Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243905
    Date Cleared
    2025-08-14

    (238 days)

    Product Code
    Regulation Number
    888.3010
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries as bone fixation cerclage, specifically:

    • Sternotomy indications including the "rewiring" of osteomized sternums;
    • Repair of long bone fractures due to trauma or reconstruction.
    Device Description

    Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries for internal fixation of discontinued bone structures. The product are not to be used in conjunction with other bone fixation implants, like plates. The surgical sutures and tapes are made of UHMWPE (ultra-high molecular weight polyethylene), the needles are made in AISI300 steel series, according to the standard ASTM F899 and cerclage is assembled on an ABS loader, additional instruments may be used as well. Those materials are identical to those used in other cleared GMReis devices like the reference device cleared under K223114. It is provided in sterile condition sterilized by Ethylene Oxide.

    AI/ML Overview

    Based on the provided FDA 510(k) clearance letter for the Stitch Cerclage - Suture System, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    Important Note: The provided document is an FDA 510(k) clearance letter, which means the device has already been cleared. This letter summarizes the regulatory decision and mentions that studies were performed, but it does not contain the detailed study protocols, results, or data itself. Therefore, I will extract all available information and explicitly state when information is not present in this document.


    Acceptance Criteria and Device Performance Study Summary

    The acceptance criteria for this device are largely implied through the comparison to a predicate device and the successful completion of specified mechanical tests. The performance study focused on demonstrating substantial equivalence to the predicate device, particularly in mechanical properties relevant to its intended use as a bone fixation cerclage.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Material Equivalence: Made of materials safe and equivalent to predicate devices.The device is made of UHMWPE (ultra-high molecular weight polyethylene) for sutures/tapes and AISI300 steel series for needles, which are identical to those used in other cleared GMReis devices (Reference Device K223114) and are considered equivalent to the predicate device.
    Design Equivalence: Shares equivalent design characteristics with predicate devices.The device has “equivalent design characteristics as well as physical dimensions” compared to the predicate. Minor differences "do not raise new issues of safety or efficacy."
    Mechanical Performance: Performance is comparable or superior to the predicate device in relevant mechanical tests (chemical composition, creep, tension static, fatigue).Performance was demonstrated through mechanical testing (chemical composition, creep test, tension static, and fatigue tests). The specific results are not provided in this document, but the conclusion states: "We therefore consider the subject device equivalent to its predicate device in the mechanical tests performed." The methodology was based on the article "Cerclage Performance Analysis – a Biomechanical Comparison of Different Techniques and Materials", L. M. Hägerich, et al, Musculoskeletal Disorders (2022).
    Sterility: Provided in a sterile condition via a validated method.The device is provided in sterile condition, sterilized by Ethylene Oxide, which is a common and accepted sterilization method.
    Biocompatibility: Materials are biocompatible for implantation.While not explicitly detailed as a separate test, the use of UHMWPE and AISI300 steel, which are "identical to those used in other cleared GMReis devices" and are common in medical devices, implies that their biocompatibility has been previously established and met.
    Substantial Equivalence: Overall device is substantially equivalent to legally marketed predicate devices in terms of indications for use, technological characteristics, and safety/effectiveness.The submission concludes that the device is "substantially equivalent" to the predicate in "intended use, indications for use, material composition, anatomical region, multiple sizes, and basic design features." The mechanical testing supported this claim, with no new issues of safety or effectiveness identified.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes (e.g., number of test specimens, number of cycles for fatigue) used for the mechanical testing. It mentions "chemical composition, creep test, tension static and fatigue tests."
    • Data Provenance: The studies were conducted by the manufacturer, GM Dos Reis Industria e Comercio Ltda, which is based in Campinas, Brazil. The data is based on mechanical testing, not clinical data, and therefore is not characterized as retrospective or prospective in the usual clinical sense. It's laboratory-based performance data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • For this type of mechanical device clearance, "ground truth" is established through engineering standards, validated test methodologies, and comparison to the known performance of predicate devices.
    • The document does not mention the use of human experts (like radiologists) for "ground truth" establishment as would be common for AI/imaging device clearances. The "experts" involved would be engineering and quality control personnel at the manufacturer and potentially external testing laboratories, ensuring compliance with test methodologies.

    4. Adjudication Method for the Test Set

    • Since this is a mechanical device performance study, no explicit "adjudication method" in the sense of clinician consensus on medical images is applicable or mentioned. The "adjudication" is inherent in the adherence to established mechanical testing protocols and statistical analysis of the results to demonstrate equivalence to the predicate.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This type of study is specifically relevant for imaging devices or AI algorithms where human readers' performance is being evaluated or augmented. The Stitch Cerclage Suture System is a physical medical device, not an imaging or AI diagnosis system.
    • The document explicitly states: "No clinical data were included in this submission."

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • Not applicable. This concept pertains to AI algorithms. The Stitch Cerclage Suture System is a physical product and does not have an "algorithm-only" performance component.

    7. Type of Ground Truth Used

    • The "ground truth" for this device type is established through objective, quantifiable mechanical properties and material composition, compared against established industry standards (where available) or the performance of a legally marketed predicate device.
    • Specifically, the study compared "chemical composition, creep test, tension static and fatigue tests" to information from a published article ("Cerclage Performance Analysis – a Biomechanical Comparison of Different Techniques and Materials", L. M. Hägerich, et al, Musculoskeletal Disorders (2022)) and to the predicate device.

    8. Sample Size for the Training Set

    • Not applicable / Not mentioned. This device is a physical product, not an AI algorithm. Therefore, there is no "training set" in the context of machine learning. The design and manufacturing process would involve internal validation and verification activities, but this is distinct from training an AI model.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no AI "training set," there is no ground truth established for one. The "ground truth" for the device's design and manufacturing is derived from engineering principles, relevant material standards (e.g., ASTM F899 for steel), and performance expectations set by the predicate device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1