Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222386
    Device Name
    Visionbody
    Manufacturer
    Date Cleared
    2023-02-24

    (200 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Visionbody System is a device intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is to be used by adults only.

    The Visionbody System is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of the training programs or operational parameters are designed or ailing muscles and its use on such muscles is contraindicated.

    The Visionbody System's electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generated mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

    Device Description

    The Visionbody System is a wireless electrical muscle stimulation device for health and fitness applications. The main components are the Visionbody Box, the Visionbody Software and the choice of a Visionbody garment, whether a Visionbody Powersuit, Visionbody PowerPants or Visionbody Power Vest, dependent on a user's personal choice and needs. The Visionbody Box delivers a safe low to mid frequency electrical current to the electrodes in a Visionbody garment. The Visionbody uses electrical muscle stimulation (EMS) and electrical muscle activation (EMA) technologies to provide a streamlined approach to a user's fitness goals, especially regarding body composition, strength and endurance. Underlying to both these technologies are safe electrical impulses, which stimulate muscle contractions equating to muscle workouts.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Visionbody system, a powered muscle stimulator. It details a comparison with a predicate device (Katalyst Training System Model 1) and two reference devices (FM-B2403A EMS and R-E1 EMS).

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative manner as one might find for diagnostic performance (e.g., sensitivity, specificity thresholds). Instead, the acceptance criteria are implicitly demonstrated through compliance with various safety and performance standards and by showing substantial equivalence to a legally marketed predicate device.

    The "reported device performance" is primarily the device's compliance with safety standards and its functional characteristics. The comparison table directly addresses this.

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Visionbody®)Comparison to Predicate/Reference
    Intended UseStimulate healthy muscles to improve/facilitate muscle performance; for adults only; not for therapy/medical conditions.Same as predicate.Similar
    ClassificationClass II, Powered muscle stimulator (21 CFR 890.5850), Product Code NGX.Class II, Powered muscle stimulator (21 CFR 890.5850), Product Code NGX.Similar
    Compliance with Voluntary StandardsCompliance with relevant IEC and ISO standards for electrical safety, EMC, and biocompatibility.Yes (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 10993-1, ISO 10993-5, ISO 10993-10, IEC 62366).Similar to predicate (both comply with core electrical safety and EMC standards).
    Software Level of ConcernModerate (failure leads to delayed medical care).Moderate level of concern.Not directly compared, but implies software validation and verification successfully addressed this.
    Maximum Output VoltageWithin safety limits of IEC 60601-2-10.60 V @ 500Ω, 60 V @ 2kΩ, 60 V @ 10kΩLower than predicate (Katalyst: 60 V @ 500 Ω, 100 V @ 2 kΩ, 100 V @ 10 kΩ). Deemed safe as both comply with IEC 60601-2-10 limits.
    Maximum Output CurrentWithin safety limits of IEC 60601-2-10.120 mA ±10% @ 500Ω, 30 mA ±10% @ 2kΩ, 6 mA ±10% @ 10kΩSimilar to predicate @ 500Ω (Katalyst: 120 mA @ 500 Ω). Lower than predicate @ 2kΩ and 10kΩ. Deemed safe/effective.
    Pulse WidthAcceptable range for muscle stimulation.150 - 500 µsDifferent from predicate (Katalyst: 250 - 375 µs). Both support adjustable pulse width, within acceptable ranges.
    Frequency RangeAcceptable range for muscle stimulation.4 - 100 HzSimilar to predicate (Katalyst: 5 - 105 Hz) and reference devices (R-E1: 1-110 Hz, FM-B2403A: <199 Hz).
    Maximum Phase ChargeWithin safety limits.30 µC @ 500 ΩLower than predicate (Katalyst: 45 µC @ 500 Ω). Similar to reference devices (R-E1: 22.8µC@500 Ohm, FM-B2403A: 35uC@500 Ohm). Deemed safe.
    Maximum Current DensityWithin safety limits.2.17 mA/cm² @ 500 ΩHigher than predicate (Katalyst: 1.15 mA/cm² @ 500Ω). Similar to reference device R-E1 (2.4 mA/cm²), and higher than FM-B2403A (0.5645 mA/cm²). Deemed safe as it's comparable to other cleared devices.
    Maximum Power DensityWithin safety limits.6.52 mW/cm² @ 500ΩLower than predicate (Katalyst: 22.68 mW/cm² @ 500Ω). Deemed safe.
    Safety FeaturesComprehensive safety mechanisms.Short-circuit, watchdog, no-load, overload trip, output current, connector, battery voltage, hardware error/manipulation monitoring, integrity/authenticity checking of binary code.More detailed list than predicate (Katalyst only lists no-load trip, overload trip, no more publicly available). Implies enhanced safety.
    EffectivenessPerforms comparably to predicate.Demonstrated comparable performance.No new questions raised regarding effectiveness due to similar intended use, functionality, operating principle, and comparable stimulation parameters to cleared devices.

    Summary of the study that proves the device meets the acceptance criteria:

    The study is a non-clinical product evaluation intended to demonstrate safety and effectiveness for substantial equivalence. It is primarily a comparison of technical characteristics and compliance with recognized standards rather than a traditional clinical trial or performance study with a test set of data.

    2. Sample size used for the test set and the data provenance

    There is no mention of a "test set" in the context of human data or any specific patient or subject data being used for device performance evaluation in this document. The evaluation is based on non-clinical testing of the device itself against established engineering and safety standards, and a comparative analysis against predicate and reference devices' technical specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As there is no human "test set" or diagnostic evaluation mentioned, there is no ground truth established by experts in that context. The "truth" here is compliance with technical specifications and safety standards.

    4. Adjudication method for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a powered muscle stimulator, not an AI-powered diagnostic imaging system that would involve human readers or AI assistance in interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The term "standalone" typically refers to the performance of an AI algorithm independent of human input or review. While the Visionbody system has software, its purpose is to control electrical muscle stimulation, not to perform diagnostic or interpretative tasks where "standalone" performance metrics for AI would apply. The device's "standalone" performance would be its functional operation in stimulating muscles according to its programmed parameters, which was evaluated through performance testing.

    The document states: "Software verification and validation testing were conducted on the Visionbody system and documentation was provided as recommended by FDA's Guidance... The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software would lead to a delayed delivery of appropriate medical care." This confirms standalone software testing was performed to ensure its functionality and safety.

    7. The type of ground truth used

    The ground truth used for this type of device (a powered muscle stimulator) is primarily:

    • Compliance with recognized international standards: Electrical safety (IEC 60601-1, IEC 60601-2-10), Electromagnetic Compatibility (IEC 60601-1-2), Biocompatibility (ISO 10993 series), and Usability/Risk Management (IEC 62366).
    • Manufacturer's own requirements and specifications: Performance testing verified the device against its own design requirements.
    • Predicate device characteristics: Similarity to a legally marketed predicate device (Katalyst Training System Model 1) serves as a benchmark for substantial equivalence in safety and effectiveness.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a "training set" of data in the typical sense for model development. The software is for device control, not for learning or prediction based on data.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1