    K Number
    K251041

    Validate with FDA (Live)

    Date Cleared
    2025-10-31

    (211 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GripTract-GI Endoscopic Tissue Manipulator (GripTract) Upper GI Model is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    GripTract Upper GI Model is indicated for use in the upper gastrointestinal tract with any standard endoscope as follows:

    Endoscope Distal Tip Outer Diameter (mm)Endoscope Working Length (cm)GripTract Model#
    9.8 - 10.5103 - 110GT-GS100
    Device Description

    The GripTract-GI™ Endoscopic Tissue Manipulator (GripTract) Upper GI Model GT-GS100 is a disposable, non-sterile accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    The GripTract Handpiece attaches to the endoscope control body just below the working channel. The soft End Cap with two integrated Fingers for tissue manipulation is placed on the distal end of the endoscope. Two Control Knobs on the Handpiece each operate a corresponding Finger, permitting the User to rotate and extend/retract the Fingers for tissue manipulation and visualization independent of the endoscope's movement or the presence of other tools in the endoscope's working channel.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242325

    Validate with FDA (Live)

    Date Cleared
    2024-10-04

    (59 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GripTract-GI Endoscopic Tissue Manipulator (GripTract) Lower GI Models are accessories intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    GripTract Lower GI Models are indicated for use in the large intestine with any standard endoscope as follows:

    EndoscopeEndoscopeGripTract
    Distal Tip Outer Diameter (mm)Working Length (cm)Model #
    12.8 - 13.3168 - 170GT-CL170
    12.8 - 13.3130 - 133GT-CL130
    11.5 - 12.0168 - 170GT-CM170
    11.5 - 12.0130 - 133GT-CM130
    Device Description

    The GripTract-GI™ Endoscopic Tissue Manipulator (GripTract) Lower GI Models are disposable, non-sterile accessories intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment. There are four (4) different GripTract Lower GI Models (GT-CL170, GT-CL130, GT-CM170, and GT-CM130) which are indicated for use in the large intestine with standard endoscopes that have specific tip outer diameters and working lengths.

    The GripTract Handpiece attaches to the endoscope control body just below the working channel. The soft End Cap with two integrated Fingers for tissue manipulation is placed on the distal end of the endoscope. Two Control Knobs in the Handpiece each operate a corresponding Finger, permitting the User to rotate and extend/retract the Fingers for tissue manipulation and visualization independent of the endoscope's movement or the presence of other tools in the endoscope's working channel.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document, addressing each of your points:

    Important Note: The provided document is a 510(k) summary for a medical device (GripTract-GI Endoscopic Tissue Manipulator Lower GI Models). This type of document focuses on demonstrating substantial equivalence to an existing predicate device rather than proving clinical efficacy or diagnostic accuracy for an AI/software device. Therefore, many of your questions, particularly those related to AI-specific studies (MRMC, standalone AI performance, ground truth establishment for training AI, etc.), are not applicable to this device and information. The device described is not an AI/software device.

    The document primarily details non-clinical performance data to support the substantial equivalence claim.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of acceptance criteria with numerical performance targets against which the device was measured like typical AI/diagnostic studies. Instead, it lists various non-clinical tests performed and states whether the device "passed" or exhibited "equivalent force transmission" or "exceeded reported maximum pull forces."

    Here's a summary derived from the "Summary of non-clinical performance testing" table and the subsequent paragraphs:

    Test PerformedAcceptance Criteria (Implied)Reported Device Performance
    Verification of Product SpecificationDevice meets all product specifications after exposure to ambient, high temperature, and low temperature conditions.Pass
    ReliabilityDevice meets all product specifications following exposure to worst-case simulated use.Pass
    Bench SafetyDevice is sufficiently safe during benchtop evaluations approximating worst-case scenarios and interactions with biological tissue.Pass
    Torsional Fatigue StrengthDevice meets all product specifications following repeated worst-case torsional load.Pass
    Torque ComparisonForce transmitted to the Finger by different models is equivalent.Equivalent force transmission
    End Cap Force ComparisonForce required to remove End Caps exceeds reported maximum pull forces in colonoscopies.Removal forces exceed reported maximum pull forces in colonoscopies
    Accelerated Shelf LifeDevice meets all product specifications after exposure to conditions simulating a two-year shelf life.Pass
    Shelf LifeDevice meets all product specifications after exposure to real-time ambient conditions for two years.Test is on-going
    BiocompatibilityDevice passes tests for cytotoxicity, sensitization, irritation, systemic toxicity, and pyrogenicity per ISO 10993-1.Pass
    Endoscope Viewing Area ComparisonThe device does not block endoscope visualization.Confirmed that GripTract Lower GI Models do not block visualization.

    2. Sample Size Used for the Test Set and Data Provenance

    As this is a physical medical device (endoscopic tissue manipulator) and not an AI/software, the concept of a "test set" in the context of diagnostic data doesn't apply directly.

    • Sample Size: The document does not specify particular "sample sizes" for the non-clinical tests in terms of number of devices tested. It refers to "worst-case simulated use," "repeated worst-case torsional load," and testing "at the dimensional extremes of the product line" for Torque Comparison. For Biocompatibility, it states "Final, finished devices tested."
    • Data Provenance: Not applicable in the context of retrospective/prospective patient data or country of origin for such data. The data provenance is from bench performance testing conducted by the manufacturer, Actuated Medical, Inc.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. "Ground truth" in the context of expert consensus on medical images or diagnostic outcomes is not relevant for this device's non-clinical performance testing. The "ground truth" for these tests would be the measured physical properties and performance characteristics against established engineering specifications and regulations.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving discrepancies among expert readers in diagnostic AI studies. This is a non-clinical, bench testing document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improvement with AI vs. Without AI Assistance

    Not applicable. This device is a physical medical instrument, not an AI or software that assists human readers in diagnostic interpretation. An MRMC study would not be performed for this type of device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm or software. It is a physical device operated by a clinician.

    7. The Type of Ground Truth Used

    For the non-clinical performance tests, the "ground truth" is defined by:

    • Engineering specifications and design requirements.
    • Industry standards (e.g., ISO 10993-1 for biocompatibility).
    • Quantitative measurements derived from bench testing (e.g., force measurements, dimensional checks, fatigue cycles).
    • Simulated worst-case scenarios and interactions.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no training set of data in that sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI training set, there's no ground truth establishment process for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231254

    Validate with FDA (Live)

    Date Cleared
    2023-11-03

    (186 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GripTract-Gl Endoscopic Tissue Manipulator (GripTract) is an accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment.

    GripTract is indicated for use in the large intestine with any standard endoscope that has a distal tip outer diameter of 11.5 - 12.0 mm and working length of 168 - 170 cm.

    Device Description

    GripTract is a single-use, non-sterile, endoscope accessory intended to ensure complete positioning of an endoscope and assist with optical visualization, diagnosis, and endoscopic treatment. It is indicated for use in the large intestine with any standard endoscope that has a distal tip outer diameter of 11.5 - 12.0 mm and a working length of 168 - 170 cm.

    The GripTract Handpiece attaches to the endoscope control body just below the working channel. The soft End Cap with two integrated Fingers for tissue manipulation is placed on the distal end of the endoscope. Two Control Knobs in the Handpiece each operate a corresponding Finger, permitting the User to rotate and extend/retract the Fingers for tissue manipulation and visualization independent of the endoscope's movement or the presence of other tools in the endoscope's working channel.

    AI/ML Overview

    The provided text describes the regulatory clearance for the GripTract-GI Endoscopic Tissue Manipulator (K231254). It outlines the device's intended use and provides non-clinical performance data to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Key Takeaway: The provided document is a 510(k) summary for a medical device (GripTract-GI Endoscopic Tissue Manipulator). It focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific numerical performance acceptance criteria typical of AI/ML device studies. Therefore, many of the requested criteria (e.g., specific quantitative metrics, sample sizes for training/test sets for AI, expert numbers for ground truth, MRMC studies) are not directly applicable or explicitly stated in this type of regulatory submission for a mechanical device.

    However, I can extract the information related to the device's performance and the studies conducted to show its safety and effectiveness.


    Acceptance Criteria and Device Performance (based on provided text)

    The acceptance criteria for this device are primarily related to its functional performance, safety, and compatibility, demonstrated through various non-clinical tests. The "performance" is reported as "Pass" for each test, indicating that the device met the pre-defined criteria for that specific test.

    Acceptance Criteria CategorySpecific Criteria / Test DescriptionReported Device Performance
    Product SpecificationsVerification of Product Specifications after exposure to ambient, high temperature, and low temperature environmental conditions.Pass
    Distribution / Packaging IntegrityConfirmation of Product Specifications following exposure to simulated distribution stress and conditions (per ASTM D4169-16).Pass
    Shelf Life (Accelerated)Confirmation of Product Specifications following exposure to accelerated conditions simulating a shelf-life of two years.Pass
    Shelf Life (Real-Time)Confirmation of Product Specifications following exposure to real-time ambient conditions for two years.Test is on-going
    ReliabilityConfirmation of Product Specifications following worst-case simulated use.Pass
    Bench SafetyAssessed safety of worst-case interactions between biological tissue and GripTract Fingers.Pass
    Usability (Set-Up)Assessment of whether the intended user population can successfully set up GripTract without serious use errors, failure to perform critical tasks, or encountering problems that affect the intended use and expected use environment.Pass
    Usability (Use)Assessment of whether the intended user population can successfully use GripTract without serious use errors, failure to perform critical tasks, or encountering problems that affect the intended use and expected use environment.Pass
    BiocompatibilityTesting of final, finished devices for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity (per ISO 10993-12, -5, -10, -23, -11).Pass
    In Vivo Safety & Effectiveness (Porcine)In vivo porcine testing of GripTract against standard of care in worst-case procedure. Acute and chronic time points examined for differences in visual and histological mucosal damage as well as presence of device-attributable adverse events. Equivalent visual and histological mucosal damage compared to endoscope-only (control).Pass
    Endoscope Viewing Area (Bench Safety)Comparison of the endoscope viewing area between GripTract and the Reference Device.Does not block visualization relative to a standard distal end cap.

    Study Details (based on provided text)

    1. Sample sizes used for the test set and the data provenance:

      • Usability Testing: "A total of 15 Users" for both "Set-Up" and "Use" usability tests.
      • Porcine Testing: Number of animals or procedures is not specified, but it's an "in vivo porcine testing" in "worst-case procedure."
      • Other Bench Tests: Sample sizes are not explicitly stated for product specification verifications, distribution, shelf-life, reliability, or bench safety (other than the endoscope viewing area comparison).
      • Data Provenance: The studies are non-clinical, primarily bench and animal (porcine) testing. No explicit mention of country of origin for the data or whether it's retrospective/prospective (these terms are more relevant for human clinical studies).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided. For non-AI mechanical devices, the "ground truth" is typically established by engineering specifications, validated test methods, and potentially expert interpretation for things like tissue damage in animal studies. The "users" in usability testing are relevant, but their "qualification" is broadly "intended user population" (medical specialists).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/mentioned for this type of non-clinical device testing.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a mechanical accessory, not an AI-powered diagnostic tool requiring human reader comparison studies.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. This is a mechanical device, not an algorithm. Bench tests evaluate the device's standalone performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Bench Performance: Engineering specifications and ASTM/ISO standards.
      • Usability: Defined user tasks and absence of serious use errors, likely determined by observation and user feedback against pre-defined criteria.
      • Biocompatibility: ISO 10993 standards and their associated pass/fail criteria (e.g., lack of cytotoxicity, sensitization).
      • Porcine Testing: Visual and histological mucosal damage assessment compared to a control (endoscope-only procedures), suggesting pathology as a component of ground truth for tissue effects.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200646

    Validate with FDA (Live)

    Device Name
    TubeClear System
    Date Cleared
    2020-12-04

    (267 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TubeClear Clearing Stem Model GJ-1422 is indicated for use ONLY and SOLELY in clearing occlusions/clogs in adults that have the following Tube type and size (French and length).

    • GJ-1422, for G-Jet®, MIC®, and MIC-KEY® Gastro-jejunostomy tubes that are size 14-22 Fr and have a jejunal length of 15-45 cm (6-18 in).
    Device Description

    The TubeClear System is comprised of a reusable Control Box and a single use Clearing Stem. Control Box Model 101 is used to actuate all Clearing Stem models. The Clearing Stem for the predicate device is TubeClear Clearing Stem Model TC-0608 (TC-0608). The Clearing Stem for the subject device is TubeClear Clearing Stem Model GJ-1422 (GJ-1422). GJ-1422 is connected to the Control Box. The Operator then coats GJ-1422 with Coconut Oil and manually inserts it into the gastro-jejunostomy (GJ) tube (Tube) and directs the progression of the Clearing Stem inside the Tube. The Control Box actuates the Clearing Stem Wire. The Clearing Stem Wire Tip mechanically clears the occlusion to restore Tube patency. Sixteen (16) Clearing Stem Models are currently marketed to accommodate different types, sizes, and materials of feeding tubes. This submission adds a seventeenth (17th) Clearing Stem model (GJ-1422) for use in a specific set of gastro-jejunostomy (GJ) tubes. Use of TubeClear Clearing Stem Model GJ-1422 requires coating the Clearing Stem with Coconut Oil prior to introduction into a patient's feeding tube.

    AI/ML Overview

    The provided text describes the submission for a new model of the TubeClear System, GJ-1422, which is essentially an updated version of the previously cleared TC-0608. The document focuses on demonstrating substantial equivalence rather than presenting an entirely new study with rigorous acceptance criteria and standalone performance for a novel AI device. Therefore, some of the requested information, particularly regarding AI-specific studies (e.g., MRMC, training set details, ground truth for training), is not applicable or not present in this regulatory submission.

    However, based on the non-clinical performance data provided, we can infer and construct a table of acceptance criteria and reported performance for the mechanical device (TubeClear System), not an AI algorithm.

    Here's the breakdown of the available information:

    Acceptance Criteria and Reported Device Performance (Non-Clinical Benchtop Testing for Mechanical Device)

    The document specifies that the acceptance criteria for the new GJ-1422 model were "the same as that used for the legally marketed TC-0608 model". It also states that both models "passed their respective a priori acceptance criteria for each test, which reflects clinically acceptable success rates." While specific numerical values for the acceptance criteria are not explicitly stated, the text indicates that the device met these criteria.

    Test CategoryAcceptance Criteria (Implied / Inferred)Reported Device Performance (GJ-1422)
    Benchtop EfficacyEffectively clear worst-case artificial clogs in indicated tubes. (Same as TC-0608, ensuring clinically acceptable success rates)Effectively cleared the required occluded Tubes (G-Jet®, MIC®, and MIC-KEY® Gastro-jejunostomy tubes, 14-22 Fr, 15-45 cm jejunal length, silicone) when used in accordance with IFU. The device met the a priori acceptance criteria, indicating substantial equivalence to TC-0608 in clearing occluded tubes.
    Clearing Stem Removal ForcesForce required to remove the Clearing Stem should be below a pre-defined maximum to prevent tube dislodgement or component strength exceedance. (Same as TC-0608)Had lower forces than the a priori acceptance criteria when used in accordance with IFU. This indicates substantial equivalence to TC-0608 regarding removal forces.
    Feeding Tube IntegrityNo signs of damage (scratching, marring) to the feeding tube after repeated use and contact with kinks.No signs of scratching or marring observed under a microscope after the test (simulated clog cleared, then 70 minutes of contact with a kink in ex vivo porcine tissue). Supports substantial equivalence to TC-0608.
    Flow Rate (Post-Use)Repeated use with Coconut Oil should not negatively alter the tube's subsequent flow rate or lead to accumulated residue.Showed a statistically significant higher amount of water collected in worst-case testing (repeated introduction of GJ-1422 coated in Coconut Oil into smallest French size/maximal length tube). This demonstrates that Coconut Oil does not lead to residue accumulation or reduce tube patency. Supports substantial equivalence to TC-0608.
    Coconut Oil Safety (USP & Toxicology)Coconut Oil should meet USP specifications and present no unacceptable adverse risks to patients.Nineteen (19) of twenty-one (21) analytes met USP 42 specification. Literature review of 8 clinical trials found no negative effects from coconut oil consumption, even at significantly higher doses than the worst-case potential administration with GJ-1422 (140-378 times higher). Indicates no unacceptable adverse risks to the patient.

    Study Details:

    This document is a 510(k) Premarket Notification, not a detailed research study publication. Therefore, some granular details typically found in a scientific paper's methodology section are not fully elaborated.

    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated as a single number. The tests involved "series of indicated Tubes" for Benchtop Efficacy, "artificially created clogged Tube" for Removal Forces, and "different geometry and materials used in indicated Tubes" (GJ tubes) with ex vivo porcine tissue for Tube Integrity. Flow Rate testing involved "feeding tube of the smallest French size and maximal length." The specific number of tubes tested for each scenario is not quantified.
      • Data Provenance: This is
        • Benchtop/Laboratory Data: The tests were performed in a controlled laboratory setting (benchtop efficacy, removal forces, tube integrity, flow rate).
        • Ex Vivo Data: Feeding Tube Integrity test used "ex vivo porcine tissue model."
        • Literature Review: Toxicity assessment relied on a "literature review found eight (8) clinical trials exploring the effects of consuming coconut oil."
        • Geographic Source: Not specified, but generally, regulatory submissions for US markets involve testing conducted in the US or in compliance with US standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is a mechanical device performance submission, not an AI diagnostic device. Therefore, no "ground truth" in the diagnostic sense (e.g., radiology interpretation) was established by human experts for a test set. The "ground truth" for these tests comes from objective measurements (e.g., cleared tube patency, measured forces, visual inspection for damage, flow rates, chemical analysis).
    3. Adjudication method for the test set:

      • Not applicable as this is not a subjective interpretation task (like image reading) requiring adjudication. The outcomes were direct measurements or observations based on predefined objective criteria.
    4. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

      • No, an MRMC study was not done. This device is a mechanical tool for clearing occlusions, not an AI-assisted diagnostic device where human reader performance is compared with and without AI.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a mechanical device, not an algorithm. The "performance" described is the ability of the physical device to clear clogs and its safety.
    6. The type of ground truth used:

      • Objective Benchtop Measurements: Patency (clearing of artificial clogs), force measurements, visual microscopic inspection for damage, flow rate measurements.
      • Chemical Analysis: USP monograph testing for Coconut Oil.
      • Literature-based Evidence: For toxicological risk assessment of Coconut Oil.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device that requires a training set. The engineering design and testing are based on traditional medical device development.
    8. How the ground truth for the training set was established:

      • Not applicable. (See #7).
    Ask a Question

    Ask a specific question about this device

    K Number
    K172556

    Validate with FDA (Live)

    Device Name
    TubeClear System
    Date Cleared
    2018-06-29

    (309 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TubeClear System Clearing Stem Models are indicated for use ONLY and SOLELY in clearing occlusions/clogs in Feeding and Decompression Tubes in adult patients that have the following Tube type and size (French and length):

    · TC-0608: for Nasoenteral and Nasogastric Tubes composed of Polyvinyl Chloride (PVC) and Polyurethane that are size 6 - 8 Fr and have a length of 38 - 140 cm (15 - 55 in);

    · TC-0812: for Nasoenteral and Nasogastric Tubes composed of Polyviny1 Chloride (PVC) and Polyurethane that are size 8 - 12 Fr and have a length of 38 - 140 cm (15 - 55 in);

    · TC-1018: for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 91 - 140 cm (36 - 55 in); and Gastrostomy and Jejunostomy Tubes that are size 10 - 18 Fr and have a length of 20 - 36 cm (8 - 14 in);

    • · NE-1036: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in);
    • NE-1042: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in);
    • NE-1043: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in);
    • NE-1045: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in);
    • · NE-1048: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in);
    • NE-1050: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in);
    • NE-1055: for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in);
    • · G-1008: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in);
    • · G-1009: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in);
    • · G-1010: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 25 cm (10 in);
    • · G-1011: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in);
    • · G-1012: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in); and
    • · G-1014: for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
    Device Description

    The TubeClear System is comprised of a reusable Control Box and a single use Clearing Stem. The Clearing Stem is connected to the Control Box. The Operator then manually inserts the Clearing Stem into the Feeding and Decompression Tube (i.e., Tube) and directs the progression of the Clearing Stem inside the Tube. The Control Box actuates the Clearing Stem Wire. The Clearing Stem Wire Tip mechanically clears the occlusion to restore Tube patency. Control Box Model 101 is used to actuate all Clearing Stem models. Sixteen (16) Clearing Stem Models are proposed to accommodate the different types, sizes, and materials of Tubes.

    AI/ML Overview

    The provided text describes the Actuated Medical, Inc. TubeClear System (K172556), a device intended for clearing occlusions in feeding and decompression tubes.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Device FunctionalityThe device is intended to clear occlusions/clogs in Feeding and Decompression Tubes.
    Substantial EquivalenceSubstantial equivalence is claimed to predicate devices K163092 (TubeClear System, Actuated Medical, Inc.) and K131052 (TubeClear System, Actuated Medical, Inc.).
    Material ChangesModified material of the Control Box O-Ring.

    2. Sample Size Used for the Test Set and Data Provenance
    The document does not explicitly state the sample size used for the test set or the data provenance (country of origin, retrospective/prospective). It only mentions "Performance bench testing used to support substantial equivalence was water ingress testing of the o-ring." This suggests that the testing was likely conducted in a lab environment.

    3. Number of Experts and Qualifications for Ground Truth
    Not applicable. The study described is bench testing (water ingress), not a clinical study involving human assessment or expert review for ground truth.

    4. Adjudication Method for the Test Set
    Not applicable. This was a bench test, not a study requiring adjudication of expert opinions.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
    No, a MRMC comparative effectiveness study was not done. The submission focuses on bench testing for a material change, not a comparative study of human readers with or without AI assistance.

    6. Standalone (Algorithm Only) Performance
    Not applicable. The TubeClear System is a physical medical device, not an AI algorithm. The performance evaluation was for the physical properties of a component (o-ring).

    7. Type of Ground Truth Used
    The ground truth for the performance testing was based on direct physical measurements: "water ingress testing of the o-ring." This is a measure of the o-ring's ability to prevent water from entering the control box.

    8. Sample Size for the Training Set
    Not applicable. This device is not an AI/algorithm-based device and therefore does not have a "training set" in the context of machine learning. The "training" for this device would refer to its design, engineering, and manufacturing processes.

    9. How Ground Truth for the Training Set Was Established
    Not applicable, as there is no "training set" in the context of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163092

    Validate with FDA (Live)

    Date Cleared
    2016-11-30

    (26 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TubeClear Clearing Stem Model TC-1018 is indicated for use ONLY and SOLELY in clearing occlusion / clogs in Feeding and Decompression Tubes in adults that have the following Tube type and size (French and length).

    · TC-1018. for Nasoenteral and Nasogastric Tubes that are size 10 - 18 Fr and have a length of 91 - 140 cm (36 - 55 in); and Gastrostomy and Jejunostomy Tubes that are size 10 - 18 Fr and have a length of 20 - 36 cm (8 - 14 in).

    TubeClear Clearing Stem Model TC-0812 is indicated for use ONLY and SOLELY in clearing occlusion / clogs in Feeding and Decompression Tubes in adults that have the following Tube type, size (French and length), and material.

    · TC-0812, for Nasoenteral and Nasogastric Tubes composed of Polyvinyl Chloride (PVC) and Polyurethane that are size 8 - 12 Fr and have a length of 38 - 140 cm (15 - 55 in).

    TubeClear Clearing Stem Model TC-0608 is indicated for use ONLY and SOLELY in clearing occlusion / cloas in Feeding and Decompression Tubes in adults that have the following Tube type, size (French and length), and material.

    • TC-0608, for Nasoenteral and Nasogastric Tubes composed of Polyvinyl . Chloride (PVC) and Polyurethane that are size 6 - 8 Fr and have a length of 38 -140 cm (15 - 55 in).
    Device Description

    The Proposed Device, the TubeClear Control Box and Clearing Stem, is comprised of Control Box Model 101 and Clearing Stem Models TC-0608, TC-0812, and TC-1018. The Control Box is reusable and the Clearing Stems are single use. The Operator manually inserts the Clearing Stem into the feeding and/or decompression tube (i.e., Tube), while the Tube remains in the Patient, and directs the Clearing Stem's progression along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear reciprocating motion. The linear reciprocating motion is transferred to the proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Wire. The motion at the Wire Tip mechanically acts on the occlusion and restores Tube patency.

    Three (3) new Clearing Stem Models are proposed. The additional Clearing Stem Models expand the specific indications for use.

    AI/ML Overview

    This document describes the TubeClear Control Box and Clearing Stem (referred to as the "Proposed Device") and its substantial equivalence to a previously marketed device, also called TubeClear (referred to as the "Predicate Device"). This is a summary from a 510(k) premarket notification to the FDA. The information provided focuses on the non-clinical performance data to demonstrate substantial equivalence, as no clinical data was collected.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table with numerical acceptance criteria and reported device performance. Instead, it uses qualitative statements about conformance and substantial equivalence.

    Acceptance Criteria CategoryReported Device Performance (Proposed Device)
    Technical Testing
    Product SpecificationsMet all product specifications.
    Wire Tip FlexibilityAssessed, results contribute to substantial equivalence.
    Forces on Clearing StemAssessed, results contribute to substantial equivalence.
    Distal Wire Pull ForceAssessed, results contribute to substantial equivalence.
    Accelerated Shelf LifeMet all acceptance criteria for 12 months storage. Ongoing testing for longer shelf life.
    Packaging Integrity / Transportation VibrationMet all acceptance criteria for US highway truck.
    Efficacy Testing
    Human Factors ValidationSuccessfully completed by certified and licensed healthcare practitioners. Results support substantial equivalence of effectiveness.
    Device EffectivenessEffectiveness during simulated use was compared to Predicate Device; results support substantial equivalence.
    Safety Testing
    Tube Integrity (scratches, nicks, tears, abrasions, punctures)Assessed; found to be substantially equivalent to Predicate Device.
    Tube HeatingFound to be negligible and substantially equivalent to Predicate Device.
    Tube MovementFound to be substantially equivalent to Predicate Device.
    General Medical Electrical Equipment Safety (IEC-60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007))Conformance tested and found by third party (Intertek Group).
    Usability Engineering (IEC 60601-1-6: 2010, IEC 62366: 2007)Conformance tested and found by third party (Intertek Group).
    Risk Management (ISO 14971-compliant system)Conformance tested and found by third party (Intertek Group).
    Electromagnetic Compatibility (IEC 60601-1-2:2001 +A1:2004)Conformance tested and found by third party (Intertek Group).
    Intestinal Lining Puncture (Animal Testing - worst case over-insertion)Results support that the Proposed Clearing Stem Design is safe for its intended use and substantially equivalent to the Predicate Device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set:
      • Technical Testing: Not explicitly stated for each test, but implied to be sufficient for verification and adherence to product specifications (e.g., "Verification confirmed that the Clearing Stem Models met product specifications").
      • Efficacy Testing (Human Factors Validation & Device Effectiveness): The number of participants for human factors validation is not specified, but they are described as "certified and licensed healthcare practitioners." The number of simulated uses or cases for device effectiveness testing is also not explicitly stated.
      • Safety Testing: Not explicitly stated for each test (e.g., number of tubes tested for integrity, heating, or movement).
      • Animal Testing: Not explicitly stated, but performed "on porcine jejunum tissue."
    • Data Provenance: The document does not specify the country of origin of the data. It indicates the testing was "in-house" for much of the technical, efficacy, and safety testing, with some safety tests conducted by a "third-party organization (Intertek Group (Boxborough, MA))" in the US.
    • Retrospective or Prospective: All testing described appears to be prospective as it's designed to evaluate the new device for market clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Human Factors Validation: "Certified or Licensed Healthcare Practitioners" were involved. The specific number or their detailed qualifications (e.g., years of experience, specific specialty) are not provided. They were the "users" in this context, and their performance likely contributed to the "ground truth" of usability.
    • Other Testing: For the bench and animal testing, the "ground truth" was established by objective measurements and assessments against predefined criteria, rather than expert consensus on interpretation. For instance, "Verification confirmed that the Clearing Stem Models met product specifications," implies internal testing against engineering specifications forming the ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any formal adjudication method (like 2+1 or 3+1 consensus) for establishing ground truth in the context of expert review, as the testing primarily involves objective performance metrics and human factors validation rather than interpretive tasks.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. This device is a mechanical tool for clearing clogs, not an AI-assisted diagnostic or interpretive system. The "Human Factors Validation and comparison of device effectiveness testing" focused on the performance of the device itself and user interaction, not on improving human "reading" or diagnostic skills with AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to the TubeClear system, as it is a mechanical device that inherently requires "human-in-the-loop" operation (an operator manually inserts and directs the clearing stem). There is no "algorithm only" component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for evaluating the TubeClear device was primarily based on:

    • Engineering Specifications/Objective Measurements: For technical aspects like product specifications, wire tip flexibility, forces, shelf life, and packaging.
    • Simulated Use Performance: For efficacy, where the ability to clear occlusions in simulated tubes was measured.
    • Direct Observation/Measurement of Physical Effects: For safety, assessing tube integrity, heating, and movement directly.
    • Controlled Animal Study: For assessing the risk of puncture in worst-case scenarios.
    • Conformance to Consensus Standards: For basic safety, usability engineering, risk management, and electromagnetic compatibility.

    8. The sample size for the training set

    This question is not applicable as the TubeClear device is a mechanical device and does not involve AI or machine learning models that require a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131052

    Validate with FDA (Live)

    Date Cleared
    2013-08-16

    (123 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TubeClear is indicated for use ONLY and SOLELY in clearing occlusions / clogs in Feeding and Decompression Tubes in adult patients that have a Tube of size 10 to 18 Fr.

    The Clearing Stem Models are indicated for use as follows:

    • NE-1036, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 91 cm (36 in).
    • NE-1042, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 107 cm (42 in).
    • NE-1043, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 109 cm (43 in).
    • NE-1045, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 114 cm (45 in).
    • NE-1048, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 122 cm (48 in).
    • NE-1050, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 127 cm (50 in).
    • NE-1055, for Nasoenteral and Nasogastric Tubes that are of size 10 18 Fr and have a length of 140 cm (55 in).
    • G-1008, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 20 cm (8 in).
    • G-1009, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 23 cm (9 in).
    • G-1010, for Gastrostomy and Jejunostomy Tubes that are of size 10 -- 18 Fr and have a length of 25 cm (10 in).
    • G-1011, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 28 cm (11 in).
    • G-1012, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 30 cm (12 in).
    • G-1014, for Gastrostomy and Jejunostomy Tubes that are of size 10 18 Fr and have a length of 36 cm (14 in).
    Device Description

    The Proposed Device, TubeClear, is comprised of Control Box Model 101 and Clearing Stem Models NE-1036, NE-1042, NE-1043, NE-1045, NE-1048, NE-1050, NE-1055, G-1008, G-1009, G-1010, G-1012, and G-1014. The Control Box is reusable and the Clearing Stems are single use. The Operator manually inserts the Clearing Stem into the feeding and/or decompression tube (i.e., Tube) and directs the Clearing Stem's progression along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear reciprocating motion. The linear reciprocating motion is transferred to the proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Mire. The motion at the Wire Tip mechanically acts on the occlusion and restores Tube patency.

    AI/ML Overview

    The provided document details the 510(k) summary for the TubeClear device, specifically focusing on the new Clearing Stem Models. The submission claims substantial equivalence to its predicate device, also named TubeClear. The core of the device is to clear occlusions in feeding and decompression tubes using a reciprocating wire.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format for each individual criterion. Instead, it describes various tests and states whether they were "successfully completed," "passed acceptance criteria," or found to be "equivalent" to the predicate device.

    Below is a synthesized table based on the descriptions provided:

    Test CategorySpecific TestAcceptance Criteria (Implied)Reported Device Performance
    Technical TestingProduct Specifications VerificationMeet product specifications"Confirmed that the Clearing Stem Models met product specifications passing all acceptance criteria."
    Transportation VibrationPass acceptance criteria for US highway truck"Transportation testing successfully passed acceptance criteria for US highway truck."
    Shelf LifePass 24-month storage"Shelf life was tested via accelerated life testing and passed 24 month storage."
    Efficacy TestingUsabilitySuccessfully reviewed by licensed healthcare practitioners"Usability testing by licensed healthcare practitioners was successfully completed."
    Effectiveness (Simulated Use)Equivalent to Predicate Clearing Stem Models"Found to be of equivalent effectiveness" to predicate devices.
    Safety TestingTube Damage (scratches, nicks, tears, abrasions, punctures)Equivalent to Predicate Clearing Stem Models"Found to be equivalent for the Proposed and Predicate Devices."
    Tube Heating during OperationNegligible and equivalent to Predicate Clearing Stem Models"Negligible and equivalent for the Proposed and Predicate Devices, thus the Devices have equivalent safety in terms of heating."
    Tube Movement during OperationSame magnitude as Predicate Clearing Stem Models"Measured and found to be of the same magnitude, thus equivalent for both Devices."
    Control Box TestingElectrical Safety (IEC 60601-1 3rd Edition)Compliance with IEC 60601-1 3rd Edition"Successfully completed in December, 2012."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for the test sets in any of the described tests (Technical, Efficacy, Safety). It mentions "Clearing Stem Models" in plural, implying multiple units were tested, but no concrete numbers.

    The data provenance is from bench testing performed by the manufacturer (Actuated Medical, Inc.) and a third-party lab (Intertek Group for electrical safety). All testing appears to be retrospective in the sense that it's performed on manufactured devices for regulatory submission. There is no mention of country of origin for the data; however, the testing entity (Intertek Group, Boxborough, MA) suggests the testing itself occurred in the US.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    For the "Usability testing," it states "Usability testing by licensed healthcare practitioners was successfully completed." However, it does not specify the number of these practitioners or their specific qualifications (e.g., years of experience, specialty). Therefore, the number and qualifications of experts for establishing ground truth are not provided.

    4. Adjudication Method for the Test Set

    The document does not describe any formal adjudication method (e.g., 2+1, 3+1, none) for the test results. The assessment of equivalence and success appears to be based on direct measurement and comparison to the predicate device or established standards, rather than expert consensus on individual cases that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly conducted or reported. This submission primarily focuses on the technical and safety equivalence of the device itself through bench testing, not on the improvement of human readers' performance with or without the device. The device is a tool used by practitioners, not an AI diagnostic aid for human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The device described, "TubeClear," is a physical medical device that requires direct human operation to insert and advance the clearing stem. It is not an algorithm or AI system that can perform without human-in-the-loop. Therefore, a standalone (algorithm-only) performance study is not applicable and was not performed. The "Control Box" provides electromechanical actuation, but the device's function is inherently human-assisted.

    7. The Type of Ground Truth Used

    The ground truth for the efficacy and safety tests appears to be based on:

    • Product Specifications: For technical verification.
    • Simulated Use Conditions: For efficacy testing, where successful clearing of occlusions in simulated tubes would be the "ground truth."
    • Measurement and Observation: For safety tests like tube damage, heating, and movement, where direct measurement of these parameters against acceptable ranges or predicate device performance constitutes the "ground truth."
    • Licensed Healthcare Practitioner Evaluation: For usability, where their feedback on the device's ease of use and performance in a simulated environment serves as the "ground truth."

    There is no mention of pathology, biological outcomes data, or patient outcomes data being used as ground truth for this submission, as it explicitly states "No clinical data was collected."

    8. The Sample Size for the Training Set

    Since "No clinical data was collected" and the device is a mechanical tool rather than an AI/ML algorithm, there is no concept of a "training set" in the context of this 510(k) submission.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML algorithm, the question of how its ground truth was established is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123659

    Validate with FDA (Live)

    Date Cleared
    2012-12-20

    (22 days)

    Product Code
    Regulation Number
    876.5980
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TubeClear™ is indicated for use only and solely in clearing occlusions / clogs in nasogastric (NG) tubes (i.e., tubes that are placed through the nose and reside in the stomach) of adult patients. Clearing Stem Model NG-1036 is indicated for use in nasogastric tubes that are of size 10 - 14 French and have a length of 36 - 42 inches (91 - 108 cm). Clearing Stem Model NG-1043 is indicated for use in nasogastric tubes that are of size 10 - 18 French and have a length of 43 - 50 inches (109 - 127 cm).

    Device Description

    The Proposed Device is TubeClear. TubeClear is composed of a reusable Control Box and single use Clearing Stem. One Control Box model. Model 101 is used to actuate all Clearing Stem models. TubeClear has two Clearing Stem models, Models NG-1036 and NG-1043.

    The Operator manually inserts the Clearing Stem into the feeding or decompression tube (i.e., Tube) and directs the Clearing Stem's advancement along the inside of the Tube. The Control Box Motor via electromechanical actuation creates a linear reciprocating motion. The linear reciprocating motion is transferred to the proximal end of the Clearing Stem which contains a Wire that also reciprocates. Because the Wire is continuous throughout the Clearing Stem, the reciprocating motion is further transferred to the distal Tip of the Wire. The motion at the Wire Tip mechanically acts on the occlusion, breaks up the occlusion and restores Tube patency.

    Because the Control Box remains outside of the patient and it functions only to provide actuation to the Clearing Stem when the two are attached, the Clearing Stem is considered to be the primary element of TubeClear.

    AI/ML Overview

    Here’s a breakdown of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Usability TestingPassed all acceptance criteria
    Retainability TestingPassed all acceptance criteria
    Operation and performance of the Control BoxUnaffected by design modification (removal of centimeter marks)
    Performance of the Clearing StemUnaffected by design modification (removal of centimeter marks)
    Shelf LifeUnaffected by design modification (removal of centimeter marks)
    BiocompatibilityUnaffected by design modification (removal of centimeter marks)
    Packaging and TransportationUnaffected by design modification (removal of centimeter marks)
    Software / FirmwareUnaffected by design modification (removal of centimeter marks)
    Electromagnetic Compatibility / Electrical SafetyUnaffected by design modification (removal of centimeter marks)

    2. Sample Size and Data Provenance for Test Set

    The document mentions "Usability and Retainability testing by end users" but does not specify the sample size for these tests or the data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth of Test Set

    The document does not specify the number of experts used to establish ground truth for the test set or their qualifications. The testing described is "Usability and Retainability testing by end users," implying a user-centric evaluation rather than expert ground truthing in a diagnostic context.

    4. Adjudication Method for Test Set

    The document does not describe an adjudication method for the test set. Given the nature of the testing (usability and retainability), a formal adjudication process as seen in diagnostic studies would likely not be relevant.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not indicate that an MRMC comparative effectiveness study was done, nor does it mention any effect size for human readers improving with or without AI assistance. This device is a mechanical medical device (TubeClear™), not an AI-powered diagnostic tool, so such a study would not be applicable.

    6. Standalone (Algorithm Only) Performance Study

    The document does not describe a standalone performance study. This device is a mechanical clearing system, not an algorithm, so a "standalone" or "algorithm only" performance study is not applicable.

    7. Type of Ground Truth Used

    The ground truth for the usability and retainability testing was based on "acceptance criteria" being met by "end users." This implies a subjective or objective assessment of the device's functionality and user interaction rather than a diagnostic accuracy ground truth like pathology or outcomes data.

    8. Sample Size for Training Set

    The document does not mention a training set as this device is a mechanical medical device and does not involve machine learning or AI models that require training data.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no training set for this type of medical device.