    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SCOPE BUDDY PLUS Endoscope Flushing Aid is an electro-mechanical device intended to pump fluids through the channels of flexible, immersible endoscopes during the manual cleaning process of endoscope reprocessing.

    Device Description

    The subject device is intended to provide fluid delivery to endoscope channels in the same manner as a manual syringe would be used during the manual cleaning phase of endoscope reprocessing as defined by the endoscope manufacturer's instructions. The subject device utilizes an external peristaltic pump and 24-hour multi-use connection tubing (tubing used for 24 hours with multiple endoscopes and then discarded) to deliver specific volumes of detergent solution and rinse water to the endoscope channels during the manual cleaning phase. The minimum fluid volumes are defined by the endoscope manufacturer. SCOPE BUDDY PLUS Endoscope Flushing Aid is intended to be used only during the manual cleaning phase of endoscope reprocessing, in conjunction with the instructions and labeling provided by the endoscope manufacturer, to deliver a volume of fluid to the endoscope channels which exceeds the minimum fluid volumes specified by the endoscope manufacturer. The subject device is not an endoscope washer-disinfector and does not bear labeling claims for direct cleaning efficacy or high-level disinfection efficacy for endoscope reprocessing.

    AI/ML Overview

    The provided text describes the acceptance criteria and a summary of performance data for the SCOPE BUDDY PLUS Endoscope Flushing Aid.

    Here's an analysis of the requested information:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance CriteriaReported Device Performance
    Fluid Delivery Flow Rate VerificationMeet manufacturers minimum requirements for individual channel flushing volumes.PASS
    Tubing Durability VerificationThe 24-hour tubing shall withstand at least 40 full automatic cycles without leaking.PASS
    Software VerificationVerification steps outlined in procedure must be met for acceptance of the protocol.PASS

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only states that "Performance testing to demonstrate substantial equivalence to the predicate has been completed."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests described (Fluid Delivery Flow Rate, Tubing Durability, Software Verification) appear to be engineering/performance tests rather than tests requiring expert human interpretation or ground truth establishment based on clinical observations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/not provided as the tests described are technical performance verifications and do not involve human adjudication for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. The device is an electro-mechanical flushing aid for endoscopes, not an AI or diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not applicable to this device. It is an electro-mechanical device with software control, not an algorithm performing a standalone diagnostic or interpretative function. The software verification confirms the software's functionality within the device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the tests appears to be engineering specifications and defined performance metrics.

    • For Fluid Delivery Flow Rate, the ground truth is the "manufacturers minimum requirements for individual channel flushing volumes."
    • For Tubing Durability, the ground truth is the specified requirement to "withstand at least 40 full automatic cycles without leaking."
    • For Software Verification, the ground truth is the "verification steps outlined in procedure."

    These are objective, measurable criteria rather than clinical ground truth established by experts.

    8. The sample size for the training set

    The document does not mention a "training set" as this device is not a machine learning or AI model.

    9. How the ground truth for the training set was established

    This question is not applicable as there is no mention of a training set.