510(k) Data Aggregation

    K Number
    K243031
    Date Cleared
    2024-10-25

    (28 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C310S, C316S and C616W are indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The displays are not intended for mammography.

    G310S and G316S are intended to be used in displaying digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The devices are not specified for digital mammography systems.

    Device Description

    3MP LCD Monitors C310S, G310S, C316S, G316S are 21.3-inch TFT LCD monitors, which are specifically designed to provide the high definition image output for general radiography. These products have been strictly calibrated so that they can meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panel, supporting resolution 1536 x 2048. The built-in brightness stabilization control circuit makes sure the brightness of these monitors is stable, so these products meet the demand of high precision medical imaging. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer. 6MP Color LCD Monitor C616W is a 30-inch TFT LCD monitor, which is specifically designed to provide the high definition image output for general radiography.

    This product has been strictly calibrated so that they can meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panel, supporting resolution 3280 x 2080. The built-in brightness stabilization control circuit makes sure the brightness of these monitors is stable, so this product meets the demand of high precision medical imaging. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

    C310S and C316S have the same color LCD panel and similar functionality. For C310S and C316S, the difference is the main board and design principle.

    G310S and G316S have the same monochrome LCD panel and similar functionality. For G310S and G316S, the difference is the main board and design principle.

    C310S and G310S have the same main board and design principle.

    C316S and G316S have the same main board and design principle.

    AI/ML Overview

    The provided text describes a 510(k) summary for several LCD medical displays (C310S, G310S, C316S, G316S, C616W) which are claimed to be substantially equivalent to a previously cleared device (K240310). The information does not detail an AI-powered device or a study involving human readers or comparative effectiveness for AI assistance. Instead, it focuses on the technical specifications and performance of the LCD monitors.

    Therefore, the following information is extracted and parts that are not relevant to an AI/ML powered device or are not provided in the text are explicitly stated as such.


    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implied by the "Description" column, indicating what was measured, and the "Test result" column, which uniformly states "PASS". This suggests that the measured performance met the predetermined criteria for each test.

    MeasurementsDescriptionTest resultAcceptance Criteria (Implied)
    1. Spatial resolutionMeasurement of spatial resolution expressed as modulation transfer function (MTF)PASSMeet predefined MTF standards
    2. Pixel defectsThe maximum number allowed for each type of pixel defects/faultsPASSNot exceed maximum allowed pixel defects/faults
    3. ArtifactsVisual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guidelinePASSAbsence of miscellaneous artifacts as per TG18
    4. Temporal responseMeasure the temporal response using the typical data provided by the panel manufacturerPASSMeet manufacturer's typical temporal response data
    5. Luminance (maximum, minimum, achievable, and recommended)Meausure the maximum, minimum, achievable, and recommended luminancePASSMeet specified luminance ranges
    6. Conformance to a grayscale-to luminance functionVerification of the conformance to DICOM GSDF as specified in Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)PASSConformance to DICOM GSDF as per TG18
    7. Color tracking (primary colors and color gamut)Measurement of Color tracking (primary colors and color gamut)PASSMeet specified color tracking and gamut standards

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The tests performed are physical laboratory tests on the display embedded software using DBI Calibration Feedback System on the monitors, not a test on a test set of medical cases.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided as the device is an LCD monitor, not an AI / image analysis software requiring expert annotation for ground truth. The tests are technical performance evaluations of the display.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided as it's not relevant for the type of device being described (an LCD monitor).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an LCD monitor, not an AI or AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone performance study was not done as this device is an LCD monitor, which is a display hardware, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The ground truth for the performance tests of the LCD monitors is based on pre-determined technical criteria and industry standards (e.g., DICOM Part 3.14, AAPM Task Group 18 guideline, manufacturer's typical data). This is not a "ground truth" derived from medical diagnosis but from established engineering and display performance metrics.

    8. The sample size for the training set

    This information is not applicable as the device is an LCD monitor and does not involve AI/ML models that require a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is an LCD monitor and does not involve AI/ML models that require a training set and ground truth generation.