510(k) Data Aggregation

    K Number
    K251660

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-27

    (28 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T3® Pro Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

    The T3® Pro Implants may also utilize immediate loading for these indications. The T3® Pro Implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function.

    Device Description

    The T3® Pro Implants with the new ZimVie Implant Packaging System. The material and design of the subject device is identical to the primary predicate device T3® Pro Implants cleared in K213672.

    The ZimVie Implant Packaging System configuration, similarly to the primary predicate device, consists of a titanium sleeve that is inserted into a polypropylene inner tray, covered with a Tyvek lid and heat-sealed. This assembly is then placed in a heat-sealed outer tray, covered with a Tyvek® lid. The outer tray is placed inside a cardboard box, and gamma sterilized. The new ZimVie Packaging System has been standardized to align the packaging across all ZimVie implant families. The subject device in the new ZimVie Packaging System has a shelf-life of two (2) years.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated documentation for the "T3 Pro Implants- ZimVie Packaging System" (K251660) do not describe acceptance criteria or a study proving the device meets those criteria in the context of an AI/human reader performance study.

    Instead, this submission is for a medical device (endosseous dental implant) and its new packaging system. The "performance data" section explicitly states that no clinical data was included in the submission for the substantial equivalence determination. The focus of the provided document is to demonstrate equivalence to a previously cleared predicate device for the dental implant itself, with the primary change being the packaging system.

    Therefore, I cannot fulfill your request to create a table of acceptance criteria and reported device performance related to an AI-assisted or standalone AI system, nor can I provide details on sample size, expert adjudication, MRMC studies, or ground truth establishment, as these concepts are not discussed or relevant to the provided documentation for this specific 510(k) clearance.

    The information provided describes non-clinical performance data related to the physical device and its packaging:

    • Sterilization validation: Gamma irradiation sterilization to a sterility assurance level of 10^-6 according to ISO 13004 and ISO 11137.
    • Aging studies: Demonstrating a shelf life of two (2) years for the implants, with packaging sterile barrier testing per ASTM F1980.
    • Packaging studies: Simulated distribution and environmental conditioning per ISTA 3A and ASTM D4169.
    • Biological evaluation: Leveraged from the primary predicate device, demonstrating no increased risks regarding biocompatibility. The material is Commercially Pure Titanium conforming to ASTM F67.
    • MR Safety testing: Conducted according to ASTM F2503-13 on the primary predicate device (identical to the subject device) for MR Conditional determination.
    • Bacterial Endotoxin Testing: Routinely performed using LAL assay in accordance with ISO 11737-3.

    The acceptance criteria for this submission would primarily revolve around demonstrating that the new packaging system maintains the sterility, integrity, and safety of the dental implant for its stated shelf life, and that the dental implant itself remains identical to the previously cleared predicate device in all other aspects (materials, design, intended use, operating principle).

    In summary, your request pertains to an AI/Software as a Medical Device (SaMD) study, while the provided document describes the 510(k) clearance process for a physical dental implant device and its new packaging.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241753

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-23

    (127 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immediate Molar Dental Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed loading, or with a terminal or intermediary abutment for fixed or removable bridgework, and to retain overdentures.

    Immediate Molar Dental Implants may also utilize immediate loading for these indications. Immediate Molar Dental Implants are intended for immediate function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. The Immediate Molar Dental Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to provide good primary stability. The Immediate Molar Dental Implants are intended for implantation in the maxillary or mandibular molar region where bone exists and the surgeon has determined that the placement of a narrower diameter implant would increase the probability of failure due to poor primary stability, or increased surgical procedures leading to complications.

    The Immediate Molar Dental Implants are also indicated for compatibility with the following OEM abutment systems:

    Abutment System NameModelsPlatform Diameters
    Abutments manufactured byTerrats Medical SLTitanium Abutments4.1, 5.0 and 5.7mm
    Abutments manufactured byZfx GmbHTitanium Abutments4.1, 5.0 and 5.7mm
    Device Description

    The Immediate Molar Implants are basic screw-type designs available in tapered body geometry. The devices are manufactured from Commercially Pure Titanium (ASTM F67) and feature a roughened apex and traditional OSSEOTITE® coronal surface. The device is packaged in a Titanium sleeve that is inserted into a polypropylene inner tray, covered with a Tyvek lid and heat-sealed. This assembly is then placed inside a larger polyethylene thermoformed outer tray, covered with a Tyvek lid and heat-sealed. The outer tray is packaged inside a box. The device is sold sterile. The shelf life of the Immediate Molar Implants is 5 years and they are intended for single use only. The device is sterilized using gamma irradiation method. The implants are available in various platform options and feature an internal hex connection for mating with associated Biomet 3i internal connection restorative components and also a TSV connection for mating with associated Zimmer Dental TSV connection restorative components. The implants are also compatible with titanium abutments manufactured by Terrats Medical SL and Zfx GmbH. The implants are offered in a variety of diameters and lengths to accommodate varying patient anatomy. The T3 PRO Immediate Molar Implants are offered in 8, 10 and 11.5mm implant lengths for each of the implant body diameter sizes of 7, 8 and 9mm. The TSX Immediate Molar Implants are offered in 6, 8, 10 and 11.5mm implant lengths for each of the implant body diameter sizes of 7 and 8mm. The TSX Immediate Molar Implants are offered in 8, 10 and 11.5mm lengths for implant body diameter size of 9mm.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Immediate Molar Implants) and does not contain information about acceptance criteria or a study proving that an AI/software device meets acceptance criteria.

    The document explicitly states: "No clinical data were included in this submission." and focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics, non-clinical testing (fatigue, MR compatibility, pull-out tests, surface area analysis, sterilization, biocompatibility, shelf life), and contractual agreements for compatible components.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving a device meets acceptance criteria using clinical or performance data, as this information is not present in the provided text. The document describes a dental implant, not an AI or software-based device that would typically have acceptance criteria presented in the manner requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241032

    Validate with FDA (Live)

    Device Name
    BellaTek Bars
    Manufacturer
    Date Cleared
    2024-07-15

    (90 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BellaTek Bars are indicated for attachments in the treatment of partially or fully edentulous jaws for the purpose of restoring chewing function. The BellaTek Bars are intended for attachment to a minimum of two (2) abutments.

    All digitally designed BellaTek Bars are intended to be sent to Biomet 3i for manufacture.

    The BellaTek Bars are indicated for compatibility with the following abutment systems:

    · Universal Multi-Unit Abutments manufactured by Terrats Medical SL., 2.9-5.7mm, Angled, max 30 degrees

    • · ZimVie Eztetic Implant System
    • · ZimVie TSX Implant System
    • · ZimVie Tapered Screw-Vent System
    • · ZimVie Low Profile Abutments, 3.4-6.0mm, Angled, max 30 degrees
      • · ZimVie Osseotite Implant System
      • ZimVie T3 and T3 PRO Implant System
    • · ZimVie Tapered Abutments, 3.5-5.7mm, Angled, max 30 degrees
      • · ZimVie TSX Implant System
      • · ZimVie Tapered Screw-Vent Implant System .
    Device Description

    The BellaTek Bars are intended to disperse the load of a prosthesis across dental implant-abutment assemblies. They also provide support for prosthetic devices such as artificial teeth, and to restore the patient's chewing function. The subject devices are identical to the reference predicate device CAM StructSURE Precision Milled Bars cleared in K080864 (SE 07/21/2008) with the only addition of the new connection geometry. The subject device has a new bar-to-abutment connection tailored to the multi-unit restorative abutment component(s) with which the device is intended to be used. Similar to the primary predicate device (K233083), the BellaTek Bars are manufactured out of commercially pure titanium per ASTM F67 or Titanium Alloy per ASTM F136. The BellaTek Bars are placed in a nylon bag and sealed in same manner as the reference predicate device (K080864). A label is applied to the bag.

    Like the primary predicate device (K233083), the BellaTek Bars are offered in two types. A Type I Bar is a Titanium Alloy bar designed for use with removable overdentures. A Type II Bar is a Pure Titanium or Titanium Alloy bar designed for fixed prosthesis. All BellaTek Bars (Dolder, Primary, Hader, Hybrid, Freeform, Canada, Wraparound and Copy Milled) are designed to match an individual patient. The bars are designed from a three-dimensional optical and/or digital scanner system that scans the casting of a patient's impression and then is machined using a CAD/CAM software system in accordance with customer specifications, within the limits of design defined in tables below. The subject devices are provided non-sterile in same manner as the predicate devices.

    AI/ML Overview

    The BellaTek Bars are dental implant abutments intended to restore chewing function in partially or fully edentulous jaws by attaching to a minimum of two abutments.

    Here's an analysis of the acceptance criteria and supporting study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the acceptance criteria are primarily aligned with demonstrating substantial equivalence to a predicate device through materials, design specifications, and performance bench testing. Specific quantitative performance targets are not explicitly listed in an "acceptance criteria" table format in the provided document, but rather implied through comparison to existing standards and predicate devices.

    Acceptance Criteria CategoryDescription and Reported Device Performance
    Intended UseSubstantially equivalent to the primary predicate device (K233083). – BellaTek Bars: Indicated for attachment to dental abutments in the treatment of partially or fully edentulous jaws for restoring chewing function. Intended for attachment to a minimum of two (2) abutments and manufactured by Biomet 3i. – Predicate Device (Osteon Precision Milled Suprastructure K233083): Indicated for attachment to dental abutments in the treatment of partially or fully edentulous jaws for restoring chewing function. Intended for attachment to a minimum of two (2) abutments and manufactured by Osteon.
    Design Principles & Technological CharacteristicsSubstantially equivalent to the primary predicate device (K233083) and reference predicate device (K080864). - Similar features: bar base material, bar-to-abutment connection, prosthetic platform sizes, mode of prosthetic retention. - Differences in prosthetic platform connection geometry are supported by identified predicate devices.
    Material and ManufacturingBar Material: Ti-6Al-4V ELI (ASTM F136) or Grade 4 CP Titanium (ASTM F67). - Meets chemical and mechanical requirements of ASTM F136 and ASTM F67. - Packaging: Nylon pouch. - Sterilization Process: Moist heat, validated according to EN ISO 17665-1:2006 to achieve a sterility assurance level (SAL) of 10-6.
    BiocompatibilityTesting performed according to ISO 10993-1:2018 and ISO 10993-5:2009 on representative Biomet 3i patient-specific bar devices using the same ASTM F136 titanium alloy and ASTM F67 unalloyed titanium materials. - Concluded that the material, chemical testing, and literature provide sufficient basis for demonstrating biocompatibility.
    MRI SafetyNon-clinical worst-case MRI review performed using scientific rationale and published literature (e.g., Woods et al., 2019). - Addressed parameters per FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque. - Reported Performance: BellaTek Bars have been labeled as MR conditional.
    Performance Bench TestingWorst-case design validated in accordance with Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, May 12, 2004. - Demonstrates that differences from the predicate are accounted for and do not render the device not substantially equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated as a number of devices or cases. The document refers to "worst case design" being validated. For sterilization, the validation process implies testing units to demonstrate SAL. For biocompatibility, "representative Biomet 3i patient-specific bar devices" were used. For MRI review, it was a "non-clinical worst-case MRI review... based on the entire system including all variations." This suggests a methodology of evaluating design limits and material properties rather than a large clinical test set.
    • Data Provenance: The studies are non-clinical (bench testing, material analysis, literature review, MRI rationale). No specific country of origin for non-clinical data is mentioned, but the standards followed (e.g., ASTM, ISO, EN ISO) are international. The studies are by nature prospective as they are conducted to support the specific device submission, but they leverage established scientific literature and standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Given that the studies are non-clinical bench tests, material characterizations, and literature reviews, the concept of "experts establishing ground truth for a test set" in a clinical diagnostic sense does not directly apply. The "ground truth" for these types of studies is established by:

    • Regulatory Standards: Adherence to recognized standards like ASTM F136, ASTM F67, ISO 10993-1, ISO 10993-5, EN ISO 17665-1, and FDA guidance documents.
    • Engineering and Scientific Expertise: The performance of bench tests, material analysis, and MRI safety assessments would be conducted by engineers, material scientists, and physicists with expertise in these areas. While individual names or specific numbers aren't listed, this expertise is inherent in conducting such tests.
    • Literature: The MRI safety assessment explicitly mentions using "published literature" as part of its scientific rationale, implying leveraging the consensus of experts in that field.

    No specific number of experts or their detailed qualifications (e.g., "radiologist with 10 years of experience") are provided in the context of establishing ground truth for a test set.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 typically refer to the process of resolving discrepancies among multiple human readers in clinical studies where subjective interpretation is involved (e.g., image reading). The studies described here are non-clinical, objective evaluations (bench testing, material analysis).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No. The provided document states: "No clinical data were included in this submission." Therefore, no MRMC study or AI assistance evaluation was conducted or reported.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical dental implant bar, not an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Regulatory Standards and Specifications: Adherence to performance parameters detailed in FDA guidance documents (e.g., "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments"), and international standards (ASTM, ISO, EN ISO).
    • Material Properties: Data derived from chemical and mechanical testing of the materials (ASTM F136, ASTM F67).
    • Biocompatibility Endpoints: Established biological safety endpoints defined by ISO 10993 standards.
    • MRI Safety Rationale: Scientific principles and published literature related to magnetic fields and medical devices.
    • Demonstration of Substantial Equivalence: Direct comparison of technological characteristics and performance to legally marketed predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI algorithms. The design and manufacturing processes are developed through traditional engineering methods and validated through the testing described.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this device.