510(k) Data Aggregation

    K Number
    K213469

    Validate with FDA (Live)

    Device Name
    VALORY
    Manufacturer
    Date Cleared
    2021-11-18

    (21 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VALORY system is a General Radiography X-ray imaging system used in hospitals, clinics and medical practices by radiographers, radiologists and physicists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adults and pediatric patients.

    Applications can be performed with the patient in sitting, standing or lying position.

    The system is not intended for use in Mammography applications

    Device Description

    VALORY is a solid state x-ray system, a direct radiography (DR) system (product code MQB) intended to capture general radiographic images of the human body. VALORY is a ceiling mounted stationary X-Ray system with digital image capture that consists of a tube and operator console with a patient table and/or wall stand. VALORY uses Agfa's NX workstation with MUSICA2 TM image processing and flat-panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI) to capture and process the digital image.

    AI/ML Overview

    The provided text describes the VALORY system, a general radiography X-ray imaging system, and its substantial equivalence to a predicate device (DR 600, K152639). However, it does not detail specific acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy) or a study proving the device meets these criteria in the way a clinical performance study would.

    Instead, the submission focuses on demonstrating substantial equivalence through:

    • Comparison of technological characteristics.
    • Bench testing for image quality and usability.
    • Software verification and validation.
    • Compliance with electrical safety, EMC, and radiation protection standards.
    • Adherence to quality management systems and guidance documents.

    Here's an attempt to extract and organize the requested information, noting where specific details are not provided in the text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not specify quantitative acceptance criteria for image quality or clinical performance metrics (like sensitivity, specificity, or accuracy) in a traditional sense. The performance is assessed relative to the predicate device.

    Acceptance Criteria (Implicit)Reported Device Performance
    Image Quality: Equivalent or better than predicate device."Image quality validation testing was conducted using anthropomorphic adult and pediatric phantoms and evaluated by qualified internal experts and external radiographers. The radiographers evaluated the VALORY X-ray system with the DR 600 (predicate device. K 152639) using XD 14 (K211790, pending 510(k) clearance) and DR 14s (K161368) flat-panel detectors comparing overall image quality. The test results indicated that the VALORY X-ray system has at least the same if not better image quality than the predicate device (DR 600 - K152639) and other flat-panel detectors currently on the market.""Additional image quality validation testing for NX 23 was completed in scope of the DX-D Imaging Package with XD Detectors and included a full range of GenRad image processing applications compared to MUSICA 2 image processing using anonymized adult and pediatric phantoms and read by eight internal experts."
    Usability: Meets safety and workflow requirements."Usability evaluations for VALORY were conducted with external radiographers. The usability studies evaluated overall product safety, including workflow functionality for adults and pediatric patients, system movements, information and support for components. The results of the usability tests, fulfillment of the validation acceptance criteria, and assessment of remaining defects support VALORY passing usability validation testing."
    Software Safety: Acceptable risk profile."Software verification testing for NX 23 was completed... For the NX 23 (NX Orion) software there are a total of 535 risks in the broadly acceptable region and 37 risks in the ALARP region with only four of these risks identified. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk."
    Compliance: Meets relevant electrical, EMC, and radiation standards.VALORY is compliant to FDA Subchapter J mandated performance standards 21 CFR 1020.30 and 1020.31. Also compliant with IEC 60601 series, ISO 13485, ISO 14971, DICOM, IEC 62366-1, and IEC 62304.
    Functional Equivalence: Same intended use as predicate.The VALORY system has indications for use that is consistent with and substantially equivalent to the legally marketed predicate device (K152639).

    2. Sample Size Used for the Test Set and Data Provenance

    The "test set" consisted of:

    • Anthropomorphic adult and pediatric phantoms.
    • No information on the exact number of phantoms or images used in these tests.
    • The data provenance is not specified as country of origin, but it is implied to be laboratory-generated phantom data as "No clinical trials were performed in the development of the device. No animal or clinical studies were performed in the development of the new device."
    • The studies were retrospective in the sense that they were conducted for submission, but not using existing clinical patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Image Quality Evaluation: "qualified internal experts and external radiographers" evaluated images from phantoms compared to the predicate device. For NX 23 testing, "eight internal experts" read images. Specific qualifications (e.g., years of experience) for these experts are not provided.
    • Usability Evaluation: "external radiographers." Specific numbers and qualifications are not provided.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method for the expert evaluations, such as 2+1 or 3+1. It states experts "evaluated" and "read" the images, implying their findings were used as the basis for comparison.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No MRMC comparative effectiveness study was done.
    • The device is a general radiography X-ray imaging system, not explicitly an AI-assisted diagnostic device for specific disease detection. Its image processing (MUSICA/MUSICA2/MUSICA3) is an established technology from the predicate device and other Agfa systems. The study focuses on demonstrating image quality equivalence, not on reader performance improvement with AI assistance.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • The VALORY system itself is an imaging system, not a standalone diagnostic algorithm.
    • The software risk assessment for NX 23 (the workstation software) involved evaluating risks without explicit human interaction for every risk listed, focusing on software failure modes. However, this is not a diagnostic "algorithm only without human-in-the-loop performance" in the sense of an AI CAD system. The image processing algorithms within MUSICA are an integral part of the system's image generation, not a separate diagnostic interpretation.

    7. The Type of Ground Truth Used

    • Image Quality: The ground truth for image quality evaluation was based on the performance of the predicate device (DR 600) and other marketed flat-panel detectors, combined with expert evaluation of images from anthropomorphic phantoms. It's a comparative assessment rather than a definitive "ground truth" of a disease state.
    • Usability: Ground truth for usability was established through the feedback and performance of external radiographers during usability studies, ensuring the system met workflow and safety requirements.
    • Software/Safety: Ground truth for software safety and compliance was against established standards (IEC, ISO, FDA regulations) and internal risk analyses.

    8. The Sample Size for the Training Set

    The document makes no mention of a "training set" in the context of machine learning or AI models. The device is being cleared based on substantial equivalence to an existing X-ray system, not as a novel AI diagnostic algorithm that requires a distinct training dataset. The "image processing algorithms" mentioned are presented as existing, cleared technologies from previous Agfa devices.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a distinct training set for an AI model, this information is not applicable and not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212145

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-31

    (53 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 800 with DSA system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures:

    • · Positioning fluoroscopy procedures
    • · Gastro-intestinal examinations
    • · Urogenital tract examinations
    • · Angiography
    • · Digital Subtraction Angiography

    It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for projection radiography of all body parts.

    In addition, the system provides the Agfa Tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-ray systems. Digital Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.

    Not intended for cardiovascular and cerebrovascular contrast studies. Not intended for mammography applications.

    Device Description

    Agfa's DR 800 with DSA medical device is a fluoroscopic x-ray system that includes digital angiography (product code JAA) intended to capture tomographic, static and dynamic images of the human body. The DR 800 is a floor-mounted radiographic, fluoroscopic and tomographic system that consists of a tube and operator console with a motorized tilting patient table. FLFS overlay and bucky with optional wall stand and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA image processing and flat-panel detectors for digital, wide dynamic range and angiographic image capture. It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.

    This submission is to add the newest version of the DR 800 with Digital Subtraction Angiography (DSA) to Agfa's radiography portfolio.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the Agfa DR 800 with DSA, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly list quantitative acceptance criteria in a table format for performance metrics. Instead, it describes a more qualitative approach, focusing on equivalence to predicate devices and confirmation through expert evaluation.

    Acceptance Criteria (Inferred from text)Reported Device Performance
    Bench Testing (General Performance)"Technical and acceptance testing was completed on the DR 800 with DSA in order to confirm the medical device functions and performs as intended. All deviations or variances are documented in a defect database and addressed in the CRD documentation and verified. All mitigations have been tested and passed. All design input requirements have been tested and passed. All planned verification activities have been successfully completed."
    Functionality and Usability"Performance functionality and usability evaluations were conducted with qualified experts. The results of these tests fell within the acceptance criteria for the DR 800 with DSA; therefore, the DR 800 supports GenRad, Full Leg/ Full Spine (FLFS), roadmapping and Digital Subtraction Angiography (DSA) workflow."
    Clinical Image Quality (DSA)"Clinical image validation was conducted using anthropomorphic phantoms and evaluated by qualified experts. The radiographers evaluated the DSA image quality on the DR 800 by comparing overall image quality with the primary predicate A device (K190373). Diagnostic confidence for DSA image quality and roadmapping on the DR 800 was between good and excellent." The document also states, "Clinical image quality evaluation is not essential in establishing substantial equivalence for the DR 800 with DSA. Adequate Bench Testing results should be sufficient to determine device safety and effectiveness." This indicates that while performed, it wasn't a strict acceptance criterion in the same vein as quantitative safety/effectiveness thresholds.
    Software Verification & Validation (Safety/Risk)"The complete device has been certified and validated. During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field." "For the NX 23 (NX Orion) software there are a total of 535 risks in the broadly acceptable region and 37 risks in the ALARP region with only four of these risks identified. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk."
    Electrical Safety and Electromagnetic Compatibility (EMC) TestingThe device is compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-54. The DR 800 is also compliant with FDA Subchapter J mandated performance standards 21 CFR 1020.30 - 1020.32.
    Quality Management, Risk Management, DICOM, Usability EngineeringThe company's in-house procedures conform to ISO 13485, ISO 14971, ACR/NEMA PS3.1-3.20 (DICOM), and IEC 62366-1. (This implies compliance with these standards as part of overall acceptance).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated in terms of number of images or cases. The document mentions "anthropomorphic phantoms" for clinical image validation.
    • Data Provenance: The study used "anthropomorphic phantoms," which are physical models designed to simulate human anatomy for imaging purposes. This indicates a laboratory/phantom study rather than real patient data. The country of origin for the phantom data is not specified, but the submission is from Agfa N.V. (Belgium). It is a prospective study in the sense that the new device was evaluated with these phantoms.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not explicitly stated. The document mentions "qualified experts" and "radiographers."
    • Qualifications of Experts: Described as "qualified experts" and "radiographers." No specific experience levels (e.g., "10 years of experience") are provided.

    4. Adjudication Method for the Test Set

    Not specified. The document states "evaluated by qualified experts" and "radiographers evaluated...by comparing overall image quality with the primary predicate A device," implying a comparative evaluation rather than a strict adjudication process for ground truth establishment.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If so, What was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted. This is not an AI-assisted diagnostic device; it's a conventional X-ray system with digital image processing and DSA capabilities. The study compared the device's image quality to a predicate device, focusing on equivalence, not human reader improvement with AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, the "Bench Testing" and "Software Verification and Validation Testing" sections describe standalone performance evaluations of the device's functions and image processing algorithms. The "Clinical image validation" with phantoms also focuses on the device's output (image quality) rather than human interaction with the device in a diagnostic workflow where the human acts as the ultimate decision-maker for the study’s performance outcome.

    7. The Type of Ground Truth Used

    The "ground truth" for the image quality evaluation was based on expert comparison and qualitative assessment of images produced by the device, specifically assessing "diagnostic confidence for DSA image quality and roadmapping" as "between good and excellent" when compared to a predicate device. This is primarily an expert consensus on image quality rather than pathology, clinical outcomes, or a gold standard.

    8. The Sample Size for the Training Set

    Not applicable. This device is an X-ray imaging system, not a machine learning or AI algorithm that requires a training set of data. The image processing algorithms are described as being "similar to those previously cleared" or "similar to the primary predicate device."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this device does not utilize a machine learning model that would require a ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211790

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-30

    (50 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Agfa's DX-D Imaging Package with XD Detectors is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of the human anatomy. The DX-D Imaging Package with XD Detectors may be used wherever conventional screen-film systems may be used.

    Agfa's DX-D Imaging Package with XD Detectors is not indicated for use in mammography.

    Device Description

    The DX-D Imaging Package, previously cleared under K142184, is a solid state x-ray system, a direct radiography (DR) system (product code MQB) intended to capture general radiographic images of the human body. It is a combination of Agfa's NX workstation with MUSICA TM image processing and one or more flat-panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS). It is capable of replacing other direct radiography, including computed radiography systems with conventional or phosphorous film cassettes.

    This submission is to add the XD Detectors (XD 10/10+, XD 14/14+ and XD 17/17+) Flat Panel Detectors to Agfa's DX-D Imaging Package portfolio. Agfa's XD Detectors are currently marketed by Vieworks Co. Ltd. as FXRD-4343VAW/VAW Plus. FXRD-3643VAW/VAW Plus. and FXRD-2530VAW/VAW Plus which is the predicate for this submission (K200418).

    The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are identical to those previously cleared in the DX-D Imaging Package - DX-D 40 (K142184reference) device and other devices in Agfa's radiography portfolio today. The addition of the offline workflow is identical to the Vivix-S VW (K200418) predicate device.

    Principles of operation and technological characteristics of the new, predicate and reference devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically similar to the predicate device (K200418) which includes the addition of an offline workflow capable of storing up to 200 images on the flat-panel detector for later viewing. It uses the same NX workstation with MUSICA™ image processing as the reference device (K142184) and the same flat panel detectors of the scintillator-photodetector type (Cesium Iodide - Csl or Gadolinium Oxysulfide - GOS) to capture and digitize the images as the predicate device (K200418). Laboratory data and image quality evaluations conducted with internal specialists confirm that performance is equivalent to the predicate. Differences in devices do not alter the intended diagnostic effect nor do they impact the safety and effectiveness of the device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Agfa’s DX-D Imaging Package with XD Detectors. The submission focuses on demonstrating substantial equivalence to a predicate device (Vieworks Vivix-S VW, K200418) and a reference device (Agfa DX-D Imaging Package -DX-D 40, K142184) rather than establishing new performance criteria or conducting extensive clinical trials of the AI component.

    Key takeaway: This submission primarily focuses on the physical components of the imaging system (detectors, workstation, image processing) and their equivalence to existing cleared devices. It does not contain details about specific acceptance criteria or performance metrics for an AI algorithm in the way one might expect for a novel AI-driven diagnostic device. The "AI" mentioned is related to existing "MUSICA image processing" which is identical to previously cleared versions and is referred to as "image processing algorithms", rather than a new AI model with specific diagnostic performance targets.

    Therefore, many of the requested details, especially those related to AI-specific performance criteria, ground truth establishment for a test set, and multi-reader studies, are not explicitly present in the provided document. The performance data presented refers to the physical detector characteristics (Spatial Resolution, DQE, MTF).

    Here's an attempt to answer the questions based on the provided text, highlighting where information is absent:


    1. A table of acceptance criteria and the reported device performance

    The document does not define explicit "acceptance criteria" in terms of clinical performance metrics for an AI algorithm. Instead, it demonstrates performance equivalence of the new detector models to existing ones.

    Performance CharacteristicDX-D 40 Flat-Panel Detector (K142184) (Reference)XD 10/10+ Wireless Detector (New Device)XD 14/14+ Wireless Detector (New Device)XD 17/17+ Wireless Detector (New Device)
    Spatial Resolution3.5 lp/mm (for 6007/100 & 6007/200)4.0 lp/mm3.5 lp/mm3.5 lp/mm
    DQE @ 1 lp/mmCsI: 0.494, GOS: 0.259XD: 0.500, XD+: 0.587XD: 0.425, XD+: 0.587XD: 0.412, XD+: 0.587
    DQE @ 2 lp/mmCsI: 0.379, GOS: 0.157XD: 0.401, XD+: 0.445XD: 0.321, XD+: 0.399XD: 0.345, XD+: 0.407
    DQE @ 3 lp/mmCsI: 0.215, GOS: 0.061XD: 0.288, XD+: 0.316XD: 0.206, XD+: 0.257XD: 0.220, XD+: 0.280
    MTF @ 1 lp/mmCsI: 0.685, GOS: 0.589XD: 0.729, XD+: 0.650XD: 0.751, XD+: 0.635XD: 0.726, XD+: 0.656
    MTF @ 2 lp/mmCsI: 0.386, GOS: 0.266XD: 0.424, XD+: 0.315XD: 0.446, XD+: 0.302XD: 0.428, XD+: 0.311
    MTF @ 3 lp/mmCsI: 0.209, GOS: 0.115XD: 0.236, XD+: 0.157XD: 0.247, XD+: 0.152XD: 0.231, XD+: 0.161

    The "acceptance criteria" appear to be met by demonstrating that the new detectors (XD series) have comparable or superior technical performance characteristics (Spatial Resolution, DQE, MTF) to the existing reference devices, and that the image processing ("MUSICA™ image processing") is identical to previously cleared devices. The document states: "The results of these tests fell within the acceptance criteria for the DX-D Imaging Package with XD Detectors." However, the quantitative thresholds for these "acceptance criteria" are not specified beyond the presented performance values.

    2. Sample size used for the test set and the data provenance:

    • Test Set (for image quality evaluation): "anthropomorphic adult and pediatric phantoms".
      • Data Provenance: Not explicitly stated (likely internal laboratory data, given "internal specialists"). This was bench testing, not clinical data from patients.
    • Software Test Iterations (NX 23): Two software iterations were tested.
    • Performance Functionality Evaluation: Not a sample size of data, but related to the number of experts (see below).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Image Quality Evaluations: "qualified internal experts" (radiographers) evaluated the overall image quality using phantoms. The exact number of radiographers is not specified.
    • Performance Functionality Evaluations: "four qualified experts". Their specific qualifications (e.g., radiologist, years of experience) are not detailed beyond "qualified experts".
    • Ground Truth: For the phantom studies, the ground truth is inherently defined by the known properties of the phantoms (e.g., specific structures, resolution targets).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not explicitly stated. The evaluations seem to be internal assessments for comparison rather than a formal human reader study with adjudication for a clinical diagnostic task.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was performed. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld."
    • The "image processing algorithms in the new device are identical to those previously cleared." This suggests that the "AI" (MUSICA image processing) is not a new or modified component requiring a new MRMC study to demonstrate clinical impact or improvement for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Performance data (Spatial Resolution, DQE, MTF) are presented for the detectors themselves, which can be considered "standalone" technical performance metrics of the hardware.
    • For the MUSICA image processing software, its "standalone" performance is implied to be equivalent to its previously cleared versions since it is "identical." No new quantitative standalone performance metrics for the algorithm itself are provided in this submission beyond this statement of identity.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the technical performance data (Spatial Resolution, DQE, MTF) related to the detectors: The ground truth is based on standard metrology and engineering principles for measuring detector performance using physical test objects (e.g., bar patterns, uniform fields) and laboratory equipment according to established standards.
    • For the image quality evaluations: The ground truth was based on "anthropomorphic adult and pediatric phantoms," meaning the content and structures within the phantoms served as the reference.
    • For the software testing: Ground truth for software verification and validation is against pre-defined requirements and design specifications, with "deviations or variances...documented in a defect database and addressed."

    8. The sample size for the training set:

    The document mentions "MUSICA™ image processing" which contains algorithms. However, this submission states these algorithms are "identical to those previously cleared" and does not describe the development or training of any new AI models. Therefore, information about a training set for a novel AI algorithm is not applicable to this 510(k) submission, as it is leveraging previously cleared technology.

    9. How the ground truth for the training set was established:

    As no new AI model training is described in this submission, information on how a training set's ground truth was established is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193262

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-09

    (104 days)

    Product Code
    Regulation Number
    892.1740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 600 system is a General Radiography X-ray imaging system used in hospitals, clinics and medical practices by radiographers, radiologists and physicists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

    In addition, the system provides the Agfa tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-ray systems. Digital tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.

    Applications can be performed with the patient in the sitting, standing or lying position.

    This system is not intended for mammography applications.

    Device Description

    The DR 600 with Tomosynthesis is a tomographic and solid state x-ray system (product codes IZF and MQB) intended to capture tomographic slices and static images of the human body. The DR 600 with Tomosynthesis is a ceiling mounted tomographic and general radiographic system that consists of a tube and operator console with a motorized patient table and/or wall stand. The DR 600 with Tomosynthesis uses Agfa's NX workstation with MUSICA2 ™ image processing and flat-panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS). It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.

    AI/ML Overview

    The provided text describes the Agfa DR 600 with Tomosynthesis device and its K193262 510(k) submission. However, it does not contain specific acceptance criteria or a detailed clinical study demonstrating the device's meeting of these criteria. The document focuses on showing substantial equivalence to predicate devices primarily through technological characteristics and bench testing, not through comparative clinical effectiveness studies with explicit acceptance criteria for diagnostic performance.

    Therefore, many of the requested details about acceptance criteria, clinical study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies are not present in the provided text. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device."

    Based on the available information, here's what can be extracted and what is missing:


    Acceptance Criteria and Device Performance (as inferred from the document's approach to substantial equivalence):

    Since no specific acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, AUC) are presented, the "acceptance criteria" for this 510(k) appear to be primarily focused on demonstrating substantial equivalence to predicate devices through technical specifications, image quality evaluations (bench testing), and compliance with various electrical safety, EMC, and software standards.

    Acceptance Criteria (Inferred from Document)Reported Device Performance
    1. Technological Characteristics are Identical/Equivalent to Predicates: - Communications (DICOM) - Flat Panel Detectors (type, material, sizes, pixel size, dynamic range) - Operator Workstation (Agfa NX) - Image Processing (MUSICA DTS, MUSICA2, MUSICA3/3+) - Operating System (Windows 7, 8, 8.1, 10) - Display System (Separately cleared medical display) - Power Supply - Generators2. Indication for Use statement is consistent/identical to predicates.3. Performance/Functionality as Intended: - Confirmed functions and performs as intended. - Supports a tomographic workflow and Smart Dr visualization (including adult and pediatric patients).4. Image Quality Equivalent to Predicate: - For both adult and pediatric patients.5. Software Validation: - Verification and validation plans confirmed. - Risk assessment shows no unacceptable risks.6. Electrical Safety and EMC Compliance: - Adherence to specified IEC standards (60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54). - Compliance with FDA Subchapter J mandated performance standard 21 CFR 1020.30 and 1020.31.7. Quality Management System Compliance: - Adherence to ISO 13485:2015, ISO 14971:2012, ACR/NEMA PS3.1-3.20 (DICOM).1. Technological Characteristics: - Communications: Same as both predicates (DICOM). - Flat Panel Detectors: Same as both predicates (Flat Panel Detectors, GOS/CsI Scintillator, various sizes 17x17, 14x17, 10x10 in., 148 µm pixel size in primary predicate, 139 µm in other, 16 bit dynamic range in primary predicate, 14 bit in other). The new device shares characteristics with both, indicating equivalence. - Operator Workstation: Same as both predicates (Agfa NX). - Image Processing: MUSICA DTS, MUSICA2, MUSICA3/3+. The addition of tomographic image processing is identical to the DR 800 (K183275) primary predicate device. - Operating System: Same as predicate K183275 (Windows 7, 8, 8.1, 10). - Display System: Same as both predicates (Separately cleared medical display K051901). - Power Supply: Same as predicate K152639 (50-60 Hz, 380/400/415/440/480V + 10%). - Generators: Same as predicate K183275 (Choice of three models: 50, 65KW, 80 KW). - Overall: "Principles of operation and technological characteristics of the new and predicate devices are the same."2. Indication for Use: "The DR 600 system is a General Radiography X-ray imaging system... In addition, the system provides the Agfa tomosynthesis option... Digital tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep." This is stated to be "virtually identical" to K152639 with the tomosynthesis addition from K183275.3. Performance/Functionality: - "Technical and acceptance testing was completed on the DR 600 in order to confirm the medical device functions and performs as intended. All deviations or variances are documented... All design input requirements have been tested and passed." - "Functionality evaluations were conducted with three qualified radiographers. Usability and TBD. The results of these tests fell within the acceptance criteria for the DR 600; therefore, the DR 600 supports a tomographic workflow and Smart Dr visualization including adult and pediatric patients."4. Image Quality: "Image quality bench tests were conducted in support of this 510(k) submission in which anthropomorphic adult and pediatric images taken with the DR 600 and the primary predicate device, DR 800 (K183275) were compared to ensure substantial equivalency. The test results indicated the image processing of the DR 600 passed the acceptance criteria and was equal to the image processing for the primary predicate, DR 800 (K183275) device for both adult and pediatric patients."5. Software Validation: "Verification and validation testing confirmed the device meets performance, safety, usability and security requirements... For the NX22 (NX Nomad) software there are a total of 342 risks in the broadly acceptable region and 27 risks in the ALARP region with only one of these risks identified. Zero risks were identified in the Not Acceptable Region."6. Electrical Safety and EMC Compliance: Document states compliance with all listed IEC standards and FDA performance standards.7. Quality Management System Compliance: Document states adherence to all listed ISO and other standards.

    Study Details (Based on the provided text):

    1. Sample sizes used for the test set and the data provenance:

      • Test Set: No specific numerical sample size is mentioned for image quality evaluations beyond "anthropomorphic adult and pediatric images." The document mentions "functionality evaluations were conducted with three qualified radiographers," but this refers to human user testing of workflow and usability, not diagnostic image performance.
      • Data Provenance: Not explicitly stated, but likely retrospective as it refers to comparisons of images taken with the new device and a predicate device. The comparison of anthropomorphic phantom images suggests a controlled laboratory setting.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document states "Laboratory data and image quality evaluations conducted with internal and independent specialists confirm that performance is equivalent to the predicates." It also mentions "clinical image quality evaluations for adults and pediatric patients" and "functionality evaluations were conducted with three qualified radiographers."
      • Number of Experts: At least "three qualified radiographers" for functionality, and "internal and independent specialists" for image quality, but exact numbers or specific qualifications (e.g., years of experience, board certification) are not detailed.
      • Qualifications: "Qualified radiographers" and "internal and independent specialists."
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • No adjudication method is described for the image quality evaluations or other performance tests. The comparison to predicates implies direct visual or quantitative comparison by specialists.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device."
      • Therefore, no effect size of human readers improving with AI assistance is provided as this type of study was not conducted. The device in question is an imaging system, not an AI-based diagnostic assistance tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The "performance data including clinical image quality evaluations for adults and pediatric patients" involved the system's output. The "image processing of the DR 600 passed the acceptance criteria and was equal to the image processing for the primary predicate, DR 800 (K183275) device for both adult and pediatric patients." This implies an evaluation of the algorithm's output (image quality) without necessarily focusing on a human-in-the-loop diagnostic task. So, in essence, standalone image quality performance was evaluated against a predicate.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for image quality evaluation appears to be comparison to a cleared predicate device's image quality, as judged by "internal and independent specialists," using anthropomorphic phantoms. There is no mention of pathological confirmation or patient outcomes for establishing ground truth, as it was not a clinical trial.
    7. The sample size for the training set:

      • This device is an X-ray system with image processing, not a machine learning/AI algorithm that requires a "training set" in the traditional sense of AI development. The software capabilities (MUSICA DTS, MUSICA2, MUSICA3/3+) are described as being identical to previously cleared versions in predicate devices. Therefore, a "training set" for a new AI model is not applicable here.
    8. How the ground truth for the training set was established:

      • Not applicable, as no new AI model training set is described. The image processing algorithms are identical to those previously cleared.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191884

    Validate with FDA (Live)

    Device Name
    DR 100s
    Manufacturer
    Date Cleared
    2019-08-09

    (25 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 100s system is a mobile X-ray imaging system used in hospitals, clinics and medical practices by radiographers and radiologists to make, process and view static X-ray radiographic images of the skeleton (including skull, spinal column and extremities), chest, abdomen and other body parts on adult, pediatric or neonatal patients.

    Applications can be performed with the patient in the sitting, standing or lying position.

    This device is not intended for mammography applications.

    Device Description

    Agfa's DR 100s is a mobile x-ray system, a direct radiography system (product code ILL) intended to capture images of the human body. The device is an integrated mobile digital radiography x-ray system. The complete DR 100s systems consists of the mobile x-ray unit with integrated x-ray generator and NX software and one or more DR detectors. The new device uses Agfa's NX workstation with MUSICA image processing and flat-panel detectors for digital image capture. It is compatible with Agfa's computed radiography systems as well.

    This submission is to add another mobile unit to Agfa's direct radiography portfolio.

    The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are identical to those previously cleared in the DX-D 100 (K103597) and other devices in Agfa's radiography portfolio today, which includes DR 600 (K152639), DR 400 (K141192) and DR 800 (K183275).

    Significant dose reduction can be achieved using the DR 100s with patented Agfa's MUSICA imaging processing and CsI flat-panel detectors. Testing with board certified radiologists determined that Cesium Bromide (CR) and Cesium Iodine (DR) detectors when used with MUSICA imaging processing can provide dose reduction between 50-60% for adult patients and up to 60% for pediatric and neonatal patients when compared to traditional Barium Fluoro Bromide CR systems (K141602).

    Principles of operation and technological characteristics of the new and predicate device are the same. The new device is virtually identical to Agfa's DX-D 100(K103597) with the exception that it has a telescopic column and ergonomic design. It uses the same flat panel detectors to capture and digitize the image. Differences in devices do not alter the intended diagnostic effect.

    AI/ML Overview

    The provided text describes the DR 100s mobile X-ray system and its substantial equivalence to a predicate device, the DX-D 100. However, it does not contain specific acceptance criteria or a detailed clinical study proving the device meets these criteria in the manner requested.

    The document states that the DR 100s uses the same image processing algorithms and flat-panel detectors as previously cleared devices, and that "Clinical image validation was conducted during testing in support for the 510(k) clearances for the flat-panel detectors (K161368 and K172784) and MUSICA software (K183275) in a previous submission." It also mentions that "Image quality bench tests were conducted in support of this 510(k) submission in which anthropomorphic adult and pediatric images taken with the DR 100s and the predicate device, DX-D 100 (K103597) were compared to ensure substantial equivalency. The test results indicated the image processing of the DR 100s passed the acceptance criteria."

    This means the primary method for demonstrating equivalence and meeting acceptance criteria was through bench testing and referencing prior clearances for components. There's no detailed mention of a specific, standalone clinical study with human patients for the DR 100s itself, nor a multi-reader multi-case (MRMC) study.

    Therefore, many of the requested details about a clinical study's methodology (sample size, data provenance, expert numbers, adjudication, MRMC results, ground truth types) cannot be extracted from this document directly for the DR 100s.

    Here's an attempt to answer the questions based only on the provided text, acknowledging where information is missing:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly list quantitative acceptance criteria in a table format for image quality or specific diagnostic performance metrics (e.g., sensitivity, specificity). Instead, it states:

    • "The test results indicated the image processing of the DR 100s passed the acceptance criteria." (for image quality bench tests)
    • "All design input requirements have been tested and passed." (for technical and acceptance testing)
    • "The results of these tests fell within the acceptance criteria for the DR 100s X-ray system therefore, the DR 100s supports a General radiographic workflow including adult and pediatric patients." (for usability and functionality)

    Without specific numerical criteria, a performance table cannot be constructed. The main "performance" metrics provided are technical specifications of the flat-panel detectors (DQE, MTF, pixel size, etc.) which are compared to predicate devices but don't represent acceptance criteria for a clinical study proving diagnostic performance.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for the image quality bench tests. The document only mentions "anthropomorphic adult and pediatric images."
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The studies are described as "bench tests" using anthropomorphic phantoms, not real patient data directly for the DR 100s itself. The "clinical image validation" mentioned refers to prior 510(k) clearances for components (detectors and software), not this specific device submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not specified.
    • Qualifications of Experts: It vaguely mentions "internal experts" for usability and functionality evaluations. For the "clinical image validation" of previously cleared components, it references validation conducted with "board certified radiologists" for dose reduction testing (K141602), but this is for a different aspect (dose reduction with MUSICA and CsI detectors) and potentially not the core image quality comparison for equivalence of the total system.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No MRMC study is explicitly mentioned for the DR 100s. The document states "No clinical trials were performed in the development of the device. No animal or clinical studies were performed in the development of the new device."
    • Effect Size: Not applicable, as no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The device is a mobile X-ray system and image processor, not an AI algorithm for diagnosis. Its performance is inherent in the image acquisition and processing. The "image processing of the DR 100s passed the acceptance criteria" refers to the system's ability to produce images comparable to the predicate. Therefore, the "standalone" performance is the image quality produced directly by the system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the image quality bench tests, the "ground truth" was likely established through objective phantom measurements and comparison to the predicate device, rather than clinical ground truth (expert consensus, pathology, or outcomes data). The document refers to "anthropomorphic adult and pediatric images" meaning images of phantoms designed to mimic human anatomy.

    8. The sample size for the training set

    • Training Set for DR 100s: The DR 100s system itself is a hardware device with integrated software for image processing (MUSICA). It's not an AI model that undergoes a "training" phase in the conventional sense (e.g., deep learning). The MUSICA image processing algorithms are stated to be "identical to those previously cleared" in other Agfa devices (K103597, K152639, K141192, K183275). Therefore, any "training" (algorithm development/tuning) would have occurred for these prior versions/devices, and no specific training set for the DR 100s is mentioned.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified as the DR 100s itself does not undergo a "training" phase like a new AI algorithm. The MUSICA algorithms were previously developed and cleared.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183275

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-02-01

    (70 days)

    Product Code
    Regulation Number
    892.1740
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures:

    • Positioning fluoroscopy procedures
    • Gastro-intestinal examinations
    • Urogenital tract examinations
    • Angiography
      It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts.
      In addition, the system provides the Agfa Tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-Ray systems. Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.
      The DR 800 is not intended for mammography applications.
    Device Description

    Agfa's DR 800 with Tomosynthesis a tomographic and fluoroscopic x-ray system (product codes IZF and JAA) intended to capture tomographic slices of the human body. The DR 800 is a floormounted radiographic, fluoroscopic and tomographic system that consists of a tube and operator console with a motorized tilting patient table and bucky with optional wall stand, FLFS overlay and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA image processing and flat-panel detectors for digital and wide dynamic range capture. It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.

    AI/ML Overview

    The Agfa DR 800 with Tomosynthesis underwent bench testing and software verification and validation to demonstrate substantial equivalence to its predicate devices, the GE Medical System's Discover XR656 with VolumeRAD (K132261) and Agfa's previous version of the DR 800 with MUSICA Dynamic (K180589). The primary focus of the testing for this submission was on the new Digital TomoSynthesis (DTS) software and its performance in generating tomographic slices.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Technical & Acceptance TestingAll deviations or variances are documented, addressed in CR&T (Corrective and Remedial Actions) documentation, and verified. All mitigations have been tested and passed. All design input requirements have been tested and passed. All planned verification activities have been successfully completed.Verification and validation testing confirmed the device meets performance, safety, usability, and security requirements. Pediatric indications were also taken into account. Results were verified and validated. Technical and acceptance testing was completed on the DR 800 with Tomosynthesis to confirm the medical device functions and performs as intended. All deviations or variances are documented in a defect database and addressed in the CR&T documentation and verified. All mitigations have been tested and passed. All design input requirements have been tested and passed. All planned verification activities have been successfully completed.
    Usability & Functionality EvaluationThe results of these tests fell within the acceptance criteria for the DR 800 X-ray system.Usability and functionality evaluations were conducted with qualified independent radiographers and internal experts. The results of these tests fell within the acceptance criteria for the DR 800 X-ray system; therefore, the DR 800 supports a radiographic, fluoroscopic, and tomosynthesis workflow including dynamic and static imaging, continuous and rapid sequence exams, tomographic slices calibration, and positioning.
    Image Quality Validation (Adults)The reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria. DTS images were suitable for diagnosis.Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers and internal experts. The image quality validation included testing a full range of applications for the DR 800 X-ray system with Tomosynthesis compared to reference images from the primary predicate GE Discovery XR656 with VolumeRAD (K132261) using anonymized adult phantoms. The test results indicated that the reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria and that the DR 800 with Tomosynthesis is capable of making DTS studies for adult patients. The test results showed MUSICA Digital TomoSynthesis (DTS) images were suitable for diagnosis for adult patients.
    Image Quality Validation (Pediatric)The reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria. Both 5x the dose and 10x the dose images were clinically sufficient and within the intended use, and DTS images were suitable for diagnosis for pediatric patients.Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers and internal experts. The image quality validation included testing using anonymized pediatric phantoms. The pediatric phantom image quality validation testing analyzed five tomographic slices at 5x the dose and five tomographic slices at 10x the dose. Both the 5x the dose and 10x the dose images are clinically sufficient and within the intended use. The test results indicated that the reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria and that the DR 800 with Tomosynthesis is capable of making DTS studies for pediatric patients. The test results showed MUSICA Digital TomoSynthesis (DTS) images were suitable for diagnosis for pediatric patients.
    Software Risk AssessmentNo risks identified in the Not Acceptable Region. The device is assumed to be safe, and the benefits of the device outweigh the residual risk.During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field. For the NX4.x.21 (NX Mentor) there are a total of 322 risks in the broadly acceptable region and 27 risks in the ALARP (As Low As Reasonably Practicable) region with only eight of these risks identified. Zero risks were identified in the Not Acceptable Region.
    Electrical Safety & EMC TestingCompliance with various IEC 60601 standards and FDA Subchapter J.The DR 800 with Tomosynthesis is compliant to the FDA Subchapter J mandated performance standard 21 CFR 1020.30 - 1020.32. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-54 was confirmed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a numerical count of cases/images. The testing involved "anthropomorphic phantoms" for image quality evaluation, including both adult and pediatric phantoms. The pediatric phantom testing analyzed "five tomographic slices at 5x the dose and five tomographic slices at 10x the dose."
    • Data Provenance: The data provenance is from bench testing using anonymized anthropomorphic phantoms. This indicates that the data is prospective in the sense that the phantoms were specifically used for this testing, but it is not from human patients. The country of origin of the data is not specified, but the manufacturer is Agfa N.V. (Belgium).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Number of Experts: "Qualified independent radiographers and internal experts" were used for usability, functionality, and image quality evaluations. The exact number of each group is not specified.
    • Qualifications of Experts: They are described as "qualified independent radiographers and internal experts" and "qualified radiologists" (in the "Descriptive characteristics and performance data including image quality evaluations by qualified radiologists are adequate to ensure equivalence" section). Specific details like years of experience or subspecialty are not provided.

    4. Adjudication Method for the Test Set

    • The document implies that the "qualified independent radiographers and internal experts" evaluated the images and that the results "fell within the acceptance criteria" or "passed the acceptance criteria," suggesting a consensus or individual assessment against predefined criteria. However, a specific adjudication method (e.g., 2+1, 3+1) is not explicitly stated.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No MRMC comparative effectiveness study was done. The study design described is a bench test comparison of the device against a predicate device's reference images using phantoms, with evaluation by human experts, rather than an assessment of human reader performance with or without AI assistance. The device itself is an imaging system, not an AI-powered diagnostic tool for interpretation assistance in the sense of comparing human performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone evaluation of the algorithm's output was done, as part of the image quality validation. The "reconstruction software of the image processing for Digital TomoSynthesis (DTS)" was evaluated to ensure the generated tomographic images were suitable for diagnosis. This is an assessment of the algorithm's output (the tomographic slices) without direct human intervention in the image generation process, beyond setting up the acquisition parameters.

    7. The Type of Ground Truth Used

    • The ground truth for the image quality evaluation was based on comparison to reference images from the primary predicate device (GE Discovery XR656 with VolumeRAD - K132261) using anthropomorphic phantoms, and expert assessment by "qualified independent radiographers and internal experts" confirming the images were "suitable for diagnosis" and "clinically sufficient." It is not pathology, or outcomes data.

    8. The Sample Size for the Training Set

    • The document does not explicitly state the sample size for the training set for the MUSICA DTS software. It mentions that "The image processing algorithms in the new device are similar to those previously cleared in the DR 800 with MUSICA Dynamic (K180589) and other devices in Agfa's radiography portfolio today... The addition of the tomographic image processing is similar to the predicate device (K132261)." This suggests leveraging existing, previously trained algorithms or development methodologies, rather than describing a specific new training dataset for this submission.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not provided in the document. As noted above, the submission emphasizes similarity to existing, cleared technologies, rather than detailing the unique training of a novel algorithm from scratch.
    Ask a Question