    K Number
    K122726

    Validate with FDA (Live)

    Device Name
    CPI RAD VISION
    Date Cleared
    2013-02-15

    (163 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CPI RAD VISION is a digital imaging system intended for the capture and display of radiographic images of human anatomy, as part of a diagnostic x-ray system. It is intended for use in general radiographic examinations and applications. The device is not intended for mammographic, fluoroscopic or angiographic applications.

    Device Description

    The CPI RAD VISION product will provide a user interface to control the digital imaging functions of a radiographic acquisition modality (acquisition, processing, storage, display, and distribution of images from a Solid State X-ray Imager (SSXI)).

    The CPI RAD VISION product will also provide either an option on the user interface to control the x-ray generator functions required by a radiographic acquisition modality or an optional hardware interface to allow the x-ray generator functions to be controlled by an independent third-party operator console while maintaining synchronization with the digital imaging functions of the product.

    CPI RAD VISION is a component of a complete radiographic x-ray system.

    AI/ML Overview

    The provided text is a 510(k) summary for the CPI RAD VISION device, a digital imaging system for radiography. The text states that the device's performance testing demonstrated its safety and effectiveness, and its equivalence to predicate devices, through "Electrical safety (IEC60601-1:2005) and Electromagnetic compatibility (IEC60601-1-2, Edition 3:2007-03), Software verification (IEC60601-1-4:2000), other performance testing and the acquisition of sample clinical images." However, the document does not include specific acceptance criteria, detailed device performance metrics, or the study specifics (like sample size, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone studies, or details on ground truth for training/test sets) requested in the prompt.

    Therefore, most of the requested information cannot be extracted from this document.

    1. A table of acceptance criteria and the reported device performance
    The document does not specify acceptance criteria or quantitative device performance metrics. It generally states that the device "demonstrated that CPI RAD VISION is safe and effective, and is equivalent to the aforementioned predicates devices" based on compliance with electrical safety, EMC, and software verification standards, and the acquisition of "sample clinical images."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not provided in the document. The document only mentions "acquisition of sample clinical images" without any details on sample size or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not provided in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not provided in the document. The device is described as a "digital imaging system," not explicitly an AI-assisted device for human reader improvement.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not explicitly stated or detailed in the document. The device is a system for image capture and display, implying human interaction for diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not provided in the document. The document mentions "acquisition of sample clinical images" but does not detail how ground truth was established for these images for evaluation purposes.

    8. The sample size for the training set
    Not provided in the document.

    9. How the ground truth for the training set was established
    Not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083224

    Validate with FDA (Live)

    Date Cleared
    2009-03-24

    (141 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPI RAD VISION is a full featured Radiographic Flat Panel Digital Imaging System for X-ray Generator and Acquisition of digital radiography. The CPI RAD VISION is configurable to any high resolution (3K x 3K) Solid State X-Ray Imager (SSXI) presently in the market. It is intended to replace conventional film screen systems.

    The CPI RAD VISION allows a qualified operator to perform digital radiographic examinations of various anatomic regions on both adult and pediatric patients. Anatomic regions of interest for diagnostic radiographic exposure include: skull, spinal column, chest, shoulder girdle, abdomen, pelvic girdle and extremities.

    The CPI RAD VISION enables a qualified operator to acquire, process, and display images with for the benefit of obtaining an optimal diagnostic product. The CPI RAD VISION system enables the qualified operator to store, hardcopy images with a laser printer or send images over a network. This device is not intended for mammographic, fluoroscopic and or angiographic applications. The CPI RAD VISION will not include the X-Ray system itself.

    Device Description

    The CPI RAD VISION is a component of a complete radiographic x-ray system. The CPI RAD VISION provides a single user interface for control of the x-ray generator and all digital imaging functions required by a radiographic acquisition modality (acquisition, processing, storage, display, and distribution of images).

    The CPI RAD VISION consists of Flat Panel Detector, ISO-BOX (Medical Grade Isolation Transformer), LCD Monitor, and a workstation (computer) with x-ray generator interface, image receptor interface, and network port for communications with supported DICOM devices.

    AI/ML Overview

    This document is a 510(k) summary for the CPI RAD VISION, a digital radiography system. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria based on performance metrics. Therefore, many of the requested details about a study and acceptance criteria are not present in the provided text.

    Based on the provided text, here's what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. The document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance studies against specific acceptance criteria. The "Conclusion drawn from comparison" section merely states that the CPI RAD VISION can be considered substantially equivalent to the CMT MEDICAL TECHNOLOGIES LTD. SMART RAD. There are no performance metrics or acceptance criteria reported.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. As explained above, the document does not detail a study with a test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC comparative effectiveness study was done. The device is described as a "Digital Radiography" system, primarily involving image acquisition, processing, storage, and display, and not an AI-assisted diagnostic tool that would typically involve human reader improvement metrics.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided. The device is an imaging system, not an algorithm being tested in isolation.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    This information is not provided.

    8. The sample size for the training set

    This information is not provided.

    9. How the ground truth for the training set was established

    This information is not provided.

    Summary of available information:

    The document serves as a 510(k) premarket notification for the CPI RAD VISION, a digital radiography system. Its primary purpose is to establish substantial equivalence to a
    predicate device (CMT MEDICAL TECHNOLOGIES LTD. SMART RAD 510(k) - K003438) based on its intended use and technological characteristics. The document does not contain specific acceptance criteria, detailed study designs, performance metrics, or information regarding ground truth establishment or expert involvement, as it is structured to fulfill the requirements of a 510(k) submission for substantial equivalence rather than a detailed clinical performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061173

    Validate with FDA (Live)

    Date Cleared
    2006-06-09

    (43 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPIVision Digital Imaging system is a high resolution, digital imaging system designed for digital videography. It is intended to replace conventional film techniques in multipurpose or dedicated applications when general fluoroscopy, interventional fluoroscopy or angiography or cardiac imaging procedures are performed.

    The CPIVision Digital Imaging system allows the operator to view and enhance digital fluoroscopic images. High resolution digital spot images may be acquired at single or rapid acquisition rates. Images may be viewed and enhanced enabling the operator to bring out diagnostic details difficult or impossible to see using conventional imaging techniques.

    The CPIVision Digital Imaging system enables the operator to hardcopy image with a laser printer or send images over a network. The major system includes: a fluoroscopic CCD camera/lens, monitors, and an image processor.

    Device Description

    The CPIVision Digital Imaging system allows the use of digital imaging to be applied to conventional X-ray system used in general fluoroscopy, interventional fluoroscopy, angiography and cardiac imaging areas. The system works by installing a CCD camera/lens on the output port of the image intensifier and digitizing the video output on the image intensifier. The digital image can be displayed on the monitor; it can be stored to the hard drive, or sent to an external device such as a laser imager or Network Storage Provider. The image can also be computer processed, including brightness and contrast, edge enhancement, zoom, and subtraction.

    The CPIVision Digital Imaging enables the operator to hardcopy image with a laser printer or send images over a network. The major system components are: a fluoroscopic CCD camera/lens, monitors, and an image processor.

    AI/ML Overview

    This document describes the CPIVision Digital Imaging System, a digital radiography device. However, this document does not contain explicit acceptance criteria or a study dedicated to proving the device meets specific performance metrics.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (INFIMED INC. ORION FLUOROSCOPIC IMAGING SYSTEM 510(k) – K012490) based on its device description and intended use. This is a common regulatory pathway for medical devices that are similar in technology and purpose to devices already on the market.

    Therefore, many of your specific questions cannot be answered from the provided text, as the focus is not on a direct performance study against a predefined set of acceptance criteria.

    Here's a breakdown of what can and cannot be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Acceptance Criteria: Not explicitly stated. The document implies that the "acceptance criteria" for regulatory approval were met by demonstrating substantial equivalence to the predicate device. This means its performance is considered acceptable if it performs comparably to the predicate for its intended use.
    • Reported Device Performance: Not directly reported in quantitative terms. The document describes the system's capabilities (e.g., high resolution, digital videography, image enhancement, rapid acquisition rates, hardcopying, networking) but does not provide specific performance metrics (e.g., spatial resolution in lp/mm, signal-to-noise ratio, frame rates under specific conditions, diagnostic accuracy for particular findings).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not applicable / Not provided. The document does not describe a performance study with a test set of images or patient data. The basis for comparison is the functional and intended use equivalence to a predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not applicable / Not provided. No specific test set or ground truth establishment process is described.

    4. Adjudication Method for the Test Set:

    • Not applicable / Not provided. No test set or adjudication process is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. The document does not describe an MRMC study. It focuses on the device's capabilities and its equivalence to an existing device, not on comparing human reader performance with and without the device.

    6. Standalone Performance Study:

    • No. The document does not describe a standalone (algorithm-only) performance study. The system described is hardware and software designed to process and display fluoroscopic images for human interpretation, not an AI algorithm making diagnoses independently.

    7. Type of Ground Truth Used:

    • Not applicable / Not provided. No specific ground truth is referenced as no performance study is described. The "ground truth" for regulatory approval in this context is that the device functions as intended and is as safe and effective as the predicate device.

    8. Sample Size for the Training Set:

    • Not applicable / Not provided. The CPIVision Digital Imaging system as described is not an AI/ML algorithm that requires a "training set" in the conventional sense. It's a digital imaging system replacing analog film techniques.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable / Not provided. As it's not an AI/ML algorithm requiring a training set, this question is not relevant to the provided text.

    Summary based on the provided text:

    The CPIVision Digital Imaging system sought 510(k) clearance by demonstrating substantial equivalence to the INFIMED INC. ORION FLUOROSCOPIC IMAGING SYSTEM (K012490). The acceptance criteria were implicitly met by showing the new device shares the same intended use and fundamentally similar technological characteristics to the predicate, ensuring it is as safe and effective. No specific performance study with a defined test set, ground truth, or statistical analysis against pre-specified acceptance metrics is described in this document.