    K Number
    K212359

    Validate with FDA (Live)

    Date Cleared
    2022-09-21

    (419 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Tissue Technologies PWD™ Platform Wound Dressing Negative Pressure System comprised of the; PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump, and the PWD™ Wound Exudate Canister is indicated in patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, sub-acute and dehisced wounds, partial thickness burns, ulcers (such as diabetic, pressure and venous insufficiencies), flaps, grafts and closed surgical incisions.

    PWD Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.

    Device Description

    The Applied Tissue Technologies PWD™ Platform Wound Dressing Negative Pressure System consists of the PWD™ Dressing, PWD™ Negative Pressure Wound Therapy Pump and the Wound exudate canister. The system provides a negative pressure wound dressing therapy in a moist environment to facilitate the wound healing process. The dressing has an access port through which fluids can be drawn from the wound site. The same port can also be utilized to administer negative pressure wound therapy using the negative pressure pump if required. The negative pressure pump is used with a collection cannister to collect the exudate. The PWD Platform Wound Dressing is supplied in five different shapes and sizes, all having the same intended use.

    AI/ML Overview

    The provided document describes the PWD™ Platform Wound Dressing Negative Pressure System and its substantial equivalence to predicate devices, focusing on performance data from bench testing.

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test NameDescriptionReference StandardsAcceptance CriteriaReported Device Performance
    Water Vapor Transmission RateMeasures the passage of water vapor through the membrane/adhesive barrierASTM 1249Must be at least equivalent to commercially available NPWT dressingPass
    Exudate Throughput Properties Under Negative PressureSimulated Use testing using various exudate types to confirm acceptable fluid flow when connected to the Pump and Canister systemN/ASystem must demonstrate the ability to remove exudates of various types from the wound bedPass
    Maintenance of Negative PressureSimulated use testing from under moist conditions to demonstrate maintenance of negative pressureNAThe PWD must maintain negative pressure at the wound bed for the test durationPass
    UsabilityUsability testing for both healthcare and home use environmentsNAUsers must demonstrate the ability to correctly use the devicePass
    Electrical SafetyComplies with applicable safety standardsIEC 60601-1, IEC 60601-1-2, IEC 60601-1-11Requirements for electrical safety per the applicable safety standards is metPass
    Electro-magnetic CompatibilityComplies with applicable safety standardsComplies with IEC 60601-2-3Requirements for electromagnetic compatibility per the applicable safety standards is metPass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states that the performance data was generated through "Bench testing" for various characteristics. This type of testing typically uses laboratory samples of the device components or simulated systems rather than human subjects or a clinical test set.

    • Sample Size for Test Set: Not specified, but implied to be laboratory samples of device components.
    • Data Provenance: Not applicable in the context of clinical data. The data is from laboratory bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document. For bench testing, ground truth is typically established by engineering specifications, validated test methods, and compliance with industry standards, rather than expert clinicians.

    4. Adjudication Method for the Test Set:

    • This information is not applicable as the described tests are bench tests, not clinical studies requiring adjudication of output from human readers or an algorithm.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. The document describes a 510(k) premarket notification process based on substantial equivalence to predicate devices and bench testing. There is no mention of a clinical trial involving human readers or a comparison of human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This device is a physical medical device (Negative Pressure Wound Therapy System), not an AI algorithm. The performance evaluation focuses on the physical and functional aspects of the system.

    7. The Type of Ground Truth Used:

    • The ground truth for the bench tests appears to be based on engineering specifications, established industry standards (e.g., ASTM, IEC), and the functional requirements for a negative pressure wound therapy system. For instance, "ability to remove exudates of various types from the wound bed" and "maintain negative pressure at the wound bed for the test duration" serve as the functional ground truths.

    8. The Sample Size for the Training Set:

    • Not applicable. This document describes the testing of a physical medical device. There is no mention of an AI component requiring a training set.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. As there is no AI component or training set mentioned, this information is not relevant to the provided document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191460

    Validate with FDA (Live)

    Date Cleared
    2019-07-02

    (29 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Tissue Technologies PWD™ Platform Wound Dressing is intended to be used in conjunction with the Invia Motion Negative Pressure Wound Therapy (NPWT) system and is indicated in patients who would benefit from a suction device (NPWT) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as: diabetic, pressure and venous insufficiencies), flaps and grafts.

    Device Description

    The Applied Tissue Technologies PWDTM Platform Wound Dressing is a single use device. Its transparency permits the healthcare practitioner to observe the progress of the wound healing without removing the wound dressing. There is an access port to administer negative pressure wound therapy. The PWD is intended to be used for a maximum of 3 days. Therapy duration may be less than indicated if clinical practice or other factors require more frequent dressing changes.

    The PWD will be marketed in five different shapes and sizes, all having the same intended use. The five sizes are identified as follows.

    PWDTM Platform Wound Dressing
    Description | Catalog/REF #
    1" Round | AT1070-01
    2" Round | AT1071-01
    3" Round | AT1072-01
    1" x 3" Oblong | AT1073-01
    3" x 5" Oblong | AT1074-01

    AI/ML Overview

    This FDA 510(k) premarket notification for the PWD™ Platform Wound Dressing indicates that it is a device, not an AI/ML algorithm, and therefore the requested information regarding acceptance criteria and studies for an AI/ML device is not applicable.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (PWD™ Platform Wound Dressing, K182409) for a new, larger size of the existing wound dressing. Substantial equivalence for medical devices typically relies on demonstrating that the new device has the same intended use, technological characteristics, and safety and effectiveness as a legally marketed predicate device.

    Here's why the AI/ML specific questions cannot be answered from this document:

    • No AI/ML Component: The device description is for a physical wound dressing used in conjunction with a negative pressure wound therapy system. There is no mention of any AI, machine learning, or software component that performs diagnostic, prognostic, or treatment guidance functions.
    • Performance Data: The "Nonclinical Performance Data" section states that "testing was conducted to demonstrate that the additional dressing size has the same performance characteristics as the predicate device." This typically refers to physical and functional testing relevant to the wound dressing itself (e.g., integrity, adhesion, ability to interface with the NPWT system, material compatibility), not to the accuracy or performance metrics of an AI/ML model against a ground truth.

    Therefore, the requested tables and details about acceptance criteria, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, and training set information are not present in this document because they are characteristic of AI/ML device submissions, not physical medical device modifications like this wound dressing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182409

    Validate with FDA (Live)

    Date Cleared
    2019-01-11

    (129 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Tissue Technologies PWD™ Platform Wound Dressing is intended to be used in conjunction with the Invia Motion Negative Pressure Wound Therapy (NPWT) system and is indicated in patients who would benefit from a suction device (NPWT) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as: diabetic, pressure and venous insufficiencies), flaps and grafts.

    Device Description

    The Applied Tissue Technologies PWD™ Platform Wound Dressing is a single use device. Its transparency permits the healthcare practitioner to observe the progress of the wound healing without removing the wound dressing. There is an access port to administer negative pressure wound therapy. The PWD is intended to be used for a maximum of 3 days. Therapy duration may be less than indicated if clinical practice or other factors require more frequent dressing changes. The PWD will be marketed in four different shapes and sizes, all having the same intended use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Applied Tissue Technologies PWD™ Platform Wound Dressing. This document asserts substantial equivalence to a predicate device, the Medela Ag Invia Foam Dressing Kit (K170088), rather than detailing an independent study to prove specific acceptance criteria for a new AI/ML device.

    Therefore, many of the requested categories are not applicable to the information contained in this FDA submission. The document focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device through performance data and a comparison of technological characteristics.

    Here's an breakdown based on the provided text, highlighting what is available and what is not:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the sense of predefined thresholds for clinical performance metrics (e.g., sensitivity, specificity, or improvement in a specific clinical outcome). Instead, it relies on demonstrating that the device's technical performance characteristics are acceptable and comparable to a predicate device.

    Performance MetricAcceptance Criteria (Implied by Comparison)Reported Device Performance (PWD™ Platform Wound Dressing)
    Device Vapor TransmissionAcceptableAcceptable test results
    Ability to Maintain Negative PressureAcceptableAcceptable test results
    Ability to Remove ExudateAcceptableAcceptable test results
    BiocompatibilityAccording to ISO 10993 standard seriesAcceptable test results
    Maximum Duration of a single dressing72 Hours (Same as predicate)72 Hours
    Negative Pressure to the wound surfacerequirements in mmHg125 mmHg (Predicate)80 mmHg (nominal) +/- 20mmHG to the wound surface. Noted as a nominal recommended setting, pump settings range from -40 to -175. This difference does not raise new questions of safety.
    Multiple Shapes for Wound Sizes up to 400cc240 - 2700cc (Predicate)400cc (within the range of the predicate device)
    Dressing ConfigurationSimilar to predicate (Clear polyurethane occlusive drape dressing with foam insert)Clear polyurethane embossed occlusive drape with tubing. (Similar materials and configuration; embossed layer allows fluid transfer instead of foam, deemed not to raise new questions of safety)
    Access PortSame as predicateFor attachment to the negative pressure pump; for removal of fluids.
    Collection PackSame as predicatePWD tubing connected to Medela Invia Motion canister/tubing set for fluid collection.
    PumpSame as predicateTo be used with the Medela Invia® Motion™ Negative Pressure Wound Therapy System.
    How SuppliedSame as predicateSterile Single Use Only
    Tissue Contact MaterialsSimilar to predicatePolymers and Adhesive (Differences do not introduce new questions of safety)
    Additives - antimicrobial, animal originNo (Same as predicate)No
    SterilizationEto (Same as predicate)Eto

    2. Sample size used for the test set and the data provenance

    The document mentions "Bench testing" and "Biocompatibility testing" but does not provide details on the sample sizes used for these tests. The data provenance is laboratory testing, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the study involves bench testing and biocompatibility, not clinical evaluation requiring expert ground truth for patient outcomes or diagnostic accuracy.

    4. Adjudication method for the test set

    Not applicable as it's not a clinical study with an adjudication process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a wound dressing, not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a wound dressing, not an AI/ML algorithm.

    7. The type of ground truth used

    For the bench tests (vapor transmission, negative pressure maintenance, exudate removal), the "ground truth" would be the measured physical properties satisfying engineering specifications. For biocompatibility, the ground truth is adherence to ISO 10993 standards and accepted biological safety profiles for the materials. This is not "expert consensus, pathology, or outcomes data."

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.