Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K192234

    Validate with FDA (Live)

    Device Name
    ActiPatch
    Date Cleared
    2020-01-31

    (165 days)

    Product Code
    Regulation Number
    890.5290
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adjunctive treatment of musculoskeletal pain

    Device Description

    The ActiPatch® device is a pulsed shortwave therapy device. The circuitry consists of low voltage (3 V) digital/analog electronics that control all timing functions to produce the therapeutic radiofrequency (RF) field, where the antenna is placed directly above the therapeutic site. This closed loop system of the antenna, low energy signal generator circuit, and battery power supply, transfers the RF energy to the target tissue as a localized therapy with no far field effects.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Device: ActiPatch® (K192234)
    Intended Use: Adjunctive treatment of musculoskeletal pain


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state "acceptance criteria" in a typical quantitative format for algorithm performance (e.g., "sensitivity must be > X%, specificity > Y%"). Instead, the substantial equivalence hinges on demonstrating that the ActiPatch is "at least as safe and effective as the predicate ActiPatch (K152432)."

    Therefore, the "acceptance criteria" can be inferred from the studies' primary effectiveness endpoints, which consistently aimed to show a statistically significant reduction in pain compared to a control group (either active-treatment or placebo). The device's performance is described by achieving these significant reductions.

    Acceptance Criteria (Inferred from Study Goals)Reported Device Performance (Summary of Findings)
    Effectiveness (Pain Reduction):
    Reduction in pain for cervical osteoarthritis (neck pain) compared to active-treatment control (NSAID)ActiPatch significantly reduced pain (measured by VAS pain) associated with COA in the device treatment group. Treatment difference between device-treatment and NSAID-treatment groups was significant (p<0.05).
    Improvement in pain for osteoarthritis of the knee compared to placeboActiPatch significantly reduced pain (measured by VAS pain) associated with KOA in the device treatment group. Treatment difference between active and placebo treatment groups was significant (p<0.05) for both VAS and WOMAC scores.
    Reduction in pain for plantar fasciitis (heel pain) compared to placeboActiPatch significantly reduced pain (measured by VAS pain) associated with PF in the device treatment group. Treatment difference between active and placebo treatment groups was significant (p<0.05).
    Safety:
    No unacceptable treatment-related adverse eventsThe studies generally mention "primary safety endpoint was all treatment-related adverse events." The conclusion states the device is "at least as safe and effective," implying the safety profile was acceptable and comparable to the predicate. Biocompatibility testing (skin sensitization, irritation, cytotoxicity) also showed no adverse reactions.
    Usability (for OTC use):
    Lay users understand indications, contraindications, and proper useUsability testing showed that lay users understand the indications for use, when not to use the device, how to turn it on, place it correctly, and how long to use it. (Specifically for RecoveryRx, implying a similar interface/device to ActiPatch for OTC use context).

    2. Sample Sizes Used for the Test Set and Data Provenance

    The "test set" in this context refers to the clinical study populations.

    • Chronic Cervical Osteoarthritis (Neck Pain) Study:
      • Sample Size: 200 intent-to-treat patients (197 completed).
      • Data Provenance: Not explicitly stated, but clinical trials are typically prospective. No country of origin is mentioned.
    • Osteoarthritis of the Knee Study:
      • Sample Size: 66 intent-to-treat patients (60 completed).
      • Data Provenance: Not explicitly stated, but clinical trials are typically prospective. No country of origin is mentioned.
    • Plantar Fasciitis (Heel Pain) Study:
      • Sample Size: 70 patients (all 70 completed).
      • Data Provenance: Not explicitly stated, but clinical trials are typically prospective. No country of origin is mentioned.
    • Usability Testing:
      • Sample Size: 46 men and women.
      • Data Provenance: Not explicitly stated, but this would be a prospective study. No country of origin is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable in the traditional sense for these types of clinical studies. The "ground truth" for pain reduction and adverse events is established by the patients themselves (self-reported pain scores like VAS, WOMAC) and clinical observation by study investigators/medical professionals for adverse events and objective measures.

    • Experts Involved: Medical professionals (physicians, nurses, study coordinators) would have been involved in diagnosing patients, conducting the study, and recording data, but they aren't "establishing ground truth" as adjudicators in the way an AI would be validated against expert labels.
    • Qualifications: Not specified for the study staff, but it's assumed they were qualified medical and research professionals appropriate for conducting clinical trials.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in diagnostic studies where multiple readers interpret images or data. This is not applicable here because the primary endpoints (pain scores, adverse events) rely on patient self-report and clinical observation, not multi-reader interpretation.

    The studies were randomized and controlled, with some being double-blinded (knee osteoarthritis and plantar fasciitis studies), which serves to mitigate bias in data collection and assessment, rather than expert adjudication of an interpretive task.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done.

    MRMC studies are typically used to assess the change in human reader performance when assisted by an AI. The ActiPatch is a medical device, and the studies evaluated its direct effect on patients, not its impact on human reader interpretations.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the primary clinical studies described are standalone performance studies of the device itself.

    The ActiPatch device is applied directly to the patient and delivers therapy. Its effectiveness in reducing pain (the primary outcome) is measured directly, without a human intermediary interpreting its output. The studies assess the performance of the device "algorithm" (its therapeutic mechanism) by observing patient outcomes.


    7. The Type of Ground Truth Used

    The ground truth used in these studies is primarily:

    • Patient-reported outcomes (PROs):
      • VAS (Visual Analog Scale) pain scores: For all three pain studies.
      • WOMAC (Western Ontario and McMaster Universities Arthritis Index) scores: For the knee osteoarthritis study.
    • Clinical observation/assessment: For adverse events.
    • Patient understanding: For usability testing.

    8. The Sample Size for the Training Set

    Not applicable. The ActiPatch is a physical medical device (pulsed shortwave therapy) and not an AI/machine learning algorithm that requires a "training set" of data. Its mechanism of action is based on established physical principles, not learned patterns from data.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190251

    Validate with FDA (Live)

    Device Name
    RecoveryRx
    Date Cleared
    2019-06-28

    (141 days)

    Product Code
    Regulation Number
    890.5290
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adjunctive treatment of postoperative pain

    Device Description

    The RecoveryRx® device is a pulsed shortwave therapy device. The circuitry consists of low voltage (3 V) digital/analog electronics that control all timing functions to produce the therapeutic radiofrequency (RF) field, where the antenna is placed directly above the therapeutic site. This closed loop system of the antenna, low energy signal generator circuit, and battery power supply, transfers the RF energy to the target tissue as a localized therapy with no far field effects.

    AI/ML Overview

    The provided text describes the regulatory filing for a medical device called "RecoveryRx®". It is a pulsed shortwave therapy device intended for the adjunctive treatment of postoperative pain. However, the document does not contain acceptance criteria and information about a study proving that the device meets specific acceptance criteria in the context of an AI/Machine Learning (ML) enabled medical device.

    Instead, the document details a traditional medical device submission (510(k)) where the manufacturer demonstrates substantial equivalence to predicate devices (ActiBand and Ivivi) by providing:

    • Technological characteristics comparison: Showing the RecoveryRx® has similar features and operates on the same principles as the predicate devices.
    • Non-clinical performance data: Electrical safety, electromagnetic safety, and biocompatibility testing, demonstrating compliance with relevant standards.
    • Clinical data: Results from two randomized, double-blinded, placebo-controlled studies to show effectiveness in reducing postoperative pain, which supports its intended use.

    Therefore, I cannot directly answer your prompt, as the requested information (acceptance criteria for AI/ML performance, test set details, expert ground truth, MRMC study, standalone performance, training set details) is specific to AI/ML device validation and is not present in this document. This document describes a physical medical device, not an AI/ML software device.

    To answer your prompt, I would need a document that describes the validation of an AI/ML-enabled medical device.

    However, I can extract the relevant information from the provided text that describes the clinical studies performed to demonstrate the device's effectiveness, which is analogous to "proving the device meets" its intended therapeutic effect (not an AI/ML performance metric).

    Here's a breakdown of what is provided regarding the device's performance validation, interpreted in light of your request:

    Study Proving Therapeutic Effect (Not AI/ML Performance)

    The document describes two clinical studies to demonstrate the effectiveness of RecoveryRx® in reducing postoperative pain.

    1. Acceptance Criteria and Reported Device Performance (Therapeutic Effect)

    Acceptance Criteria (Implied)Reported Device Performance
    Reduction in postoperative pain compared to placebo (statistical significance).Study 1 (Breast Augmentation): Statistically significant treatment effect (reduction in postoperative pain) compared to placebo (p<0.05).
    Study 2 (Cesarean Section): Statistically significant treatment effect (reduction in postoperative pain) compared to placebo (p<0.05).

    2. Sample Size and Data Provenance

    • Study 1 (Breast Augmentation):
      • Sample Size: 18 women
      • Data Provenance: Not explicitly stated, but clinical studies are generally prospective. Implied to be from trials conducted by the manufacturer or their partners.
    • Study 2 (Cesarean Section):
      • Sample Size: 96 women
      • Data Provenance: Not explicitly stated, but clinical studies are generally prospective. Implied to be from trials conducted by the manufacturer or their partners.

    3. Number of Experts and Qualifications for Ground Truth

    • N/A: These were clinical trials measuring patient pain levels, not studies establishing ground truth for AI/ML image or signal interpretation. The "ground truth" was the patient's reported pain score.

    4. Adjudication Method for the Test Set

    • N/A: The studies were "randomized, double-blinded, placebo-controlled," which is a standard method for clinical trials to reduce bias. There wasn't an "adjudication method" in the sense of expert consensus on a diagnostic outcome.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No: This was a clinical trial evaluating a physical therapy device, not an AI-assisted diagnostic tool. Therefore, an MRMC study (relevant for AI in radiology, for example) was not performed.

    6. Standalone (Algorithm Only) Performance

    • N/A: This is a physical device, not an algorithm. The "standalone performance" is its therapeutic effect as demonstrated in the clinical trials.

    7. Type of Ground Truth Used

    • Patient-reported outcomes: Pain was measured on a 0-10 visual analog pain scale. This is a subjective patient-reported measure.

    8. Sample Size for the Training Set

    • N/A: This is a physical device, not an AI/ML model that requires a training set. The "training" for such a device comes from engineering design and pre-clinical testing, followed by clinical validation.

    9. How Ground Truth for the Training Set Was Established

    • N/A: (See above)

    In summary, the provided document details the regulatory pathway for a conventional medical device, not an AI/ML-enabled one. The "proof" of meeting criteria relates to its safety, effectiveness, and substantial equivalence to existing devices, demonstrated through engineering tests and clinical trials, rather than AI model performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152432

    Validate with FDA (Live)

    Device Name
    ActiPatch(R)
    Date Cleared
    2017-02-03

    (526 days)

    Product Code
    Regulation Number
    890.5290
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adjunctive treatment of musculoskeletal pain related to: (1) plantar fasciitis of the heel; and (2) osteoarthritis of the knee.

    Device Description

    The ActiPatch® device is a pulsed shortwave therapy device. The circuitry consists of low voltage (3 V) digital/analog electronics that control all timing functions to produce the therapeutic radiofrequency (RF) field, where the antenna is placed directly above the therapeutic site. This closed loop system of the antenna, low energy signal generator circuit, and battery power supply, transfers the RF energy to the target tissue as a localized therapy with no far field effects.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the ActiPatch® device. This type of regulatory submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel safety and effectiveness. Therefore, the information provided focuses on comparative performance and safety rather than establishing acceptance criteria as would be seen for a new medical device. The document does not explicitly state "acceptance criteria" but rather presents clinical study results as evidence of comparable safety and effectiveness to predicate devices.

    Here's an attempt to extract and organize the requested information based on the document's content:

    1. Table of Acceptance Criteria and Reported Device Performance

    As noted, the document doesn't define explicit "acceptance criteria" in the traditional sense of pre-defined thresholds. Instead, it presents clinical study results to demonstrate effectiveness and safety comparable to predicate devices. The "reported device performance" below summarizes these clinical findings.

    Criterion Type (Implicit)Acceptance Criteria (Not explicitly stated, but implied by study results)Reported Device Performance (ActiPatch®)
    Effectiveness - Knee OAImprovement in pain levels (VAS and WOMAC scores) comparable to or better than placebo/no treatment, and potentially supporting reduction in pharmacological therapy for osteoarthritis of the knee.Osteoarthritis of the knee study: - 36% of the treatment group reported a >30% decrease in VAS pain, compared to 9% for the placebo group. - 18% of the treatment group reported a >30% decrease in total WOMAC pain, compared to 3% for the placebo group. - 26% of the treatment group stopped pharmacological therapy, whereas 33% of the placebo group started new pharmacological therapy.
    Effectiveness - Plantar FasciitisReduction in pain levels (daily morning VAS score) comparable to or better than placebo/no treatment for plantar fasciitis.Plantar fasciitis study: - Average reported pain reduction (day 1 AM to day 7 AM VAS) for the treatment group was 40%, compared to 7% for the control group.
    SafetyNo significant adverse events related to the device.Osteoarthritis of the knee study: No adverse events were recorded. Plantar fasciitis study: Primary safety endpoint was all treatment-related adverse events; the document does not explicitly state if any were recorded but implies a positive safety profile through the "safe and effective as its predicate devices" conclusion.
    Usability (OTC Use)Lay users must understand how to use the device safely and effectively for pain relief in the specified indications.Usability testing: - 46 men and women (age >17, wide education levels) demonstrated understanding of indications for use, contraindications, how to turn on, correct placement, and duration of use.
    Non-Clinical PerformanceConformity to relevant electrical safety, electromagnetic compatibility, and biocompatibility standards. Performance as intended under anticipated conditions of use regarding output power, pulse characteristics, absorption rates, and electromagnetic fields.Non-Clinical/Performance Data: - Conformed to IEC/EN 60601-1-2:2012, IEC 60601-1:2005+A1:2012, EN 60601-1:2006. - Biocompatibility testing (ISO 10993-10:2010 for sensitization and irritation, ISO 10993-5:2009 for cytotoxicity) showed no evidence of sensitization, irritation, or in vitro cytotoxicity. - Special controls testing (per October 13, 2015 Final Reclassification Order) demonstrated intended performance, characterizing peak output power, pulse characteristics, duty cycle, average output power, specific absorption rates, and electrical/magnetic fields in a saline gel test load.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Osteoarthritis of the knee study:
      • Sample size: 66 intent-to-treat patients, with 60 patients completing the four-week study.
      • Data Provenance: Not explicitly stated (e.g., country of origin, specific locations), but described as "double-blind randomized controlled study" implying a prospective design.
    • Plantar fasciitis study:
      • Sample size: 70 patients completed the study.
      • Data Provenance: Described as "double-blind, multicenter, randomized, placebo-controlled study," implying a prospective design. Country of origin not specified.
    • Usability testing:
      • Sample size: 46 men and women.
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective), but the description ("These subjects demonstrated use...") implies a prospective, interactive study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" for the test set in the context of image interpretation or diagnostic accuracy studies. The clinical studies evaluated patient-reported outcomes (VAS and WOMAC scores) and adverse events, which are direct measures from the participants themselves, rather than interpretations by experts.


    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    The document describes the clinical studies as "double-blind and placebo-controlled randomized controlled trials." This implies that blinding was used for both participants and researchers/assessors regarding treatment allocation. However, it does not specify an adjudication method for outcome measures like VAS or WOMAC scores, which typically do not require such adjudication as they are direct patient responses. For adverse events, it states "No adverse events were recorded" for the knee study; the method for reviewing or adjudicating AEs in the plantar fasciitis study is not detailed beyond "primary safety endpoint was all treatment-related adverse events."


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The ActiPatch® is a physical medical device for pain relief, not an AI or imaging interpretation device that would typically involve human readers.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The ActiPatch® is a physical device, not an algorithm. Its performance is inherent in its physical and electrical characteristics and its interaction with the patient's body. The clinical studies evaluate the device's effect directly on patients.


    7. The Type of Ground Truth Used

    For the clinical studies, the "ground truth" was primarily based on patient-reported outcomes:

    • Pain levels: Measured using validated scales like the Visual Analog Scale (VAS) and Western Ontario and McMaster Universities Arthritis Index (WOMAC) scores.
    • Adverse events: Self-reported by patients or observed during the study.

    For usability testing, the "ground truth" was established by observing lay user comprehension and ability to perform tasks (turn on, place correctly, understand indications/contraindications).

    For non-clinical testing, the "ground truth" was conformance to established engineering and biological standards (e.g., IEC, ISO standards for electrical safety, biocompatibility, and specific power output measurements).


    8. The Sample Size for the Training Set

    The document does not describe the use of a "training set" as it would for an algorithm or AI model development. The clinical studies described are for evaluating the final device's performance, not for training a model.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was explicitly mentioned or used in the context of algorithm development. The document describes clinical trials and usability studies of a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1