Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    DEN190023

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-21

    (309 days)

    Product Code
    Regulation Number
    866.5970
    Type
    Direct
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AmplideX® Fragile X Dx & Carrier Screen Kit is an in vitro diagnostic device that uses polymerase chain reaction (PCR) and capillary electrophoresis to detect and identify the number of cytosine-guanine (CGG) repeats in the fragile X mental retardation-1 (FMR1) gene using genomic DNA isolated from peripheral whole blood specimens. It is solely intended as an aid in the post-natal diagnosis of fragile X syndrome, and fragile Xassociated disorders [i.e., fragile X-associated tremor/ataxia syndrome (FXTAS) or fragile Xassociated primary ovarian insufficiency (FXPOI)], and for carrier testing in adults of reproductive age. Assay results are solely intended to be interpreted by healthcare professionals who are board certified in molecular genetics and to be used in conjunction. with other clinical and diagnostic findings, consistent with professional standards of practice. Reflex testing, clinical genetic evaluation, and genetic counseling should be offered as appropriate. The test is for use on the 3500Dx Series Genetic Analyzer.

    This test is not indicated for use for fetal diagnostic testing, newborn screening or for standalone diagnostic purposes.

    Device Description

    The AmplideX Fragile X Dx & Carrier Screen Kit (hereafter referred to as the AmplideX Kit) includes reagents sufficient for 100 reactions and are for use on the Applied Biosystems® 3500 Dx Series Genetic Analyzer (8 and 24 capillary) with AmplideX reporter software.

    AI/ML Overview

    The AmplideX® Fragile X Dx & Carrier Screen Kit is an in vitro diagnostic device used to detect and identify the number of cytosine-guanine (CGG) repeats in the FMR1 gene, which is associated with fragile X syndrome and related disorders. The device uses PCR and capillary electrophoresis.

    Here's a breakdown of the acceptance criteria and the studies proving the device's performance:

    1. Acceptance Criteria and Reported Device Performance

    The device's performance was evaluated against specific criteria for genotype categorization and allele size precision.

    CriterionAcceptance CriteriaReported Device Performance
    Precision (CGG Repeat Length)- 1-70 CGG repeats: ± 1 - 71-120 CGG repeats: ± 3 - 121-199 CGG repeats: ± 5% - ≥ 200 CGG repeats: N/A (reported as >200)Three-site Reproducibility: - For 10 out of 11 samples, >97% genotype agreement was observed for categorical calls. Sample ASGN-018 (Full Mutation mosaic) showed 54.7% agreement due to the mosaic allele not being found in 81/179 replicates, prompting a limitation statement in IFU. - Reproducibility of expected allele lengths (excluding mosaic alleles) showed >98% within target precision range. Lot-to-Lot Reproducibility: - >99% categorical genotype agreement was observed for all samples (excluding invalids). Samples ASGN-112 and ASGN-016 were not 100% due to third alleles in a few replicates. - Reproducibility of all alleles (excluding mosaic alleles) showed 100% within target precision specifications. Mosaicism Reproducibility/LoD: - Greater than 95% hit rate demonstrated for claimed LoD for mosaic alleles (Table 9 shows various MAFs from 2.0% to 10.0% achieved 94.4% to 100% detection).
    Diagnostic Performance (Full Mutation)PPA, NPA, and OPA ≥ 95%Full Mutation vs. Southern Blot: - PPA: 95.71% (95% CI: 88.1-98.5) - NPA: 99.30% (95% CI: 96.0-99.9) - OPA: 97.6% (95% CI: 94.5-99.0) - One discordant sample (194-repeat AmplideX, >200 Southern Blot) was borderline.
    Diagnostic Performance (Premutation)PPA, NPA, and OPA ≥ 95%Premutation vs. Normal or Intermediate (Southern Blot): - PPA: 100% (95% CI: 94.7-100.0) - NPA: 97.10% (95% CI: 90.0-99.2) - OPA: 98.6% (95% CI: 94.9-99.6)
    Carrier Screening PerformanceOverall agreement ≥ 95% for Full Mutation/Premutation, Intermediate, or Normal genotypes.AmplideX vs. FMR1 Dual-PCR Reference Method: - Premutation/Full Mutation: 100% (95% CI: 94.7-100.0) - Intermediate: 85.7% (95% CI: 75.7-92.1) - Normal: 98.6% (95% CI: 92.2-99.7) - Note: When considering assay precision, 8 borderline samples (54/55 CGG) were reclassified as concordant, increasing Intermediate agreement to 97.1%.
    LoD for mosaic alleles> 95% hit rate for claimed LoD- Intermediate on Normal: 2.0% MAF, 100.0% detected - Premutation on Normal: 2.0% MAF, 94.4% detected - Full Mutation on Normal: 6.1% MAF, 100.0% detected - Normal on Premutation: 2.0% MAF, 100.0% detected - Premutation on Premutation: 5.0% MAF, 97.1% (low), 100.0% (high) detected - Full Mutation on Premutation: 7.0% MAF, 100.0% detected - Normal on Full Mutation: 2.0% MAF, 100.0% detected - Premutation on Full Mutation: 10.0% MAF, 100.0% detected
    DNA Input Range (20-80 ng)100% categorical genotype agreement for valid samples, all expected allele peaks within target precision range, and acceptable QC failure rates within the recommended range.- 100% categorical genotype agreement and all expected allele peaks within target precision for all valid samples and all input levels. - QC failure rates: 1 ng (6.25%), 10 ng (1.88%), 20 ng (0.6%), 40 ng (1.3%), 80 ng (0%), 160 ng (0.63%). These results support the recommended 20-80 ng range.
    DNA Extraction Methods100% categorical genotype agreement and 100% of all measured alleles within target precision range for common commercial DNA extraction methods.- 100% categorical genotype agreement and 100% of all measured alleles within target precision range for the three common commercial DNA extraction methods tested (solution precipitation, manual silica spin column, and manual magnetic bead).
    Thermal Cycler Equivalence100% genotype category calls agreement and 100% of all identified independent alleles within their respective target precision ranges across all thermal cycler types.- 100% genotype category calls agreement and 100% of all identified independent alleles within their respective target precision ranges across three separate thermal cyclers.
    Interfering Substances (Hemoglobin, EDTA)For processable samples, >95% genotype category agreement and 100% of alleles within expected precision range. (Limitation statement regarding impact on assay failure rate for excess hemoglobin/EDTA is included in IFU).- Effects on DNA concentration yield observed (hemoglobin reduced, EDTA enhanced). - For samples that could be processed, genotype category agreement was >95% and 100% of alleles within expected precision range. (Limitation statement for IFU).
    Specimen Handling Stability100% genotype category agreement and all expected allele peaks within target precision for DNA extracted from whole human blood stored at 2-8°C for up to 2 weeks.- 100% genotype category agreement and all expected allele peaks within target precision for all timepoints across all samples (up to 2 weeks storage at 2-8°C).
    DNA Freeze/Thaw Stability100% genotype category agreement and all expected allele peaks within target precision for extracted DNA across (b)(4) freeze/thaw cycles.- 100% genotype category agreement and all expected allele peaks within target precision for up to (b)(4) freeze/thaw cycles.
    PCR & CE Reagent Product StabilityPCR Product: 100% categorical genotype agreement and >95% of measured alleles within target precision for PCR products stored up to 3 days at 2-8°C. CE Product: 100% categorical genotype agreement and 100% of measured alleles within target precision for samples prepared for CE analysis and stable at ambient temperature for up to 24 hours.- PCR Product: 100% categorical genotype agreement for all timepoints across all samples and >95% of all measured alleles within target precision range (up to 3 days at 2-8°C). - CE Product: 100% categorical genotype agreement for all timepoints across all samples and 100% of all measured alleles within target precision range (up to 24 hours at ambient temperature).
    Real Time Kit Stability100% categorical genotype agreement and 100% of measured alleles within target precision. (Supports a minimum of 1 year, with ongoing study for 24 months).- 100% categorical genotype agreement for all timepoints across all samples and 100% of all measured alleles within target precision range. (Supports at least 1-year storage stability at present).
    Freeze/Thaw Kit Stability100% categorical genotype agreement and 100% of measured alleles within target precision for up to eight freeze/thaw cycles.- 100% categorical genotype agreement for all timepoints across all samples and 100% of all measured alleles within target precision range (up to eight freeze/thaw cycles).
    Kit Shipping Stability100% of measured alleles within target precision range and 100% genotype category agreement across different shipping configurations.- 100% of their measured alleles within the target precision range and 100% genotype category agreement across all different shipping configurations (based on ISTA 7D, 24-hour summer profile).
    WHO Standards Testing AccuracyCorrectly classifies and sizes international Fragile X reference standards.- The test correctly classified and sized the 5 WHO International Standard Fragile X Syndrome Reference Panel samples (NIBSC code: 08/158) with 100% of measured values falling within the expected CGG range, even when reported as ">200" for full mutations.

    2. Sample Sizes and Data Provenance (Test Set)

    • Precision (Reproducibility) Studies:
      • Three-site reproducibility: 11 samples (7 clinical whole blood specimens, 4 contrived from cell lines spiked into leukocyte-depleted whole human blood). This yielded 1980 total sample measurements (excluding controls).
      • Lot-to-lot reproducibility: 11 samples (same panel as above). This yielded 2376 total sample measurements (excluding controls).
      • Mosaicism reproducibility/LOD: Panel of mixed cell line DNA with known genotypes and a panel of 30 samples involving mixing DNA from clinical specimens. This yielded 36 replicate measurements per specimen.
    • DNA Input Study: 8 samples (6 clinical, 2 contrived). This resulted in 960 sample measurements (20 replicates per dilution level per sample).
    • Clinical Performance Studies:
      • Diagnostic Performance: 207 leftover clinical specimens (111 female, 96 male) obtained from patient samples submitted for routine FMR1 5'UTR mutation testing across multiple clinical sites.
      • Carrier Screening Performance: (b)(4) specimens from females ((b)(4) years of age) enrolled across (b)(4) clinical sites, resulting in 207 evaluable subjects.
      • Mosaic Allele Calling Accuracy: 49 specimens identified as mosaic by AmplideX across the two clinical studies (27 from diagnostic, 22 from carrier screening). An additional 40 randomly selected samples were further evaluated for mosaicism resolution after initial discordant cases were addressed.
      • WHO Standards Testing: 5 samples from the WHO International Standard Fragile X Syndrome Reference Panel (NIBSC code: 08/158). This yielded 135 sample measurements (excluding control).

    Data Provenance: The data primarily appears to be retrospective for the clinical performance studies, using "leftover specimens" and "patient samples submitted for routine FMR1 5'UTR mutation testing" or "enrolled across clinical sites." The analytical performance studies used a mix of clinical specimens and contrived samples (cell lines spiked into blood). The document does not explicitly state the country of origin for the clinical data.

    3. Number of Experts and Qualifications (Ground Truth for Test Set)

    The document does not explicitly state the number of experts or their specific qualifications for establishing the ground truth for the clinical test sets (Southern Blot and FMR1 Dual-PCR Reference Method). However, it mentions that the interpretation of assay results is "solely intended to be interpreted by healthcare professionals who are board certified in molecular genetics."

    4. Adjudication Method (Test Set)

    The document does not describe a formal adjudication method for the test set.

    • In the diagnostic performance study, results were compared directly against the reference methods (Southern Blot or FMR1 Dual-PCR). Discordant samples were noted, and in one case for full mutation, a borderline sample was identified.
    • For mosaicism, there was a manual evaluation of orthogonal PCR method results for cases discordant with AmplideX.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study to assess the effect size of human readers improving with AI vs. without AI assistance. This device is a diagnostic kit for laboratory use, and its performance is evaluated against reference methods, not typically in a human-in-the-loop context.

    6. Standalone Performance Study (Algorithm Only)

    Yes, the studies described are for standalone performance (algorithm only, as applied to the output of the PCR and capillary electrophoresis). The AmplideX Fragile X Reporter Software performs automated analysis and categorization of FMR1 alleles based on size. The performance is assessed by comparing the device's output (categorical calls and CGG repeat lengths) directly against established reference methods (Southern Blot, FMR1 Dual-PCR) or known values (WHO standards, contrived samples).

    7. Type of Ground Truth Used

    • Clinical Performance (Diagnostic):
      • Southern Blot: For Full Mutation classification and Premutation vs. Normal or Intermediate assessment. Southern Blot is a molecular biology method considered a gold standard for detecting large FMR1 expansions.
    • Clinical Performance (Carrier Screening):
      • FMR1 Dual-PCR Test: An alternate orthogonal independently validated PCR assay was used as the comparator for carrier screening, as Southern blot is noted to not have accurate sizing in the premutation range.
    • Analytical Performance (Precision, LoD):
      • Clinical Samples: Whole blood specimens from patients with known FMR1 genotypes.
      • Contrived Samples: Genomic DNA extracted from cell lines with known gender, genotype, and expected allele sizes, sometimes spiked into leukocyte-depleted whole human blood.
      • Mixed Cell Line DNA: With known genotypes and mosaic allele frequencies for LoD studies.
    • WHO Standards Testing:
      • WHO International Standard Fragile X Syndrome Reference Panel (NIBSC code: 08/158): These are internationally established reference materials with characterized mean repeat lengths and ranges.

    8. Sample Size for the Training Set

    The document does not provide a specific sample size for a "training set." This device is a diagnostic assay with software for interpretation, but the context here suggests a traditional verification and validation study rather than a machine learning model where a distinct training set (for model parameters) and test set (for performance evaluation) would be explicitly separated and reported. The analytical and clinical studies serve as the validation of the device's performance.

    9. How the Ground Truth for the Training Set Was Established

    As mentioned above, there isn't an explicit "training set" in the context of a machine learning-based device. The ground truth for the reference materials and clinical samples used in the validation studies was established through:

    • Reference Methods: Southern Blot analysis and an independently validated FMR1 Dual-PCR assay are considered the ground truth for clinical performance.
    • Known Characteristics: For contrived samples and cell lines, the genotypes and allele sizes are "known" based on prior characterization.
    • International Standards: The WHO/NIBSC standards have established and published characterizations from a study consortium.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN160003

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-22

    (185 days)

    Product Code
    Regulation Number
    866.6060
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuantideX qPCR BCR-ABL IS Kit is an in vitro nucleic acid amplification test for the quantitation of BCR-ABL1 and ABL1 transcripts in total RNA from whole blood of diagnosed t(9,22) positive Chronic Myeloid Leukemia (CML) patients expressing BCR-ABL1 fusion transcripts type e13a2 and/or e14a2. The QuantideX qPCR BCR-ABL IS Kit is a reverse transcription-quantitative PCR performed on the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument and is intended to measure BCR-ABL1 to ABL1, expressed as a log molecular reduction (MR value) from a baseline of 100% on the International Scale, in t(9;22) positive CML patients during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs).

    The test does not differentiate between e13a2 or e14a2 fusion transcripts and does not monitor other rare fusion transcripts resulting from t(9:22). This test is not intended for the diagnosis of CML.

    Device Description

    The QuantideX qPCR BCR-ABL IS Kit reagents are adapted for use on the ABI 7500 Fast Dx Real-Time PCR Instrument. The assay includes reagents sufficient for 60 reactions. A description of the reagents provided is described below in Table 1.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes both analytical and clinical performance acceptance criteria.

    Analytical Performance - Precision/Reproducibility

    Performance MetricAcceptance Criteria (SD)Reported Device Performance (Max Total SD)Met?
    Within-lab Precision
    MR Value < 3.5≤ 0.210.128 (at MR 1)Yes
    MR Value 3.5-4.25≤ 0.290.111 (at MR 3.5)Yes
    MR Value > 4.25≤ 0.360.134 (at MR 4)Yes
    Site-to-Site Reproducibility
    MR Value < 3.5≤ 0.210.117 (at MR 3)Yes
    MR Value 3.5-4.25≤ 0.290.169 (at MR 3.5)Yes
    MR Value > 4.25≤ 0.360.167 (at MR 4)Yes

    Analytical Performance - Linearity

    Performance MetricAcceptance Criteria (Slope/Intercept for Regression)Reported Device Performance (Slope/Intercept/Max SD)Met?
    Linear RegressionSlope close to 1 (not explicitly defined but implied typical regression expectations), intercept close to 0 (implied)e13a2: Slope 1.01, Intercept -0.11, Max SD 0.17 e14a2: Slope 1.01, Intercept -0.05, Max SD 0.17Yes
    Range of Linearity(Implied based on clinical use range)MR 0.3 (50% IS) to MR 4.7 (0.002% IS) for both transcriptsYes

    Analytical Performance - Limit of Quantitation (LoQ)

    Performance MetricAcceptance Criteria (SD)Reported Device Performance (MR Range/SD Range)Met?
    LoQ≤ 0.36 at MR 4.5 or greaterMR values 4.6 to 4.87, SD values 0.23 to 0.34Yes

    Analytical Performance - Interfering Substances

    Performance MetricAcceptance Criteria (Mean difference from control)Reported Device PerformanceMet?
    InterferenceMean of test samples ± 0.5 MR of the controlAll met acceptance criteriaYes

    Analytical Performance - Primer Specificity

    Performance MetricAcceptance CriteriaReported Device PerformanceMet?
    SpecificitySpecimens without a major BCR-ABL1 breakpoint reported as negative or below LoD in > 8/9 replicatesMet acceptance criteria (e.g., cell lines and IVT controls without e13a2/e14a2 were <LOD in 8/9 or 9/9 replicates)Yes

    Analytical Performance - Specimen Carryover Contamination

    Performance MetricAcceptance CriteriaReported Device PerformanceMet?
    Carryover> 95% of CML-negative samples tested negative or below LoDAll negative specimens (25/25) reported as "Negative (sufficient ABL1)"Yes

    Analytical Performance - RNA Input

    Performance MetricAcceptance Criteria (SD criteria from Table 3)Reported Device Performance (RNA input range with met criteria)Met?
    RNA Input RangeSame as Table 3 for corresponding MR values750 to 6000 ng from MR 1 to MR 4.5Yes

    Analytical Performance - Traceability

    Performance MetricAcceptance Criteria (Regression to WHO panel)Reported Device Performance (Slope/Intercepts for 3 lots)Met?
    TraceabilitySlope 1.0 to 1.1, Intercept 0.02 - 0.11Lot1: y=-0.11+1.1x; Lot2: y=0.02+1x; Lot3: y=-0.059+1.1xYes

    Analytical Performance - Specimen Stability (Whole Blood)

    Performance MetricAcceptance Criteria (MR unit difference from baseline)Reported Device Performance (Duration and storage conditions)Met?
    Whole Blood StabilityAll results within 0.5 MR units from baselineUp to 72 hours at 2-8°CYes

    Clinical Performance

    Performance MetricAcceptance CriteriaReported Device PerformanceMet?
    Event-Free Survival (EFS) Difference at 36 monthsStatistically significantly different AND point estimates differing by at least 10 percentage points (for MR < 3 vs MR ≥ 3 groups)EFS difference: 22.2% (95% CI: 2.0%-42.4%), p = 0.0279Yes
    Event Status Known at 36 months95% Wilson score confidence interval for the difference > 0MetYes

    2. Sample Size Used for the Test Set and Data Provenance

    The primary test set for clinical performance involved a retrospective study with:

    • Sample Size: 137 evaluable samples from 96 subjects (out of 139 samples from 98 patients initially collected).
    • Data Provenance: Retrospective, collected at 2 clinical sites in the US (OHSU and Hospital of the University of Pennsylvania).

    For analytical performance studies, various sample compositions and quantities were used:

    • Precision (within-lab & site-to-site): 25 samples formulated from 5 human RNA specimens positive for BCR-ABL1 diluted into RNAs from CML-negative blood. The site-to-site reproducibility study involved 1200 measurements from this 25-sample pool.
    • Linearity/Reportable Range: 2 separate RNA specimens (one e13a2, one e14a2) diluted into RNAs from CML-negative whole blood (ranging from MR 0.1 to MR 4.8).
    • Limit of Blank (LoB): 30 non-leukemic human RNA specimens.
    • Limit of Detection (LoD): 4 separate human RNA specimens positive for BCR-ABL1, serially diluted to 28 levels.
    • Limit of Quantitation (LoQ): 6 specimens derived from 6 human RNA positive for BCR-ABL (diluted to target MR 4.7).
    • Interfering Substances: Residual CML-positive blood diluted to approximately MR 4.0 into RNA from CML-negative whole blood (tested in 9 replicates).
    • Primer Specificity: 11 leukemic specimens (CML, AML, ALL) and 2 non-leukemic RNA specimens.
    • Specimen Carryover Contamination: Serially diluted total RNA from residual clinical CML-positive blood into RNA from CML-negative whole blood in a checkerboard pattern (25 high positive, 25 negative).
    • RNA Isolation: CML-positive white blood cells serially diluted across 4 levels (MR 1-4) into CML-negative human anti-coagulated whole blood.
    • RNA Input: 30 samples derived from 2 primary RNA samples diluted into non-leukemic human RNA (total 216 evaluable observations).
    • Traceability: WHO Reference Panel (4 panel members tested in duplicate).
    • Reagent Stability: 5 samples diluted to approximately MR 1, 2, 3, 4, and < LoD (15 replicates per time point).
    • Specimen Stability (Whole Blood): 13 CML positive specimens and 3 CML negative specimens.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their specific qualifications (e.g., radiologist with X years of experience) used to establish the ground truth for the test set.

    For the clinical study, the "ground truth" for event-free survival (EFS) was based on clinical outcomes defined by specific criteria:

    • Death by any cause
    • Development of accelerated-phase or blast crisis CML
    • Loss of complete hematologic response
    • Loss of complete cytogenetic response
    • Appearance of mutation
    • Change in TKI treatment not due to toxicity and not due to the results of another BCR-ABL assay

    These outcomes would typically be determined by treating physicians and medical records, often relying on established clinical guidelines and diagnostic criteria. While not explicitly stated as "experts establishing ground truth," the medical documentation and clinical assessments would serve this role.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1, 3+1, none) for a test set, particularly in the context of expert review. The ground truth for the clinical study was based on observed clinical outcomes and treatment changes, which are derived from patient records and physician decisions rather than expert consensus on image or assay interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a quantitative in vitro diagnostic assay, not an imaging device or a device requiring human interpretation in the loop in the same way an MRMC study evaluates. The clinical study assessed the device's ability to predict clinical outcomes (EFS) based on its quantitative output (MR value), rather than comparing human reader performance with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done

    Yes, the primary clinical performance and all analytical performance studies represent a standalone assessment of the device. The QuantideX qPCR BCR-ABL IS Kit is an in vitro diagnostic test that provides quantitative results (MR values) directly. There is no human-in-the-loop interpretation of the assay's raw output described; the device itself calculates the BCR-ABL1 to ABL1 ratio and MR value. The clinical study then correlated these device-generated MR values with patient outcomes.

    7. The Type of Ground Truth Used

    • Analytical Ground Truth: For many analytical performance studies (precision, linearity, LoD, LoQ, interference), the ground truth was established by known concentrations or dilutions of BCR-ABL1 and ABL1 transcripts, or by using characterized RNA specimens (e.g., CML-positive or CML-negative clinical samples, or cell lines with specific translocations) to create spiked samples or panels. Traceability was established against the First (1st) WHO International Genetic Reference Panel.
    • Clinical Ground Truth: For the clinical performance study, the ground truth was patient outcomes data, specifically "event-free survival" at 32-40 months after TKI treatment. "Events" were defined by specific clinical criteria related to disease progression, death, or changes in therapy.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set size. As an in vitro diagnostic kit, the "training" analogous to machine learning would be the extensive analytical validation and optimization performed by the manufacturer during product development, which informs the assay's design, protocols, and reagent formulations. The clinical and analytical studies described are primarily for validation and verification of the final device, not for training a specific algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Since a distinct "training set" in the machine learning sense is not explicitly mentioned or applicable for this type of IVD kit, the concept of establishing ground truth for a training set does not directly apply. Instead, the analytical characteristics (like limit of detection, linearity, precision) are established using controlled experiments with known concentrations and reference materials, and clinical utility is demonstrated by correlating the device's measurements with observed patient clinical outcomes. These outcomes in the clinical study are determined by medical definitions and patient records, as described in point 3.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113420

    Validate with FDA (Live)

    Device Name
    RNARETAIN
    Manufacturer
    Date Cleared
    2012-03-16

    (119 days)

    Product Code
    Regulation Number
    866.4070
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RNARetain® device is a single-use, prefilled container intended for the collection, storage, and transportation of fresh breast tissue specimens for subsequent RNA isolation and further molecular diagnostic testing. For Professional Use Only.

    Device Description

    The RNARetain® device consists of an aqueous, hypertonic tissue preservation solution that is provided in a single-use, non-sterile vial intended to serve as the container for the collection, storage, and transport of breast tissue specimens. The ability of the solution formulation to preserve fresh tissue and nucleic acids within the tissue is due to its rapid permeation of tissue to protect cellular nucleic acids from nucleases that would otherwise rapidly degrade the nucleic acids within the specimen. The demonstrated inhibitory effect of the RNARetain® solution on the growth of bacteria, yeast and mold protects the specimen from inadvertent microbial contamination during storage or usage of the device.

    AI/ML Overview

    The RNARetain® device is a single-use, prefilled container intended for the collection, storage, and transportation of fresh breast tissue specimens for subsequent RNA isolation and further molecular diagnostic testing. The device's performance was evaluated through several non-clinical analytical studies, including precision/reproducibility, stability of MammaPrint® outcomes, comparison with a predicate device, and internal analytical performance by Asuragen.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Reproducibility of RNA Isolation & MammaPrint® Index: No statistically significant difference in MammaPrint® risk group assignment or MammaPrint® index between replicate RNA isolations.Precision/Reproducibility (K070675): Five previously analyzed tumor samples (one borderline, two high risk, two low risk) were isolated in duplicate. No statistically significant difference in MammaPrint® risk group assignment or MammaPrint® index was observed between the two separate RNA isolations. (Table 3)
    Stability of MammaPrint® outcomes in RNARetain®: MammaPrint® indices from samples stored in RNARetain® should be stable and comparable to immediately snap-frozen samples.Stability of MammaPrint® outcomes in RNARetain® (K070675): Tumors shipped in RNARetain® showed greater stability in MammaPrint® index (Std dev 0.022) compared to immediately frozen samples (Std dev 0.042). An unpaired T-test showed no significant difference in MammaPrint® indices (p=0.24) between RNARetain® and frozen samples. The difference (Δ 0.027) was within acceptable limits.
    Equivalence to Flash-Frozen Method for RNA Quality: RNARetain® preserved RNA yield, purity, and integrity should be equivalent to flash-frozen tissue.Comparison Studies (K070675): A set of 33 breast tumor samples showed a median difference of 0.070 in MammaPrint® Indices between RNARetain®-preserved and snap-frozen samples. Pearson correlation was 0.94 and regression analysis showed R=0.90, indicating high similarity. This was comparable to the variation observed between two snap-frozen samples (median difference 0.105), with no statistically significant difference (t-test, p=0.57) between the two comparison series.
    RNA Purity (A260/A280): ≥ 1.6Analytical Performance Asuragen (Table 4): Subject 1: RNARetain® AVG 1.97, Frozen AVG 1.92Subject 2: RNARetain® AVG 2.03, Frozen AVG 2.01Subject 3: RNARetain® AVG 1.96, Frozen AVG 1.96All values met the criteria.
    RNA Integrity (28S:18S): ≥ 1.0 or RIN ≥ 7.0Analytical Performance Asuragen (Table 4): Subject 1: RNARetain® AVG 1.5, Frozen AVG 1.6Subject 2: RNARetain® AVG 1.6, Frozen AVG 1.5Subject 3: RNARetain® AVG 1.5, Frozen AVG 1.5All values met the criteria (implied, as 28S:18S ratios are ≥ 1.0).
    RNA Yield: Comparable to flash-frozen specimens.Analytical Performance Asuragen (Table 4): Subject 1: RNARetain® AVG 148, Frozen AVG 131Subject 2: RNARetain® AVG 62, Frozen AVG 43Subject 3: RNARetain® AVG 101, Frozen AVG 81Yields were comparable or higher in RNARetain® samples.
    Sample Stability in RNARetain®: Acceptable RNA yield, purity, and integrity over various storage conditions and durations.Sample Stability (Asuragen): Samples passed acceptance criteria up to 3 days at 35-39 °C, 15 days at 18-25 °C, 60 days at 2-8 °C, and up to 3 years at -15 to -30 °C.
    RNARetain® Reagent Stability: Reagent stability at room temperature.RNARetain® Reagent Stability (Asuragen): 8mL vial (6 mL volume) stable up to 20 months at room temperature. 6 mL vial (5 mL volume) and 2 mL vial (1 mL volume) configurations stable up to 36 months. All passed acceptance criteria.

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Reproducibility (K070675): 5 tumor samples, each isolated in duplicate (total 10 isolations). Origin: Not explicitly stated, but part of Agendia's MammaPrint® submission. Retrospective.
    • Stability of MammaPrint® outcomes in RNARetain® (K070675): One tumor, with sections preserved in RNARetain® and immediately snap-frozen. Both sections were labeled 5 times. Origin: Not explicitly stated, but part of Agendia's MammaPrint® submission. Retrospective.
    • Comparison Studies (K070675):
      • RNARetain® vs. snap-frozen: 33 breast tumor samples. One part of each sample stored in RNARetain®, another part snap-frozen. Origin: Collected in 2003 as a pilot study for the Dutch Raster clinical trial. Retrospective.
      • Frozen vs. frozen: 18 tumors, with two parts of each immediately snap-frozen. Origin: Collected in the same time period as the 33 breast tumor samples. Retrospective.
    • Repeatability and Reproducibility (Asuragen): 10 adjacent pairs of breast tissue from 3 subjects (total 30 tissue samples). Origin: Asuragen, within the US. Prospective (implied, as it was specifically collected for this study).
    • Sample Stability (Asuragen for storage conditions): Human breast cancer cell line MCF-7. Quantitative sample size for each condition not explicitly stated (e.g., how many replicates per condition), but various time points and temperatures were tested. Origin: Asuragen labs.
    • RNARetain® Reagent Stability: 3 lots of the 8mL vial (6 mL volume); unspecified number of lots for 6 mL vial (5 mL volume) and 2 mL vial (1 mL volume). Origin: Asuragen.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Comparison Studies (K070675): "HE stained sections were re-examined by a pathologist to confirm invasive ductal carcinoma and sufficient tumor cell content." Specific number and qualifications of pathologists are not given beyond "a pathologist".

    4. Adjudication Method for the Test Set:

    • Not applicable as the reported studies primarily involve analytical performance metrics (RNA purity, integrity, yield, MammaPrint® index correlation) rather than subjective assessments by multiple readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs. without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is a pre-analytical system for tissue preservation, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance studies described are essentially standalone analytical performance evaluations of the RNARetain® device's ability to preserve RNA, mimicking the "algorithm only" concept in the context of a pre-analytical device. The device's performance is measured directly through RNA metrics and comparison to a gold standard (flash-frozen tissue) or established assay (MammaPrint®).

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    • MammaPrint® Index: The MammaPrint® assay itself serves as a reference, with the "original index" used for reproducibility studies (Table 3), implying an established ground truth for those samples.
    • Pathology: For tissue samples in the comparison studies, "HE stained sections were re-examined by a pathologist to confirm invasive ductal carcinoma and sufficient tumor cell content." This indicates pathology review was used to establish the suitability of the tissue samples.
    • Reference Methods: The "gold standard" for tissue preservation (flash-frozen tissue) was used as a comparator/ground truth for RNA yield, purity, and integrity measurements. Fresh Frozen cells were used as a comparator for sample stability studies.

    8. The Sample Size for the Training Set:

    • Information on a separate "training set" for the RNARetain® device, in the context of an algorithm or learnable system, is not applicable. RNARetain® is a chemical preservation solution and does not involve AI or machine learning that would require a distinct training set. The studies described are validation studies of its chemical and physical performance.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set in the context of an AI/ML algorithm for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1