Search Filters

Search Results

Found 86 results

510(k) Data Aggregation

    K Number
    K252740

    Validate with FDA (Live)

    Date Cleared
    2025-11-04

    (68 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voyant Open Fusion device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures where the ligation and division of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

    The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    Voyant® Open Fusion is an advanced bipolar instrument that uses radiofrequency (RF) energy, provided by the Voyant Electrosurgical Generator (Voyant ESG), to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. Once the tissue is fused, the sealed tissue may be divided with a user-actuated blade to transect the tissue at the seal site.

    The Voyant Open Fusion is comprised of three sections: jaws, shaft, and handle (including the cord and device plug). The subject device features a curved jaw with a tapered jaw design. The updated tapered jaw design does not change the device's intended function or user interface. The pistol-grip style handle features a knob, fuse button, trigger, and blade lever. The knob enables 180° rotation of Voyant Open Fusion device jaws. The trigger opens and closes the device jaws and can latch when the jaws are in the closed position. Squeezing the trigger again will release the latch and allow the jaws to open. The fuse button on the back of the handle activates RF energy. When the fuse button is pressed and held for the full seal cycle, RF output will automatically stop upon a complete seal, an audible tone will be heard and a message is displayed on the Voyant ESG screen. If the user releases the button prior to seal completion, RF output will be interrupted, an alarm tone will be heard and a message is displayed on the ESG screen. To transect the tissue, the user pulls the blade lever while the jaws are latched in the closed position.

    The device connects to the Voyant ESG via a device plug and cord. The cord and device plug are attached to the device handle. The cord facilitates the transfer of RF energy from the ESG, through the handle, along the shaft, and to the jaws. The device plug has a distinct shape which enables a proprietary connection to the Voyant ESG.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252442

    Validate with FDA (Live)

    Date Cleared
    2025-10-31

    (88 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kii Structural Balloon Access System is indicated for use in patients undergoing laparoscopic surgery requiring a path of entry and/or tissue retraction of the extraperitoneal space.

    Device Description

    Applied Medical's Kii Structural Balloon Access System provides a path of entry and/or tissue retraction for laparoscopic procedures in the extraperitoneal space. The system is provided sterile.

    The system is composed of four main components:

    • An obturator that facilitates insertion of the system through an incision.
    • A cannula with bolster and attached balloon, encased within a perforated sheath.
    • A seal which maintains insufflation.
    • An inflation bulb that is used to manually inflate and deflate the balloon.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252412

    Validate with FDA (Live)

    Date Cleared
    2025-10-31

    (91 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Advanced Bipolar: The Voyant electrosurgical generator is indicated for use with advanced bipolar Voyant devices in open and minimally invasive procedures where the ligation of vessels up to and including 7mm in diameter and tissue bundles is desired.

    NOTE: For indications specific to each Voyant device used with the Voyant generator, refer to each device's Instructions for Use (IFU).

    Monopolar/Bipolar: Intended to provide energy to standard monopolar and bipolar surgical accessories.

    Device Description

    The Applied Medical Voyant Electrosurgical Generator (EA030) is a reusable, electrically isolated, microprocessor-controlled power supply intended to deliver radiofrequency (RF) energy for electrosurgical applications. The system provides energy output to compatible monopolar, bipolar, and Voyant advanced bipolar instruments for the cutting, coagulation, and sealing of vessels and tissue bundles. The Voyant EA030 operates outside the sterile field and interfaces with active surgical instruments that operate within the sterile field. The system comprises integrated mechanical, electrical, and software components, constructed from various metals and polymers. It features a front-panel LCD touchscreen that enables users to view, navigate and adjust operating modes and system settings.

    Voyant EA030 is designed to function as part of a system that includes optional accessories such as active devices, footswitches, and patient return electrodes. These components work together to ensure safe and effective energy delivery during surgical procedures.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243152

    Validate with FDA (Live)

    Date Cleared
    2025-10-02

    (367 days)

    Product Code
    Regulation Number
    884.1640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments and maintain insufflation during laparoscopic-assisted and vaginal gynecological procedures. The device may be used for sentinel pelvic lymph node dissection occurring in the retroperitoneal space. The device is also a conduit for the extraction of specimens

    Device Description

    The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device consists of an intraperitoneal Alexis retractor and retroperitoneal Alexis retractor that are placed transvaginally to create a pathway to the peritoneum and retroperitoneum. A GelSeal cap attaches to the retractors at the opening of the vagina. The cap allows for insufflation and smoke evacuation. Sleeves inserted through the cap allow for passage of 5mm to 12mm laparoscopic instrumentation. Visualization is achieved via introduction of an endoscope through a sleeve. The device is provided sterile.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K232880

    Validate with FDA (Live)

    Date Cleared
    2023-11-01

    (44 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Applied Medical's Inzii Ripstop Redeployable Retrieval System is indicated for use as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgical procedures.

    Device Description

    The Inzii Ripstop Redeployable Retrieval System is a single-use device designed for the containment and extraction of multiple tissue specimens. The device consists of a flexible specimen bag and an introducer tube which can be redeployed for multiple uses within a single procedure. The product will be available in two models, a 10 mm model with a 265 ml specimen bag and a 15 mm model with a 1600 ml specimen bag size. The device is provided sterile and is constructed of a Ripstop nylon specimen bag, various polymers and stainless steel. The major differences between the predicate and the subject device include the subject device featuring a stronger Ripstop nylon specimen bag and the ability to be redeployed multiple times within a single surgical procedure.

    AI/ML Overview

    The Inzii Ripstop Redeployable Retrieval System is a single-use device designed for the containment and extraction of multiple tissue specimens. The study aims to demonstrate substantial equivalence to its predicate device, the Inzii Tissue Retrieval System (K060051), by performing non-clinical bench tests and simulated use tests.

    Acceptance Criteria and Reported Device Performance:

    Performance TestAcceptance CriteriaReported Device Performance
    Specimen Retrieval System Functionality testingThe device must demonstrate proper deployment, retraction (for redeployment), and cinching mechanisms. It must effectively contain and allow for the extraction of specimens.The subject device demonstrated proper deployment, the ability to be retracted for redeployment multiple times within a single surgical procedure, and effective cinching for final retrieval. It successfully contained and allowed for extraction of specimens, meeting the functional requirements.
    Specimen Retrieval System Leak testingThe specimen bag must be liquid-tight and prevent leakage of contents during simulated use and retrieval, ensuring no spillage of bodily fluids or extracted materials.The specimen bags of the Inzii Ripstop Redeployable Retrieval System demonstrated liquid-tight integrity. No leakage was observed during simulated use or retrieval, confirming its ability to prevent spillage as required.
    Specimen Bag Closure testingThe bag closure mechanism (cord loop) must reliably and securely close the bag, preventing accidental opening or spillage of contents during retrieval.The cord loop closure mechanism was found to be reliable and secure, effectively closing the bag and preventing accidental opening or spillage of contents during simulated retrieval procedures.
    Specimen Bag Tear Resistance TestThe Ripstop nylon specimen bag must exhibit superior tear resistance compared to the predicate device's polyurethane bag, ensuring the integrity of the bag during specimen collection and retrieval.The Ripstop nylon specimen bag of the subject device demonstrated significantly higher tear resistance compared to the predicate device, as confirmed by additional testing referenced from K100959. This ensures enhanced integrity and durability during use.
    Specimen Bag Strength testingThe bag must withstand forces encountered during specimen retrieval, ensuring it does not rupture under typical surgical loads.The specimen bag demonstrated adequate strength, withstanding the forces typically encountered during specimen retrieval without any rupture or compromise to its integrity.
    Specimen Bag Pressure testingThe specimen bag must endure internal pressure from contained specimens without bursting or leaking.The specimen bag successfully endured internal pressure from contained specimens without bursting, leaking, or showing signs of structural failure.
    Specimen Bag Puncture testingThe specimen bag must resist accidental punctures from instruments or sharp edges of extracted specimens, maintaining its barrier function.The specimen bag exhibited resistance to accidental punctures from surgical instruments and sharp edges of extracted specimens, effectively maintaining its barrier function and preventing compromise of the contained material.
    Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity)The device must pass all biological endpoints as per EN ISO 10993-1; 2020 for an externally communicating device with limited contact duration (up to 24 hours), contacting tissue/bone/dentin.The device successfully passed all required biological endpoints: Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity (AST), and Material Mediated Pyrogenicity (MMP), confirming its biocompatibility for the intended use and contact duration.

    Study Information:

    1. Sample sizes used for the test set and data provenance:

      • The document does not explicitly state the exact sample sizes for each specific performance test (e.g., number of devices tested for leak, strength, etc.). It refers to "sample size" generally in the context of the performance data summary.
      • The data provenance is from non-clinical bench tests and simulated use tests conducted by Applied Medical Resources Corporation. There is no indication of country of origin for data or whether it's retrospective or prospective, as these are bench tests.
    2. Number of experts used to establish the ground truth for the test set and qualifications:

      • This information is not applicable as the document describes performance testing of a physical medical device (specimen retrieval system), not an AI/software device that requires expert-established ground truth. The 'ground truth' here is the objective outcome of the physical/mechanical and biological tests.
    3. Adjudication method for the test set:

      • This is not applicable for the type of device and testing described. Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving reader interpretations, not for bench testing of physical device performance.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study:

      • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic/interpretive devices where human readers' performance is augmented by AI. The Inzii Ripstop Redeployable Retrieval System is a physical surgical tool.
    5. Standalone (algorithm only without human-in-the-loop performance) study:

      • No, a standalone study was not done. This concept is specific to AI algorithms. The device described functions with human interaction during surgery.
    6. Type of ground truth used:

      • For the performance testing, the "ground truth" used was based on objective measurements and observations from the non-clinical bench tests and simulated use tests. This includes parameters like liquid tightness, resistance to tearing/puncture, and mechanical functionality, as well as established ISO standards for biocompatibility.
    7. Sample size for the training set:

      • This information is not applicable as the document describes performance testing of a physical medical device, not an AI/machine learning model that requires a training set.
    8. How the ground truth for the training set was established:

      • This information is not applicable as there is no training set for a physical medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222284

    Validate with FDA (Live)

    Date Cleared
    2022-10-14

    (74 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Voyant Maryland Fusion Device with Single-Step Activation is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open and laparoscopic procedures where the ligation of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.

    The device has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures.

    Device Description

    The Voyant Maryland Fusion device with Single Step is an advanced bipolar instrument that uses RF energy, provided by the Voyant Electrosurgical Generator (K182244), to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The device features a mechanical, user-actuated blade for the division of sealed tissue.

    AI/ML Overview

    The medical device in question is the Voyant® Maryland Fusion Device with Single-Step Activation. It is an electrosurgical device intended for use with the Voyant Electrosurgical Generator to seal and divide vessels and tissue bundles up to 7mm in diameter.

    Based on the provided text, here's an analysis of the acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a direct table with specific numerical acceptance criteria and corresponding reported device performance values. Instead, it states that the devices "met all acceptance criteria" or that results were "sufficient to represent the subject device is safe and effective."

    Acceptance Criteria CategoryReported Device Performance
    EMC, Electrical Safety, and Mechanical TestingThe predicate device met all acceptance criteria in accordance with relevant standards of the IEC 60601 series. Additional electrical safety testing on the subject device also met all acceptance criteria.
    Simulated Repeated-Use TestingThe subject device met the predetermined acceptance criteria for continued performance over multiple device activations.
    System Testing (Burst Pressure)(Performed on predicate device) Representative vessels were sealed, and burst pressure was recorded. The results are considered representative of system performance for the subject device. (No specific threshold or value is provided, but implies satisfactory burst pressure).
    System Testing (Thermal Spread)(Performed on predicate device) Thermal spread damage was evaluated. The results are considered representative of system performance for the subject device. (No specific thermal spread limit is provided, but implies acceptable thermal effects).
    Animal Testing (Long-term Seal Quality, Adverse Effects)(Performed on predicate device) Evaluated long-term seal quality, device performance, and potential for adverse effects on adjacent structures in large porcine animal models. Vessels were sealed and evaluated for hemostasis and signs of hematoma. The results are sufficient to represent the subject device is safe and effective.
    Software VerificationUnit, integration, and system level software testing were conducted to evaluate design, implementation, and performance. (Implies all tests passed the defined acceptance criteria for software).

    Note: The document generally indicates that the device "met all acceptance criteria" without specifying the exact numerical thresholds for many of these criteria. This type of summary is common in 510(k) submissions where the focus is on demonstrating substantial equivalence rather than presenting a detailed performance report.


    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes for all test sets. However, it provides some information:

    • Burst pressure testing: "vessels representative of the devices' indications" were used. (No specific number mentioned).
    • Thermal spread testing: (No specific number mentioned).
    • Simulated repeated-use testing: (No specific number mentioned for activations or devices).
    • Animal testing: "large porcine animal models" were used. The number of animals or vessels tested is not specified.

    Data Provenance:

    • Country of Origin: Not explicitly stated, but typically these studies are conducted by the manufacturer or contract research organizations, often in the country where the manufacturer is based (USA, in this case).
    • Retrospective or Prospective: The testing described (simulated repeated-use, system testing, animal testing, software verification) are all prospective studies, designed and executed to evaluate the new or modified device. The reference to the predicate device's data suggests that previously collected prospective data from the predicate was leveraged.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide any information about the number or qualifications of experts used to establish ground truth for any of the described performance tests. For electrosurgical devices, "ground truth" often refers to objective measurements (e.g., burst pressure, thermal spread measurements, histopathological analysis in animal studies) rather than subjective expert interpretations like in imaging studies.


    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method. Given that most of the described tests are objective measurements (e.g., electrical parameters, burst pressure, thermal spread, software testing), an adjudication method as typically understood in studies involving human interpretation (like imaging reads) would not be applicable. For animal studies, evaluation of hemostasis and hematoma would typically follow pre-defined histological or gross pathology criteria.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The device is a direct-use surgical instrument, not an AI-based diagnostic tool requiring human interpretation/reading. Therefore, the concept of human readers improving with AI assistance is not applicable here.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    While the term "standalone" typically refers to AI algorithms, in the context of this electrosurgical device, the device itself is a "standalone" instrument. The performance tests (electrical, mechanical, system, animal) described are evaluating the performance of the device only (or the device in combination with its specified generator), without a human-in-the-loop actively assisting in output. The software verification tests are also standalone evaluations of the algorithm's performance within the device's control systems.


    7. The Type of Ground Truth Used

    The ground truth used for various tests includes:

    • Objective Measurements: For burst pressure (quantifiable pressure values), thermal spread (measurable tissue damage), and electrical safety/EMC (compliance with quantifiable standards).
    • Biological/Pathological Outcomes: For animal studies, this would involve macroscopic and potentially microscopic examination of sealed vessels and surrounding tissues to assess hemostasis, presence of hematoma, and tissue damage, serving as the "ground truth" for seal quality and safety.
    • Software Design Specifications: For software verification, the "ground truth" is adherence to predefined design requirements and functional specifications.

    8. The Sample Size for the Training Set

    The document describes performance testing for an electrosurgical device, not an AI/ML product. Therefore, the concept of a "training set" (used for training machine learning models) is not applicable to this submission.


    9. How the Ground Truth for the Training Set Was Established

    As the concept of a "training set" is not applicable, the question of how its ground truth was established is also not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220969

    Validate with FDA (Live)

    Date Cleared
    2022-09-16

    (165 days)

    Product Code
    Regulation Number
    884.1630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GelPOINT V-Path Vaginal Access System (VAS) is intended to be inserted in the vagina to allow for entry of minimally invasive instruments while maintaining insufflation for vaginal gynecological procedures.

    Device Description

    The GelPOINT V-Path Vaginal Access System consists of an Intravaginal Alexis retractor and two access channels that are placed vaginally to create a pathway to gynecological organs. A GelSeal cap attaches to the retractor at the opening of the vagina. The cap allows for insufflation and smoke evacuation. Sleeves inserted through the cap allow for passage of 5mm to 12mm instrumentation. Visualization is achieved via introduction of an endoscope through a sleeve. The device is provided sterile.

    AI/ML Overview

    This document is a 510(k) summary for the GelPOINT® V-Path Vaginal Access System. It describes the device, its intended use, and the studies conducted to demonstrate its substantial equivalence to a predicate device.

    Here's an analysis of the provided text in relation to your request:

    1. Table of acceptance criteria and the reported device performance:

    The document describes functional testing but doesn't explicitly state quantitative acceptance criteria or provide specific numerical device performance metrics in a table format. Instead, it describes general successful outcomes.

    Acceptance Criteria (Inferred from study goals)Reported Device Performance (Summary of results)
    Ability to maintain transvaginal access.The studies confirm that the GelPOINT V-Path Vaginal Access System has the ability to establish and maintain a path of entry for minimally invasive instruments.
    Ability to facilitate instrument access to the surgical site.The studies confirm that the GelPOINT V-Path Vaginal Access System has the ability to establish and maintain a path of entry for minimally invasive instruments.
    Ability to allow for device removal.The studies confirm that the GelPOINT V-Path Vaginal Access System has the ability to allow for device removal. (Inferred from "Device removal" being an assessed function).
    Ability to maintain device retention/anchoring.The studies confirm that the GelPOINT V-Path Vaginal Access System can remain anchored in the patient while withstanding conditions under normal clinical use.
    Ability to maintain insufflation.The studies confirm that the integrity and sealing capability of the device system are adequate to maintain insufflation as a means to provide visualization of the surgical space.
    Biocompatibility (non-cytotoxic, non-sensitizer, non-irritant, non-systemically toxic, non-pyrogenic).The biocompatibility test results demonstrate the GelPOINT V-Path Vaginal Access System is biocompatible and has met the criteria to be classified as non-cytotoxic, non-sensitizer, non-irritant, non-systemically toxic, and non-pyrogenic.
    Shelf-life of 3 years.Results from testing demonstrated that the devices could maintain their specifications over the stated shelf-life duration of 3 years.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document mentions "bench testing" and "Functional bench testing" but does not provide specific sample sizes (e.g., number of devices tested, number of trials).
    • Data Provenance: The studies were internal company studies ("A protocol was developed by Applied Medical"). There is no mention of country of origin of data, and these are laboratory bench tests, not patient data, so the retrospective/prospective distinction is not applicable in the usual sense for clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The studies are described as functional bench tests, suggesting engineering or quality assurance personnel would evaluate the performance rather than clinical experts establishing a ground truth in the context of diagnostic accuracy.

    4. Adjudication method for the test set:

    This information is not provided. Given the nature of bench testing, it's unlikely an adjudication method like 2+1 or 3+1 (typically used for clinical imaging assessments) would be applied.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No.
    • AI Assistance: The document describes a medical device for surgical access, not an AI-powered diagnostic or assistive tool. Therefore, a study on human readers improving with AI assistance is not applicable and was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No. This is a physical medical device, not an algorithm, so "standalone algorithm performance" is not applicable.

    7. The type of ground truth used:

    For the functional tests, the "ground truth" would be the successful demonstration of the device's intended functions (e.g., maintaining an open pathway, retaining its position, maintaining insufflation). For biocompatibility, the ground truth is established by meeting the pass/fail criteria outlined in the referenced ISO standards and USP <151>. For shelf-life, the ground truth is the device maintaining specifications after accelerated aging.

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. As above, this is not an AI/machine learning device.

    In summary:

    This 510(k) summary focuses on demonstrating the substantial equivalence of a physical surgical access device through functional bench testing, biocompatibility testing, and shelf-life studies. It does not involve clinical studies with patient data, expert review of images, or artificial intelligence algorithms. Therefore, many of your requested points related to AI performance, reader studies, and large-scale data sets are not relevant to this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211043

    Validate with FDA (Live)

    Date Cleared
    2022-03-31

    (357 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Alexis® Contained Extraction System is indicated to contain and isolate tissue during, or prior to, surgical removal and/or extracorporeal manual morcellation via an abdominal or vaginal approach.

    Device Description

    The Alexis® Contained Extraction System is a sterile, single use tissue bag with a guard system component. The polyurethane tissue bag consists of a flexible ring, tabs, strap with a snap fastener, and an attached tether It is available in two models GTB14 and GTB17 with the following dimensions and volumes:

    ModelRing DiameterVolume
    GTB1414 cm3,400 mL
    GTB1717 cm6,500 mL

    The tissue bag is used to contain and isolate specimens for surgical removal and/or manual morcellation. The strap with snap fastener maintains the folded bag and flexible ring in a collapsed state during insertion. After the device is fully inserted and deployed into the abdominal or pelvic cavity, the opening of the bag returns to its original, circular shape, facilitating placement of the specimen in the bag. When the specimen is ready for removal and/or manual morcellation, the tether, tabs, and strap are used to maneuver the ring to the surface of the extraction site.

    If the specimen requires manual morcellation. the ring may be repeatedly flipped to shorten the bag and consequently bring the specimen closer to the extraction site.

    The guard system consists of a guard and a self-retaining retractor. The guard is composed of a coiled polyethylene material that conforms to the extraction site. The self-retaining retractor consists of two polyurethane rings connected by a polyurethane sheath and a tether attached to the inner ring. It helps anchor the guard at the extraction site. The guard system provides a robust barrier between the bag and sharp instruments.

    The Alexis® Contained Extraction System can be inserted and retrieved either through an abdominal incision or through the vaginal canal and colpotomy. In both cases, manual morcellation is performed extracorporeally at the extraction site.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Alexis® Contained Extraction System. This notification focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report with specific performance metrics and acceptance criteria for a new AI/software device.

    Therefore, the requested information regarding acceptance criteria, reported device performance, sample size for test sets (including provenance, expert qualifications, and adjudication methods), MRMC studies, standalone performance, ground truth establishment, and training set details cannot be extracted from the provided text.

    The document details engineering tests related to the physical device's characteristics (biocompatibility, sterilization, shelf life, and functional performance) to support its substantial equivalence to an existing device, as opposed to clinical performance data against defined metrics for a diagnostic or AI-driven system.

    The "Discussion of Performance Data" section refers to:

    • Biocompatibility: Evaluation in accordance with ISO 10993-1 and FDA guidance.
    • Sterilization/Shelf Life: Validation performed in accordance with ISO 11135-2.
    • Functional Performance: Design Verification testing for various physical attributes (e.g., integrity of tissue bag, tether, guard system, resistance to penetration, ability to maintain a closed system, specimen capture). This also included "simulated use testing with the subject device in both an abdominal bench model and a vaginal bench model."

    However, these are engineering and safety tests typical for a physical medical device, not performance studies as would be conducted for a diagnostic algorithm or AI system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202049

    Validate with FDA (Live)

    Date Cleared
    2021-09-07

    (410 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Python catheters are indicated for removal of thromboemboli from the peripheral arterial system, and for occlusion and infusion of fluids into a vessel.
    The Over-the-Wire Latis Graft Cleaning Catheter is indicated for the removal of thromboemboli from vascular grafts, and for occlusion and infusion of fluids into a graft.

    Device Description

    The Python Embolectomy Catheter & Over-the-Wire Latis Graft Cleaning Catheters are sterile, single use vascular balloon catheters intended for use in the removal of thromboemboli, occlusion of vessels, and the infusion of fluids. The devices are intended for use in arterial vessels and artificial vascular grafts respectively. The body of the catheters utilize a dual lumen design with one inflation lumen used to inflate the balloon, and one through lumen allowing for insertion over a guide wire or infusion of fluids. The outer diameter of the body is designed to accommodate use with a standard percutaneous access sheath. The Over-the-wire Latis Graft Cleaning Catheter includes a braided mesh which surrounds the balloon and is designed to increase engagement with wall of artificial vascular graft.

    AI/ML Overview

    The provided text describes a medical device's 510(k) submission to the FDA, demonstrating substantial equivalence to predicate devices. It focuses on the physical and material characteristics of the catheters and the bench testing performed. However, it does not contain information pertinent to the typical acceptance criteria or study design for AI/ML-based medical devices, such as a test set size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details.

    Therefore, I am unable to provide the requested information based on the input text.

    The document discusses the following:

    • Device Name: Python Catheter and Over-the-Wire Latis Graft Cleaning Catheter
    • Intended Use: Removal of thromboemboli from the peripheral arterial system/vascular grafts, and for occlusion and infusion of fluids into a vessel/graft.
    • Testing Performed:
      • Engineering/Bench Testing: Balloon Pull Force, Inflation & Guidewire Hub Pull Force, Catheter Body Tensile Strength, Catheter Balloon Cycling Test, Balloon Burst Pressure Test, Inflated Balloon Diameter Inspection, Guidewire & Inflation Hub Leak Test, Simulated Use Bend Test, Torsion Test, Kink Test.
      • Biocompatibility Testing: Cytotoxicity, Intracutaneous Irritation, Sensitization, Acute Systemic Toxicity, Hemolysis, Materials Mediated Pyrogenicity, Platelet and Leukocyte Counts, Partial Thromboplastin Time, Complement Activation, Thrombogenicity.
    • Standards Referenced: ISO 10555-1:ed2:2017 and EN 13868.

    This information relates to the physical and biological safety and performance of the catheter itself, not to the performance of an AI/ML component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200021

    Validate with FDA (Live)

    Date Cleared
    2021-01-06

    (366 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Medical Anoscope is intended for transanal use to provide physicians access to the anal sphincter, anus, and rectum and to perform various diagnostic and therapeutic procedures using additional accessories.

    Device Description

    The Applied Medical Anoscope consists of a one-piece polycarbonate half round channel with a tapered closed tip. The device retracts the anal sphincter and provides access to the anorectal anatomy during transanal procedures.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the Applied Medical Anoscope. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, the document does not contain information about the acceptance criteria or a study that proves the device meets specific acceptance criteria related to its performance beyond general biocompatibility and functional performance tests.

    The 510(k) summary (page 4) states: "The anoscope is a simple one-piece device that provides a passage into the anal canal. There are no recognized performance standards for access devices of this kind. Therefore, Applied Medical devised criteria by which to assess safety and efficacy, including a computer-aided simulation and a compression test."

    This indicates that while some criteria and tests were performed, their details (specific quantitative acceptance criteria, the results of these tests, sample sizes, ground truth establishment, expert involvement, etc.) are not included in this publicly available 510(k) clearance document.

    Therefore, I cannot fulfill your request for the following information based solely on the provided text:

    1. A table of acceptance criteria and the reported device performance: Not detailed in the document.
    2. Sample sizes used for the test set and the data provenance: Not detailed for the described "computer-aided simulation and a compression test."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as the testing described is physical, not interpretive, and not detailed in the document.
    4. Adjudication method for the test set: Not applicable based on the described tests.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a physical anoscope, not an AI or imaging device with human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a physical anoscope.
    7. The type of ground truth used: Not applicable for the described physical tests.
    8. The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
    9. How the ground truth for the training set was established: Not applicable.

    The document focuses on demonstrating substantial equivalence to a predicate device through:

    • Comparison of technological characteristics (page 4), noting minor differences like the pre-existence of a dedicated light source channel in the predicate vs. the subject's larger lumen and use of external light sources.
    • Biocompatibility testing (page 4) against ISO 10993-1 endpoints (cytotoxicity, intracutaneous irritation, sensitization), which the device passed.
    • Functional performance testing (page 4), which involved a "computer-aided simulation and a compression test." The details of these tests and their results are not provided.

    In summary, this document is a regulatory clearance letter, not a detailed scientific study report. It states that the manufacturer "devised criteria" and conducted "computer-aided simulation and a compression test" to assess safety and efficacy, but it does not provide the specific quantitative acceptance criteria or the results of those tests.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 9