Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K241642

    Validate with FDA (Live)

    Date Cleared
    2024-08-02

    (56 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    Diode Laser Hair Removal System (Model: MBT-Diode Laser) is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). The Dermatological Diode Laser Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handle. The emission laser is activated by footswitch or handle button.

    The Diode Laser Hair Removal System consists of the following major components

    • The main console unit that incorporates the main CPU board, power supply modules, laser power supply, laser device, cooling system and switching module
    • Handpiece
    • Footswitch
    AI/ML Overview

    The provided text is a 510(k) Summary for a Diode Laser Hair Removal System (MBT-Diode Laser). It describes the device, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not outline specific, quantified acceptance criteria for the new device as would be seen in a clinical study. Instead, it demonstrates substantial equivalence to a predicate device by comparing technical specifications and adherence to recognized standards.

    The table below summarizes the technical comparison provided, where "acceptance criteria" can be interpreted as demonstrating equivalence or justification for differences compared to the predicate/reference devices:

    FeatureAcceptance Criteria (Comparison to Predicate/Reference)Reported Device Performance (Proposed Device)
    General
    Device NameNA (for comparison)Diode Laser Hair Removal System
    Classification RegulationSAME as predicate (21 CFR 878.4810)21 CFR 878.4810
    Classification PanelSAME as predicate (General & Plastic Surgery)General & Plastic Surgery
    ClassSAME as predicate (II)II
    Product CodeSAME as predicate (GEX)GEX
    Indications for UseSAME as predicate: Hair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.Same as criterion
    Performance (Technical)
    Laser TypeSAME as predicate (Diode Laser)Diode Laser
    Laser ClassificationSAME as predicate (Class IV)Class IV
    Laser WavelengthSAME as predicate (808nm)808nm
    Spot SizeSIMILAR to predicate (15mm x 15mm vs. 2.25cm², 1.44cm²)15mm x 15mm
    FluenceDifferent from predicate (0-120 J/cm²), but justified as lower (55.96 J/cm²) and similar to reference device (48 J/cm²) for effectiveness. Safety ensured by compliance with standards.1-55.96 J/cm²
    FrequencySAME as predicate (1-10Hz)1-10Hz
    Pulse DurationDifferent from predicate (10-400ms), but within predicate range (2-240ms) and considered a minor difference not affecting effectiveness/safety. Safety ensured by compliance with standards.2-240ms
    Power SupplyDifferent from predicate (220/110 VAC/50Hz-60Hz), but electrical safety and EMC tests conducted ensure normal operation.110 VAC/50Hz-60Hz
    DimensionDifferent from predicate (112 cm x 42 cm x 60 cm), but mechanical performance accepted by IEC 60601-1.650mm x 580mm x 1210mm
    WeightDifferent from predicate (63Kg), but mechanical performance accepted by IEC 60601-1.68Kg
    Safety
    Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22 (SAME as predicate)Complies
    EMCComply with IEC 60601-1-2 (SAME as predicate)Complies
    Laser SafetyComply with IEC 60601-2-22, IEC 60825 (SAME as predicate)Complies
    CytotoxicityNo Cytotoxicity (SAME as predicate; aligns with ISO 10993-5)No Cytotoxicity
    SensitizationNo evidence of sensitization (SAME as predicate; aligns with ISO 10993-10)No evidence of sensitization
    IrritationNo evidence of irritation (SAME as predicate; aligns with ISO 10993-23)No evidence of irritation

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical study is included in this submission."
    Therefore, there is no "test set" in the context of clinical performance data. The evaluation relies on non-clinical (bench) testing and comparison to predicate devices, not human subject data.

    Since no clinical study was conducted, there is no data provenance (country of origin, retrospective/prospective) for a test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As no clinical study was conducted and no clinical test set was used, there was no ground truth for a test set established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical study was conducted and no clinical test set was used, there was no adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states: "No clinical study is included in this submission." Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a hardware product (Diode Laser Hair Removal System), not an algorithm or software-only device. Therefore, the concept of "standalone algorithm performance" as typically applied to AI/software as a medical device (SaMD) is not applicable here. The performance is evaluated based on its physical/electrical characteristics and safety standards.

    7. The Type of Ground Truth Used

    For the non-clinical tests regarding electrical safety, EMC, laser safety, and biocompatibility, the "ground truth" is defined by compliance with internationally recognized standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 10993 series). The device's performance is measured against the requirements of these standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device submission, not an AI/ML algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170281

    Validate with FDA (Live)

    Date Cleared
    2017-10-06

    (249 days)

    Product Code
    Regulation Number
    868.1860
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV 1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.

    Device Description

    MSA100BT Peak Flow Meter is a hand-held pulmonary function measuring medical device that measures patient's maximum possible exhalation which is called peak expiratory flow (PEF) and forced expiratory volume in 1 second (FEV1). The accuracy meets American Thoracic Society (ATS) Standard 2005 Revision. The MSA100BT is designed for pediatric to adult patients for home use and for single user, and is intended for monitoring respiratory conditions such as asthma. The MSA100BT has an automatic memory of 250 recordings and can connect to cell phone by Bluetooth. Patients can transfer the records to phone via BLE4.0, then using app to help save and track long term. The MSA100BT is capable of measuring PEF from 50 to 900 L/min and FEV1 from 0.01 to 9.99 L. Its measuring accuracy is ± 20 L/min or ± 10% of the reading for PEF, and ± 0.050 L or ± 3% of the reading for FEV1. The MSA100BT is composed of main unit, turbine sensor, and mouth piece. After each use, the mouthpiece and turbine of the device should be cleaned by soap solution and water within 30 minutes.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the MSA100BT Peak Flow Meter. It focuses on demonstrating substantial equivalence to a predicate device, the MSA100 Peak Flow Meter, rather than detailing a study that establishes acceptance criteria for a novel AI device and proves its performance.

    Therefore, the specific information requested about acceptance criteria, a study proving device meeting criteria, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set details for an AI-powered device cannot be extracted directly from this document. This document pertains to a physical medical device (peak flow meter) and its performance validation using non-clinical testing and comparison to an existing device, not an AI algorithm.

    However, I can extract the performance criteria and the testing performed to demonstrate that the device complies with certain standards, which can be interpreted as fulfilling "acceptance criteria" in the context of this device's submission.

    Here's the closest information I can provide based on the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Requirement)Reported Device Performance
    Accuracy:
    PEF: ± 20 L/min or ± 10% of the readingPEF: ± 20 L/min or ± 10% of the reading (Stated as identical to predicate and meets ATS Standard 2005 Revision)
    FEV1: ± 0.05 L or ± 3% of the readingFEV1: ± 0.05 L or ± 3% of the reading (Stated as identical to predicate and meets ATS Standard 2005 Revision)
    Biocompatibility: (According to ISO 10993-1:2009/(R)2013)
    Non-cytotoxicTest article confirmed non-cytotoxic (Complied with ISO 10993-5:2009)
    Non-irritating (oral mucosa)Test article confirmed non-oral-mucosal-irritating (Complied with ISO 10993-10:2010)
    Non-sensitizingTest article confirmed non-sensitizing (Complied with ISO 10993-10:2010)
    Electrical Safety: (According to IEC 60601-1)Device complies with IEC 60601-1:2012
    Electromagnetic Compatibility (EMC): (According to IEC 60601-1-2)Device complies with IEC 60601-1-2:2014
    Wireless Compliance: (According to FDA Guidance on RF Wireless Tech)Device complies with FDA's guidance: "Guidance Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff."
    Performance Parameters: (According to Standards of Spirometry 2005)All of the tested parameters (accuracy, intra instrument repeatability, PEF inter instrument repeatability, performance of flow resistance) meet the requirements in the standards. The performance of the proposed device is demonstrated to be comparable with the predicate device. The accuracy meets American Thoracic Society (ATS) Standard 2005 Revision. (Range: PEF 50 to 900 L/min, FEV1 0.01 to 9.99 L; Measuring solution: PEF 1 L/min; FEV1 0.01 L).
    Software Verification & Validation: (According to FDA Guidance)Software validation demonstrated that the software functions as specified in the software requirement specifications. (For "moderate" level of concern software)

    Regarding the specific questions about AI/ML device studies, the document does not contain this information as it is not about such a device.

    1. Sample size used for the test set and the data provenance: Not applicable. The testing described is for a physical device, not an AI algorithm evaluated on a data test set from specific provenance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The ground truth for performance is established by reference standards for spirometry or laboratory tests for biocompatibility/electrical safety.
    3. Adjudication method: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this is not an AI-assisted diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is not an AI algorithm.
    6. The type of ground truth used:
      • Performance: Reference standards specified by Standards of Spirometry (2005 Revision) and American Thoracic Society (ATS) Standard 2005 Revision.
      • Biocompatibility: Laboratory testing results against ISO 10993 standards.
      • Electrical Safety & EMC: Compliance with IEC 60601 standards.
      • Wireless: Compliance with FDA guidance.
      • Software: Software requirement specifications.
    7. The sample size for the training set: Not applicable, as this is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133975

    Validate with FDA (Live)

    Date Cleared
    2014-10-03

    (281 days)

    Product Code
    Regulation Number
    868.1860
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV 1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatic to adult patients. The device is intended for monitoring respiratory conditions such as asthma.

    Device Description

    The MSA100 Peak Flow Meter is a medical device that measures patient's maximum possible exhalation which is called "peak flow" (referred to in medical terms as PEF-Peak Expiratory Flow) and "forced expiratory volume" (commonly known as FEV1) in one second. Regular monitoring of the peak flow is useful for monitoring diseases of the airways such as asthma, C.O.P.D, or chronic bronchitis, is usually used for patient home use.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Predicate Device)Reported Device Performance (Proposed MSA100 Peak Flow Meter)
    PEF Measuring Range50 to 900 l/min50 to 900 l/min
    FEV1 Measuring Range0.01 to 9.99 liters0.01 to 9.99 liters
    PEF Accuracy± 25 l/min or 12% of the reading± 20 l/min or ± 10% of the reading
    FEV1 Accuracy± 0.1 l or ± 5% of the reading± 0.05 l or ± 3% of the reading
    PEF Measuring Resolution1 l/min1 l/min
    FEV1 Measuring Resolution0.01 l0.01 l
    Data SafetyData rememorized by EEPROMData rememorized by Flash memory
    Memory240 measurements with date/time300 measurements with date/time
    Power Source2 batteries of 1.5V, size AAAAAA 1.5x2 alkaline batteries
    Electrical SafetyIEC 60601-1 and IEC 60601-1-2IEC 60601-1 and IEC 60601-1-2
    Biocompatibility (In vitro cytotoxicity, irritation, sensitization)Met acceptance criteria (implied by ISO standards)Met acceptance criteria (ISO 10993-5: 2009; ISO 10993-10: 2010)
    Performance TestingATS 2005 updateATS 2005 update

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions that "The sponsor has performed bench tests to demonstrate the MSA100 Peak Flow Meter performs within specifications." This suggests the testing was laboratory-based rather than involving human subjects for performance evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided. The testing appears to be based on engineering and laboratory standards, not clinical ground truth established by medical experts for individual cases.

    4. Adjudication method for the test set:

    Not applicable. The testing described involves bench tests against specified performance metrics and adherence to international standards, not subjective expert adjudication of results.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a measurement instrument (Peak Flow Meter) and does not involve AI or human "readers" interpreting data in a typical MRMC study context.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device is a standalone instrument for measuring PEF and FEV1. The "algorithm only" concept doesn't directly apply as it's a physical device with embedded measurement technology (turbine transducer, photoelectric conversion). The performance described is its standalone measurement accuracy.

    7. The type of ground truth used:

    The ground truth used for specific performance metrics (like accuracy for PEF and FEV1) would be established by reference standards or calibrated equipment used in the bench testing. For biocompatibility and electrical safety, the ground truth is adherence to the specified ISO and IEC standards.

    8. The sample size for the training set:

    Not applicable. This device is a hardware measurement instrument tested against established engineering and performance standards, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1