Search Results
Found 1 results
510(k) Data Aggregation
(218 days)
The device is intended to be used with endoscopic accessories such as a biopsy forceps and Electronic Endoscope Image Processor for endoscopy and endoscopic surgery within urinary tract and interior of the ureter and kidney.
The Flexible Ureterorenoscope System consists of a sterile Flexible Ureterorenoscope and an Electronic Endoscope Image Processor with its accessories power adapter. The Flexible Ureterorenoscope is provided sterile (sterilized by EO) and is single use.
The Flexible Ureterorenoscope consists of a head section, bending section, insertion section, operating section, electrical connection section, and built-in LED light source. The Electronic Endoscope Image Processor is powered by the main line.
The Flexible Ureterorenoscope is an endoscope instrument and is intended to be used with the Electronic Endoscope Image Processor during endoscopic procedures. The handle section features several key components, including a control lever, a handle for manipulation, a button for capturing photos or video, and a Luer port for the insertion of accessories or irrigation to the working channel.
The insertion section houses a working channel and wiring to transmit the image signals to the image processor. The distal bending section of the insertion section contains a camera lens for real-time image capture, an LED light source for illumination, and a distal outlet for the working channel.
The Electronic Endoscope Image Processor is a video processor which is intended and designed for use during endoscopic procedures. It receives and processes image signals from the endoscope and outputs signals to an external medical monitor for display.
The Electronic Endoscope Image Processor has integrated software to control the operation of the device and is a reusable device that does not require sterilization before use, as there is no direct/in-direct patient contact material. The image processor is intended to be cleaned and disinfected before use, the cleaning and disinfection method is validated and recommend in the User Manual.
The equipment is for use in a hospital or qualified medical institution. It is only to be used by skilled physicians trained in clinical endoscopic techniques and procedures.
N/A
Ask a specific question about this device
Page 1 of 1