Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K240621

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-30

    (56 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K193487

    Validate with FDA (Live)

    Device Name
    Agilon Strip
    Manufacturer
    Date Cleared
    2020-03-30

    (104 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Agilon Strip is indicated for use in voids or gaps of the skeletal system. i.e., the extremities, pelvis, and posteroateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities and pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.

    Device Description

    Agilon Strip is a flexible, resorbable, wicking, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules bound by type I collagen fibers to facilitate shaping and containment of the implant. The osteoSPAN granules in Agilon Strip are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate, with a thin laver of calcium phosphate throughout its entire porosity.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Agilon Strip," a resorbable calcium salt bone void filler. The submission aims to demonstrate substantial equivalence to a predicate device, Morpheus (K142828).

    However, the provided text does not contain any information about acceptance criteria or a study that proves the device meets those criteria in the context of device performance metrics like accuracy, sensitivity, or specificity. Instead, the document focuses on demonstrating substantial equivalence through:

    • Comparison of technological characteristics: Comparing the composition and properties of Agilon Strip to the predicate.
    • Bench testing: Verifying that the addition of collagen does not affect granule chemistry, crystallinity, porosity, pore diameter, and pore interconnectivity, along with sterilization and shelf-life studies.
    • Biocompatibility testing: Following ISO 10993-1 for biological effects such as cytotoxicity, sensitization, systemic toxicity, genotoxicity, and implantation.
    • In vivo testing: Using critically sized defect models and a clinically relevant spinal fusion model to assess bone formation, resorption, remodeling, and fusion rates compared to the predicate and controls.

    The "Performance Data" section discusses various tests and their outcomes but does not specify quantitative acceptance criteria for these tests (e.g., "fusion rate must be X% or higher"). The conclusion is based on the device being "substantially equivalent" to the predicate, implying that its performance in these tests was comparable and acceptable, but no explicit thresholds are provided.

    Therefore, I cannot populate the requested table and details as the necessary information (explicit acceptance criteria, detailed quantitative performance metrics against those criteria, data provenance, expert qualifications for ground truth, sample sizes for test/training sets with detailed ground truth establishment methods, MRMC studies, or standalone algorithm performance) is not present in the provided text.

    The document is a regulatory submission for a bone void filler, not an AI/ML device that would typically have performance metrics like sensitivity, specificity, or accuracy, which necessitate the kind of detailed study design outlined in the prompt (e.g., multi-reader multi-case studies, expert adjudication for ground truth).

    Ask a Question

    Ask a specific question about this device

    K Number
    K193168

    Validate with FDA (Live)

    Device Name
    Agilon Moldable
    Manufacturer
    Date Cleared
    2020-02-13

    (90 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Agilon Moldable is indicated for use in voids or gaps of the skeletal system, i.e., the extremities, pelvis, and posterolateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. The product may be used alone in the extremities and pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Moldable resorbs and is replaced with bone during the healing process.

    Device Description

    Agilon Moldable collagen enhanced bone graft substitute is a moldable, resorbable osteoconductive bone void filler composed of 1-2mm osteoSPAN granules suspended in a biocompatible orqanic binder to facilitate implant mixing, shaping, and containment. The osteoSPAN granules in the product are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate, with a thin layer of calcium phosphate throuqhout its entire porosity. The orqanic binder is a combination of a biocompatible polymer and type I collagen fibers to provide enhanced intraoperative handling. The polymer is rapidly absorbed in-situ, leaving behind osteoSPAN granules and collagen fibers as an osteoconductive scaffold.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Agilon Moldable device. This document is a premarket notification to the FDA, asserting substantial equivalence to a predicate device. It primarily focuses on the device's characteristics, indications for use, and a summary of performance data rather than a detailed study protocol for acceptance criteria.

    The information necessary to fully answer all parts of your request (especially concerning specific acceptance criteria and detailed study designs for proving those criteria) is not comprehensively present in this type of FDA letter. However, I can extract the available information and indicate where details are missing.

    Here's a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA letter and associated documents for a 510(k) submission typically don't explicitly list "acceptance criteria" in a table format with numerical targets. Instead, the demonstration of substantial equivalence is based on comparing the subject device's characteristics and performance to a legally marketed predicate device. The performance data presented focuses on equivalence to a predicate device and positive control rather than predefined numerical thresholds for the new device.

    From the text, the key performance indicator mentioned is "Fusion rates."

    Acceptance Criteria (Implied)Reported Device Performance
    Fusion rates (equivalence to predicate and autograft)Fusion rates were the same between all treatment groups (Agilon Moldable, predicate, positive control (autograft)) at each time point and consistent with published literature.
    Absence of adverse reactionsNo adverse reactions were noted at the implant site or in distant organs.
    Bone formation, remodeling, and implant resorptionNew bone formation, bone remodeling, and implant resorption for the test materials were confirmed with time.

    2. Sample size used for the test set and the data provenance

    • Test set sample size: Not explicitly stated. The study involved multiple time points and comparisons between Agilon Moldable, a predicate device (Morpheus), and a positive control (autograft) in an in vivo animal model. However, the exact number of subjects or samples per group is not provided in this summary.
    • Data provenance: Prospective in vivo animal testing in a validated, clinically relevant, single-level posterolateral spinal fusion model. The country of origin of the data is not specified, but the submission is to the U.S. FDA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not explicitly stated. The ground truth would likely be established by trained veterinary histopathologists and radiologists/imaging specialists for the animal study. The exact number and qualifications are not detailed in this summary.

    4. Adjudication method for the test set

    Not explicitly stated. For animal studies, evaluation usually involves standardized scoring by expert histopathologists for histology and morphometry, and radiologists for x-ray and micro-CT. Whether these involve independent assessment or consensus is not mentioned.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a bone void filler, a physical implant, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to human reader performance with or without AI is irrelevant to this device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. As noted above, this is a physical implant, not an algorithm.

    7. The type of ground truth used

    The ground truth for the in vivo study included:

    • Mechanical analysis: Objective quantitative assessments of bone strength/fusion.
    • Histology and Histomorphometry: Microscopic examination and quantitative measurements of tissue samples, evaluating new bone formation, tissue integration, and implant resorption.
    • X-ray and Micro-CT analyses: Imaging techniques to assess fusion status, bone volume, and structural characteristics.
    • Observation of adverse reactions: Clinical and pathological assessment for any negative responses to the implant.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that would require a distinct "training set." The in vivo study involved a test set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device. The ground truth for the in vivo performance study was established through the various analyses mentioned in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190371

    Validate with FDA (Live)

    Device Name
    Morpheus-C
    Manufacturer
    Date Cleared
    2019-08-09

    (175 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOGENNIX Morpheus-C is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

    Device Description

    Morpheus-C is a moldable, resorbable osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules suspended in a biocompatible organic binder to facilitate shaping and containment of the implant. The osteoSPAN granules in Morpheus-C are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate with a thin laver of calcium phosphate throughout its entire porosity. The organic binder in Morpheus-C is a combination of a biocompatible polymer and type I collagen fibers. The polymer is rapidly absorbed in-situ, leaving behind osteoSPAN granules and collagen fibers as an osteoconductive scaffold. The collagen in Morpheus-C provides improved intraoperative handling.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called Morpheus-C, a resorbable calcium salt bone void filler. This document is a regulatory submission to the FDA, and as such, it focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a performance study with acceptance criteria in the way one might for an AI/ML medical device.

    Therefore, the requested information regarding acceptance criteria, sample sizes for test/training sets, expert adjudication, MRMC studies, standalone performance measures, and ground truth establishment cannot be found or inferred from this document.

    This document describes a traditional medical device (a bone void filler) where performance is typically assessed through:

    • Biocompatibility testing: As detailed in Section VII, "Biocompatibility evaluation of Morpheus-C was conducted in accordance with 'Use of International Standard ISO 10993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.'" This included tests for cytotoxicity, sensitization, acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, genotoxicity, and implantation.
    • In vivo testing: "In vivo testing of Morpheus-C was conducted using a critically sized defect model. Device performance was evaluated at multiple time points against the primary predicate and negative controls using histology, histomorphometry, x-ray, and micro-CT analyses."
    • Bench testing: "Bench testing results for various physical and chemical characteristics of the primary predicate apply to the subject device since the addition of the type I collagen does not affect the granule chemistry, crystallinity, porosity, pore interconnectivity, or density."
    • Sterilization validation and shelf-life aging studies.

    These tests are designed to show the device is safe and effective for its intended use and performs similarly to legally marketed predicate devices, not to meet "acceptance criteria" in the sense of a machine learning model's performance on a dataset.

    In summary, none of the specific information requested in points 1-9 is available in this 510(k) document because it pertains to a different type of device (a physical implant) and a different type of evidentiary standard (substantial equivalence) than what is typically asked for AI/ML devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181337

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-18

    (120 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sypher Spacer System is a standalone intervertebral body fusion system for use in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use with autogenous and/or allogenic bone graft (composed of cancellous and/or corticocancellous bone) at one or two contiguous levels of the lumbar spine (L2-S1) for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Supplemental fixation is required whenever fewer than 3 bone screws are utilized.

    Device Description

    The Sypher Spacer System is a standalone intervertebral body fusion system intended for the lumbar spine. The Sypher Spacer System implants include cylinder shaped blocks manufactured from PEEK-OPTIMA™ LT1 conforming to ASTM F2026, tantalum markers conforming to ASTM F560, titanium alloy self-tapping, self-drilling bone screws conforming to ASTM F136, and titanium alloy cage lock assemblies conforming to ASTM F136.

    The implants are available in a variety of footprints. heights, and lordotic angles. The device features a hollow center to accommodate autogenous and/or allogenic bone graft (composed of cancellous and/or corticocancellous bone). The implants have integrated anterior screw holes to allow for medial placement of screws, and a titanium alloy cage lock assembly for securing the screws in place. The superior and inferior surfaces of the implant have a pattern of teeth to prevent movement.

    AI/ML Overview

    This document is a 510(k) summary for the Sypher Spacer System, an intervertebral body fusion device. It primarily focuses on demonstrating substantial equivalence to a predicate device based on design, materials, and mechanical testing. As such, the document does not contain the detailed information typically found in an AI/ML device submission regarding acceptance criteria, performance studies involving human readers, or detailed ground truth establishment.

    Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is absent:


    Acceptance Criteria and Study Details for the Sypher Spacer System

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Testing:Static Compression (ASTM F2077): Performance not explicitly stated (results deemed sufficient for substantial equivalence)
    Static compression per ASTM F2077Static Shear Compression (ASTM F2077): Performance not explicitly stated (results deemed sufficient for substantial equivalence)
    Static shear compression per ASTM F2077Expulsion Testing (ASTM Draft Standard F-04.25.02.02): Performance not explicitly stated (results deemed sufficient for substantial equivalence)
    Expulsion testing per ASTM Draft Standard F-04.25.02.02Finite Element Analysis (FEA): Results deemed sufficient for substantial equivalence
    Finite Element Analysis (FEA)

    Explanation: The document states that "Substantial Equivalence is supported by the results of mechanical testing and Finite Element Analysis (FEA)." However, it does not provide the specific acceptance criteria thresholds or the quantitative results from these tests. It only lists the types of tests performed.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The document describes mechanical testing of a medical device (intervertebral spacer), not an AI/ML diagnostic or predictive tool that would utilize a "test set" in the context of data. The mechanical tests would typically involve a certain number of manufactured units, but this sample size is not disclosed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable/provided. The device is a physical implant, and its performance is assessed through mechanical testing against standards, not by an expert review of a "test set" of patient data for ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert labeling of data, which is not relevant to the mechanical testing of a physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Reader Improvement with AI vs. Without AI Assistance

    This information is not applicable/provided. An MRMC study is relevant for AI-powered diagnostic tools to assess the impact on human reader performance. This document describes a physical medical implant, not an AI system.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This information is not applicable/provided. The device is a physical intervertebral spacer. It does not have an "algorithm-only" performance as it is not an AI/ML software.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by engineering standards and specifications for mechanical properties and biocompatibility, as evidenced by the mention of ASTM standards (e.g., ASTM F2077, ASTM F560, ASTM F136). The device's design also ensures it can accommodate autogenous and/or allogenic bone graft (composed of cancellous and/or corticocancellous bone) for fusion.

    8. The Sample Size for the Training Set

    This information is not applicable/provided. The device is a physical implant and does not involve a "training set" in the context of AI/ML models.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/provided. As above, there is no "training set" for this type of device. The design and manufacturing process would adhere to established quality systems and design controls rather than data-driven ground truth establishment.


    Summary of what the document does provide:

    • Device Type: Intervertebral body fusion system (Sypher Spacer System).
    • Regulatory Status: 510(k) clearance (K181337) based on substantial equivalence.
    • Predicate Devices: Biogennix Sypher Spacer System (K141798) as primary, Zimmer Biomet Solitaire (K081395, K081501) and Zimmer Biomet ROI-A® ALIF Cage System (K153495) as reference predicates.
    • Performance Study: Mechanical testing including static compression, static shear compression (both per ASTM F2077), expulsion testing (per ASTM Draft Standard F-04.25.02.02), and Finite Element Analysis (FEA).
    • Conclusion: The device demonstrated substantial equivalence based on indications for use, technological characteristics, and comparison with predicate devices supported by mechanical testing and FEA.

    The provided document is a 510(k) summary for a traditional medical device (implant), not an AI/ML-powered device. Therefore, many of the requested criteria related to AI/ML device validation are not present or applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141798

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-02-19

    (231 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sypher Spacer System is an intervertebral body fusion system indicated for intervertebral body fusion procedures in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use with autogenous bone graft at one or two contiguous levels of the lumbar spine (from L2 to S1) for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation to facilitate fusion.

    Device Description

    The Sypher Spacer System implants are cylinder shaped blocks made of PEEK Optima® LT1 (Polyether ether ketone per ASTM F2026), with tantalum alloy radiological position markers (per ASTM F560), titanium alloy self-drilling and self-tapping bone screws (per ASTM F136), and titanium alloy cage lock assemblies (per ASTM F136).

    The implants are available in a variety of footprints, heights and lordotic angles. The shape of the Sypher product allows for a larger implant (height and width) to be used allowing for more surface area contact. The Sypher Spacer System is offered in a closed graft space design. The implants incorporate integrated anterior screw holes to allow for medial placement of screws, as well as a titanium alloy cage lock assembly for securing the screws once in place. Additional or other supplemental fixation may be used, as patient needs dictate. The superior and inferior surfaces of the devices have a pattern of teeth to provide increased stability and to prevent movement of the implants. The Sypher Spacer System is intended to be used with autologous bone graft.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Sypher Spacer System, an intervertebral body fusion device. While it mentions performance testing, it does not include detailed acceptance criteria or a study that specifically proves the device meets those criteria in the way you've outlined.

    Here's what can be extracted based on the provided text, and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Specific quantitative criteria for intervertebral fusion devices related to clinical outcomes or device durability (e.g., fusion rates, subsidence limits, stability thresholds)Not explicitly stated in the provided document. The document refers to "sufficient strength for its intended use" without quantifiable acceptance criteria.
    Material conformance to standardsPEEK Optima® LT1 conforms to ASTM F2026. Tantalum conforms to ASTM F560. Titanium alloy conforms to ASTM F136.
    Mechanical Test StandardsDevice was tested according to: - Static axial compression bending per ASTM F2077-11 - Static compressive shear per ASTM F2077-11 - Static subsidence per ASTM F2267-04 and ASTM F2077-11 - Dynamic axial compression per ASTM F2077-11 - Dynamic compressive shear per ASTM F2077-11 - Expulsion per ASTM Draft Standard F-04.25.02.02
    Comparative StrengthThe device's strength is considered "substantially equivalent to legally marketed predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document only mentions that the device "has been tested" in various modes. It refers to "non-clinical testing," which typically means bench testing or cadaver studies, not human clinical trials.
    • Data Provenance: The testing is "non-clinical," implying it was conducted in a laboratory setting, likely in the US where the company and consultants are located. It is not clinical data (retrospective or prospective from patients).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Not applicable. This document describes non-clinical, mechanical performance testing of an implantable medical device, not a diagnostic device or a study requiring expert interpretation of results to establish ground truth. The "ground truth" for mechanical testing is typically the adherence of the device's physical properties to engineering standards and comparison to predicate devices, measured by instruments.

    4. Adjudication Method for the Test Set

    • Not applicable. As this is mechanical testing, there is no "adjudication method" in the sense of expert consensus on clinical or imaging readings. The results are quantitative measurements against established engineering standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This type of study (MRMC) is typically done for diagnostic devices where human readers interpret images or other data, and the goal is to assess the impact of an AI tool on reader performance. The Sypher Spacer System is an implantable surgical device, not a diagnostic one, and the reported tests are mechanical performance tests.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    • Not applicable. The Sypher Spacer System is a physical intervertebral fusion device, not a software algorithm. Its "performance" is its mechanical integrity and ability to facilitate fusion when implanted, which is assessed through bench testing and, eventually, clinical outcomes (though clinical trial data is not presented here, as this is a 510(k) submission based on substantial equivalence to predicates).

    7. Type of Ground Truth Used

    • For the non-clinical mechanical testing, the "ground truth" is defined by established ASTM (American Society for Testing and Materials) standards for intervertebral body fusion devices, and comparison to the mechanical properties of legally marketed predicate devices. The goal is to demonstrate that the Sypher Spacer System's mechanical characteristics are equivalent or superior to these benchmarks.

    8. Sample Size for the Training Set

    • Not applicable. There is no "training set" in the context of mechanical performance testing of a physical medical device. Training sets are relevant for machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set.

    In summary: The provided document is a 510(k) clearance letter and summary for a physical medical device. It focuses on demonstrating substantial equivalence to predicate devices primarily through mechanical (bench) testing, adherence to material standards, and similar indications for use. It does not contain information about clinical trials, diagnostic performance, or AI algorithms, which are the contexts in which many of your questions would typically apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132377

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-20

    (293 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BIOGENNIX osteoSPAN Morpheus is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

    Device Description

    osteoSPAN Morpheus is a moldable, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules suspended in a resorbable organic binder. The osteoSPAN granules used in osteoSPAN Morpheus are a composite of calcium carbonate with a thin calcium phosphate layer, in the form of hydroxyapatite, coating on all the surfaces of the interconnected porosity. The interconnected pores are approximately 500 microns in diameter and occupy approximately 65% of the volume. The biocompatible, organic binder facilitates placement and containment of the implant and is rapidly resorbed in-situ, leaving behind the osteoSPAN granules without affecting their osteoconductive or resorbable properties.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called osteoSPAN Morpheus, a resorbable calcium salt bone void filler. The submission does not include information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML device.

    Instead, this document focuses on demonstrating substantial equivalence to predicate devices based on:

    1. Technological characteristics comparison: A table comparing osteoSPAN Morpheus to a primary predicate (osteoSPAN granules) and a material-reference predicate (Actifuse ABX E-Z-fil Putty) across several characteristics (Calcium Salt, Granule Size, Porosity, Polymer Binder, Osteoconductive).
    2. Safety and performance testing: "Chemical and physical testing confirmed acceptable findings and were consistent with prior osteoSPAN predicate device qualifications. Biogennix osteoSPAN Morpheus was thoroughly evaluated for biocompatibility in accordance with ISO 10993 appropriate parts and for performance, through multiple non-clinical in-vivo and ex-vivo studies."

    Since this is a conventional medical device submission rather than an AI/ML device, the specific questions regarding AI/ML device performance (like sample size for test set, number of experts, adjudication method, MRMC studies, standalone performance, training set details) are not applicable to this document.

    Therefore, I cannot provide the requested information from the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093342

    Validate with FDA (Live)

    Device Name
    BIOGENNIX RPC
    Manufacturer
    Date Cleared
    2010-07-29

    (276 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Biogennix RPC is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Biogennix RPC is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Biogennix RPC bone void filler may be used as a bone graft extender for posterolateral spine fusion when mixed in a one to one ratio with autogenous bone graft. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

    Device Description

    Biogennix RPC is an osteoconductive, open cell, resorbable ceramic. The ceramic is a composite of calcium salts. The open pores and porosity provide space and structure for bone and vascular ingrowth. The calcium salts resorb over time. The product is available in block and granular forms.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Biogennix RPC bone void filler, which is a medical device. This type of regulatory document focuses on establishing substantial equivalence to a predicate device, rather than defining novel acceptance criteria and conducting a new study to prove it.

    Therefore, many of the requested categories (2, 3, 4, 5, 6, 7, 8, 9) are not applicable or cannot be extracted directly from this specific document, as they pertain to the evaluation of a new device's performance against predefined criteria, a Multi-Reader Multi-Case (MRMC) study, or a standalone algorithm performance study.

    Here's the information that can be extracted and a clear indication of what is not applicable based on the provided text:

    Acceptance Criteria and Device Performance

    The "acceptance criteria" for a 510(k) submission like this are primarily demonstrating substantial equivalence to a predicate device. This means showing that the new device is as safe and effective as a legally marketed device and does not raise new questions of safety or effectiveness. The device performance relates to its ability to meet the characteristics of the predicate device.

    Acceptance Criteria (Demonstrates Substantial Equivalence)Reported Device Performance (Biogennix RPC)
    Same Intended Use as the predicate device.Biogennix RPC has the same intended use: filling bony voids or gaps not intrinsic to bony structure stability, and as a bone graft extender for posterolateral spine fusion when mixed with autogenous bone graft.
    Same Technological Characteristics or differences do not raise new questions of safety or effectiveness.Biogennix RPC has the same technological characteristics (osteoconductive, open cell, resorbable ceramic, composite of calcium salts, provides space for bone/vascular ingrowth, resorbs and replaced by bone). Bench testing (x-ray diffraction, FTIR, IPS-MS, biomechanical, morphometry, in vitro dissolution, biocompatibility) and animal implantation studies demonstrated substantial equivalence to the predicate device (Pro Osteon 500R).
    Conformity to Applicable Standards (e.g., biocompatibility, sterilization).Conforms to ISO 10993 series (Biological evaluation of medical devices) and ANSI/AAMI/ISO 11137 (Sterilization of Health Care Products for Radiation Sterilization).
    Biocompatibility consistent with legally marketed devices.Biocompatibility testing was included in the studies demonstrating substantial equivalence.
    Mechanical Properties (e.g., structural integrity, handling) comparable to predicate.Biomechanical and morphometry studies were conducted to demonstrate substantial equivalence.
    Resorption and Bone Replacement characteristics (for resorbable materials).In vitro dissolution and animal implantation in long bone defects and posterolateral spine demonstrated the device's resorption and replacement with bone during the healing process, consistent with the predicate.

    Study Information

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not Applicable in this context. This document describes a 510(k) submission, which relies on demonstrating substantial equivalence to a predicate device through various tests (bench, animal) rather than a clinical "test set" in the context of diagnostic performance. The studies mentioned are:
        • x-ray diffraction, FTIR, IPS-MS, biomechanical, morphometry, in vitro dissolution (likely laboratory-based, not involving human subjects for a "test set").
        • Biocompatibility (likely in vitro and animal studies).
        • Animal implantation (in long bone defects and posterolateral spine).
      • Details on the sample sizes for these specific tests (e.g., number of animals, number of specimens for bench tests) are not provided in this summary.
      • Data provenance (country of origin, retrospective/prospective) is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not Applicable. As this is not a diagnostic device submission focusing on human interpretation of data, there is no "ground truth" established by experts in the sense of image interpretation or clinical diagnosis. The "truth" in this context is the performance of the device in various physical, chemical, and animal models.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable. See point 2.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. The device is a bone void filler (a physical implant), not a diagnostic tool or AI software that assists human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. The device is a physical bone void filler, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" in this context is the established scientific and engineering principles for material performance, biocompatibility, and biological response in animal models. For example, for biomechanical tests, the ground truth would be the measured physical properties; for animal studies, it would be histological analysis of bone ingrowth and resorption. This is not "expert consensus" on diagnostic findings or patient outcomes in a clinical trial, but rather expert assessment of experimental results.
    7. The sample size for the training set:

      • Not Applicable. This is not an AI/machine learning device that would require a "training set."
    8. How the ground truth for the training set was established:

      • Not Applicable. See point 7.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1