Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K240253

    Validate with FDA (Live)

    Device Name
    Reach Needle
    Manufacturer
    Date Cleared
    2024-03-21

    (51 days)

    Product Code
    Regulation Number
    868.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reach Needle is used for in injecting local anesthetics into a patient to provide regional anesthesia.

    Device Description

    The Reach Needle is a sterile, single use, 4-inch-long needle used for administering anesthetic solutions for regional anesthesia. The Reach Needle is 27-qauge needle that is reinforced to reduce needle flex during administration of anesthesia solutions. The needle tip is a short bevel quincke style tip that extends approximately 2mm past the reinforcing shaft, providing a distal stop for consistent injection depth. The distal end of the needle has a slight bend to help access difficult to reach areas such as nasal anatomy. The needle hub is compatible with luer lock connectors. The Reach Needle is intended for use in adults (≥ 22 years).

    AI/ML Overview

    The provided text is a U.S. FDA 510(k) summary for the "Reach Needle," an anesthesia conduction needle. It outlines the device's characteristics, intended use, and comparison to a predicate device.

    Crucially, the document explicitly states under the section "Summary of clinical test": "Clinical testing was not necessary for this device."

    This means that the device's acceptance was based on non-clinical testing (e.g., in vitro or bench testing) and a demonstration of substantial equivalence to a legally marketed predicate device, rather than a clinical study involving human patients or complex AI algorithms requiring extensive ground truth establishment and expert adjudication.

    Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria, as the nature of the device (a medical needle, not an AI or imaging device) and the regulatory pathway chosen (510(k) based on substantial equivalence and non-clinical data) did not involve such a study.

    The questions you asked are highly relevant for AI/ML-enabled medical devices or devices where performance needs to be validated against complex human assessment or outcomes, but they do not apply to this specific medical needle as per the provided FDA documentation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221907

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-29

    (29 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RhinAer® Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the destruction of soft tissue in the nasal airway, including in posterior nasal nerve regions in patients with chronic rhinitis.

    Device Description

    The RhinAer Stylus is functionally unchanged from the predicate in design and intended use to generate and deliver bipolar RF energy to treat tissue in ENT procedures. The design of the Stylus shaft has been changed to improve physician visualization and access during the procedure. The Stylus shaft diameter has been reduced and is more malleable. The treatment tip shape has been re-designed to change its position relative to the shaft to facilitate tissue apposition. The subject and predicate devices have the same material composition. The device and packaging system has been tested and validated with shelf-life studies out to 6 months. The optional custom treatment parameters (power, temperature, time) have been removed.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called the RhinAer® Stylus (FG1393). This summary focuses on demonstrating substantial equivalence to a predicate device (RHIN1 Stylus, K192471) rather than presenting a study where specific acceptance criteria are defined for a performance outcome and then proven.

    In the context of this 510(k) submission, "acceptance criteria" are generally related to demonstrating that the modified device maintains the safety and effectiveness of its predicate, and that changes to the device (shaft diameter, malleability, treatment tip design) do not introduce new safety or effectiveness concerns. The "study" here refers to the non-clinical tests performed to support this claim.

    Given the information, here's how we can address your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria for a "device performance" in the sense of a clinical outcome, but rather for engineering and safety aspects to demonstrate substantial equivalence.

    Acceptance Criteria CategorySpecific Criteria (Implicit or Explicit in Document)Reported Device Performance
    Functional PerformanceAdequate stylus strength (force load testing)Met all performance testing requirements.
    Thermocouple accuracyMet all performance testing requirements.
    Thermocouple response timeMet all performance testing requirements.
    UsabilityNo new use errors causing serious harmAll study objectives met; no new use errors causing serious harm identified.
    No further improvement to device/user interface necessaryAll study objectives met; device safe to use.
    BiocompatibilityEquivalent to predicate (due to similar material composition)Predicate device's testing applicable.
    Electrical SafetyEquivalent to predicate (due to similar design and intended use)Predicate device's testing applicable.
    Sterility ValidationEquivalent to predicate (due to similar design and intended use)Predicate device's testing applicable.
    Shelf-lifeMaintaining integrity and functionality for 6 monthsPassed all aging and transit testing acceptance criteria for 6-month shelf life.
    Material CompositionSame as predicate deviceSubject and predicate devices have the same material composition.
    RF Generator CompatibilitySame as predicate (Aerin Console, Model FG226)Same
    Energy TypeSame as predicate (Bipolar radiofrequency)Same
    Tissue TemperatureConsistent with predicate's operating range (50-70°C)60°C (within predicate's range)
    PowerConsistent with predicate's operating range (3-5W)4 W (within predicate's range)

    2. Sample Size Used for the Test Set and Data Provenance

    The document only references "non-clinical tests" and "bench functional testing" and "usability testing." It does not provide specific sample sizes for these tests beyond implying a sufficient number were performed to meet test objectives.

    • Sample Size (Test Set): Not specified beyond general statements about "bench functional testing," "usability testing," and "aging and transit testing."
    • Data Provenance: Not explicitly stated, but as these are non-clinical (bench and usability) tests conducted for a medical device submission, they would typically be performed in a controlled laboratory or simulated clinical environment. There is no indication of country of origin of patient data, as no clinical studies were performed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This section is not applicable (N/A) because the reported "studies" are non-clinical engineering and usability tests, not studies requiring expert-adjudicated ground truth for a clinical outcome (e.g., image interpretation or disease diagnosis).
    • Usability testing likely involved healthcare professionals or trained users, but their role was to evaluate the device interface and identify potential use errors, not to establish a clinical "ground truth."

    4. Adjudication Method for the Test Set

    • N/A. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in clinical studies involving interpretation (e.g., radiology reads). The non-clinical tests described (functional, usability, shelf-life) do not involve such adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    • N/A. This device is an electrosurgical cutting and coagulation device, not an AI or imaging diagnostic device. Therefore, an MRMC study related to AI assistance is not relevant to this submission.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • N/A. This device is a physical medical instrument, not an algorithm, so the concept of standalone algorithm performance does not apply.

    7. The Type of Ground Truth Used

    • N/A for clinical ground truth. For the non-clinical tests, the "ground truth" was established by engineering specifications, safety standards, and validated test methods (e.g., force load measurements, thermocouple calibration, sterility validations, package integrity standards).

    8. The Sample Size for the Training Set

    • N/A. Training sets are relevant for machine learning algorithms. This device is a physical medical instrument and does not involve a training set as understood in AI/ML contexts.

    9. How the Ground Truth for the Training Set Was Established

    • N/A. This question is not applicable for the reasons mentioned in point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192471

    Validate with FDA (Live)

    Device Name
    RHIN1 Stylus
    Manufacturer
    Date Cleared
    2019-12-20

    (101 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RHIN I Stylus is indicated for use in otothing of the destruction of soft tissue in the nasal airway, including in posterior nasal nerve regions in patients with chronic rhinitis.

    Device Description

    The RHIN1 Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft, and treatment tip. The treatment tip consists of an array of bipolar electric es and a temperature sensor that allows for monitoring of tissue temperature during RF energy delivery. The shaft is malleable to allow the user to bend the shaft in order to more easily access the desired treatment area(s). The Stylus is designed for use with the Aerin Console. It includes features to allow compatibility with and authentication by the Aerin Console. The Stylus connects to the Aerin Console via a flexible cable.

    The RHIN1 Stylus is used to treat patients experiencing chronic rhinitis. During a treatment procedure, the clinician inserts the shaft of the RHIN1 Stylus into a patient's nostril to position the treatment tip and deliver low power RF energy to the target tissue of the nasal airway. The low-power radiofrequency energy generates heat within the tissue, causing destruction of the unwanted soft tissue. The shaft of the RHIN1 Stylus is then removed from the nostril.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the RHIN1 Stylus, a radiofrequency probe for treating chronic rhinitis. While the document details the device's technical characteristics, regulatory classification, and a summary of non-clinical and clinical tests, it does not include information about an AI/algorithm-based device. Instead, the study outlined is a clinical trial assessing the effectiveness and safety of the RHIN1 Stylus itself in human subjects for a medical procedure.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets those criteria for an AI/algorithm-based device cannot be extracted from this document. The document pertains to a physical medical device and its performance in a clinical setting.

    To answer your request, I would need a document describing an AI/algorithm-based medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1