Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222151
    Date Cleared
    2023-03-23

    (246 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used up to 365 days post sterilization.

    Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:

    a) 100% ethylene oxide (EO) with a concentration of 725 - 735 mg/l at 131 F/55°C and 40%-80% relative humidity for 60 minutes.

    b) Exhausting the EO gas vacuum depth: 157 mBar to 160 mBar

    c) Aeration time: 12 hours

    d) Aeration temperature: 55℃

    e) Aeration pressure : 841 mBar to 864 mBar

    Device Description

    Sterilization Wrap is single use, non-sterile provide. It is divided into single and double layers, with double layer edges closed by ultrasonic suture. The sterilization Wrap is made from 100% polypropylene spunbond- meltblown - spunbond (SMS), not made with natural rubber latex.

    AI/ML Overview

    The provided text describes the regulatory clearance of a "Sterilization Wrap" device (K222151) and its comparison to a predicate device (K181174). The information focuses on the physical, chemical, and sterilization properties of the wrap. The document does not contain information related to a study involving an AI-powered medical device or human readers for image interpretation. Therefore, I cannot provide details on the sample size for a test set, expert qualifications, or MRMC studies, as these concepts are not applicable to the provided document.

    However, I can extract the acceptance criteria and performance data for the Sterilization Wrap based on the provided text, as it relates to the device's functional and safety aspects.

    Here's the analysis based on the provided text:


    Analysis of Sterilization Wrap Acceptance Criteria and Performance

    The "Sterilization Wrap" device (K222151) has undergone non-clinical testing to demonstrate its performance and establish substantial equivalence to a predicate device (K181174). The acceptance criteria and reported device performance are generally based on meeting recognized industry standards for medical device sterilization wraps.

    1. Table of Acceptance Criteria and Reported Device Performance

    ItemAcceptance CriteriaProposed Device PerformancePredicate Device (K181174) PerformanceRemark (from document)
    Sterilant Penetration and Sterilization EfficacyAchieving a 10⁻⁶ sterility assurance level (e.g., negative for growth)Negative for growthNegative for growthSAME
    Maintenance of package sterilityMaintain sterility for up to 365 DaysNegative for growth (tested for 365 days)Negative for growth (tested for 365 days)SAME
    Residuals (Ethylene Oxide)Meet ISO 10993-7Meet ISO 10993-7 (None detected)Meet ISO 10993-7SAME
    Biocompatibility (Cytotoxicity)Non-cytotoxic (ISO 10993-5)Non-cytotoxicNon-cytotoxicPass
    Biocompatibility (Irritation)Non-irritant (ISO 10993-10)Non-irritantNon-irritantPass
    Biocompatibility (Sensitization)Non-sensitizer (ISO 10993-10)Non-sensitizerNon-sensitizerPass
    Material Compatibility (EO sterilization process)Compatible to EO sterilization processCompatibleCompatibleSAME
    Tensile StrengthComplies with selected physical properties (ASTM D 5034 -09)Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    Bursting StrengthComplies with selected physical properties (ASTM D3786/D3786M-18)Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    Water Resistance Hydrostatic Pressure TestComplies with selected physical properties (AATC127-18)Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    Air PermeabilityComplies with selected physical properties (ASTM D737-18)Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    Mass per Unit AreaComplies with selected physical properties (ASTMD37776/D3776M-20)Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    Tearing StrengthComplies with selected physical properties (ASTM D5587-15)Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    DimensionMeet ISO 11607:2019Passed (before and after EO sterilization)(Not explicitly stated but implied by "SAME" for material properties)Pass
    Lint Generation TestingCoefficient of linting log₁₀ ≤ 4.0 (EN 13795-1:2019)≤4.0 (before and after EO sterilization)(Not explicitly stated)Pass
    Microbial Barrier TestProduct has good anti-bacterial permeability of the packaging materialNegative for growth(Not explicitly stated but implied by "Maintenance of package sterility")Pass
    Shelf LifeShelf Life 1 yearAfter 1 year accelerated aging, meets requirements and no significant change from 0 year's test results.Not obtained (for predicate)Passed

    2. Sample size used for the test set and the data provenance

    The document refers to various standard tests (e.g., ASTM, ISO standards) for physical properties, sterilization validation, and biocompatibility. The specific sample sizes for each of these non-clinical tests are not provided in the document.

    The data provenance is from non-clinical tests conducted by the manufacturer of the proposed device, Kenpax International Limited. The country of origin of the data is not explicitly stated beyond the manufacturer's address in Hong Kong and the contact person's address in Beijing, China. The studies are prospective in nature, as they involve testing the device to verify its performance against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The "ground truth" for this device is established through quantitative measurements and adherence to international testing standards (e.g., sterility assurance level, physical property thresholds) rather than expert consensus on interpretation, as might be the case for an AI diagnostic device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation or subjective assessments, not for non-clinical, objective performance testing of a physical product like a sterilization wrap.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device is a sterilization wrap, not an AI-powered diagnostic tool, and therefore, an MRMC study with human readers is irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical product, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device is based on pre-defined, quantitative performance metrics and established international standards. For example:

    • Sterility: Achieved a 10⁻⁶ sterility assurance level (objective microbial growth testing).
    • Physical Properties: Meeting specific tensile strength, bursting strength, water resistance, air permeability, mass per unit area, tearing strength, and dimension parameters (objective physical measurements).
    • Biocompatibility: Absence of cytotoxicity, irritation, and sensitization (objective biological tests against ISO standards).
    • Residuals: Levels of ethylene oxide residuals below specified thresholds (objective chemical analysis).

    8. The sample size for the training set

    This question is not applicable. The device is a physical product, and there is no "training set" in the context of machine learning or AI models. Its design and manufacturing process would involve internal validation and quality control, but not a "training set" for an algorithm.

    9. How the ground truth for the training set was established

    This question is not applicable. As there is no training set for an AI model, the concept of establishing ground truth for it does not apply.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1