Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K011855

    Validate with FDA (Live)

    Date Cleared
    2002-01-08

    (209 days)

    Product Code
    Regulation Number
    866.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wampole HS-CRP ELISA is an in vitro diagnostic enzyme-linked immunosorbent assay (ELISA) intended for the quantitative measurement of C-reactive protein (CRP) levels in human serum and plasma. CRP measurements aid in the evaluation of the extent of injury to body tissues.

    Device Description

    The Wampole™ HS-CRP ELISA is a Class II in vitro medical diagnostic device that utilizes conventional enzyme-linked immunosorbent assay technology to quantitatively measure the level of C-reactive protein (CRP) in a serum or plasma specimen. The assay kit consists of plastic microwell strips, containing adsorbed monoclonal anti-CRP antibody. CRP in the specimen is captured by the immobilized antibody, forming immobilized immune complexes. After washing away excess specimen, polyclonal anti-CRP antibody that has been conjugated to horseradish peroxidase is added, forming a "sandwich" of monoclonal antibody, CRP antigen, and conjugated polyclonal antibody. After washing, the immobilized CRP sandwich is exposed to tetremethlybenzidine (TMB), a chromogenic substrate that turns from clear to blue in the presence of peroxidase reaction products. The peroxidase reaction is stopped by the addition of dilute sulfuric acid, which also changes the color of the solution from blue to yellow. The optical density of the yellow end product is directly proportional to the amount of bound CRP antigen, and is quantitated on an ELISA reader at 450 nm. Specimen CRP levels are then determined by interpolation from a standard curve of optical density versus CRP concentration.

    AI/ML Overview

    The provided text does not contain explicit acceptance criteria tables or a detailed study description with all the requested information for the Wampole HS-CRP ELISA device. The document is primarily a 510(k) summary and an FDA clearance letter, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a full performance study report with specific acceptance criteria that the device must meet in a formal table with reported performance.

    However, based on the information provided, I can infer some aspects and highlight what is missing.

    Here's an attempt to answer your request given the limitations of the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not contain a table of acceptance criteria and reported device performance in the typical format you requested (e.g., analytical precision, accuracy, limit of detection, etc., against specific numerical targets). The document states the device has "the same technological features, characteristics and intended use as the predicate device," implying that its performance is expected to be comparable to or meet the standards established for the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not explicitly stated in the provided text. The document mentions "human serum and plasma" specimens, but not the number of samples used in any specific test set or their origin (country, retrospective/prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not explicitly stated in the provided text. For an in vitro diagnostic (IVD) like an ELISA, ground truth is typically established by reference methods or clinical diagnosis rather than expert consensus on images.

    4. Adjudication Method for the Test Set

    This information is not applicable or explicitly stated in the provided text, as the device is an ELISA for measuring CRP levels, not a diagnostic imaging device requiring expert adjudication of interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret images, often with and without AI assistance. This device is an in vitro diagnostic assay.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The Wampole HS-CRP ELISA is a standalone in vitro diagnostic assay. Its performance is evaluated biochemically and quantitatively, not as an algorithm assisting a human interpreter. Therefore, its performance is inherently "standalone" in the context of laboratory testing. The document focuses on the assay's ability to quantitatively measure CRP.

    7. The Type of Ground Truth Used

    For an ELISA like the Wampole HS-CRP ELISA, the "ground truth" would typically be established by:

    • Reference materials/calibrators: Known concentrations of CRP used to generate standard curves.
    • Comparison to established reference methods: Testing patient samples and comparing results to a gold standard CRP measurement method or a highly accurate predicate device.
    • Clinical outcomes/diagnosis: While not directly mentioned for establishing the "ground truth" for the assay's analytical performance, CRP levels themselves aid in the "evaluation of the extent of injury to body tissues," which relates to clinical outcomes.

    The document states "Specimen CRP levels are then determined by interpolation from a standard curve of optical density versus CRP concentration," indicating the use of reference materials/calibrators as part of establishing the measurement ground truth. The comparison to the predicate device (Dade Behring N High Sensitivity CRP - K991385) implies that the performance of the Wampole ELISA was benchmarked against an existing, legally marketed device that serves as a standard for CRP measurement.

    8. The Sample Size for the Training Set

    This information is not explicitly stated in the provided text. For ELISA assays, a "training set" in the context of machine learning isn't directly applicable. Instead, the assay is developed and optimized using a variety of samples and conditions to establish its analytical characteristics (e.g., linearity, precision, accuracy). The document does not specify the number of samples used during this development and validation phase.

    9. How the Ground Truth for the Training Set Was Established

    This information is not explicitly stated in the provided text, and the concept of a "training set" with ground truth as defined for AI/ML models is not directly applicable to a traditional ELISA. The development and optimization of an ELISA involve using known concentrations of purified analyte (CRP standards) and well-characterized human serum and plasma samples to ensure the assay performs as expected across its analytical range. These known concentrations and characterized samples serve as the basis for establishing the assay's analytical performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1