Search Results
Found 125 results
510(k) Data Aggregation
(102 days)
The Alexis Lighted Wound Protector-Retractor is a device that allows the surgeon to access the abdominal cavity through an atraumatically retracted wound, providing maximum exposure with minimum incision size. In addition to incision retraction, it is intended to provide supplemental illumination and protect against wound contamination during both laparoscopic and open surgery. The Alexis Lighted Wound Protector-Retractor is capable of temporarily closing an incision to maintain pneumoperitoneum during laparoscopic surgery, or to serve as an additional trocar port site. In addition, the device may be used to access the thoracic cavity or other soft tissue retraction during cardiac and general surgical procedures.
Applied Medical's Alexis Lighted Wound Protector-Retractor provides wound protection and retraction during laparoscopic and open surgery while providing supplemental illumination. The device is provided sterile, and the reusable power supply is provided non-sterile. The device and power supply are used in the clinical setting.
The system includes five main components:
- A sheath that protects the wound and provides retraction with tension applied.
- An outer ring that is rolled to create tension on the sheath and anchor the device.
- An inner ring with embedded LEDs that anchors the device and provides illumination.
- A power cord that connects to the reusable power supply.
- A reusable power supply that connects to a grounded receptable to provide power to the LEDs.
N/A
Ask a specific question about this device
(186 days)
The Explant Express™ is a single patient, single use suction device used to assist in the removal of one intracapsular ruptured silicone breast implant. Not intended for en bloc removal. Not intended to remove residual silicone or be applied directly to tissue.
The Explant Express™ consists of a container with a nozzle and a port for applied suction. The scientific concept that forms the basis for the device is suction. The components of this device consist of two separable container halves and an O-ring.
N/A
Ask a specific question about this device
(68 days)
The Voyant Open Fusion device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures where the ligation and division of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device.
The device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.
Voyant® Open Fusion is an advanced bipolar instrument that uses radiofrequency (RF) energy, provided by the Voyant Electrosurgical Generator (Voyant ESG), to seal vessels up to and including 7mm in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. Once the tissue is fused, the sealed tissue may be divided with a user-actuated blade to transect the tissue at the seal site.
The Voyant Open Fusion is comprised of three sections: jaws, shaft, and handle (including the cord and device plug). The subject device features a curved jaw with a tapered jaw design. The updated tapered jaw design does not change the device's intended function or user interface. The pistol-grip style handle features a knob, fuse button, trigger, and blade lever. The knob enables 180° rotation of Voyant Open Fusion device jaws. The trigger opens and closes the device jaws and can latch when the jaws are in the closed position. Squeezing the trigger again will release the latch and allow the jaws to open. The fuse button on the back of the handle activates RF energy. When the fuse button is pressed and held for the full seal cycle, RF output will automatically stop upon a complete seal, an audible tone will be heard and a message is displayed on the Voyant ESG screen. If the user releases the button prior to seal completion, RF output will be interrupted, an alarm tone will be heard and a message is displayed on the ESG screen. To transect the tissue, the user pulls the blade lever while the jaws are latched in the closed position.
The device connects to the Voyant ESG via a device plug and cord. The cord and device plug are attached to the device handle. The cord facilitates the transfer of RF energy from the ESG, through the handle, along the shaft, and to the jaws. The device plug has a distinct shape which enables a proprietary connection to the Voyant ESG.
N/A
Ask a specific question about this device
(91 days)
Advanced Bipolar: The Voyant electrosurgical generator is indicated for use with advanced bipolar Voyant devices in open and minimally invasive procedures where the ligation of vessels up to and including 7mm in diameter and tissue bundles is desired.
NOTE: For indications specific to each Voyant device used with the Voyant generator, refer to each device's Instructions for Use (IFU).
Monopolar/Bipolar: Intended to provide energy to standard monopolar and bipolar surgical accessories.
The Applied Medical Voyant Electrosurgical Generator (EA030) is a reusable, electrically isolated, microprocessor-controlled power supply intended to deliver radiofrequency (RF) energy for electrosurgical applications. The system provides energy output to compatible monopolar, bipolar, and Voyant advanced bipolar instruments for the cutting, coagulation, and sealing of vessels and tissue bundles. The Voyant EA030 operates outside the sterile field and interfaces with active surgical instruments that operate within the sterile field. The system comprises integrated mechanical, electrical, and software components, constructed from various metals and polymers. It features a front-panel LCD touchscreen that enables users to view, navigate and adjust operating modes and system settings.
Voyant EA030 is designed to function as part of a system that includes optional accessories such as active devices, footswitches, and patient return electrodes. These components work together to ensure safe and effective energy delivery during surgical procedures.
N/A
Ask a specific question about this device
(88 days)
Kii Structural Balloon Access System is indicated for use in patients undergoing laparoscopic surgery requiring a path of entry and/or tissue retraction of the extraperitoneal space.
Applied Medical's Kii Structural Balloon Access System provides a path of entry and/or tissue retraction for laparoscopic procedures in the extraperitoneal space. The system is provided sterile.
The system is composed of four main components:
- An obturator that facilitates insertion of the system through an incision.
- A cannula with bolster and attached balloon, encased within a perforated sheath.
- A seal which maintains insufflation.
- An inflation bulb that is used to manually inflate and deflate the balloon.
N/A
Ask a specific question about this device
(367 days)
The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device is intended to be inserted transvaginally to establish a path of entry for minimally invasive instruments and maintain insufflation during laparoscopic-assisted and vaginal gynecological procedures. The device may be used for sentinel pelvic lymph node dissection occurring in the retroperitoneal space. The device is also a conduit for the extraction of specimens
The GelPOINT V-Path Transvaginal Access Platform with Retroperitoneal Access Device consists of an intraperitoneal Alexis retractor and retroperitoneal Alexis retractor that are placed transvaginally to create a pathway to the peritoneum and retroperitoneum. A GelSeal cap attaches to the retractors at the opening of the vagina. The cap allows for insufflation and smoke evacuation. Sleeves inserted through the cap allow for passage of 5mm to 12mm laparoscopic instrumentation. Visualization is achieved via introduction of an endoscope through a sleeve. The device is provided sterile.
N/A
Ask a specific question about this device
(253 days)
The AMT Low-Profile Suprapubic Catheter & Drainage Set is indicated for patients 7 years of age and older with lower urinary tract dysfunctions such as neurogenic bladder dysfunction or urethral obstruction or who are severely disabled or requiring constant care. The AMT Low-Profile Suprapubic Catheter & Drainage Set is indicated to be used for temporary suprapubic urinary diversion and drainage for less than or equal to four weeks. The AMT Low- Profile Suprapubic Catheter is intended to be placed directly into the bladder through a secured (initial placement) or formed (replacement) stoma.
The Applied Medical Technology, Inc. (AMT) Low-Profile Suprapubic Catheter & Drainage Set (LPSPC) consists of two primary assemblies. The LPSPC is identical in materials and manufacturing processes to the Low-Profile Balloon Feeding Device cleared in K161413 for a different indication.
The Suprapubic Catheter is used to provide urinary diversion and drainage through the lower abdomen, placed and maintained in a percutaneously prepared cystostomy opening. The Drainage Set is intended to allow a connection between the suprapubic catheter device and a urinary collection bag system. The device is designed for bladder drainage by way of a secured (initial placement) or formed (replacement) cystostomy tract and may be used for irrigation. Patients may include: any user with urethral trauma, those who require long-term catheterization and urethral catheterization is intolerable, or those with a spinal cord injury and an inability to self-catheterize who may find the suprapubic method simpler to manage.
The LPSPC is a urinary drainage device made principally of silicone elastomer. The device consists of an internal retention balloon and a flexible external bolster with a balloon fill-valve, interlocking irrigation port, and safety plug. The catheter is inserted through the stoma and into the urinary bladder and is secured by the internal retention balloon. The device is offered in various sizes to accommodate different stoma diameters and lengths. On the shaft near the end is located a balloon which, when properly inflated, acts as an internal stabilizer preventing outward displacement through the stoma in the abdominal wall. The balloon is inflated through a separate inflation valve located in the external portion of the device.
The external profile of the device acts as a stabilizer to prevent inward migration and affords the user the ability to connect a provided Drainage Set.
The Drainage Set is identical in material and manufacturing process to certain configurations of the Irrigation Set included with the AMT Bowel Management Device previously cleared in K180026. The right-angle connector of the Drainage Set is compatible with the interlock on the LPSPC and is used to direct fluids drained from the bladder, through the catheter, and into a separate urinary collection bag.
The AMT Low-Profile Suprapubic Catheter & Drainage Set may optionally be marketed with a Foley Insertion Tray accessory. If included, the accessory would be a legally marketed Foley Insertion Tray, such as:
- AMSure Foley Insertion Tray (previously cleared Class II, Product Code FCM, K030664)
- Bardia Foley Insertion Tray (registered Class II, Product Code OHR, 510(k) exempt)
- Dynarex Foley insertion Tray without Catheter (registered Class II, Product Code OHR, 510(k) exempt)
The device is sold as sterile and can be initially placed in an outpatient-setting by a health care professional. The device may be replaced in an outpatient-setting or home setting by a health care professional or care giver.
The provided text describes a 510(k) premarket notification for the "AMT Low-Profile Suprapubic Catheter & Drainage Set." This document focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a standalone study with specific acceptance criteria and performance metrics in the format requested.
Therefore, much of the requested information cannot be extracted directly from this document. The document primarily describes the types of testing performed and concludes that the device met acceptance criteria, but generally does not specify these criteria or report quantitative performance results against them.
Here's an attempt to answer the questions based on the available information:
1. A table of acceptance criteria and the reported device performance
The document states that the device "met or exceeded all the acceptance criteria and does not raise new questions of safety or effectiveness when compared to the predicate/reference devices." However, specific numerical acceptance criteria and precise performance data are not detailed in this summary. The performance is generally stated as meeting standards.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Biocompatibility | Met acceptance criteria for permanent contact (>30 days) with mucosal membrane and breached/compromised surfaces, based on ISO 10993 standards. |
| Sterilization | Achieved a Sterility Assurance Level (SAL) of 10-6. Complies with ISO 11135 and ISO 10993-7. |
| Shelf Life | Packaging complies with ASTM F1980-21, ISO 11607-1 & 2, and ISTA 3A 2018. |
| Bench Testing (General) | Met or exceeded all predetermined specifications for performance principles. |
| Bench Testing (Specific Examples - criteria not detailed, assumed to be 'met') | Balloon Assembly Bond Peel/Tear Strength, Balloon Burst, Fill Valve Blow Out, Fill Valve Pullout, Flow Rate, Leak Test, Tubing Tensile Strength, Minimum Overmold Bond Strength, Stoma Pullout, Tubing Cyclic and Tensile Test, Main Strap Tensile Strength, Side Strap Tensile Strength. |
| MR Compatibility | Complies with ASTM F2052-21 (magnetically induced displacement force), ASTM F2213-17 (magnetically induced torque), ASTM F2119-07 (Reapproved 2013) (MR image artifacts), ASTM F2182-19e2 (radio frequency induced heating). |
| Enteral Feeding Systems Design & Testing | Complies with ISO 20695 First edition 2020-03. |
| Enteral Feeding Devices with Retention Balloon | Complies with ASTM F2528-06 (Reapproved 2023). |
| Small-bore Connectors | Complies with ISO 80369-1 Second edition 2018-11. |
| Clinical Performance (based on literature review) | Determined to function clinically as intended and benefits outweigh risks in the target user population. Risks (hematuria, wound infection, UTI, leakage, poor fit, discomfort, bladder spasms) are similar to those seen with this device type. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size:
- Bench Testing: Not specified. The document lists the types of tests but not numbers of samples.
- Clinical Performance (Literature Review): Clinical experience was reported on 34 adults and 188 pediatric patients across 22 published articles.
- Data Provenance: The clinical data is from a systematic literature review (retrospective). The countries of origin are not specified, only that the literature from "independent organizations" was reviewed.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This was a systematic literature review, not a study where experts established ground truth for newly collected patient data. The "ground truth" for the clinical performance was derived from the findings reported in the published literature.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The clinical evidence was gathered through a systematic literature review. The "reporting of this systematic review was guided by the standards of the Preferred Reporting Items for Systematic Review and Meta-Analysis (PRISMA) Statement, as well as the Cochrane Handbook for Systematic Reviews of Interventions. Assessment of the included literature was completed using the critical appraisal tools published by the Joanna Brigges Institute." This describes the methodology for reviewing existing literature, not an adjudication process for a new test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a suprapubic catheter, not an AI-assisted diagnostic or therapeutic device. The submission does not mention any AI components or MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the clinical performance aspects, the "ground truth" was derived from "clinical experience... reported on 34 adults and 188 pediatric patients" across published literature, guidelines from organizations like the Pediatric Urologists of Canada and the Canadian Urological Association, operative textbooks, and information from the International Continence Society (ICS). This can be broadly categorized as published clinical outcomes data and expert-derived guidelines/clinical experience.
8. The sample size for the training set
Not applicable. This is a physical medical device submission, not a machine learning model. There is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(74 days)
The AMT Suture Passer is indicated to pass suture through soft tissues of the body during endoscopic/laparoscopic surgery and interventional radiology procedures in child, adolescent and adult populations.
The AMT Suture Passer is a sterile, single-case use, hand-held suture grasping device. The device features a hypodermic needle through which a suture can be passed and retrieved through another. The retrieval component may be offered in two configurations: a non-magnetic version, and a magnetically assisted version, using a grasper with a magnetized arm. Using either configuration, the AMT Suture Passer provides more flexibility in stitch geometry and approach than traditional U-stitches. The preferred method to introduce the suture is through a separate introducer needle; the AMT Suture Passer Kit with Magnet Assist instrument set bundles the magnetically assisted configuration of the AMT Suture Passer with an introducer needle and two magnetic sutures to create a non-procedure-specific kit. For the magnetically assisted configuration, the magnetized arm of the grasper works in conjunction with a magnetic suture to reduce the technical difficulty of intracorporeal suture retrieval under laparoscopic, endoscopic, radiologic, or ultrasound guidance. This magnet-assist technology is derived from the ATLAS Suture Delivery System (K193612), also manufactured by Applied Medical Technology, which performs the same clinical function and is used in similar clinical applications. Both configurations of the AMT Suture Passer feature identical components, except for the end-effectors, and make use of Luer-lock compatible hubs that allow a syringe to be attached for the purpose of administering contrast through the device. The contrast enters through the Luer hub and exits from the distal end of the needle, which allows the needle position to be verified prior to passing the suture during interventional radiology procedures.
This document is an FDA 510(k) clearance letter for a medical device called the "AMT Suture Passer." It outlines the regulatory review process and the basis for the FDA's determination of substantial equivalence to previously cleared devices.
The request asks for details on acceptance criteria and a study proving the device meets these criteria. However, the provided document does not describe a study involving an AI/Machine Learning algorithm, human readers, or a test set with ground truth established by experts.
Instead, the document details the performance data for a physical medical device (AMT Suture Passer) through bench testing, biocompatibility testing, sterilization validation, and shelf-life testing. It explicitly states:
- "B. Software: There are no software components related in any way to this device. Therefore, Software Validation is not applicable to this device."
- "E. Animal Study: Animal testing was NOT performed."
- "F. Clinical Study: Clinical testing was NOT performed."
Therefore, I cannot provide the requested information regarding an AI study, test sets, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or training set details from this document. The information requested aligns with studies typically performed for AI/ML-driven medical devices, which the AMT Suture Passer is not.
However, I can extract the acceptance criteria and performance data for the physical device as presented in the document:
Acceptance Criteria and Device Performance (Based on the Provided Document for a Physical Device)
The "AMT Suture Passer" is a physical medical device, not an AI/ML diagnostic tool. Therefore, the acceptance criteria and performance data relate to its physical properties, sterility, biocompatibility, and functionality as a surgical instrument. There is no mention of AI performance metrics like sensitivity, specificity, AUC, or reader improvement.
Here's a summary of the performance data and acceptance criteria as described:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Test/Standard Applied | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| Biocompatibility | ISO 10993-1, -5, -7, -10, -11, -23; USP-NF <85>, <161> | Met acceptance criteria for limited contact (≤24 hours) with tissue/bone/dentin. Suture component: Met acceptance criteria for prolonged contact (>24 hours to 30 days) with tissue/bone/dentin. | "It was determined that the AMT Suture Passer met the acceptance criteria for limited contact (24 hours or less) with tissue/bone/dentin. The suture component of the subject device met the acceptance criteria for prolonged contact (greater than 24 hours to 30 days) with tissue/bone/dentin." |
| Software | N/A (Device has no software) | N/A | "There are no software components related in any way to this device. Therefore, Software Validation is not applicable to this device." |
| Electromagnetic Comp./Electrical Safety | N/A (Device has no electronic components) | N/A | "There are no electronic components related in any way to this device." |
| Sterilization | ISO 10993-7 | Sterility Assurance Level (SAL) of 10-6; compliance with standards. | "The subject device is ethylene oxide sterilized, and has been validated to confirm a Sterility Assurance Level (SAL) of 10-6. The sterilization processing complies with the standards." |
| Shelf Life | ISO 11607-1 & 2; ISTA 3A 2018 | Validated shelf life demonstrated. | "Testing indicates that the subject device has a validated shelf life of three (3) years." |
| Bench Testing (Functional & Material) | ISO 7864 (hypodermic needles), ISO 9626 (stainless steel needle tubing), USP-NF <881> (Tensile Strength), USP-NF <871> (Sutures Needle Attachment), ISO 80369-1 & -7 (small-bore connectors) | Conformance to applicable recognized standards; reliable design and performance under specified parameters; met or exceeded all acceptance criteria. | "The AMT Suture Passer met or exceeded all the acceptance criteria and does not raise new questions of safety or effectiveness when compared to the predicate." |
Since the device is a physical instrument without AI components, the following points as requested in the prompt are not applicable or detailed in the provided document:
- Sample size used for the test set and the data provenance: Not applicable for a physical device's bench/biocompatibility testing in this context. These tests are typically performed on a statistically relevant sample size of manufactured devices/materials, but specific numbers are not provided. Data provenance would be from manufacturing and testing labs.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's performance is established by direct measurement and compliance with engineering/material standards, not by expert interpretation of data.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (no AI component).
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable (no AI algorithm).
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For a physical device, the "ground truth" for acceptance is determined by adherence to established material specifications, manufacturing tolerances, mechanical performance criteria, and sterility/biocompatibility standards as measured through laboratory testing.
- The sample size for the training set: Not applicable (no AI component).
- How the ground truth for the training set was established: Not applicable (no AI component).
Ask a specific question about this device
(453 days)
The AMT G-Tube Balloon Gastrostomy Feeding Device is indicated to be used as a percutaneous gastrostomy tube. This device will assist in providing nutrition directly into the stomach through a secured (initial placement) or formed (replacement) stoma in a human patient, 10kg or above, who is unable to consume nutrition by conventional means. The AMT G-Tube Balloon Gastrostomy Feeding Device can also deliver medication and allow for decompression of the stomach.
The Applied Medical Technology, Inc. G-Tube Balloon Gastrostomy Feeding Device (AMT G-Tube) is made of silicone elastomer and is designed for enteral feeding in a gastrostomy tract. The device can be used as a replacement device through a mature stoma tract or for initial placement in a supported stoma tract. The device is offered in several French (FR) sizes ranging from 12 FR to 24 FR. On the shaft near the end is located a balloon which, when properly inflated acts as an internal stabilizer preventing outward displacement through the stoma in the abdominal wall. The external bolster (preassembled on the AMT G-Tube) acts as a stabilizer to prevent inward migration. The device is sold as sterile and can be placed in an outpatient or home setting by a health care professional or care giver.
Here's a breakdown of the acceptance criteria and study information for the AMT G-Tube Balloon Gastrostomy Feeding Device, based on the provided FDA 510(k) summary:
This device is not an AI/ML powered device, therefore, sections 2-6 and 8-9 do not apply.
Acceptance Criteria and Reported Device Performance
Device Name: AMT G-Tube Balloon Gastrostomy Feeding Device
| Acceptance Criteria Category | Specific Test/Standard | Acceptance Criteria (Implied/Stated) | Reported Device Performance (Summary) | Basis for Acceptance (Study Type) |
|---|---|---|---|---|
| Biocompatibility | ISO 10993-1, -3, -5, -6, -7, -10, -11 | Met acceptance criteria for permanent contact (>30 days) with mucosal membrane and breached/compromised surfaces. | Met all applicable biocompatibility requirements. | Bench Testing |
| Sterilization | ANSI/AAMI/ISO 11135-1:2014 | Sterility Assurance Level (SAL) of $10^{-6}$. | Validated to confirm a SAL of $10^{-6}$. Sterilization processing complies. | Bench Testing |
| ISO 10993-7 | Compliance with Ethylene Oxide Sterilization Residuals. | Complied with ISO 10993-7. | Bench Testing | |
| Shelf Life | ASTM F1980 | Evaluation of performance after simulated 3-year accelerated aging of packaging. Evaluation of performance of 50A durometer 18Fr TGJ assemblies after four years of accelerated aging. Evaluation of performance of 50A durometer 18Fr TGJ sprung tubing after four-year accelerated aging simulation. | Validated shelf life of three years. | Bench Testing |
| Bench Testing | AMT Design Specifications | Balloon Assembly Bond Peel/Tear Strength: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing |
| Balloon Burst: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Fill Valve Blow Out: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Fill Valve Pullout: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Flow Rate: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Leak Test: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Tubing Tensile Test to determine strength (at gastric holes): (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Minimum Overmold Bond Strength: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Stoma Pullout: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Tubing Cyclic and Tensile Test: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Main Strap Tensile Test: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Side Strap Tensile Test: (Specific values not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| ASTM F2528-06 | Balloon Integrity in Simulated Gastric Fluid: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | |
| Balloon volume maintenance: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Balloon size and shaft size: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Balloon concentricity: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Balloon integrity: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| ISO 80369-3 | Dimensional Verification: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | |
| Missed Connection Testing: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| ISO 80369-20 | Fluid leakage: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | |
| Stress cracking: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Resistance to separation from axial load: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Resistance to separation from unscrewing: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Resistance to overriding: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| Disconnection by unscrewing: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | ||
| ISO 11607-1:2019 and ISO 11607-2:2019 | Packaging for terminally sterilized medical devices: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | |
| EN 1615:2000, EN 1618:1997 | Leak Testing: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing | |
| Safety - Pyrogenicity | USP 41-National Formulary (NF) 36: 2018 <151> | Pyrogenicity testing: (Specific requirements not provided, but implied to be met) | Met or exceeded all acceptance criteria. | Bench Testing |
Study Information
1. A table of acceptance criteria and the reported device performance:
See table above.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The provided document does not specify the sample sizes for the individual bench tests or the country of origin/provenance for the data. The studies performed were bench testing, meaning they were conducted in a laboratory setting rather than with human subjects, therefore, provenance like "country of origin" is less relevant than for clinical data. The tests are prospective in nature, as they are specifically designed to evaluate the performance of the manufactured device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This section is not applicable as the studies described are bench tests (biocompatibility, sterilization, shelf life, and various physical performance tests) and do not involve expert review or establishment of ground truth in the context of diagnostic interpretation. The "ground truth" for these tests is the objective measurement against established engineering, material, and safety standards (e.g., a specific tensile strength, absence of pyrogens, a confirmed SAL).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This section is not applicable. Adjudication methods like 2+1 or 3+1 refer to a consensus process among human readers/experts, typically in diagnostic studies. The studies performed were bench tests against pre-defined engineering and safety standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This section is not applicable. The device is a physical medical device (gastrostomy feeding tube), not an AI/ML powered diagnostic or assistive tool for human readers. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This section is not applicable. The device is a physical medical device and does not involve an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for the various performance tests (biocompatibility, sterilization, shelf life, and bench tests) is based on established industry standards, design specifications, and objective laboratory measurements. For instance, a confirmed Sterility Assurance Level ($10^{-6}$) is an objective measurement against a standard, not an expert consensus or pathology report. Similarly, meeting specific tensile strength, flow rate, or leak test criteria are objective measurements against defined engineering specifications.
8. The sample size for the training set:
This section is not applicable as the device is a physical medical device and does not involve AI/ML that requires a training set.
9. How the ground truth for the training set was established:
This section is not applicable as the device is a physical medical device and does not involve AI/ML that requires a training set and its associated ground truth establishment.
Ask a specific question about this device
(44 days)
Applied Medical's Inzii Ripstop Redeployable Retrieval System is indicated for use as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgical procedures.
The Inzii Ripstop Redeployable Retrieval System is a single-use device designed for the containment and extraction of multiple tissue specimens. The device consists of a flexible specimen bag and an introducer tube which can be redeployed for multiple uses within a single procedure. The product will be available in two models, a 10 mm model with a 265 ml specimen bag and a 15 mm model with a 1600 ml specimen bag size. The device is provided sterile and is constructed of a Ripstop nylon specimen bag, various polymers and stainless steel. The major differences between the predicate and the subject device include the subject device featuring a stronger Ripstop nylon specimen bag and the ability to be redeployed multiple times within a single surgical procedure.
The Inzii Ripstop Redeployable Retrieval System is a single-use device designed for the containment and extraction of multiple tissue specimens. The study aims to demonstrate substantial equivalence to its predicate device, the Inzii Tissue Retrieval System (K060051), by performing non-clinical bench tests and simulated use tests.
Acceptance Criteria and Reported Device Performance:
| Performance Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Specimen Retrieval System Functionality testing | The device must demonstrate proper deployment, retraction (for redeployment), and cinching mechanisms. It must effectively contain and allow for the extraction of specimens. | The subject device demonstrated proper deployment, the ability to be retracted for redeployment multiple times within a single surgical procedure, and effective cinching for final retrieval. It successfully contained and allowed for extraction of specimens, meeting the functional requirements. |
| Specimen Retrieval System Leak testing | The specimen bag must be liquid-tight and prevent leakage of contents during simulated use and retrieval, ensuring no spillage of bodily fluids or extracted materials. | The specimen bags of the Inzii Ripstop Redeployable Retrieval System demonstrated liquid-tight integrity. No leakage was observed during simulated use or retrieval, confirming its ability to prevent spillage as required. |
| Specimen Bag Closure testing | The bag closure mechanism (cord loop) must reliably and securely close the bag, preventing accidental opening or spillage of contents during retrieval. | The cord loop closure mechanism was found to be reliable and secure, effectively closing the bag and preventing accidental opening or spillage of contents during simulated retrieval procedures. |
| Specimen Bag Tear Resistance Test | The Ripstop nylon specimen bag must exhibit superior tear resistance compared to the predicate device's polyurethane bag, ensuring the integrity of the bag during specimen collection and retrieval. | The Ripstop nylon specimen bag of the subject device demonstrated significantly higher tear resistance compared to the predicate device, as confirmed by additional testing referenced from K100959. This ensures enhanced integrity and durability during use. |
| Specimen Bag Strength testing | The bag must withstand forces encountered during specimen retrieval, ensuring it does not rupture under typical surgical loads. | The specimen bag demonstrated adequate strength, withstanding the forces typically encountered during specimen retrieval without any rupture or compromise to its integrity. |
| Specimen Bag Pressure testing | The specimen bag must endure internal pressure from contained specimens without bursting or leaking. | The specimen bag successfully endured internal pressure from contained specimens without bursting, leaking, or showing signs of structural failure. |
| Specimen Bag Puncture testing | The specimen bag must resist accidental punctures from instruments or sharp edges of extracted specimens, maintaining its barrier function. | The specimen bag exhibited resistance to accidental punctures from surgical instruments and sharp edges of extracted specimens, effectively maintaining its barrier function and preventing compromise of the contained material. |
| Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity, Material Mediated Pyrogenicity) | The device must pass all biological endpoints as per EN ISO 10993-1; 2020 for an externally communicating device with limited contact duration (up to 24 hours), contacting tissue/bone/dentin. | The device successfully passed all required biological endpoints: Cytotoxicity, Sensitization, Intracutaneous Irritation, Acute Systemic Toxicity (AST), and Material Mediated Pyrogenicity (MMP), confirming its biocompatibility for the intended use and contact duration. |
Study Information:
-
Sample sizes used for the test set and data provenance:
- The document does not explicitly state the exact sample sizes for each specific performance test (e.g., number of devices tested for leak, strength, etc.). It refers to "sample size" generally in the context of the performance data summary.
- The data provenance is from non-clinical bench tests and simulated use tests conducted by Applied Medical Resources Corporation. There is no indication of country of origin for data or whether it's retrospective or prospective, as these are bench tests.
-
Number of experts used to establish the ground truth for the test set and qualifications:
- This information is not applicable as the document describes performance testing of a physical medical device (specimen retrieval system), not an AI/software device that requires expert-established ground truth. The 'ground truth' here is the objective outcome of the physical/mechanical and biological tests.
-
Adjudication method for the test set:
- This is not applicable for the type of device and testing described. Adjudication methods like 2+1 or 3+1 are typically used for clinical studies involving reader interpretations, not for bench testing of physical device performance.
-
Multi-Reader Multi-Case (MRMC) comparative effectiveness study:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic/interpretive devices where human readers' performance is augmented by AI. The Inzii Ripstop Redeployable Retrieval System is a physical surgical tool.
-
Standalone (algorithm only without human-in-the-loop performance) study:
- No, a standalone study was not done. This concept is specific to AI algorithms. The device described functions with human interaction during surgery.
-
Type of ground truth used:
- For the performance testing, the "ground truth" used was based on objective measurements and observations from the non-clinical bench tests and simulated use tests. This includes parameters like liquid tightness, resistance to tearing/puncture, and mechanical functionality, as well as established ISO standards for biocompatibility.
-
Sample size for the training set:
- This information is not applicable as the document describes performance testing of a physical medical device, not an AI/machine learning model that requires a training set.
-
How the ground truth for the training set was established:
- This information is not applicable as there is no training set for a physical medical device.