Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K191119

    Validate with FDA (Live)

    Date Cleared
    2019-08-23

    (119 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EkoSonic® Endovascular System is indicated for the:

    • Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
    • Infusion of solutions into the pulmonary arteries.
    • Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
    Device Description

    The EkoSonic Endovascular System consists of an EkoSonic Endovascular Device and the Control System (Control Unit and Connector Interface Cables). The EkoSonic Endovascular Device consists of a single-use, disposable infusion catheter with removable ultrasound core. The infusion catheter contains multiple side holes distributed over the length of the treatment zone. The ultrasound core contains up to 30 ultrasound elements, evenly spaced over the treatment zone. Thermal sensors in the treatment zone monitor catheter temperature. The Control System generates and controls the delivery of radiofrequency energy to the ultrasound core while monitoring and controlling the temperature of the treatment zone.

    This notification is being made due to the modification of the MSD hub portion of the EkoSonic Endovascular Device.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the EkoSonic Endovascular Device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes several device-level and system-level tests performed to demonstrate substantial equivalence, particularly focusing on modifications to the MSD hub portion of the device. The reported performance for all tests is "Pass," indicating that the device met the established acceptance criteria for each test.

    Product SpecificationPurposeAcceptance Criteria (Implied)Reported Device Performance (T=0)Reported Device Performance (Artificially Aged (T=3 Years))
    Device Level Testing
    Tensile StrengthRedesign of shaft-to-cable and Luer-to-cover joints (replaces lap joint with soldered PCB, Luer-to-cover epoxy with snap-on cover).Maintain structural integrityPassPass
    Catheter InterlockRedesigned catheter interlock, replacing watertight epoxy seal with copolyester overmold.Maintain watertight sealPassPass
    ImpedanceRedesigns connection between shaft-to-cable, which is the RF energy path.Maintain electrical conductivityPassPass
    Resonant Frequency(Associated with Impedance testing)Maintain optimal frequencyPassPass
    LuersRedesigned Luer barb to facilitate snap-on cover.Maintain proper Luer functionPassPass
    Electrical IsolationRedesigns leakage current pathway from connector to external fluid.Maintain electrical safetyPassPass
    SterilizationDevice undergoes shipping, storage, sterilization, and shelf life simulation to provide representative test articles.Maintain sterile barrier/integrityPassPass
    Shipping(Part of simulating representative test articles)Withstand shipping stressesPassPass
    Storage(Part of simulating representative test articles)Withstand storage conditionsPassPass
    Shelf Life(Part of simulating representative test articles)Maintain function over shelf lifePassPass
    Burst StrengthRedesigns Luer-to-cover interface, affecting resistance to high static pressure.Maintain resistance to burstPassPass
    Fluid ImmersionRedesigns and replaces primary seal with overmold; assessed continuity before and after fluid immersion.Maintain fluid integrity/continuityPassPass
    Bending ForceReplaces Luer and strain relief with new Luer barb and snap-on cover; assessed bending force of snap-on cover.Maintain structural flexibilityPassPass
    System Level Testing
    System Integration with PT-3BMSD hub changes affect connection between MSD and Control System.Maintain system compatibilityPassN/A
    System Integration with CU4.0MSD hub changes affect connection between MSD and Control System.Maintain system compatibilityPassvalidation testing.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states "Testing has confirmed that the EkoSonic Endovascular Device functions as intended and is substantially equivalent to the predicate device." However, it does not explicitly state the sample sizes used for each of the various tests (e.g., tensile strength, burst strength). It also does not specify the data provenance (e.g., country of origin, retrospective or prospective). The nature of the tests (physical and electrical performance) generally implies laboratory testing rather than clinical data provenance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The study described is focused on the physical and electrical performance of the device components due to a modification. It is not a diagnostic device study that would typically involve establishing ground truth based on expert consensus of medical images or patient outcomes. Therefore, there is no mention of experts or their qualifications for establishing ground truth in the context of these engineering performance tests. The "ground truth" here is the expected physical and electrical performance as defined by engineering specifications and standards.

    4. Adjudication Method for the Test Set:

    Given that this is a technical performance study of device components, an "adjudication method" in the clinical sense (like 2+1 or 3+1 consensus for medical image interpretation) is not applicable and not mentioned. The results of these engineering tests ("Pass") would be determined by whether the device meets pre-defined quantitative or qualitative engineering specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not mentioned or performed. The device is an EkoSonic Endovascular Device used for ultrasound-facilitated, controlled, and selective infusion of fluids. It's a therapeutic device, not a diagnostic imaging device typically evaluated with MRMC studies comparing human readers with and without AI assistance.

    6. Standalone Performance Study:

    The study is inherently about the standalone performance of the modified device components and its integration with the control unit. The listed tests (tensile strength, impedance, electrical isolation, etc.) evaluate the device's intrinsic physical and electrical characteristics. There is no "algorithm only" aspect as it's a physical medical device. It's not an AI-driven diagnostic tool that would have a separate standalone algorithm performance.

    7. Type of Ground Truth Used:

    The ground truth for this type of performance study is based on engineering specifications, design requirements, and established industry standards for medical device safety and functionality. For example, burst strength would have a defined pressure threshold, and electrical isolation would have maximum leakage current limits. The "Pass" results indicate that the device met these predefined engineering criteria.

    8. Sample Size for the Training Set:

    This document describes a study for a physical medical device. It does not involve a "training set" in the context of machine learning or artificial intelligence. The performance data presented are for validation testing of the modified device.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no training set for this type of device performance study, this question is not applicable. The "ground truth" for evaluating the test items relates to engineering specifications as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183361

    Validate with FDA (Live)

    Date Cleared
    2019-04-05

    (122 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EkoSonic Endovascular Device with Control Unit 4.0 is indicated for the:

    • Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
    • Infusion of solutions into the pulmonary arteries.
    • Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
    Device Description

    The EkoSonic Endovascular Device with CU 4.0 consists of an EkoSonic Endovascular Device and the Control System 4.0 (Control Unit 4.0 and Connector Interface Cables). The EkoSonic Endovascular Device consists of a disposable infusion catheter with removable ultrasound core. The infusion catheter contains multiple side holes distributed over the length of the treatment zone. The ultrasound core contains up to 30 ultrasound elements, evenly spaced over the treatment zone. Thermal sensors in the treatment zone monitor catheter temperature. The Control System 4.0 generates and controls the delivery of radiofrequency energy to the ultrasound core while monitoring and controlling the temperature of the treatment zone.

    It is the modifications to the Control System 4.0 that are the subject of this submission.

    AI/ML Overview

    The EkoSonic Endovascular Device with Control Unit 4.0 is indicated for ultrasound-facilitated infusion of fluids, including thrombolytics, into the vasculature for pulmonary embolism treatment, infusion into pulmonary arteries, and controlled infusion into the peripheral vasculature.

    The device's substantial equivalence to its predicate (EKOS EkoSonic Endovascular Device with CU 4.0, K162771) was established based on performance data provided in the submission. The modifications in this submission specifically relate to the Control System 4.0.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" with numerical thresholds for performance. Instead, it lists various hardware and software tests conducted, with the reported performance being "Pass" for all of them, indicating compliance with internal standards for the modifications made to the Control System 4.0.

    Test CategorySpecific TestReported Performance
    HardwareSecondary Cell Battery TestsPass
    Reliability PredictionPass
    Battery Hardware Design VerificationPass
    Electrical SafetyPass
    Hardware InspectionPass
    EMC/EMIPass
    RF Board Hardware TestsPass
    System IntegrationPass
    CIC Hardware TestsPass
    Shipping Package TestsPass
    Design Verification Hardware TestsPass
    UI Board Hardware TestsPass
    Environmental TestsPass
    Power Management Module TestsPass
    Short Circuit ResponsePass
    Patched Fastener Design VerificationPass
    EkoSonic Device ReliabilityPass
    SoftwarePeriodic Self-TestPass
    User InterfacePass
    Firmware UpgradePass
    Event LoggingPass
    System Connect ValidationPass
    Single Channel RFPass
    GraphingPass
    Device CompatibilityPass
    Dual Channel RFPass
    Therapy SupportPass
    System StartupPass
    Processor CommunicationPass
    Constant ParameterPass
    Source Code InspectionPass
    AC and Battery PowerPass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes used for the various hardware and software tests, nor does it provide details on data provenance (e.g., country of origin, retrospective/prospective). The tests were likely conducted internally by the manufacturer, EKOS Corporation, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not mention the use of experts to establish a "ground truth" for the test set in the context typically seen for diagnostic or AI-based devices. The performance data here pertains to engineering and software validation of device functionality, not clinical outcome or diagnostic accuracy. Therefore, the concept of expert-established ground truth as a comparison point is not applicable to the type of testing described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The reported tests are engineering and software verification and validation, not clinical studies requiring adjudications by experts for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an endovascular catheter system with a control unit, not an AI-assisted diagnostic or imaging device that would typically undergo MRMC studies with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a physical medical device (catheter and control unit) that requires physician involvement. The "standalone" performance in this context refers to the functionality of the device components as tested in isolation or in simulated environments, as reflected in the hardware and software tests. The document confirms that "Testing has confirmed that the EkoSonic Endovascular Device with CU 4.0 functions as intended."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" in the context of clinical outcomes or pathology is not directly applicable to the performance data presented. The "ground truth" for these engineering and software tests would be the established design specifications, functional requirements, and recognized industry standards for safety and performance of such medical devices. The predicate device (K162771) serves as the basis for demonstrating substantial equivalence based on technological characteristics and intended use.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML device that requires a training set in the conventional sense. The "training" for the device, if any, would be its design and manufacturing processes, which are verified through the listed tests.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182324

    Validate with FDA (Live)

    Date Cleared
    2018-11-26

    (91 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EkoSonic Endovascular System is indicated for the:

    • Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
    • Infusion of solutions into the pulmonary arteries.
    • Controlled and selective infusion of physician-specified fluids, into the peripheral vasculature.
    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter for the EkoSonic MACH 4 Endovascular Device with EkoSonic Control Unit. It primarily focuses on the regulatory aspects and indications for use, and does not contain information regarding detailed acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert qualifications.

    Therefore, I cannot provide the requested information from the provided text. To answer your questions, I would need a different document, such as a summary of safety and effectiveness (SSE) or a clinical study report that details the performance evaluation of the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172035

    Validate with FDA (Live)

    Date Cleared
    2018-03-24

    (262 days)

    Product Code
    Regulation Number
    882.5550
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EKOS Ultrasound EVD System is intended to be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) as a means of reducing intracranial pressure and CSF volume.

    Device Description

    Not Found

    AI/ML Overview

    This is a 510(k) premarket notification for the EKOS Ultrasound EVD System, which is a ventricular catheter. The document does not describe the acceptance criteria or a study proving the device meets those criteria. It is a regulatory letter from the FDA confirming substantial equivalence to legally marketed predicate devices and outlining general controls and regulations.

    Therefore, I cannot provide the requested information from the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162771

    Validate with FDA (Live)

    Date Cleared
    2017-02-24

    (144 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EkoSonic® Endovascular System is indicated for the:

    • · Ultrasound facilitated, controlled and selective infusion of physician-specified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.
    • · Infusion of solutions into the pulmonary arteries.
    • · Controlled and selective infusion of physician-specified fluids, into the peripheral vasculature.
    Device Description

    The EkoSonic Endovascular System consists of an EkoSonic Device and the Control Unit 4.0 (CU 4.0). The EkoSonic Device consists of a disposable infusion catheter with removable ultrasound core. The infusion catheter contains multiple side holes distributed over the length of the treatment zone. The ultrasound core contains up to 30 ultrasound elements, evenly spaced over the treatment zone. Thermal sensors in the treatment zone monitor catheter temperature. The CU 4.0 generates and controls the delivery of radiofrequency energy to the ultrasound core while monitoring and controlling the temperature of the treatment zone.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the EkoSonic® Endovascular System with Control Unit 4.0. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with clinical endpoints and detailed acceptance criteria for an AI/ML-driven medical device. Therefore, much of the requested information regarding acceptance criteria, sample sizes for test/training sets, ground truth establishment, expert qualifications, and MRMC studies is not present in the provided text.

    Based on the information provided, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied to be that the new Control Unit 4.0 (CU 4.0) functions as intended and is substantially equivalent to the predicate device, the EkoSonic Control Unit (PT-3B), without raising new questions of safety or effectiveness. The performance data presented consists of a series of hardware and software tests, all of which "Pass."

    CategorySpecific TestAcceptance Criteria (Implied)Reported Device Performance
    Hardware TestingSystem Tip TestPassPass
    Secondary Cell Battery TestPassPass
    Reliability PredictionPassPass
    Battery Life PredictionPassPass
    Electrical SafetyPassPass
    Hardware InspectionPassPass
    EMC/EMIPassPass
    RF Board TestsPassPass
    System IntegrationPassPass
    CIC Hardware TestsPassPass
    Shipping Package TestsPassPass
    Design Verification Hardware TestsPassPass
    UI Board Hardware TestsPassPass
    Environmental TestsPassPass
    Power Management Module TestsPassPass
    Software TestingPeriodic Self-TestPassPass
    User InterfacePassPass
    Firmware UpgradePassPass
    Event LoggingPassPass
    System Connect ValidationPassPass
    Single Channel RFPassPass
    GraphingPassPass
    Device CompatibilityPassPass
    Dual Channel RFPassPass
    Therapy SupportPassPass
    WorkflowPassPass
    System StartupPassPass
    Processor CommunicationPassPass
    Constant ParameterPassPass
    Source Code InspectionPassPass
    AC and Battery PowerPassPass

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of a typical AI/ML algorithm evaluation (e.g., a set of medical images or patient data). The evaluation appears to be based on engineering verification and validation testing of the hardware and software components of the device itself. Therefore, sample sizes for data (e.g., patient data) are not applicable or provided. The data provenance is internal testing performed by the manufacturer, EKOS Corporation, Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This submission is for a hardware/software device modification that does not involve diagnostic interpretation using AI/ML with a ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The tests performed are engineering-based, not clinical evaluations requiring adjudication of expert opinions.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC study is not mentioned as this device is a continuous flush catheter system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable. The device is a physical system (catheter, control unit) that performs a therapeutic function (ultrasound-facilitated infusion), not a standalone algorithm. The "performance" refers to the device's functional integrity as a system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for the engineering tests performed would be the design specifications and expected functional behavior of the hardware and software components.

    8. The sample size for the training set

    Not applicable. This document describes a device modification, not the development or training of an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable. There is no mention of a "training set" in the context of AI/ML algorithm development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140151

    Validate with FDA (Live)

    Date Cleared
    2014-05-21

    (119 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EkoSonice Endovascular System is indicated for the ultrasound facilitated, controlled and selective infusion of physicianspecified fluids, including thrombolytics, into the vasculature for the treatment of pulmonary embolism.

    Device Description

    Not Found

    AI/ML Overview

    I apologize, but the provided text does not contain information about the acceptance criteria, device performance, or any study details. The documents are standard FDA 510(k) clearance letters for the EkoSonic Endovascular System, primarily focused on administrative changes, regulatory classification, and indications for use. There is no information in these documents about how the device meets acceptance criteria or details of a study proving its performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1