Found 2 results

510(k) Data Aggregation

    K Number
    K061516

    Validate with FDA (Live)

    Date Cleared
    2007-02-06

    (250 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Staodyn® Max Preset Transcutaneous Electrical Nerve Stimulator Device is used for the symptomatic relief and management of chronic intractable pain and relief of pain associated with arthritis. It is also used as an adjunctive treatment in the management of post-surgical and post-traumatic pain.

    Device Description

    The Staodyn® Max Preset is a Transcutaneous Electrical Nerve Stimulator (TENS) that transmits small electrical pulses through the skin to the underlying peripheral nerves. It is battery operated using 3-AAA batteries and includes two controllable output channels. It generates and delivers electrical impulses whose duration, rate and modulation can be altered by selection of 24 preset programs. The amplitude of the impulses can be increased or decreased with the use of two intensity toggle buttons, one for each output channel. The device has an on/off button that enables the device to be turned on off. In addition, the on/off button can be used to pause a treatment program during use. The stimulator is housed in a molded portable plastic case with a viewable LCD display, accessible keypad, and accessible battery storage compartment.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Staodyn® Max Preset TENS Device, focusing on acceptance criteria and study information:

    Key Takeaway: The provided document explicitly states that clinical testing was not required to demonstrate substantial equivalence for the Staodyn® Max Preset TENS Device. Therefore, many of the requested elements related to clinical study design and outcomes are not present.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of the submission (510(k) for a TENS device) and the explicit statement that clinical testing was not required, there isn't a table of specific clinical acceptance criteria with corresponding device performance metrics in the document. Instead, the "acceptance criteria" are implied by the non-clinical testing and the comparison to an already marketed predicate device.

    Acceptance Criteria (Implied by Equivalence)Reported Device Performance
    Non-clinical Electrical Performance:
    - Symmetric biphasic pulseVirtually identical to predicate: Symmetric biphasic pulse
    - Peak voltage up to 50 Volts on 500 ohms loadUp to 50 Volts on 500 ohms load (matches predicate)
    - User-selectable preset programs24 preset programs available
    - Adjustable intensity (1-20)Intensity adjustable from 1 to 20
    - Two controllable output channelsTwo controllable output channels
    - Lead open detectionHas lead open detection
    - Auto shut-off featureHas Auto shut-off feature
    Non-clinical Mechanical Performance:
    - Meets product specifications over operating and storage conditionsVerification and validation testing conducted to show device meets product specifications and user needs.
    Intended Use Equivalence:
    - Symptomatic relief and management of chronic intractable painStates "used for symptomatic relief and management of chronic intractable pain"
    - Relief of pain associated with arthritisStates "relief of pain associated with arthritis"
    - Adjunctive treatment in post-surgical and post-traumatic painStates "adjunctive treatment in the management of post-surgical and post-traumatic pain"
    Safety and Effectiveness:
    - Safe and effectiveNon-clinical testing demonstrates device is "safe, effective, and performs as well as the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable. No clinical test set data is reported. The non-clinical testing would have involved engineering samples and lab tests, not a patient test set in the clinical sense.
    • Data Provenance: Not applicable for clinical data. The non-clinical testing was performed by the manufacturer/OEMs.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not applicable. No clinical ground truth was established, as no clinical study was performed or required. The "ground truth" for the device's performance was its ability to meet engineering specifications and demonstrate equivalence to a predicate device through non-clinical means.

    4. Adjudication Method for the Test Set

    • Not applicable. No clinical test set, thus no adjudication method needed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a TENS stimulator, not an AI-assisted diagnostic or therapeutic tool that involves human readers interpreting data. No MRMC study was conducted or relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a medical device for pain relief, not an algorithm. Its performance is inherent to its physical and electrical characteristics.

    7. The Type of Ground Truth Used

    • Engineering Specifications and Equivalence to Predicate: The "ground truth" for this submission is that the device's electrical and mechanical characteristics meet its design specifications and are substantially equivalent to a legally marketed predicate TENS device. The intended use, function, and performance were compared to the American Imex Premier Plus - K032003.

    8. The Sample Size for the Training Set

    • Not applicable. As this is not an AI/ML device, there is no "training set" in the computational sense. The device's design is based on established TENS technology and the predicate device.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. No training set was used. The device's functionality is based on its hardware design and the established principles of transcutaneous electrical nerve stimulation, which have been validated through decades of use in this product category.
    Ask a Question

    Ask a specific question about this device

    K Number
    K050046

    Validate with FDA (Live)

    Date Cleared
    2005-04-13

    (93 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Interferential stimulation can be used for the following applications:

    • Relieve acute pain .
    • . Relieve and manage chronic pain
    • Relax muscle spasms .
    • Maintain and increase the range of motion .
    • Increase local blood circulation .

    Neuromuscular stimulation can be used for the following applications:

    • Relax muscle spasms .
    • Prevent or retard disuse atrophy .
    • Increase local blood circulation .
    • Re-educate muscles .
    • Maintain or increase the range of motion .
    • Prevent venous thrombosis using immediate postsurgical stimulation on calf . muscles

    Pulsed direct current stimulation can be used for the following applications:

    • Reduction of edema (under negative electrode) t
    • Reduction of muscle spasm .
    • Influencing local blood circulation (under negative electrode) .
    • Retardation or prevention of disuse atrophy .
    • Facilitation of voluntary motor function .
    • Maintenance or increase of range of motion .

    Interferential and Neuromuscular combination stimulation can be used for the following applications:

    • Reduction of edema (under negative electrode) .
    • Reduction of muscle spasm .
    Device Description

    The IF 3 Wave is a non-invasive medium/low frequency interferential muscle rie it -> wave is a intended for treating patients that have acute and chronic pain, edema, tight musculature, muscle spasms and muscular weakness due to disuse atrophy. The IF 3 Wave stimulator is a microprocessor controlled dual channel Interferential electro-stimulator with neuro-muscular and pulse direct current micronital expabilities. It will offer 4 separate electro-therapy stimulation modalities. Stimulation capacificies: 17 Will on P), neuromuscular electrical stimulation (NMES), r neso direct current (PDC) electrical stimulation, and a combination of interferential and neuromuscular electrical stimulation (IF/NMES).

    The IF 3 Wave will be capable of storing device settings, use, and compliance data for up to 90 days. It will be capable of down loading the above compliance date through a up to 90 days. which is contained within the battery charger base. Once per phone mic modelin, will be instructed to download the stored data from their device, month, the patient will over compliance report which is forwarded to their physicians for which will ereate a paysician will review the report and provide adjustment to the patient treatment regimen and device setting accordingly.

    Accessories provided with the IF 3 Wave include lead wires, electrode pads, Accessories proTracu wik, battery charger/modem, phone cable, power supply, and a users manual.

    AI/ML Overview

    This K510(k) submission describes the IF 3Wave Interferential Muscle Stimulator System, Model 7110S. The submission focuses on establishing substantial equivalence to previously cleared predicate devices rather than reporting on a new clinical study with acceptance criteria.

    Therefore, many of the requested sections related to acceptance criteria and study data are not applicable to this 510(k) submission.

    Here's an analysis based on the provided text, addressing the applicable points:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a 510(k) summary demonstrating substantial equivalence, there are no explicit "acceptance criteria" in the sense of predefined thresholds for clinical or diagnostic performance metrics. Instead, the device's technical characteristics are compared to those of its predicate devices to demonstrate equivalence. The "reported device performance" is essentially the listed technical specifications of the IF 3Wave.

    CharacteristicPredicate Device (IF-II Model 7100S)IF 3Wave (Model 7110S)Predicate Device (EMS+2 Model 6840S)IF 3Wave (Model 7110S)Predicate Device (SporTX Model 7700S)IF 3Wave (Model 7110S)
    Interferential Modality
    Patient Leakage Current (Normal)Unknown< 0.1 microamperesUnknown< 0.1 microamperesUnknown< 0.1 microamperes
    Patient Leakage Current (Single Fault)Unknown2.4 microamperesUnknown2.4 microamperesUnknown2.4 microamperes
    Output Modes131313
    Automatic Overload TripNoYes, 5 kohmsNoSameNoSame
    Continuous Treatment Time Selection20 minutes, fixed10 - 60 min, selectable (5 min steps)15, 30, or 60 min, or Continuous, selectable10 - 60 min, selectable (5 min increments)38 min or Continuous, selectable30 - 60 min, selectable (5 min increments)
    Number of Variable Sweep Programs34N/AN/AN/AN/A
    Number of Fixed Parameter Sweep Programs07N/AN/AN/AN/A
    Max Charge per Pulse120 microcoulombsSame30 microcoulombsSame6 microcoulombsSame
    Average DC CurrentN/AN/AN/AN/A266 microamperes300 microamperes
    Compliance with Voluntary StandardsUnknownYesUnknownYesUnknownYes
    NMES Modality(see above)
    Output ModesN/AN/A13N/AN/A
    Pulse WidthN/AN/A50 - 300 us, adjustable70 - 300 us (10 us steps)N/AN/A
    Pulse FrequencyN/AN/A4 - 80 Hz, adjustable10 – 80 Hz (5 Hz steps)N/AN/A
    Pulse AmplitudeN/AN/A0 – 100 mA (1000 Ohm load), adjustable0 - 100 mA (1000 Ohm load) (1 mA steps)N/AN/A
    Cycle ON TimeN/AN/A0.5 - 30 seconds, adjustable1 - 30 seconds (1 second increments)N/AN/A
    Cycle OFF TimeN/AN/A0.1 - 60 seconds, adjustable1 - 60 seconds (1 second increments)N/AN/A
    Ramp-Up TimeN/AN/A0.1 - 6, adjustable0 - 10 seconds (1 second increments)N/AN/A
    Ramp-Down TimeN/AN/A0.1 - 6 seconds, adjustable0 - 10 seconds (0.5 second increments)N/AN/A
    PDC Modality(see above)
    Output ModesN/AN/AN/AN/A13
    Number of Preset ProgramsN/AN/AN/AN/A03
    Pulse Width (Biphasic/Triphasic)N/AN/AN/AN/ABiphasic: 60 us / Triphasic: 60 usSame
    Pulse Frequency (Biphasic/Triphasic)N/AN/AN/AN/ABiphasic: 278 Hz / Triphasic: 222 HzSame
    Pulse AmplitudeN/AN/AN/AN/A0 - 100 mA (1000 Ohm Load), adjustable0 - 100 mA (1000 Ohm Load) (1 mA increments)

    Note: "Same" in the table indicates that the characteristic matches the predicate device. "N/A" means the characteristic is not applicable to that specific modality for the predicate device as presented in the provided text.

    The rationale for substantial equivalence is that the differences in the IF 3Wave (e.g., three output modes instead of one, automatic overload trip, wider range of treatment times and programming options, compliance with voluntary standards, lithium-ion battery) do not raise new questions of safety or effectiveness.

    2. Sample Size for the Test Set and Data Provenance:

    Not applicable. This is a 510(k) submission based on substantial equivalence to predicate devices, focusing on technical specifications and indications for use, not a clinical study with a "test set" of patients.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. No ground truth for a test set was established as no clinical study was performed for this 510(k) submission.

    4. Adjudication Method for the Test Set:

    Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. No MRMC study was conducted. This type of study demonstrates the clinical utility and impact of AI, which is not relevant for a device cleared through substantial equivalence to existing stimulators.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. This device is a powered muscle stimulator, not an AI algorithm. Its performance is evaluated based on its physical and electrical characteristics and comparison to predicate devices, not on algorithmic outputs.

    7. Type of Ground Truth Used:

    Not applicable. Instead of "ground truth," the basis for equivalence is the well-established safety and effectiveness of the predicate muscle stimulator devices. The "truth" is that these types of devices are already legally marketed and have known performance characteristics.

    8. Sample Size for the Training Set:

    Not applicable. There is no AI model or algorithm being trained as part of this submission. The device itself is a hardware product.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.