Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K051712

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-09-01

    (66 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Model 2800 is indicated for use with pulmonary function testing equipment, to filter air between the patient's exhaled air and the testing equipment. Single patient use.

    Device Description

    The EMS PFT Filter is a compact, electrostatic filter with various end-fitting adaptable to various pulmonary function testing circuits. It has 75 ml deadspace of 0.7 cm H2O at 720 lpm per ATS spirometry guidelines or 0.5 cm H2O @ 60 lpm. There are various connectors to allow connection to various PFT equipment. Single patient use.

    AI/ML Overview

    Air Safety Model 2800 PFT Filter Acceptance Criteria and Study Findings

    This document outlines the acceptance criteria for the Air Safety Model 2800 PFT Filter and the study that demonstrates its conformance.

    1. Acceptance Criteria and Reported Device Performance

    The Air Safety Model 2800 PFT Filter underwent testing to demonstrate substantial equivalence to the predicate device, the Engineered medical Systems - K013123. The acceptance criteria were based on the performance characteristics of this predicate device.

    AttributeAcceptance Criteria (Predicate EMS K013123)Reported Device Performance (Model 2800)
    Intended useFilter air between patient's exhaled air and PFT equipmentMeets criteria
    Intended for single patient useYesYes
    PrescriptionYesYes
    Intended populationAny patientAny patient
    Intended Environment of UseHospital, sub-acuteHospital, sub-acute, physician office
    Can be used with several different PFT machinesYesYes
    Compact housingYesYes
    Various end-fittingsYesYes
    Dead Space (ml)75 ml75 ml
    Resistance to flow at 720 lpm (ATS standard)$0.7 cm H_2O$$0.7 cm H_2O$
    Resistance to flow at 60 lpm$0.5 cm H_2O$$0.5 cm H_2O$
    Bacterial filtration99.9999%99.9999%
    Viral filtration99.999+%99.999+%
    Weight40 gm40 gm
    Housing materialPolystyrenePolystyrene
    Filter mediaElectrostatic polypropyleneElectrostatic polypropylene

    The study demonstrates that the Air Safety Model 2800 PFT Filter meets all specified acceptance criteria, showing identical performance characteristics to the predicate device for all listed attributes.

    2. Sample Size and Data Provenance

    • Sample Size for the Test Set: Not explicitly stated for each specific test (BFE, VFE, flow resistance). However, the narrative indicates that "BFE and VFE testing has been performed by Nelson Laboratories to demonstrate substantial equivalence to the predicate device."
    • Data Provenance: The data was generated by Nelson Laboratories, a third-party testing facility. The country of origin for the testing is not explicitly stated, but the submission is from Air Safety Ltd. in England. The study appears to be retrospective in the sense that it evaluates the performance of the manufactured device against pre-established criteria based on the predicate.

    3. Number of Experts and Qualifications for Ground Truth

    This type of device (PFT filter) does not typically involve expert human interpretation or "ground truth" in the diagnostic sense (e.g., image interpretation by radiologists). Instead, its performance is objectively measured against physical and biological metrics. Therefore, the concept of "experts establishing ground truth" as it applies to medical imaging algorithms is not relevant here. The "ground truth" for this device's performance is derived from standardized testing protocols and the established performance of the predicate device.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, the evaluation relies on objective laboratory testing against defined performance metrics, not on human interpretation or adjudication of ambiguous cases.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not applicable and therefore not performed. This device is a passive filter and does not involve human readers or AI in its operation. Its effectiveness is measured through physical and microbiological performance characteristics.

    6. Standalone Performance Study

    A standalone performance study was done for the device. The study involved BFE (Bacterial Filtration Efficiency) and VFE (Viral Filtration Efficiency) testing performed by Nelson Laboratories. This testing was conducted on the Air Safety Model 2800 PFT Filter independently of human interaction to objectively measure its filtration capabilities and other physical properties (e.g., dead space, resistance to flow). The results of these standalone tests were then compared directly to the established performance characteristics of the predicate device.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the Air Safety Model 2800 PFT Filter was objective performance data from standardized laboratory testing. This includes:

    • Bacterial and viral filtration efficiency percentages.
    • Measured resistance to flow at specified rates (cm H2O).
    • Measured dead space (ml).
    • Material composition and physical dimensions.

    This "ground truth" was established by comparing the results of the Model 2800's tests against the known, documented performance parameters of the legally marketed predicate device (EMS K013123), demonstrating substantial equivalence.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical component, not a software algorithm that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050953

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-06-29

    (75 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Filtration of exhaled gases in the ventilator circuit.

    Model 3000/06 to be placed in the exhalation pathway of the ventilator circuit where filtration of the expired gases is desired. Filter is reusable up to 100 times, extended use as determined by the increase in airflow resistance, which is determined during self testing (SST) mode of the ventilator. For use with the Series 700 and 7200 NPB ventilators.

    Cleaning method -- autoclaved -- 30 minutes with 3 minutes at 134°C.

    Environment of Use --Hospital, Sub-acute Institutions

    Device Description

    The Air Safety Model 3000/06 HEPA filter is a multi-use, reusable ventilator in-line filter which The An Bareey Indust side of the PB 700 series and 7200 ventilators. It can be cleaned by steam ecomeets in the ennation are 1 year service life. It has standard 22 mm tubing connectors and it has a clear housing. It contains the Air Safety standard Multipleat media which has been HEPA tested for enting performance according to BS 3928 Sodium Flame for Air filters and in accordance to DOE 3202-97 and ASTM D2986 - DOP. In addition BFE and VFE testing has been performed by 1902 5202 77 and in the substantial equivalence to the predicate devices.

    AI/ML Overview

    This response describes the acceptance criteria and the study that indicates the device meets the acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    AttributeAcceptance Criteria (Predicate Devices)Reported Device Performance (Air Safety Model 3000/06)
    Indications for useTo filter expired gases.Same
    Extended useUp to 15 days continuous, reusable up to 100 times via autoclave or 1 year service lifeYes (Matches predicate)
    PrescriptionYesYes (Matches predicate)
    Intended populationAny patient on a ventilator - PB 700 series and 7200Any patient on a ventilator - PB 700 series and 7200
    Intended Environment of UseHospital, sub-acute institutionHospital, sub-acute institution
    Placement in exhalation limb of circuitYesYes (Matches predicate)
    Reusable housing and materialsYesYes (Matches predicate)
    Clean by autoclaveYesYes (Matches predicate)
    Standard 15/22 mm connectorsYesYes (Matches predicate)
    Dead Space (ml)~ 350 ml~ 350 ml
    Resistance to flow< 1.5 cm H₂O at 60 lpm and < 2.5 cm H₂O at 100 lpm< 1.5 cm H₂O at 60 lpm and < 2.5 cm H₂O at 100 lpm
    HEPA - Bacterial filtration (BFE)99.999%99.999%
    HEPA - Viral filtration (VFE)99.999%99.999%
    Housing materialPolycarbonatePolycarbonate (Matches predicate)
    Filter mediaHEPA (Paper fiber)Paper fiber (Matches predicate)
    Performance Standards MetISO 5356-1 Conical 15/22; DOE 3202-97 and ASTM D2986 - DOP ( > 99.97% of 0.3 micron DOP particle)Yes (Matches predicate)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific "test set" sample size in terms of the number of devices or units tested for each performance characteristic. Instead, it refers to the performance of the predicate devices and implies that the Air Safety Model 3000/06 HEPA Filter matches these established performance characteristics.

    The provenance of the data is primarily from Nelson Lab for the Bacterial Filtration Efficiency (BFE) and Viral Filtration Efficiency (VFE) testing. The document also refers to testing performed according to BS 3928 Sodium Flame for Air filters and in accordance with DOE 3202-97 and ASTM D2986 - DOP. The context suggests these are retrospective references to established standards and predicate device performance, rather than a prospective, independently designed clinical trial specifically for the 3000/06 model.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    The document does not specify the number or qualifications of experts used to establish a "ground truth" for a specific test set. The data presented is engineering and laboratory performance data (e.g., filtration efficiency, resistance to flow) measured against established industry standards and specifications, rather than clinical outcomes requiring expert consensus on a patient-level "ground truth."

    4. Adjudication Method for the Test Set

    Not applicable. The evaluation relies on direct measurements against predefined technical specifications and standards, not on subjective assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is typically used for diagnostic imaging devices where human readers interpret patient cases, and the AI's impact on their performance is evaluated. The Air Safety Model 3000/06 is a physical filter, not a diagnostic tool that involves human interpretation of cases.

    6. Standalone Performance Study

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was effectively done. The performance metrics listed (e.g., BFE, VFE, resistance to flow, dead space) are inherent properties of the device itself, measured independently of human interaction in typical clinical use. The "study" here refers to the laboratory testing conducted to ascertain these physical and filtration properties against established standards.

    7. Type of Ground Truth Used

    The "ground truth" for the performance claims is based on established engineering and laboratory standards and predicate device specifications. For example:

    • Bacterial and Viral Filtration Efficiency: Determined by testing performed by Nelson Lab, likely against a standardized challenge (e.g., specific bacterial/viral aerosols) under controlled laboratory conditions.
    • Resistance to Flow: Measured using standardized methodologies relevant to ventilator circuits (e.g., flow rates of 60 lpm and 100 lpm).
    • HEPA testing: Performance according to BS 3928 Sodium Flame for Air filters and DOE 3202-97 and ASTM D2986 - DOP, which specify testing procedures and filtration efficiency requirements for particulate matter.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical filter being evaluated against quantifiable performance metrics, not an AI model that requires a "training set" for machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040712

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-06-03

    (77 days)

    Product Code
    Regulation Number
    868.5895
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Model 3500/01 is a replacement part for the ventilator. The Inlet Filter removes particles of 0.3 micron size (or larger) from entering the ventilator. It is a HEPA grade particulate filter for use with the NPB Achieva or LP 10 ventilators. The filter is in the gas flow pathway. As a HEPA particulate filter for the inlet air to be supplied to the patient from the ventilator.

    Device Description

    Model 3500/01 is a replacement part for the ventilator. The Inlet Filter removes particles of 0.3 micron size (or larger) from entering the ventilator. It is a HEPA grade particulate filter for the NPB Achieva or LP 10 ventilators, which is in the gas flow pathway. Screw-in cartridge to fit the inlet air filter port of the ventilator housing.

    AI/ML Overview

    The Air Safety HEPA Model 3500 Filter is an inlet filter designed to remove particles of 0.3 micron size or larger from entering ventilators, specifically the NPB Achieva and LP 10 models. It is a HEPA grade particulate filter.

    1. Table of Acceptance Criteria and Reported Device Performance:

    AttributeAcceptance CriteriaReported Device Performance
    Resistance to flow≤ 0.9 cm H2O @ 40 Lpm≤ 0.9 cm H2O @ 40 Lpm
    ≤ 1.25 cm H2O @ 60 Lpm≤ 1.25 cm H2O @ 60 Lpm
    HEPA – particulate filtration efficiency (retention)≥ 99.97% of 0.3 micron DOP particle at 60 Lpm> 99.97% of 0.3 micron DOP particle at 60 Lpm
    ≥ 99.97% of 0.3 micron DOP particle at 100 Lpm> 99.97% of 0.3 micron DOP particle at 100 Lpm

    2. Sample size used for the test set and the data provenance:

    The document describes performance testing based on "standard DOP aerosol testing methods" and references specific standards (DOE 3025-99, DOE 3020-97, and ASTM D2986 - DOP). However, it does not specify the sample size for the test set used to generate the reported performance data. The data provenance is implied to be from internal testing conducted by Air Safety Ltd. in England. It's a prospective testing for the device's performance characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This device is a physical filter, and its performance (filtration efficiency and resistance to flow) is assessed through objective laboratory testing using recognized standards. Therefore, the concept of "experts establishing ground truth" in the way it applies to image analysis or clinical diagnoses is not relevant in this context. The ground truth is the result of the objective measurements obtained through the specified test methods.

    4. Adjudication method for the test set:

    Not applicable. As described above, the performance is determined by objective, standardized measurements, not through expert judgment or consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or imaging devices where human interpretation is involved. This device's performance is quantifiable through laboratory testing.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, the performance testing of the HEPA filter can be considered a standalone assessment. The filtration efficiency and resistance to flow are measured attributes of the device itself, independent of human interaction in its primary function (filtering air).

    7. The type of ground truth used:

    The ground truth used for this device's performance assessment is based on objective measurements obtained through standardized laboratory protocols. Specifically, it involves:

    • DOP aerosol testing for particulate filtration efficiency (retention).
    • Measurement of resistance to flow at specified flow rates.

    These are established engineering and scientific methods to quantify the physical properties of filters.

    8. The sample size for the training set:

    Not applicable. This device is a physical product, not an algorithmic model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033008

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-13

    (230 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Anesthesia / Respiratory Filters
    For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired.
    Model – HEPA filter 3000/04 – Single patient use for exhalation limb of circuit on NPB 700 series ventilators
    Models – HEPA filters - 6500/01, 6888/01, 8222/01, 8444/01 Model – Non-HEPA filters - 4000/01 Single patient use up to 24 hours. Patient tidal volumes > 150 ml, when applicable.

    Device Description

    The Air Safety HEPA and Non-HEPA filters are available in multiple sizes and shapes, and incorporate standard 15 / 22 mm connectors with or without a gas sampling luer port. Some models adapt to fit ventilator exhalation limb only. The depth (HEPA filtration) filter uses a pleated paper fiber for filtration. Filters are tested for rating performance according to EN 13328 Salt for Breathing System filtration performance. The "HEPA" performance was also tested in accordance to DOE-3025-99, DOE-3020-97 and ASTM D2986 - DOP. The electrostatic (non-HEPA filtration) filters are tested by Nelson Laboratories for BFE and VFE.

    AI/ML Overview

    This document describes the safety and effectiveness of Air Safety HEPA and Non-HEPA Filters. However, it does not fit the typical format of a study designed to prove a device meets acceptance criteria, especially in the context of an AI/ML powered device. This is a 510(k) premarket notification for traditional medical devices (filters), and therefore, many of the requested categories are not applicable.

    Here's an attempt to answer the questions based on the provided text, with clear indications where the information is not available or not relevant to this type of device submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present "acceptance criteria" in a pass/fail format with direct device performance comparison in a single table, as might be found in a study for an AI/ML device. Instead, it lists technical characteristics and performance standards that the device meets.

    However, based on the General Technical Characteristics table, we can infer performance criteria and compare them to the device's reported performance:

    Attribute/Performance StandardAcceptance Criteria (Inferred from Predicate/Standards)Reported Device Performance (Air Safety Filters)
    Dead Space (ml)Comparable to predicate / within acceptable range for intended use45 to 84 ml (various models), 209 mm for Model 3000/04
    Resistance to flow< 3.4 cm H2O @ 60 Lpm (Likely a standard for breathing circuits)< 3.4 cm H2O @ 60 Lpm
    HEPA Filtration Efficiency (for listed HEPA models)Specified by standards (e.g., DOE-3025-99, DOE-3020-97, ASTM D2986 - DOP for > 99.97% of 0.3 micron DOP particle @ 60 Lpm)99.99999% (Models 6500/01, 6888/01, 8222/01, 8444/01) Model 3000/04 - 99.9999%
    Bacterial Filtration Efficiency (BFE) - Nelson (for listed HEPA models)Standard requirement for bacterial filters (implied by predicate and testing)99.99975% (Models 6500/01, 6888/01, 8222/01, 8444/01) Model 3000/04 - 99.9999%
    Viral Filtration Efficiency (VFE) - Nelson (for listed HEPA models)Standard requirement for viral filters (implied by predicate and testing)99.99996% (Models 6500/01, 6888/01, 8222/01, 8444/01)
    Bacterial Filtration Efficiency (BFE) - Nelson (for Non-HEPA Model 4000/01)Standard requirement for bacterial filters (implied)99.99996%
    Viral Filtration Efficiency (VFE) - Nelson (for Non-HEPA Model 4000/01)Standard requirement for viral filters (implied)99.99925%
    Filtration Performance (General)According to EN 13328 Salt for Breathing System filtration performance."Filters are tested for rating performance according to EN 13328 Salt for Breathing System filtration performance." (Implied compliance)
    DOP Test Performance (for HEPA)> 99.97% of 0.3 micron DOP particle @ 60 Lpm> 99.97% of 0.3 micron DOP particle @ 60 Lpm
    Standard 15/22 mm connectorsRequired for compatibilityYes
    Gas sampling portOptionalOptional
    Single patient use up to 24 hoursYesYes
    ISO 5356-1 Conical 15/22Compliance with standardYes
    ISO 594-2 Luer FittingsCompliance with standardYes

    Summary of "Study" to Prove Device Meets Criteria:

    The document details the device characteristics and states that the devices were tested to specific performance standards. This is not a "study" in the clinical trial sense but rather a series of engineering and laboratory performance tests.

    • Testing Methodologies:
      • Filtration Performance: Tested according to EN 13328 Salt for Breathing System filtration performance.
      • HEPA Performance (Specific models): Tested in accordance with DOE-3025-99, DOE-3020-97, and ASTM D2986 - DOP.
      • Electrostatic (non-HEPA) Filtration (BFE and VFE): Tested by Nelson Laboratories.
    • Conclusion: The submission concludes that "The data within the submission demonstrates that the proposed devices when compared to the legally marketed predicate devices are safe and effective and substantially equivalent." This implies that the performance characteristics met or exceeded those of the predicate devices and applicable standards.

    Information Not Applicable / Not Provided for this Device Type:

    The following requested information is primarily relevant for AI/ML-powered diagnostic or predictive devices, or those requiring clinical studies with human assessors. The provided document concerns a physical filter, for which these criteria are generally not applicable or not detailed in this type of 510(k) submission.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Applicable / Not provided in this context. For physical filters, testing typically involves laboratory samples of the filters themselves, not patient data in the sense of AI/ML. The number of filter units tested for each specific performance claim (e.g., BFE, VFE, resistance) is not specified. Data provenance like "country of origin" for test data is not a typical requirement for device performance testing in this manner.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. "Ground truth" established by experts (like radiologists for an image analysis AI) is not relevant for a breathing circuit filter. Performance is measured objectively through standardized laboratory tests (e.g., particle counts, pressure measurements).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. Adjudication methods are used to resolve discrepancies in expert opinions, typically for establishing ground truth in clinical datasets. This is not relevant for the objective performance testing of a physical filter.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a physical filter, not an AI/ML-powered diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical filter, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not Applicable. The "ground truth" for a filter's performance is derived from standardized physical and biological challenge tests (e.g., aerosolized particles, bacterial/viral suspensions) and measurements, not clinical expert consensus or pathology.

    8. The sample size for the training set:

    • Not Applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1