Search Results
Found 1 results
510(k) Data Aggregation
(78 days)
i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications.
- Repair of Root Perforation
- Repair of Root Resorption
- Root End Filling
- Apexification
- Pulp Capping
i-MTA BP Bioceramic Root Canal Sealer and Repair Materials is a ready-to-use premixed bioceramic paste developed for permanent root canal sealer and repair and surgical applications. i-MTA BP is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. i-MTA BP does not shrink during setting and demonstrates excellent physical properties. i-MTA BP is packaged in a preloaded syringe.
Here's a breakdown of the acceptance criteria and the study details for the Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP), based on the provided FDA 510(k) summary:
This device is not an AI/ML device, therefore, some of the requested information (like effect size of human readers with vs. without AI, number of experts for ground truth, adjudication methods, details on training set) are not applicable. The information provided relates to a traditional medical device's performance against established standards.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Requirement) | Standard/Guideline | Reported Device Performance (i-MTA BP) |
|---|---|---|
| Physical/Chemical Properties: | ||
| Setting time | ISO 6876:2012 | 2.5h (within 2h-6h) |
| Solubility | ISO 6876:2012 | 2.80% (< 3%) |
| Radio-opacity | ISO 6876:2012 | > 3mm Al (> 3mm Al) |
| Composition (visual foreign objects) | Technical Specifications | Compliant with requirements |
| Content (of indicated loading capacity) | Technical Specifications | Compliant with requirements (>93%) |
| Size change after curing | Technical Specifications (ISO6876:2001) | -0.50% (< 1.0%) |
| pH | Technical Specifications | 12.29 (within 11.5-13.5) |
| Acid-soluble arsenic content | Technical Specifications | Compliant with requirements (< 2mg/kg) |
| Acid-soluble lead content | Technical Specifications | Compliant with requirements (< 100mg/kg) |
| Heavy metal content | Technical Specifications | < 100 µg/g (< 100 µg/g) |
| Biocompatibility: | ||
| Cytotoxicity | ISO 7405:2018 | Compliant |
| Sensitization | ISO 10993-10:2021 | Compliant |
| Irritation | ISO 10993-10:2021 | Compliant |
| Acute Systemic Toxicity | ISO 10993-11:2017 | Compliant |
| TK Gene Mutation Test | ISO 10993-3:2014 | Compliant |
| Bacterial Reverse Mutation Test | ISO 10993-3:2014 | Compliant |
| Subchronic Systemic Toxicity | ISO 10993-11:2017 | Compliant |
| Implantation | ISO 10993-6:2016 | Compliant |
| Pulp Capping Testing | ISO7405:2018 | Compliant |
| Endodontic Usage Test | ISO 7405:2018 | Compliant |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific performance test (e.g., how many samples were tested for setting time, solubility, etc.). The provenance of the data is not specified beyond being generated from "Performance testing - Bench" and "Biocompatibility Testing" which implies laboratory testing directly on the subject device. There is no indication of retrospective or prospective clinical data from specific countries for these bench tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable to the type of device and testing presented. The "ground truth" for this device is based on objective measurements against international and technical standards, not expert consensus on medical images or diagnoses.
4. Adjudication Method for the Test Set
Not applicable. The tests involve objective measurements against predefined criteria, not subjective interpretation requiring adjudication among experts.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is typically for diagnostic imaging devices where human readers interpret medical images. This device is a dental material, and its performance is evaluated by bench testing and biocompatibility.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the device's performance was evaluated in a standalone manner (without human-in-the-loop performance) through bench testing and biocompatibility studies. As it is a physical material, there is no "algorithm" involved, but its intrinsic properties were tested directly.
7. The Type of Ground Truth Used
The ground truth used for performance validation is objective measurement against established international standards (e.g., ISO 6876:2012, ISO 10993 series, ISO 7405:2018) and internal Technical Specifications.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1