(175 days)
The BD Texium™ Closed Male Luer (CML) is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer and administration of hazardous drugs when paired with the SmartSite™ NFC.
The BD Texium™ Closed Male Luer (CML) is indicated for use by trained healthcare professionals within healthcare facilities who prepare and/or administer non-hazardous drugs for adults, pediatrics and neonates.
The BD Texium™ CML is an airtight, leak-free and drip-free closed system drug transfer device (CSTD). When paired with devices containing a SmartSite™ NFC, the BD Texium™ CML mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the BD Texium™ CML/SmartSite™ NFC connection, thereby minimizing individual and environmental exposure to drugs, leaks and spills. The BD Texium™ CML is a sterile single-use CSTD intended for the reconstitution, transfer and administration of hazardous and non-hazardous drugs when paired with the SmartSite™ NFC.
This document describes the premarket notification (510(k)) for the BD Texium™ Closed Male Luer (CML). It is a sterile, single-use closed system drug transfer device (CSTD) intended for the reconstitution, transfer, and administration of hazardous and non-hazardous drugs when paired with the SmartSite™ Needle-Free Connector (NFC).
Here's an analysis of the acceptance criteria and study data:
Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria alongside performance data in a dedicated section. However, the "Technological Characteristics and Substantial Equivalence" table outlines differences and similarities between the subject device (BD Texium™ CML) and the predicate device (Alaris® Safety Male Luer, K053049) and indicates where testing was conducted to verify new claims or differences. The "Substantial Equivalence Discussion" and "Discussion of Non-Clinical Tests" sections detail the types of tests performed and state that "All test results met their acceptance criteria".
Below is a reconstructed table based on the provided text, indicating acceptance criteria implicitly by reference to the predicate or standards, and the reported performance as meeting those criteria.
| Feature / Test Category | Acceptance Criteria (Implicit from Predicate/Standards) | Reported Device Performance |
|---|---|---|
| FDA Regulation Number | 21 CFR 880.5440 (Intravascular Administration Set) | Same as Predicate |
| FDA Regulation Name | Intravascular Administration Set | Same as Predicate |
| FDA Class | Class II | Same as Predicate |
| Principle of Operation | Airtight, leak-free, drip-free, mechanically prohibits transfer of environmental contaminants and escape of drug vapor (similar to predicate) | Equivalent; air leakage, vacuum leakage, fluid leakage, and residual fluid testing verified new claims. |
| Device Compatibility | Use with SmartSite™ Needle-Free Connector (similar to predicate's SmartSite™ Needle Free Valve port) | Equivalent; device to be used with SmartSite™ Needle-Free Connector. |
| Method of Administration | Closed system drug transfer device (CSTD) | Same as Predicate |
| Non-DEHP | Yes | Same as Predicate |
| Device Components/Materials | Biocompatible and functionally equivalent materials (predicate materials: Polycarbonate, Polypropylene, TPE, Silicone, Fluorosilicone Fluid, Acetal, PTFE). Verification through biocompatibility testing, ISO 80369-7, air/vacuum/fluid leakage. | Different materials, but biocompatibility testing conducted to assess new materials; female luer body design changes verified through ISO 80369-7; actuator and male luer body changes verified through air/vacuum/fluid leakage. |
| Packaging | Maintain sterility and integrity (similar to predicate packaging). Verification through packaging validation. | Different packaging, but packaging validation verifies new webs. |
| No natural rubber latex | Yes | Same as Predicate |
| Sterilization Method | Irradiation | Same as Predicate |
| Sterilization Claim | Content Sterile (predicate claims Fluid Path Sterile). Verification through package integrity testing. | Different claim, but package integrity testing (seal strength, corner thickness, seal width, air volume, microbial barrier, dye test, bubble leak testing) verified sterile barrier claim. |
| Biocompatibility | Biocompatible for intended use per ISO 10993-1 and related sub-standards (similar to predicate). | Same as Predicate; specific testing per ISO 10993-1, -3, -4, -5, -10, -11, -17, -18, -23. |
| Non-Pyrogenic | Yes | Same as Predicate |
| Duration of Use | 7 days | Same as Predicate |
| Disinfect with 70% IPA | Disinfect with 70% Isopropyl Alcohol | Same as Predicate |
| Priming Volume | ≤ 0.2 ml maximum (predicate specification) | 0.17 ml (within predicate specification) |
| Flow Rate | ≥ 4756 ml/hr (predicate specification) | >4280 ml/hr, when activated with min 3.2 mm insertion depth (flow rate conducted to verify rate). Implicitly meets internal specification, though slightly lower than predicate. |
| Shelf Life | 3 Years | Same as Predicate; data supports 3 years. |
| Particulate Testing | Meets requirements of USP, General Chapter <788>, Particulate Matter in Injections. | Device meets particulate requirements. |
| Microbial Ingress | Performance demonstrated according to FDA guidance. | Additional testing conducted for harsh infusates. |
Study Details
The document describes non-clinical testing to demonstrate substantial equivalence to a predicate device.
-
Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
- The document does not specify the exact sample sizes for each non-clinical test (e.g., number of devices tested for leakage, biocompatibility, flow rate).
- The data provenance is not explicitly stated regarding country of origin or whether it was retrospective or prospective. Given it's a premarket submission for a new device design and material changes, the testing is prospective and conducted by the manufacturer, CareFusion/BD.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- This is a non-clinical device test for physical and chemical properties, not a clinical study involving diagnosis or interpretation by human experts. Therefore, the concept of "experts establishing ground truth" as it applies to AI/clinical diagnostic studies is not relevant here. Ground truth is established by standardized test methods and measurement equipment.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable as this is not a study involving human interpretation or clinical adjudication. Test results are objective measurements against predefined criteria from standards.
-
If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done. This is a non-clinical device submission, not an AI/software as a medical device (SaMD) product.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm or AI.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the non-clinical tests is established by recognized international and national standards (e.g., ISO, USP, ASTM, ANSI/AAMI) and the performance of a legally marketed predicate device. For example:
- Biocompatibility: Conformance to ISO 10993 series of standards.
- Sterilization: Conformance to ISO 11137 series and USP chapters.
- Packaging: Conformance to ISO 11607 series and ASTM standards.
- Performance (Leakage, Flow Rate, etc.): Conformance to ISO 80369 series and ISO 8536-4, as well as showing performance equivalent to or better than the predicate device.
- Particulate: Conformance to USP <788>.
- The "ground truth" for the non-clinical tests is established by recognized international and national standards (e.g., ISO, USP, ASTM, ANSI/AAMI) and the performance of a legally marketed predicate device. For example:
-
The sample size for the training set:
- Not applicable. This is a physical device, not an AI/ML model that requires a training set. The device design and materials are developed and then validated through testing.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set for a physical device.