Found 13 results

510(k) Data Aggregation

    K Number
    K030258

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-02-25

    (32 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FIRST RESPONSE® 1-Step Pregnancy Test is an in vitro diagnostic test device intended for the early detection of pregnancy by the lay user up to four (4) days prior to the expected menses

    Device Description

    The device detects the presence of hCG in the urine of a pregnant woman by a series of immunochemical reactions via component reagents striped onto a chromatographic strip contained within plastic housing.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the FIRST RESPONSE® Pregnancy Test, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state quantitative acceptance criteria in the typical format of a performance study (e.g., "sensitivity must be > X%"). Instead, the acceptance criterion for this 510(k) essentially revolved around demonstrating that the device could reliably detect hCG earlier than its predicate while maintaining comparable performance.

    The reported device performance is described in terms of its ability to detect hCG at various days before the expected period. While specific percentages are mentioned, they are presented as a revision/reinterpretation of existing data rather than new study outcomes against defined criteria.

    Acceptance Criteria (Implicit)Reported Device Performance
    Ability to detect hCG as early as four (4) days before the expected menses.The device demonstrated the ability to detect hCG 4 days, 2 days, and 1 day before the expected period. (Specific percentages are not provided in this summary but are referenced as revised labeling based on reinterpreted data from K992232).
    Performance comparable to predicate devices (FIRST RESPONSE® Pregnancy Test - K992232 and Clearblue Easy® Early Result - K013372).The subject device and the predicate version of FIRST RESPONSE® are identical except for the number of early detection days. It is stated to be "very similar" to Clearblue Easy®, with "similar performance (per analytical and consumer study data previously provided)" in K992232. The reinterpreted data for the current submission also includes percentages for 3 days before and 2 days before the expected period (which differ due to a change in the basis of calculation, not underlying data changes).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document does not explicitly state a sample size for a new test set in this 510(k) submission. It mentions that the sponsor "reinterpreted the same data provided in K992232." Therefore, the test set data originates from the studies performed for the predicate device, K992232.
    • Data Provenance: The data is retrospective as it was previously acquired for K992232 and re-analyzed for the current submission. The country of origin is not specified in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Number of Experts: This information is not provided in the document.
    • Qualifications of Experts: This information is not provided in the document. Given the nature of a pregnancy test (detecting hCG in urine), ground truth would typically be established by confirmed pregnancy outcomes or quantitative laboratory hCG levels, rather than interpretation by clinical experts in the traditional sense.

    4. Adjudication Method for the Test Set:

    • The document does not specify an adjudication method. For a qualitative pregnancy test, ground truth is usually established by objective criteria (e.g., quantitative hCG levels, subsequent pregnancy confirmation), rather than subjective expert adjudication of the test result itself.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done as described in this document. The device is an in vitro diagnostic test, not an imaging or interpretive device that typically involves human readers in the same way clinical algorithms do. The "consumer study data" mentioned from the predicate (K992232) might refer to usability or non-expert interpretation, but not a formal MRMC study comparing human readers with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

    • Yes, this submission describes standalone performance. The device itself is a standalone, over-the-counter diagnostic test that provides a result without requiring human interpretation beyond reading the result line(s). The performance data cited relates directly to the device's ability to detect hCG. There is no "human-in-the-loop" aspect to its primary function of detection.

    7. Type of Ground Truth Used:

    • The ground truth used was based on quantitative hCG levels and likely confirmed pregnancy outcomes (as implied by "percentage of women in which the device detected the hormone X days before... expected period" and "conceptive cycles"). The predicate test's claims were based on "the percentage of cycles having quantitative hCG levels approximating the claimed sensitivity level or greater."

    8. Sample Size for the Training Set:

    • The document does not provide a specific sample size for a training set. Since this submission reinterpreted data from a prior 510(k) (K992232), any "training" or development data would have originated from that earlier submission, and its details are not present here.

    9. How the Ground Truth for the Training Set Was Established:

    • Similar to the test set, the ground truth for any underlying development/training from the K992232 submission would have been established by quantitative hCG levels and/or confirmed pregnancy outcomes. The text specifically mentions "hCG levels approximating the claimed sensitivity level or greater."
    Ask a Question

    Ask a specific question about this device

    K Number
    K023405

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-12-16

    (67 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TROJAN® MINT TINGLE™ Latex Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted diseases, STDs).

    Device Description

    The TROJAN® MINT TINGLE™ Brand Latex Condom is a male condom consisting of a sheath of natural rubber latex with a colored pigment and a flavored water-based lubricant. The condom is a straight-walled, nipple-end condom with a nominal length of 180 mm and an approximate width of 52 mm.

    AI/ML Overview

    This document is a 510(k) summary for a TROJAN® MINT TINGLE™ Brand Latex Condom, seeking clearance from the FDA. It does not include acceptance criteria or the study details you've requested. The 510(k) process is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving a device meets specific performance criteria through new studies in the same way a PMA (Premarket Approval) might require.

    Therefore, I cannot provide the information you asked for based on the provided text, as it is not present in this 510(k) summary. I can only confirm that this document pertains to a condom and its intended uses for contraception and STD prevention, and that it was deemed substantially equivalent to a predicate device.

    To give you the information you're looking for, the document would need to include details about:

    1. Specific performance tests conducted (e.g., burst strength, leak test, freedom from holes, dimensions).
    2. Quantitative acceptance criteria for these tests (e.g., "99.9% freedom from holes").
    3. The actual results from testing the TROJAN® MINT TINGLE™ condom against these criteria.
    4. Sample sizes used for these performance tests.
    5. Ground truth establishment, expert qualifications, adjudication methods, and details about MRMC or standalone studies are generally not relevant for 510(k) submissions of devices like condoms, as they typically don't involve algorithms or image analysis. Their performance is assessed through physical and mechanical property testing and adherence to standards like ASTM D-3492-97.
    Ask a Question

    Ask a specific question about this device

    K Number
    K001212

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-05-09

    (25 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K000748

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-04-06

    (29 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K992232

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-12-21

    (172 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The First Response® 1-Step Pregnancy Test is an at-home pregnancy test marketed over-the-counter (OTC) to lay consumers. The First Response® I-Step Pregnancy Test detects the presence of hCG in the urine as an aid in the carly diagnosis of pregnancy. The test, which can be used anytime of the day, can detect hCG as early as three days before the expected period.

    Device Description

    The First Response®1-Step Pregnancy Test is a Class II in vitro diagnostic medical device product that functions by way of a double antibody immunochromatographic assay in detecting the presence of hCG in the urine as an aid in the early diagnosis of pregnancy. It consists of a plastic stick, which contains an absorbent tip that protrudes from the end of the device and collects and delivers urine to reagents on a chromatographic strip contained within the device. The test is performed by placing the absorbent tip of the device in the urine stream for 5 seconds or by immersing the absorbent tip into a container of urine for 5 seconds. The absorbent section of the strip allows the urine sample to move chromatographically along the reagent strip reconstituting the diffusible reagents placed strategically along the strip and delivering them to the appropriate capture zones for visualization of the test results. The detection of hCG (pregnant) in the urine sample is indicated by the appearance of two pink lines in the test window. If there is no hCG (not-pregnant) in the urine, one pink line will appear.

    AI/ML Overview

    The acceptance criteria for the First Response® 1-Step Pregnancy Test and the studies proving it meets these criteria are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Consumer accuracy in reading low levels of hCGEvaluated in a performance study
    Ability of lay consumers to perform the test and interpret the resultEvaluated in a performance study
    Ability to understand the package insert, including use before the expected periodEvaluated in a quantitative market research study
    Clinical sensitivity using samples collected prior to missed mensesEvaluated and confirmed in a laboratory study

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not explicitly stated in the provided text for individual studies.
    • Data Provenance: Not explicitly stated, but the studies were conducted by Carter-Wallace, Inc., in the United States.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable as the ground truth was established through laboratory testing for clinical sensitivity and consumer studies for user performance and comprehension, rather than expert consensus on diagnostic images or conditions.

    4. Adjudication Method

    • Not applicable for the types of studies described. Clinical sensitivity would rely on laboratory reference methods, and consumer studies would involve direct observation and survey results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned. The studies focused on the device's standalone performance and consumer interaction rather than comparing human readers with and without AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, standalone performance studies were done. The text describes a laboratory study to evaluate and confirm the clinical sensitivity of the First Response® 1-Step Pregnancy Test using samples collected prior to missed menses. This assesses the device's ability to detect hCG independently. Additionally, consumer accuracy in reading low levels of hCG suggests a standalone assessment of the device's output interpretation by users.

    7. Type of Ground Truth Used

    • Clinical Sensitivity: Likely laboratory reference methods for hCG detection (e.g., quantitative serum hCG assays) or established clinical diagnosis of pregnancy.
    • Consumer Accuracy/Performance/Comprehension: Actual presence or absence of hCG in urine samples (for accuracy) and direct observation or survey responses regarding user actions and understanding.

    8. Sample Size for the Training Set

    • Not applicable. This device is an in-vitro diagnostic test, not an AI or machine learning algorithm that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K973352

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-10-27

    (52 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brand Name 1-Step Pregnancy Test is an at-home pregnancy test marketed over-the-counter (OTC) to lay consumers. The Brand Name 1-Step Pregnancy Test detects the presence of hCG in the urine as an aid in the early diagnosis of pregnancy. The test, which can be used anytime of the day, can detect hCG as early as the first day of the missed menses.

    Device Description

    The Brand Name 1-Step Pregnancy Test is a Class II in vitro diagnostic medical device product that functions by way of a double antibody immunochromatographic assay in detecting the presence of hCG in the urine as an aid in the early diagnosis of pregnancy. It consists of a plastic stick, which contains an absorbent tip that protrudes from the end of the device and collects and delivers urine to reagents on a chromatographic strip contained within the device. The test is performed by placing the absorbent tip of the device in the urine stream for 5 seconds or by immersing the absorbent tip into a container of urine for 5 seconds. The absorbent section of the strip allows the urine sample to move chromatographically along the reagent strip reconstituting the diffusible reagents placed strategically along the strip and delivering them to the appropriate capture zones for visualization of the test results. The detection of hCG (pregnant) in the urine sample is indicated by the appearance of two pink lines in the test window. If there is no hCG (not-pregnant) in the urine, one pink line will appear.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and supporting studies for the "Brand Name 1-Step Pregnancy Test":

    1. Acceptance Criteria and Reported Device Performance

    The provided document details various performance studies but does not explicitly state numerical acceptance criteria for sensitivity, specificity, or agreement. Instead, it describes general objectives and outcomes. However, it implicitly demonstrates that the device met the criteria for substantial equivalence to the predicate device.

    Acceptance Criteria (Implicit)Reported Device Performance
    Laboratory AccuracyTwo laboratory studies compared the Brand Name 1-Step Pregnancy Test to the predicate device, First Response® 1-Step Pregnancy Test, to determine its performance.
    SensitivityA third laboratory study was performed to evaluate and confirm the sensitivity of the Brand Name 1-Step Pregnancy Test.
    Lay Consumer PerformanceThree separate studies evaluated the ability of consumers to perform and interpret the Brand Name 1-Step Pregnancy Test using both midstream and dip methods.
    Substantial EquivalenceThe FDA determined the device is substantially equivalent to legally marketed predicate devices, implying that its performance aligns with existing standards for such devices.
    Intended Use FulfillmentThe device successfully detects hCG in urine as an aid in the early diagnosis of pregnancy and can be used by lay consumers anytime of the day, as early as the first day of missed menses.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes for the test sets in any of the described studies (laboratory accuracy, sensitivity, or consumer use).

    Regarding data provenance, the studies are described as "laboratory studies" and "studies" evaluating consumer ability. Given the context of a 510(k) submission for a new product, these would almost certainly be prospective studies conducted specifically for this submission. The document does not mention the country of origin of the data, but
    given the submitter is Carter-Wallace in Cranbury, New Jersey, it is highly probable the
    studies were conducted in the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not specify the number of experts used or their qualifications for establishing ground truth in the laboratory studies. For a hCG test, the ground truth for positive/negative results would typically be established by:

    • Quantitative laboratory assays: Using a highly accurate reference method to measure hCG levels.
    • Clinical assessment: In some cases, confirming pregnancy status through other clinical means (though less common for in vitro diagnostic ground truth).

    For the consumer studies, the "ground truth" for the interpretation ability would be derived from comparison to the device's actual performance (i.e., whether the consumer's interpretation matched the correct result as determined by the laboratory).

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method (e.g., 2+1, 3+1, none) for the test sets. For IVD devices like pregnancy tests, adjudication among human readers or experts is less common for establishing the analytical ground truth (presence or absence of hCG) as the results are typically objectively measured by quantitative methods. However, for the consumer interpretation studies, it's possible internal protocols were used to resolve discrepancies if consumers disagreed on the interpretation of a test result, but this is not detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document describes laboratory studies comparing the new device to a predicate device, and consumer studies to assess user interpretation. These are not MRMC studies designed to assess the improvement of human readers with AI assistance, as this is a standalone diagnostic test, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a form of standalone performance was assessed. The "Brand Name 1-Step Pregnancy Test" is an in vitro diagnostic device, not an algorithm. Its "standalone" performance refers to its ability to detect hCG accurately on its own. The laboratory studies ("evaluation of the laboratory accuracy" and "evaluate and confirm the sensitivity") directly assess this standalone performance – how well the device itself, without human interpretation of ambiguous signals, determines the presence or absence of hCG. The consumer studies then assess the "human-in-the-loop" aspect: whether lay users can correctly interpret the device's output.

    7. The Type of Ground Truth Used

    The ground truth for the analytical performance studies (laboratory accuracy and sensitivity) would be based on quantitative biochemical assays for human chorionic gonadotropin (hCG) levels in urine samples. This would be the definitive measure of hCG concentration, allowing for the determination of true positive and true negative samples.

    For the consumer interpretation studies, the ground truth for correct interpretation would be established by comparing the consumer's visual reading of the test lines against the known analytical result (positive or negative for hCG) of the sample used.

    8. The Sample Size for the Training Set

    The concept of a "training set" is typically associated with machine learning or artificial intelligence models. As the "Brand Name 1-Step Pregnancy Test" is an immunochromatographic device and not an AI algorithm, there is no training set in the conventional sense. The device's characteristics are inherent to its design and chemical reagents, not learned from data.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set for this type of device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973310

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-10-27

    (54 days)

    Product Code
    Regulation Number
    862.1485
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TBD One-Step Ovulation Predictor Test is an at-home in vitro diagnostic use test to be marketed over-the-counter (OTC) to lay consumers. The test device product is intended for the detection of human luteinizing hormone (hLH) in urine as an aid in the prediction of ovulation.

    Device Description

    The TBD One-Step Ovulation Predictor Test is a simple to use, plastic stick, in vitro device. The top of the stick is clear to allow the user to visualize the migration of the test reagents. The stick contains a chromatographic test strip which has all the necessary reagents to perform the test. An absorbent tip collects and delivers the urine sample to the reagents on the chromatographic strip. The test is performed by placing the absorbent tip into the urine stream for 5 seconds. Alternatively, the test may be performed by immersing the absorbent tip into a container of urine for 5 seconds. The urine sample from the absorbent tip migrates by capillary action along the chromatographic strip, reconstituting the reagents placed strategically along the strip. Visualization of the test results occurs when the appropriate reagents react producing one or two pink-purple lines which appear in a window in the hull of the test stick. The luteinizing hormone (LH) surge is indicated if the test line is similar to or darker than the reference line. If there is no test line or the test line is lighter than the reference line, then there has been no LH surge.

    AI/ML Overview

    The provided text describes the TBD One-Step Ovulation Predictor Test and its similarity to a predicate device. However, it does not include detailed acceptance criteria or the specific study that would directly prove the device meets these criteria in the format requested.

    Here's an analysis based on the information available in the text, highlighting what is missing for a complete answer to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that the TBD One-Step Ovulation Predictor Test was found to have "equivalent performance to that of the predicate device" in a direct laboratory comparison, and in consumer use studies, it was "at least as accurate as the predicate device with regard to consumers' accuracy in performing the test."

    Missing Information:

    • Specific quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy thresholds).
    • Detailed quantitative performance metrics for the TBD One-Step Ovulation Predictor Test.
    • Quantitative performance metrics for the predicate device to compare against.

    2. Sample Size for the Test Set and Data Provenance:

    The document mentions "direct laboratory comparison" and "consumer use studies."

    Missing Information:

    • Sample Size: The number of participants or samples used in these studies is not specified.
    • Data Provenance: The country of origin of the data is not mentioned. It is implied that the studies were conducted by Carter-Wallace, Inc., an American company. The description refers to "consumer use studies," which suggests prospective data collection, but this is not explicitly stated. It's unclear if any retrospective data was used.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    The document does not mention the involvement of experts in establishing ground truth for the performance studies. Since it's an at-home ovulation predictor test intended for lay consumers, the "accuracy in performing the test" likely refers to the consumers' ability to correctly interpret the results based on the package inserts, rather than expert interpretation of complex clinical data.

    Missing Information:

    • Details on expert involvement in establishing ground truth are entirely absent.

    4. Adjudication Method:

    Given the nature of an at-home ovulation test, the concept of an adjudication method as typically applied to expert review of medical images or complex diagnoses is not directly applicable or described.

    Missing Information:

    • No adjudication method is mentioned or implied.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC study is not relevant to this type of device (an at-home diagnostic test for a single user).

    Missing Information:

    • An MRMC study was not conducted or described.

    6. Standalone Performance:

    The "direct laboratory comparison" and "consumer use studies" would likely represent the standalone performance of the device as it would be used by consumers, without human-in-the-loop assistance beyond the consumer following instructions.

    Missing Information:

    • While implied, the term "standalone performance" isn't explicitly used, and specific metrics for this are absent (as noted in point 1).

    7. Type of Ground Truth Used:

    The ground truth for an ovulation predictor test would typically involve confirming actual ovulation, likely through other clinical methods (e.g., daily blood hormone levels, ultrasound). However, the document focuses on the prediction of the LH surge and refers to "consumers' accuracy in performing the test." This suggests the ground truth for consumer studies might be tied to how well users followed instructions and achieved a correct reading. For the "direct laboratory comparison," the ground truth for LH levels would be laboratory-measured LH concentrations.

    Missing Information:

    • The exact method for establishing ground truth (e.g., comparison against a gold standard lab assay, confirmation of actual ovulation through other means) for the performance studies is not explicitly detailed.

    8. Sample Size for the Training Set:

    The concept of a "training set" applies to machine learning algorithms. This device is a biochemical in-vitro diagnostic test, not an AI or machine learning product. Therefore, there wouldn't be a training set in that context.

    Missing Information:

    • Not applicable as this is not an AI/ML device.

    9. How Ground Truth for the Training Set Was Established:

    As above, this is not an AI/ML device, so a training set and its associated ground truth establishment are not applicable.

    Missing Information:

    • Not applicable as this is not an AI/ML device.

    In summary, the provided text serves as a 510(k) summary demonstrating substantial equivalence to a predicate device, focusing on technological characteristics and overall performance claims ("equivalent performance," "at least as accurate"). It lacks the granular detail about specific study designs, acceptance criteria, and quantitative results that would be expected in a comprehensive study report to fulfill all aspects of your requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K963318

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-03-03

    (192 days)

    Product Code
    Regulation Number
    866.3300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    for the qualitative detection of total antibodies (IgG and IgM) 10 Legionella pneumophila serogroups 1-6 in serum from patients with clinical suspicion of Legionella Disease.

    Device Description

    The Legionella pneumophila IgG/IgM ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of total antibodies (IgG and IgM) 10 Legionella pneumophila serogroups 1-6 in serum from patients with clinical suspicion of Legionella Disease.

    The Legionella pneumophila IgG/IgM ELISA test is an enzyme linked immunosorbent assay to detect IgG/IgM antibodies to Legionella. Purified Legionella pneumopilla antigen (serogroups 1, 2, 3, 4, 5, 6) is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG/IgM is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Legionella pneumophila IgG/IgM ELISA Test Kit, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state numerical "acceptance criteria" in the typical sense (e.g., a pre-defined threshold that must be met for approval). Instead, it presents "Performance Characteristics" and compares the device's performance to a predicate device (BioWhitttaker's Legionella STAT test) and a reference method (IFA). The reported performance values are those observed in these comparative studies.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    Performance MetricImplied Acceptance Criterion (from predicate comparison)Reported Device Performance
    Relative Sensitivity (Vs IFA)Comparable to predicate or reference method90.00% (95% CI: 79.0% - 100%)
    Relative Specificity (Vs IFA)Comparable to predicate or reference method98.53% (95% CI: 95.6% - 100%)
    Relative Agreement (Vs IFA)Comparable to predicate or reference method98.57% (95% CI: 95.7% - 100%)
    Seroconversion Sensitivity (CDC Panel)Demonstrate ability to detect seroconversion93.5% (29/31 seroconversions detected)
    Precision (Intra-assay CV)Not explicitly stated, but generally <15-20%Ranges from 2.72% to 15.5% (Site 1) and 5.00% to 11.4% (Site 2)
    Precision (Inter-assay CV)Not explicitly stated, but generally <15-20%Ranges from 7.50% to 15.9% (Site 1) and 7.14% to 9.54% (Site 2)
    Precision (Inter-site CV)Not explicitly stated, but generally <20%Ranges from 7.50% to 19.2%

    Note: The phrase "substantially equivalent to BioWhitttaker's Legionella STAT test" implies that the acceptance criteria are met if the performance characteristics are similar or non-inferior to the predicate device. The document does not provide specific numerical criteria for the predicate device.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Test Set 1 (Commercial R&D Lab - Maryland):
      • Sample Size: 33 single IFA positive sera (from an outbreak and routine submissions).
      • Data Provenance: United States (Maryland), Retrospective (samples routinely submitted or from an outbreak)
    • Test Set 2 (Clinical Laboratory - Pennsylvania):
      • Sample Size: 72 prospective serum samples for Legionella testing.
      • Data Provenance: United States (Pennsylvania), Prospective
    • CDC Panel (Seroconversion Study):
      • Sample Size: 31 serum pairs (total 62 individual serum samples).
      • Data Provenance: United States (CDC-confirmed panel), Retrospective (samples submitted to CDC for titer confirmation)
    • Precision Studies (Various Sera):
      • Sample Size:
        • 7 different sera at two sites, tested 10 times each on 3 different days (7 sera x 2 sites x 30 tests/site = 420 tests).
        • An additional 3 sera at Site 1, tested 10 times each on 3 different days (3 sera x 30 tests = 90 tests).
        • Control samples (HPC, CAL, LPC, NC) also included.
      • Data Provenance: Unspecified, likely United States (commercial R&D lab in Maryland and clinical lab in Pennsylvania).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the primary comparative studies was established using Legionella IFA (Immunofluorescence Assay) results. This is a laboratory-based assay, not typically requiring human expert interpretation in the same way as imaging studies.

    • The IFA results for the first test set were "Legionella positive sera from an outbreak and samples routinely submitted for Legionella testing."
    • The IFA results for the second test set were "prospective serum for Legionella testing."
    • For the CDC panel, the IFA titers were "confirmed to be serologically positive for an increase in titer from < 1:256 to >1:256."

    The document does not explicitly state the number of human experts involved in interpreting the IFA results or their specific qualifications (e.g., medical technologists, clinical pathologists). However, IFA is a standardized laboratory test, and its results are generally considered objective.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method in the context of human readers interpreting data. The comparisons are made against established laboratory test results (IFA). Equivocal results from the ELISA device were "not included in the above calculations" for sensitivity and specificity, which acts as a form of handling ambiguous results rather than human adjudication of ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This study focuses on the performance of a diagnostic kit (ELISA) compared to another diagnostic kit (IFA), not on human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are standalone performance evaluations of the Legionella pneumophila IgG/IgM ELISA Test Kit. The results (Relative Sensitivity, Specificity, Agreement, Seroconversion Sensitivity, and Precision) reflect the performance of the assay itself, without any human-in-the-loop interaction influencing the calculation of these metrics.

    7. The Type of Ground Truth Used

    The primary ground truth used for performance evaluation was:

    • Legionella IFA (Immunofluorescence Assay) results: This is a laboratory-based method for detecting Legionella antibodies and is used as the reference standard in the comparative studies.
    • CDC-confirmed IFNA titer changes: For the seroconversion study, the ground truth was a confirmed greater than 4-fold increase in IFA titer from the CDC.

    The document explicitly states: "There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease." This means the ground truth is based on a serological reference method, not clinical disease outcomes (e.g., pathology confirmation or patient follow-up).

    8. The Sample Size for the Training Set

    The document does not mention a training set or machine learning/AI development. This is a traditional in vitro diagnostic (IVD) device, and its performance is evaluated through clinical laboratory studies comparing it to established methods, rather than through a machine learning development pipeline involving training and test sets in the AI sense.

    9. How the Ground Truth for the Training Set was Established

    As there is no mention of a training set for an AI/ML algorithm, this question is not applicable to the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K962558

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-02-24

    (238 days)

    Product Code
    Regulation Number
    866.3120
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chlamydia IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative detection of IgG antibodies in human serum to Chlamydia for the determination of immunological experience.

    Device Description

    The Chlamydia IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Chlamydia. Purified Chlamydia antigen (strain LGV II) is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided text, formatted as requested:

    Acceptance Criteria and Device Performance for Chlamydia IgG ELISA Test Kit

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Chlamydia IgG ELISA)
    Relative Sensitivity92.1% (95% CI: 86.3% - 97.8%)
    Relative Specificity98.0% (95% CI: 96.2% - 99.8%)
    Relative Agreement96.5% (95% CI: 94.5% - 98.5%)
    Precision (CV)Intra-assay CV: typically <15% (e.g., 2.57% to 19.95% in Study 1, 1.69% to 53.59% in Study 2). Inter-assay CV: 3.41% to 17.75%. Inter-site CV: 4.63% to 78.91%
    Reproducibility (% Agreement between sites)97.1% (excluding equivocals)
    Seroconversion Detection (Sensitivity vs. CF titer)44% (4/9) for demonstrating seroconversion

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 355 sera were used for the relative sensitivity and specificity analysis.
    • Data Provenance: The sera were from "normal individuals of various ages, gender, and geographical areas." The study was conducted at "R&D laboratories at commercial companies located in Maryland and New York and affiliated with the manufacturing of the kit." This indicates a prospective and likely geographically diverse dataset, but specifically not from patients with documented Chlamydia infections.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for the relative sensitivity and specificity study was established by a predicate device (commercial IFA kit), not by human experts interpreting images or clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. The performance was compared against a predicate device, not against a consensus of human reviewers. Equivocal results from the comparison were excluded from the calculations of sensitivity, specificity, and agreement.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a diagnostic immunoassay kit, not an AI-assisted diagnostic tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies on relative sensitivity, specificity, and precision assess the standalone performance of the Chlamydia IgG ELISA test kit. There is no human interpretation component in the ELISA assay itself once the sample is processed.

    7. The Type of Ground Truth Used

    The ground truth for the relative sensitivity and specificity study was defined by the results of a commercial IFA kit (Immunofluorescence Assay). The document explicitly states: "Please be advised that 'relative' refers to the comparison of this assay's results to that of a similar assay. There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease."
    For the seroconversion study, the ground truth was a "greater than 4 fold increase in CF titer or seroconversions by CF" as determined by a Complement Fixation (CF) assay.

    8. The Sample Size for the Training Set

    Not applicable. This is a non-AI diagnostic kit; there is no "training set" in the context of machine learning. The "training" would be the process of developing and optimizing the assay itself, which is not detailed in terms of a specific sample size.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as explained in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K963645

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-02-04

    (145 days)

    Product Code
    Regulation Number
    866.3305
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Herpes Group IgG ELISA kit is an Enzyme-Linked Immunosorbent Assay (ELISA) for qualitative determination of IgG antibodies in human serum to Herpes simplex virus. The Herpes Group IgG ELISA kit may be used to evaluate paired sera for the presence seroconversions of IgG as an aid in the diagnosis of Herpes simplex virus infection.

    Device Description

    The Herpes Group IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Herpes simplex virus. Purified Herpes Group antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled anti-human IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Relative Sensitivity (compared to Clark HSV 1 & HSV 2 ELISA)Overall: 98.9% (95% CI: 97.9% - 100%) Site 1: 99.1% Site 2: 97.9% Site 3: 99.1% Site 4: 100.0%
    Relative Specificity (compared to Clark HSV 1 & HSV 2 ELISA)Overall: 96.7% (95% CI: 94.2% - 99.1%) Site 1: 94.7% Site 2: 100% Site 3: 94.7% Site 4: 96.2%
    Relative Agreement (compared to Clark HSV 1 & HSV 2 ELISA)Overall: 98.1% (95% CI: 97.0% - 99.2%) Site 1: 97.2% Site 2: 98.6% Site 3: 97.7% Site 4: 98.9%
    Precision (Inter-site, %CV)Generally <15% CV for most sera Serum 1: 9.21% Serum 2: 12.6% Serum 3: 9.08% Serum 4: 13.5% Serum 5: 15.1% Serum 6: 121% Serum 7: 150%
    Seroconversion Sensitivity (compared to CF)100% for serum meeting paired sera criteria (evaluating 11 pairs)
    Cross-Reactivity (with EBV, CMV, VZV IgG antibodies)No cross-reactivity observed (values for Herpes Group IgG were low, while other antibodies were present)
    Agreement with CDC Panel96.9% total agreement (excluding 2 equivocals) 95.7% agreement with positive specimens 100% agreement with negative specimens

    Study Information

    1. Sample sizes used for the test set and the data provenance:

      • Total Sample Size (Overall Comparison Study): 603 sera (excluding equivocals from individual calculations).

      • Individual Site Sample Sizes:

        • Site 1: 187 sera (181 used in calculations)
        • Site 2: 152 sera (145 used in calculations)
        • Site 3: 176 sera (170 used in calculations)
        • Site 4: 88 sera (88 used in calculations)
      • Precision Study: 7 sera, each assayed 10 times at 3 different sites (total 90 assays per serum).

      • CF Paired Serum Study: 20 serum pairs, with 11 evaluable pairs.

      • Cross-Reactivity Study: 9 serum samples containing IgG antibodies to EBV, CMV, and VZV.

      • CDC Panel: Not explicitly stated, but described as containing 72% positive and 28% negative samples.

      • Data Provenance:

        • Site 1: R&D laboratory at a commercial company in Maryland. Frozen sera from normals, ages 12-83, various genders, geographical areas. Retrospective.
        • Site 2: R&D laboratory at a commercial company in New York. Frozen sera from normals, ages 17-59, various genders, geographical areas. Retrospective.
        • Site 3: Clinical laboratory in Pennsylvania. Prospective samples sent for Herpes antibody testing. Prospective.
        • Site 4: Clinical laboratory in Wisconsin. Frozen random normal samples. Retrospective.
        • CF Paired Serum Study: From patients suspected of having acute Herpes simplex infection.
        • Cross-Reactivity Study: Serum samples containing specific antibodies.
        • CDC Panel: Serum panel obtained from the CDC.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The ground truth for the comparison study (relative sensitivity/specificity/agreement) was established by the Clark HSV 1 and HSV 2 ELISA assays (predicate device), not human experts.
      • For the CF Paired Serum Study, the ground truth was Complement Fixation (CF), a laboratory method, not human experts.
      • For the CDC panel, the ground truth was "masked, characterized serum panel" with established positive and negative samples, but the method of characterization (and thus involvement of experts) is not specified.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • There was no stated adjudication method involving human experts for the test set. The comparison was directly against the results of the predicate device (Clark HSV 1 and HSV 2 ELISA).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study evaluates an ELISA test kit, which is an automated diagnostic assay, and not an AI or human-assisted image interpretation tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, this study represents a standalone performance evaluation of the Wampole Herpes Group IgG ELISA kit. This is an "algorithm only" (or assay-only, in this context) device designed for qualitative determination of IgG antibodies. Its performance is measured directly against a predicate device and other established lab methods, without human interpretation as part of the primary diagnostic step.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Comparison Studies: The "ground truth" was established by the predicate device, the Clark HSV 1 and HSV 2 ELISA assays. The report explicitly states: "There was not an attempt to correlate the assay's results with disease presence or absence. No judgment can be made on the comparison assay's accuracy to predict disease." This means the ground truth is relative to another diagnostic test, not a definitive clinical diagnosis or pathology.
      • CF Paired Serum Study: The ground truth was based on Complement Fixation (CF) results for seroconversion.
      • Cross-Reactivity Study: The ground truth was based on known presence/absence of other specific antibodies in the serum.
      • CDC Panel: "Masked, characterized serum panel" with established positive and negative results, though specific methodology for characterization is not detailed.
    7. The sample size for the training set:

      • The document does not provide information on a training set. For an ELISA kit, which is a biochemical assay rather than a machine learning algorithm, the concept of a "training set" in the AI sense is generally not applicable. Batch validation and standardization would be part of manufacturing, but not in the same way as training an AI model.
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set (in the context of AI/machine learning) is described.