properties and user experience.
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Performance as intended (mechanical stability, insertion, pull-out) | Performance Testing (Insertion, Cyclic and Ultimate Pull Displacement Resistance) completed. |
| User validation for surgical procedures | User Validation (Cadaver Medial Collateral Ligament/Lateral Ligament) completed. |
| Packaging and labeling suitability | Packaging and Labeling User Validation completed. |
2. Study that proves the device meets the (implied) acceptance criteria:
The study refers to a variety of tests and validations performed:
- Performance Testing: This would typically involve laboratory tests to measure parameters like insertion force, resistance to cyclic loading, and ultimate pull-out strength. The document doesn't provide specific quantitative results but states these tests were "completed."
- User Validation (Cadaver Medial Collateral Ligament/Lateral Ligament): This indicates hands-on testing of the device's usability and effectiveness in a cadaveric model, simulating surgical conditions for the intended use (reattaching soft tissue to bone).
- Packaging and Labeling User Validation: This would involve confirming that the packaging maintains sterility and integrity, and that the labeling is clear and accurate for users.
Data Provenance and Sample Size for Test Set:
- Sample Size: The specific sample sizes for these tests (e.g., number of anchors tested, number of cadavers, number of users for validation) are not provided in this summary.
- Data Provenance: Not specified, but generally, such performance and cadaveric testing would be conducted in a laboratory or simulated surgical environment. It is not "data" in the sense of patient records from a specific country.
In summary, the provided document is a regulatory submission for a physical medical device. It does not contain the specific information requested about acceptance criteria and study design as it pertains to the performance validation of an AI/software device.
Ask a specific question about this device
(125 days)
BioBrace® is intended for use in surgical procedures for reinforcement of soft tissue where weakness exists. BioBrace® is also intended for reinforcement of soft tissues that are repaired by suture or other fixation devices during tendon and ligament repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, spring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extraarticular ligaments. BioBrace® is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, quadriceps tendon, medial collateral ligament, lateral collateral ligament, soring ligament, deltoid ligament, ulnar collateral ligament or other tendons or extra-articular ligaments. Sutures, used to repair the tear, and sutures or other fixation devices, used to attach the tissue to the bone, provide mechanical strength for the repair.
The BioBrace® implant is a bioresorbable and bioinductive scaffold composed of a highly-porous collagen sponge made from insoluble bovine tendon type-1 collagen, and reinforced with poly- L-lactic-acid (PLLA) multifilament yarn (75 denier, 15 micron filament diameter). The BioBrace implant is 80% porous, average density of 0.2 grams/cm3, and median pore diameter of 19 microns. The highly porous collagen sponge comprises the majority of implant surface area (0.7 m2/gram) versus the PLA filaments alone (0.2 m2/gram), creating a large biologic matrix for cellular ingrowth, tissue regeneration, and healing. BioBrace implants are approximately 3 mm thick, provided in multiple sizes, and provide for soft tissue and tendon augment, and clinically relevant strengthening of the surgical repair. The BioBrace implant is single-use and supplied sterile with SAL of 10-6.
This document is a 510(k) Summary for the BioBrace® surgical mesh. It details the device, its intended use, and comparative studies for substantial equivalence. However, it does NOT contain the specific information required to answer your prompt about acceptance criteria and the study that proves the device meets those criteria.
The 510(k) summary focuses on demonstrating "substantial equivalence" to predicate devices, rather than establishing specific performance acceptance criteria for a new, AI-based device and then proving them. The prompt asks about an AI/algorithm-based device and its performance metrics, such as sensitivity, specificity, or reader improvement. This document describes a surgical mesh, which is a physical implant, not an AI or software device.
Therefore, I cannot provide the requested information based on this document. The document primarily discusses:
- Device Description: Bioresorbable and bioinductive scaffold made of collagen and PLLA yarn.
- Intended Use/Indications for Use: Reinforcement of soft tissue, particularly in tendon and ligament repair, including expansion to extra-articular ligaments.
- Technological Comparison: Emphasizes that the subject device is identical to the primary predicate (K203267) in design, materials, manufacturing, and sterilization, with new sizes and expanded indications for use being the primary changes.
- Non-Clinical and/or Clinical Tests Summary & Conclusions: Mentions cadaver-based studies, an in-vivo study in sheep, and literature reviews to support the expanded indications and performance characteristics like bioinductivity, strengthening, and stiffness.
There are no details related to an AI/algorithm device, its acceptance criteria (like sensitivity, specificity, AUC), test set characteristics (sample size, data provenance), ground truth establishment by experts, or MRMC studies, as would be expected for an AI device submission.
Ask a specific question about this device
(201 days)
The CONMED PlumeSafe® X5™ Smoke Management System is designed to remove and filter smoke and aerosols from a surgical site produced during electrosurgical and laser procedures.
The PlumeSafe® X5™ Smoke Management System is a portable, stand-alone, capital unit, that when paired with a reusable filter and disposable accessories makes up a Smoke Management System. The smoke unit is accompanied by a reusable filter and is available in two different configurations, a 22mm or 25mm size. Each filter is designed to capture components of surgical smoke and return clean air to the environment and can connect to different accessories which include pencils, adapters, fluid traps, and tube sets. Each filter is a seff-contained device that is completely enclosed to protect health care personnels from potential contamination during filter changes. Each filter is comprised of filtration (gross particulate pre-filter, Ultra-Low Penetration Air (ULPA), coconutbased carbon, impregnated alumina, and carbon dust capture media) that produce capture efficiency ratings equal to or better than 99.999% for particles ranging in size from 0.01μm to 0.2μm.
PlumeSafe® X5™ is comprised of a vacuum motor, aluminum, steel, and plastic components combined with sound reducing insulation. The vacuum motor is used to draw the surgical site, through an accessory, and into the PlumeSafe® X5™ filter. The PlumeSafe® X5™ Smoke Management System provides multiple modes (open, laparoscopic, and pencil) and settings to adjust the flow to the clinical need.
Here's a breakdown of the acceptance criteria and study information for the CONMED PlumeSafe® X5™ Smoke Management System, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Test Performed) | Reported Device Performance (Result) | Applicable Standards |
|---|---|---|
| Sound | Verify sound pressure and noise level characteristics per ISO Standards. | Pass |
| Initial Filter Efficiency | Verify filter configurations maintained the ULPA efficiency rating to remove 99.999% of the most penetrating particle sizes (MPPS) at set flow rate and concentration factor per ISO Standards. This criteria is met if the reported filter efficiency is 99.9997% at 0.01 to 0.20 micron particle size. | Pass |
| Final Filter Efficiency | Verify filter configurations maintained the ULPA efficiency rating to remove 99.999% of the most penetrating particle sizes (MPPS) at set flow rate and concentration factor over a period per ISO Standards. | Pass |
| Continuous Use | Verify continuous operation and continuous surgical plume evacuation. | Pass |
| Intermittent Use | Verify the device runs intermittently over a specified amount of time without the motor stopping or overheating at a set airflow. | Pass |
| Motor Ramp-up | Verify the device can reach 90% of maximum suction within 2 seconds of activation when used with a surgical pencil. | Pass |
| Detection | Verify the device detects occlusion. | Pass |
| Reliability | Verify lifetime of motor at 350 hours. | Pass |
| Transportation | Verify the Filter and Console packaging configurations per ASTM D4169. | Pass |
| Flow | Verify airflow is maintained in Open, Pencil, and Laparoscopic modes. | Pass |
| Software verification and validation | Software testing completed per IEC 62304. | Pass |
| Electrical safety | Electrical Safety testing completed per IEC 60601-1 | Pass |
| Electromagnetic Compatibility (EMC and Immunity) | Electromagnetic Capability testing completed per IEC 60601-1-2. | Pass |
2. Sample size used for the test set and the data provenance:
- The document describes "benchtop testing" and "non-clinical performance testing." These are physical device tests, not tests on a clinical dataset.
- Therefore, there is no "test set" in the sense of a collection of patient data or images. The "samples" would be the physical devices, filters, and components themselves, tested under specific laboratory conditions.
- The provenance of data is not applicable in the context of clinical data, as this is solely benchtop testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable. "Ground truth" in the context of expert review typically refers to a clinical diagnosis or labeling of medical images/data. Since this is benchtop device testing, there is no "ground truth" established by clinical experts for a test set. The acceptance criteria are based on engineering specifications and international standards, and the "ground truth" is measured by laboratory equipment and procedures.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This question is not applicable. Adjudication methods are used for resolving discrepancies in expert interpretations of clinical data. For benchtop testing, results are typically objective measurements against predefined thresholds, not subjective interpretations requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a smoke management system, not an AI-powered diagnostic tool. The testing performed is related to the physical performance and safety of the device in filtering smoke and aerosols.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- This question is not applicable. The device is a physical smoke management system, not an algorithm or software that operates standalone for diagnostic or interpretive purposes. Its performance is measured independently of human interaction in terms of its ability to filter air, maintain flow, etc., but this is inherent to the device's function, not "standalone algorithm performance."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the benchtop tests, the "ground truth" is established by adherence to recognized international standards (e.g., ISO, IEC, ASTM) and engineering specifications. The performance metrics (e.g., efficiency, flow rates, sound levels, motor lifetime) are objectively measured using calibrated laboratory equipment. There is no clinical "ground truth" in the form of expert consensus, pathology, or outcomes data for this type of device testing.
8. The sample size for the training set:
- This question is not applicable. The device is a hardware product, not a machine learning model; therefore, there is no "training set" in the context of artificial intelligence.
9. How the ground truth for the training set was established:
- This question is not applicable, as there is no training set for this device.
Ask a specific question about this device
(28 days)
Ask a specific question about this device
(53 days)
Argo Knotless® GENESYS™ SP Anchor
The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ SP Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaqed soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.
Argo Knotless® GENESYS™ Anchor
The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.
Argo Knotless® GENESYS ™ SP Anchor: The Self-Punching (SP) Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied for single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the broaching tip (suture eyelet) is manufactured of titanium material. Each size features a single-use driver, the threaded anchor, a Titanium suture eyelet, a UHMWPE 1.0mm non-absorbable retention suture, and loader tab. The retention suture holds the titanium eyelet in place on the driver assembly and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures.
Argo Knotless® GENESYS ™ Anchor: The Non-Self-Punching Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The anchor configuration requires a pre-prepared bone hole. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK material. Each size features a single use driver, a threaded anchor, a PEEK suture eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.
Argo Knotless® GENESYS ™ Anchor: The Preloaded Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK material. The eyelet is held on with a #2 UHMWPE, nonabsorbable retention suture and supplied with either a white/black or blue/blue colored non-absorbable Hi-Fi suture tape, a retention suture, and loader tab. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.
Principle of Operation - soft tissue to bone fixation
This document is an FDA 510(k) clearance letter and summary for a medical device (Argo Knotless GENESYS Anchor). It does not describe a study involving an AI/software as a medical device (SaMD) or diagnostic imaging. Therefore, the requested information about acceptance criteria, study design, ground truth establishment, and human-in-the-loop performance is not applicable to this document.
The document focuses on demonstrating substantial equivalence of a physical medical device (suture anchor) to a previously cleared predicate device. The "Performance Data" section lists various engineering and material tests conducted on the physical device, not clinical performance or diagnostic accuracy studies.
Here's why the questions are not applicable:
- No AI/Software Component: The device described is a physical implantable suture anchor, not an AI or software-based diagnostic tool.
- No Diagnostic Performance: The performance data mentioned (e.g., "Performance Testing," "Magnetic Resonance," "Transportation," "Corrosion Susceptibility") relate to the mechanical, material, and sterility aspects of the anchor, not its diagnostic accuracy in interpreting images or data.
- No Human Reader Interaction: There's no scenario where human readers (like radiologists) would "improve with AI vs. without AI assistance" as this is not an imaging or diagnostic AI product.
- No Ground Truth for AI: Ground truth for a physical device like a suture anchor is typically established through engineering specifications, material properties, and in-vitro or in-vivo (animal or human cadaver) biomechanical testing, not through expert consensus on images or pathology.
Therefore, I cannot populate the table or answer the specific questions related to AI/SaMD performance criteria, study design, and ground truth. The provided text simply doesn't contain that type of information.
Ask a specific question about this device
(132 days)
CONMED CleanSeal Vessel Sealers are intended to be used in open and mimimally invasive procedures for the ligation and division of tissue bundles, lymphatics, and vessels up to and including 6 mm vessels (arteries, veins, pulmonary arteries, pulmonary veins). The handpiece is indicated for use in gynecological and general surgical procedures, including urologic, thoracic, and vascular. These procedures include, but are not limited to, hysterectomies, colectomies, Nissen fundoplication, and adhesiolysis. CleanSeal vessel sealer has not been shown to be effective for ubal sterilization or tubal coagulation for sterilization procedures. Do not use CleanSeal vessel sealer for these procedures.
CleanSeal Advanced Bipolar Vessel Sealer Maryland is designed to be used, with a compatible MEP-1 electrosurqical generator, in open and minimally invasive procedures for the ligation and division of tissue bundles, lymphatics, and vessels up to and including 6 mm arteries and veins. They are sterile, single use bipolar vessel sealing hand pieces featuring double action non-stick coated jaws, three shaft length options, a one-hand controlled handle with latchable lever. The instrument creates a seal by application of radio frequency (RF) electrosurgical energy to vascular structures or tissue bundles interposed between the jaws.
The provided text describes a 510(k) premarket notification for a medical device called "CleanSeal Advanced Bipolar Vessel Sealer Maryland." This document is a regulatory submission to the FDA, demonstrating the device's substantial equivalence to previously cleared predicate devices.
However, the information provided does not contain any details about acceptance criteria, study design for a test set, sample sizes for training or test sets, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, or ground truth establishment relevant to AI/ML device performance.
The document primarily focuses on:
- Device Description and Intended Use: Explains what the device is, how it works, and its surgical applications for sealing and dividing tissue and vessels.
- Performance Testing: Mentions that the device has been tested according to FDA guidance for electrosurgical devices.
- Biocompatibility Testing: States that biocompatibility testing was conducted and met acceptance criteria.
- Chronic Animal Studies: Describes two chronic animal studies (porcine and canine models) comparing the performance of the CleanSeal Maryland with its predicate devices for effective vessel sealing. These are biological/physiological studies, not studies of an AI/ML algorithm's performance.
- Shelf Life Study: Indicates a shelf-life study was performed.
- Substantial Equivalence Argument: Compares the new device's features (Model numbers, intended use, polarity, dimensional design, rated voltage, shelf-life, sterilization method, and disposability) to its predicate and reference devices to argue for substantial equivalence.
Therefore, based solely on the provided text, it is not possible to describe the acceptance criteria and the study that proves the device meets those criteria in the context of an AI/ML device. The document refers to a physical electrosurgical device, and the "performance testing" and "chronic animal studies" relate to its mechanical and biological effectiveness, not an AI/ML algorithm's analytical performance on data.
If this device were an AI/ML device, the required information (acceptance criteria for AI metrics, test set details, ground truth, etc.) would be explicitly mentioned in sections describing its software performance and validation. Since those details are absent, it confirms that this is not an AI/ML device submission from which such information could be extracted.
Ask a specific question about this device