Search Results
Found 22 results
510(k) Data Aggregation
(30 days)
Ask a specific question about this device
(7 days)
Ask a specific question about this device
(27 days)
Ask a specific question about this device
(51 days)
Ask a specific question about this device
(16 days)
Ask a specific question about this device
(77 days)
The Original Perry Style 42 Surgeon's Glove intended use is to be worn by operating room personnel to protect a surgical wound from contamination.
The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) meet all of the requirements of ASTM D 3577-99, Type 1. The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) meet all the current specifications for ASTM D 3577-99 Rubber Surgical Gloves. The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.
The provided text describes a 510(k) summary for medical gloves and does not contain information about a device that uses AI or requires a study involving ground truth, expert consensus, or training sets as outlined in the request.
The device in question, "The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim)," is a medical glove. The acceptance criteria and "study" described are based on compliance with existing ASTM (American Society for Testing and Materials) standards for surgical gloves and FDA hole requirements.
Therefore, many of the requested fields cannot be filled as they are not applicable to the type of device and submission described.
Here's a breakdown of the available information based on your request, highlighting what is not applicable:
1. A table of acceptance criteria and the reported device performance
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimensions | Meets ASTM D 3577-99 | Meets ASTM D 3577-99 |
| Physical Properties | Meets ASTM D 3577-99, Type 1 | Meets ASTM D 3577-99, Type 1 |
| Freedom from holes | Meets ASTM D 3577-99 | Meets ASTM D 3577-99 |
| Meets ASTM D 5151-92 | Meets ASTM D 5151-92 | |
| Protein Label Claim | 60 micrograms or less of total water extractable protein per gram. Meets ASTM D 5712-95 Standard Test Method for Analysis of Protein in Natural Rubber and Its Products | This latex glove contains 60 micrograms or less of total water extractable protein per gram. Meets ASTM D 5712-95 Standard Test Method for Analysis of Protein in Natural Rubber and Its Products |
| Primary Skin Irritation | Passes | Passes |
| Guinea Pig Sensitization | Passes | Passes |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not specified in the provided text. Compliance with ASTM standards typically involves specified sample sizes for testing, but these are not explicitly detailed here.
- Data Provenance: Not specified. The testing is described as "performance test data of the non clinical tests," implying laboratory testing rather than data from a specific country or clinical setting. It's prospective testing against a standard.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This device is a physical product (medical glove) and its performance is evaluated against defined physical and chemical standards, not through "ground truth" established by human experts in the context of diagnostic or interpretive tasks.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are relevant for subjective evaluations, typically in clinical studies or expert reviews. The testing described here is objective and based on established ASTM test methods.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is a medical glove, not an AI-powered diagnostic device. No human reader or AI assistance is involved in its evaluation as described.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This is a physical product, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable. The "ground truth" for this device's performance is adherence to defined physical and chemical properties and biocompatibility tests as specified in ASTM standards.
8. The sample size for the training set
- Not applicable. This device is not an AI algorithm; there is no training set.
9. How the ground truth for the training set was established
- Not applicable. This device is not an AI algorithm; there is no training set or associated ground truth establishment process.
In summary: The provided document describes a medical glove's compliance with established industry standards through non-clinical performance testing. It does not involve AI, human expert evaluation for "ground truth," or clinical studies of the type typically associated with AI-powered diagnostic devices.
Ask a specific question about this device
(58 days)
A device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Encore Orthopaedic Powder Free Surgical Gloves (Protein Labeling Claim), meet all of the requirements of ASTM D 3577, Type 1. Encore Orthopaedic Powder Free Surgical Gloves (Protein Labeling Claim) meet all the current specifications for ASTM D 3577 Rubber Surgical Gloves. Encore Orthopaedic Powder Free Surgical Gloves (Protein Labeling Claim) are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.
The provided document is a 510(k) premarket notification for "Encore Orthopaedic Powder Free Surgical Gloves (Protein Labeling Claim)." The document outlines the device's characteristics and its compliance with established standards, but it does not describe a study in the traditional sense of a clinical trial or a formal performance study with a test set, ground truth experts, or multi-reader multi-case analysis.
Instead, the "study" demonstrating the device meets acceptance criteria is primarily an assertion of compliance with existing ASTM standards and FDA requirements for surgical gloves.
Here's a breakdown based on the provided input:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimensions | Meets ASTM D 3577 | Meets ASTM D 3577 |
| Physical Properties | Meets ASTM D 3577, Type 1 | Meets ASTM D 3577, Type 1 |
| Freedom from holes | Meets ASTM D 3577 & ASTM D 5151 | Meets ASTM D 3577 & ASTM D 5151 |
| Powder-Free | Meets described test in Attachment VI | Not more than 2 mg residue by mass |
| Protein Label Claim | Meets ASTM D 6124 | 50 micrograms or less of total water extractable protein per gram |
| Biocompatibility: Primary Skin Irritation in Rabbits | Passes | Passes |
| Biocompatibility: Guinea Pig Sensitization | Passes | Passes |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated as a separate "test set" for a research study. The performance is based on the inherent manufacturing and testing procedures required for compliance with ASTM standards. These standards typically involve sampling from production lots. The document implies that samples from the manufactured gloves were tested to ensure compliance.
- Data Provenance: Not explicitly stated. Likely refers to internal quality control and testing conducted by Ansell Perry during the manufacturing process. Retrospective, in that the samples were already produced, and data collected on them.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. The "ground truth" here is the established ASTM standards and FDA regulations for surgical gloves. Compliance is determined by direct measurement and testing against these predefined criteria, not by expert consensus on individual cases.
4. Adjudication method for the test set
- Not Applicable. As there's no "test set" requiring expert interpretation or consensus, no adjudication method was used. Performance is based on objective measurements against standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a medical device (surgical gloves), not an AI-powered diagnostic or assistive tool. No MRMC study was performed, and there is no AI component.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a medical device (surgical gloves). There is no "algorithm only" performance to evaluate. The device itself is the physical product.
7. The type of ground truth used
- Established Standards and Regulations: The ground truth used is objective; it's the specific numerical and qualitative requirements outlined in:
- ASTM D 3577 (Standard Specification for Rubber Surgical Gloves)
- ASTM D 5151 (Standard Test Method for Detection of Holes in Medical Gloves)
- ASTM D 6124 (Standard Test Method for Residual Powder on Medical Gloves)
- FDA hole requirements
- Specific thresholds for protein labeling claim (50 micrograms or less)
- Biocompatibility testing protocols (Primary Skin Irritation in Rabbits, Guinea Pig Sensitization).
8. The sample size for the training set
- Not Applicable. There is no "training set" in the context of this device. This is a manufactured product subject to quality control and compliance testing, not a machine learning model that requires training data.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set, there is no ground truth established for one.
Ask a specific question about this device
(124 days)
A sterile disposable device intended for medical purpose that is worn on the examiners hand to prevent contamination between patient and examiner.
SYNSATION® Sterile Powder Free, Polyvinyl Chloride Medical Examination Gloves examination gloves are sterile disposable device intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
The SYNSATION® Sterile Powder Free, Polyvinyl Chloride Medical Examination Gloves undergo a performance study to demonstrate that they meet the acceptance criteria. The study ensures that the gloves conform to the specified standards and are safe and effective for their intended medical purposes.
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimensions | Meets ASTM D 5250 | Meets ASTM D 5250 |
| Physical Properties | Meets ASTM D 5250 | Meets ASTM D 5250 |
| Freedom from holes | Meets ASTM D 5250 | Meets ASTM D 5250 |
| Meets ASTM D 5151 | Meets ASTM D 5151 | |
| Powder-Free | Not more than 2 mg residue by mass | Not more than 2 mg residue by mass |
| Meets ASTM D 6124 and ASTM D 5250 | Meets ASTM D 6124 and ASTM D 5250 | |
| Biocompatibility | ||
| Primary Skin Irritation in Rabbits | Passes | Passes |
| Guinea Pig Sensitization | Passes | Passes |
2. Sample size used for the test set and the data provenance:
The document does not explicitly state the specific sample size used for each test. However, the performance is reported against ASTM standards, which would imply that testing was conducted according to the sampling plans outlined within those standards. The data provenance is not specified in terms of country of origin, but the testing is non-clinical. The studies are assumed to be prospective, specifically designed to test the glove's performance against the established standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
This information is not applicable. The device is a medical examination glove, and its performance is evaluated against established material and safety standards (ASTM standards and biocompatibility tests), not against expert clinical ground truth like in a diagnostic device. The "ground truth" here is the pass/fail criteria defined by the objective standards themselves.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not applicable as the evaluation involves objective measurements against predefined standards rather than subjective expert interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. The device is a medical examination glove, which is a physical barrier device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This information is not applicable. The device is a medical examination glove, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the performance evaluation of these gloves is defined by:
- Established ASTM Standards: These are objective measurement criteria for physical properties (dimensions, tensile strength, elongation), freedom from holes, and powder residue.
- Biocompatibility Test Standards: These are established protocols for assessing primary skin irritation in rabbits and sensitization in guinea pigs, yielding objective pass/fail results.
8. The sample size for the training set:
This information is not applicable. The device is a physical medical glove and does not involve AI algorithms that require a "training set."
9. How the ground truth for the training set was established:
This information is not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(57 days)
A disposable device intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
Nitra-Touch™ Sterile examination gloves are disposable device intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
This document describes the acceptance criteria and the study results for the Nitra-Touch™ Sterile Powder-Free Nitrile Medical Examination Glove by Ansell Perry.
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimensions | Meets ASTM D 6319 | Meets ASTM D 6319 |
| Physical Properties: | ||
| Tensile Strength, minimum | Meets ASTM D 6319 (14 MPa) | Exceeds ASTM D 6319 (14 MPa) |
| Freedom from holes | Meets ASTM D 6319, Meets ASTM D 5151 | Meets ASTM D 6319, Meets ASTM D 5151 |
| Powder-Free | Not more than 2 mg residue by mass, Meets ASTM D 6124 | Not more than 2 mg residue by mass, Meets ASTM D 6124 |
| Biocompatibility | ||
| Primary Skin Irritation (Rabbits) | Acceptable (Passes) | Passes |
| Guinea Pig Sensitization | Acceptable (Passes) | Passes |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes used for each individual test (e.g., for dimensions, tensile strength, freedom from holes, powder-free, or biocompatibility). However, it indicates that the tests were conducted according to established ASTM standards (D 6319, D 5151, D 6124). These standards often prescribe minimum sample sizes for testing.
The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective." However, the tests are "non-clinical tests" conducted to demonstrate conformity with recognized standards like ASTM, which are generally performed in a controlled laboratory setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable and not provided in the document. The "ground truth" for medical examination gloves is established by objective, quantifiable measurements against defined performance standards (ASTM standards) and biocompatibility tests, rather than expert consensus on subjective interpretations.
4. Adjudication method for the test set
This information is not applicable. The performance is measured against objective standards, not through subjective adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. This device is a medical examination glove, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI-related effect sizes are irrelevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used
The ground truth used for evaluating the Nitra-Touch™ gloves is defined by:
- Established Industry Standards: ASTM Standard D 6319 (Nitrile Examination Gloves for Medical Application), ASTM D 5151 (Freedom from Holes), and ASTM D 6124 (Powder Residue).
- Biocompatibility Test Pass/Fail Criteria: Standard methods for assessing Primary Skin Irritation and Guinea Pig Sensitization.
- FDA Hole Requirements: The gloves also meet FDA hole requirements.
These ground truths are objective and quantitatively measurable.
8. The sample size for the training set
This information is not applicable. There is no concept of a "training set" for a physical medical device like an examination glove. These products are manufactured and then tested for compliance with specifications.
9. How the ground truth for the training set was established
This information is not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(65 days)
A device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination and for use handling chemotherapy drugs.
Derma Prene® Powder-Free Synthetic Surgical Gloves (Chemotherapy Use), meet all of the requirements of ASTM D 3577, Type 2.
Derma Prene® Powder-Free Synthetic Surgical Gloves (Chemotherapy Use) meet all the current specifications for ASTM D 3577 Rubber Surgical Gloves.
Derma Prene® Powder-Free Synthetic Surgical Gloves (Chemotherapy Use) are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination and for use handling chemotherapy drugs.
Derma Prene® Powder-Free Synthetic Surgical Gloves (Chemotherapy Use) are summarized with the following technological characteristics compared to ASTM or equivalent standards.
Characteristics: Dimensions, Physical Properties, Freedom from holes, Powder-Free, Biocompatability Primary Skin Irritation in Rabbits, Guinea Pig Sensitization
Standard: Meets ASTM D 3577, Meets ASTM D 3577, Type 2, Meets ASTM D 3577, Meets ASTM D 5151, Meets described test in Attachment VI of K980929, Meets ASTM D 6124 Not more than 2 mg residue by mass., Passes, Passes
Here's an analysis of the provided text, focusing on the acceptance criteria and the study used to prove the device meets those criteria:
Device: Derma Prene® Powder-Free Synthetic Surgical Gloves (Chemotherapy Use)
1. Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimensions | ASTM D 3577 | Meets ASTM D 3577 |
| Physical Properties | ASTM D 3577, Type 2 | Meets ASTM D 3577, Type 2 |
| Freedom from holes | ASTM D 3577, ASTM D 5151 | Meets ASTM D 3577, ASTM D 5151 |
| Powder-Free | Described test in Attachment VI of K980929 | Not more than 2 mg residue by mass |
| Biocompatibility | Primary Skin Irritation in Rabbits (Passes) | Passes |
| Biocompatibility | Guinea Pig Sensitization (Passes) | Passes |
2. Sample Size and Data Provenance for the Test Set
- Sample Size: The document does not explicitly state the sample size used for the performance tests. It mentions "performance test data of the non clinical tests," but no specific number of units tested.
- Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. It implies the studies were conducted by Ansell Perry based on their submission.
3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)
Not applicable. The ground truth for this device (surgical gloves) is established through adherence to recognized international standards (ASTM) and specific performance tests, not through expert consensus on interpretation.
4. Adjudication Method for the Test Set
Not applicable. As the tests involve objective measurements against established standards (e.g., dimensional checks, physical property tests, hole detection), there is no need for an adjudication method by experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This type of study is typically used for diagnostic devices where human readers interpret images or data. Surgical gloves are a medical device with objective performance criteria.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This device is a physical product (surgical glove), not an algorithm or AI system. Its performance is measured directly through laboratory testing.
7. Type of Ground Truth Used
The ground truth used is primarily established international standards (ASTM) and objective laboratory test results (e.g., measurements for dimensions, tensile strength, elasticity, freedom from holes, powder residue, and biological response in animal models for biocompatibility).
8. Sample Size for the Training Set
Not applicable. This device is not an AI/ML product; therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no AI/ML component, there's no training set or ground truth establishment in this context.
Ask a specific question about this device
Page 1 of 3