    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CONE-TROL Liquid Chemistry Control is a human liquid control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.

    Device Description

    CONE-TROL Liquid Chemistry Control is prepared from human serum to which purified biochemical material (human and animal origin), chemicals, drugs, preservatives, and stabilizers have been added. The control is provided in liquid form for user convenience.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined acceptance criteria for performance metrics. However, we can infer the implicit "acceptance criteria" and how the device's performance aligns with them based on the stability testing described. The primary "performance" being evaluated here is the stability of the control material over time under various storage conditions.

    Acceptance Criteria (Inferred)Reported Device Performance (CONE-TROL Liquid Chemistry Control)
    Closed Vial Stability (2-8°C): Control material remains stable for at least 30 days at 2-8°C (matching predicate).Acceptable stability through 33 days at 2-8°C (real-time testing).
    Open Vial Stability (2-8°C): Control material remains stable for at least 14 days at 2-8°C after opening (matching predicate).Acceptable stability through 15 days at 2-8°C after opening and simulating handling (real-time testing).
    Shelf Life (Frozen <-20°C): Control material remains stable until expiration date at <-20°C (matching predicate).Acceptable stability through 254 days at <-20°C (real-time testing). Real-time studies are ongoing to support the shelf life.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Value Assignment: 6 randomly selected vials from each manufacturing lot were assayed in duplicate across 2 or more instruments for each Abaxis Xpress panel. "Endogenous analytes" were value assigned using Kamiya and Sekisui reagents, but the specific number of vials/measurements for this is not explicitly stated, although it implies a similar process.
      • Performance (Stability) Testing: The document does not specify the exact number of vials or lots used for the real-time stability studies. It mentions "CONE-TROL Liquid Chemistry Control was stored..." (singular phrasing), which could imply a limited number of samples, but it doesn't give a concrete count.
    • Data Provenance: The studies were conducted by Cone Bioproducts, so the data is likely generated within the United States. The studies are described as "real-time stability testing," indicating they are prospective in nature, as the control material was monitored over time under specified conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (liquid chemistry control) does not typically involve expert review for "ground truth" in the same way an imaging or diagnostic AI device would. The "ground truth" for a chemistry control is established through:

    • Assigned Values: These are determined by assaying the control material on validated instruments (Abaxis Piccolo Express Analyzer, Roche Cobas Mira Plus, Mindray BS-200) using established methods and reagents.
    • Stability: This is determined by comparing measured values at different time points to initial values, adhering to established analytical chemistry principles.

    Therefore, there were no "experts" in the traditional sense involved in establishing a ground truth for interpretation based on subjective judgment. The "truth" is derived from analytical measurements performed by trained technicians or laboratory personnel.

    4. Adjudication Method for the Test Set

    Not applicable. As explained above, the "ground truth" for this type of device is based on quantifiable analytical measurements and stability parameters, not on subjective interpretations requiring adjudication among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a medical device for quality control in chemistry testing, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a chemical quality control material, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" used for this device is based on analytical measurements and stability data against predetermined ranges or expectations for various analytes as measured by laboratory equipment. It ensures that the control material maintains its integrity and assigned values over time under specified conditions.

    8. The Sample Size for the Training Set

    The concept of a "training set" is not applicable to this device. This is a manufactured product (chemistry control) whose performance is evaluated against established analytical methods and specifications, not an algorithm that learns from data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121534

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-20

    (27 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CONE-TROL Hemoglobin A1c Control Set is intended for use as quality control material to monitor the performance and precision of Hemoglobin A 1 c determination methods.

    Device Description

    The CONE-TROL Hemoglobin A 1c Control Set is prepared from human blood to which chemicals, preservatives, and stabilizers are added. The control is provided in liquid form for user convenience.

    AI/ML Overview

    The provided documentation describes the acceptance criteria and study for the CONE-TROL Hemoglobin A1c Control Set, a quality control material, primarily focusing on its stability. It is not an AI/ML device, so many of the requested fields (e.g., MRMC studies, human reader improvement, multi-reader consensus) are not applicable.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criterion for the CONE-TROL Hemoglobin A1c Control Set is that the analyte activity (concentration) must fall within ±10% of the initial value at various storage conditions and time points.

    AnalyteAcceptance Criteria (Concentration Range Accepted by Company)Initial ValueReported Performance (Concentration/Percent Change) at different conditions
    Accelerated Closed Vial Stability (37°C, 30 hours)
    HbA1c-Level 14.2-6.4% (implies ±10% around 5.3%)5.3%5.4% (1.89% change)
    HbA1c-Level 28.6-13.0% (implies ±10% around 10.8%)10.8%10.8% (0% change)
    Real Time Closed Vial Stability (-20°C, 210 days)
    HbA1c-Level 14.2-6.4% (implies ±10% around 5.3%)5.3%5.4% (1.89% change)
    HbA1c-Level 29.5-14.3% (implies ±10% around 11.9%)11.9%12.1% (1.68% change)
    Open Vial Stability (2-8°C, 210 days)
    HbA1c-Level 14.2-6.4% (implies ±10% around 5.3%)5.3%5.0% (5.7% change)
    HbA1c-Level 29.5-14.3% (implies ±10% around 11.9%)11.9%11.5% (3.4% change)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • For stability studies, the "sample" typically refers to the number of vials or batches tested. The document mentions two vials were assayed in triplicate for value assignment, and data points were collected over several days (minimum 6 data points per level for value assignment). For stability, specific numbers of individual vials tested at each time point are not explicitly stated, but it follows a standard protocol of testing representative aliquots from the prepared control sets.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the studies were conducted by the manufacturer, Cone Bioproducts, located in Seguin, TX, USA, for regulatory submission to the FDA. The stability studies are prospective, as they track the product over time.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This device is a quality control material whose "truth" is its chemical concentration, determined by analytical instrumentation. It does not involve human interpretation or expert opinions to establish ground truth in the way an AI diagnostic device would.

    4. Adjudication Method for the Test Set

    Not applicable. Ground truth for this chemical control is established through objective analytical measurements, not through adjudication of expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is not an AI/ML device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense, the performance of the control material is assessed in a "standalone" manner. Its stability and value assignment are determined purely through analytical measurements on laboratory instruments (Tosoh G7 HPLC Analyzer). There is no "human-in-the-loop" component in determining the performance of the control material itself; it's an objective measurement of its chemical properties over time.

    7. The Type of Ground Truth Used

    The ground truth used is the measured concentration of HbA1c in the control material, determined by objective analytical methods using the Tosoh G7 HPLC Analyzer. This is a form of measurement data/analytical truth.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, there is no training set for this type of device. The "ground truth" for the control material's initial value (value assignment) is established by assaying Levels 1 and 2 in triplicate on the applicable instrument system, with a minimum of 6 data points for each level. The mean, standard deviation, and coefficient of variance are then determined.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052010

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-10-04

    (71 days)

    Product Code
    Regulation Number
    862.1660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hemoglobin Alc Linearity is intended for use as quality control material to demonstrate linearity throughout the reportable range of Hemoglobin A1c (HbA1c%) for Immunoassay and HPLC test methods using protocols established in individual laboratories.

    Device Description

    The HbAIc Linearity Set is prepared from human blood to which stabilizers are added. The control is provided in liquid form for user convenience.

    AI/ML Overview

    This submission describes a medical device, the Cone Bioproducts' HbA1c Linearity Set, which is a quality control material. As such, the concept of "acceptance criteria" and "device performance" in the context of clinical efficacy (as with an AI diagnostic tool) does not directly apply. The submission focuses on demonstrating substantial equivalence to a predicate device for its intended use as a quality control material to verify linearity for HbA1c testing.

    Therefore, many of the requested items (e.g., sample size for test set, ground truth experts, MRMC study, standalone performance) are not relevant to this type of device and its regulatory submission.

    Instead, the "acceptance criteria" here are implicitly defined by the demonstration of substantial equivalence to the predicate device, LiniCAL Enzyme Calibration Verifiers (K040535), based on comparison of intended use and physical properties.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    As this is a quality control material seeking substantial equivalence, the "acceptance criteria" are not based on diagnostic performance metrics but rather on characteristics that establish its equivalence to a predicate. The "reported device performance" is essentially the detailed description of the subject device's characteristics compared to the predicate.

    CharacteristicAcceptance Criteria (Implicitly from Predicate)Reported Device Performance (Subject Device)
    Intended UseQuality control material to verify calibration/linearity of laboratory assays.Hemoglobin A1c Linearity is intended for use as quality control material to demonstrate linearity throughout the reportable range of Hemoglobin A1c (HbA1c%) for Immunoassay and HPLC test methods.
    AnalyteMultiple analytes (Alkaline Phosphatase, Alanine Aminotransferase, etc.)Single Constituent: HbA1c
    Methodology/ AnalyzersBeckman Coulter SynchronCompatible with Immunoassay and HPLC HbA1c test methods.
    MatrixHuman and Bovine SerumHuman Blood
    Control FormLiquidLiquid
    Levels5 levels (A-E)4 levels (1-4)
    Storage2°C to 8°C-20°C
    Unopened Vial Stability3 Years (at 2-8°C)2 Years (at -20°C)
    Opened Vial Stability14 days (at 2-8°C)14 days (at 2-8°C)

    Notes on the table: The "acceptance criteria" column is derived from the characteristics of the predicate device. The submission argues that even with differences (e.g., single vs. multiple analytes, specific methodologies), the subject device is substantially equivalent for its stated intended use (HbA1c linearity) as a quality control material.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and not provided in the submission. The submission is for a quality control material, not a diagnostic or screening device that would typically involve a "test set" of patient data. The demonstration of equivalence relies on comparing product characteristics and intended use, not on a clinical performance study using patient samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided in the submission. There is no "test set" or "ground truth" establishment in the context of clinical interpretation by experts for this type of device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided in the submission. There is no "test set" requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided in the submission. This is a quality control material, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided in the submission. This is a quality control material, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not applicable and not provided in the submission. For a quality control material, the "truth" is established by the known concentration or characteristic of the control material itself, and its ability to demonstrate linearity is assessed through its performance on a laboratory analyzer, not by comparison to external clinical ground truth.

    8. The sample size for the training set

    This information is not applicable and not provided in the submission. The device is a manufactured product (quality control material), not a machine learning algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable and not provided in the submission. As above, there is no "training set" for this type of device.