Search Results
Found 1 results
510(k) Data Aggregation
(169 days)
The Broncho videoscope system is designed to be used for endoscopic diagnosis and therapies within the respiratory system such as trachea, bronchi, and lungs. The Broncho videoscope System is for use in a hospital environment.
This Endoscopic Image Processor is used for endoscopic diagnosis and therapies. It connects to the electronic endoscopes, displaying the images on the monitor detected within the field of view from the body cavity.
This Single-use Broncho Videoscope is intended to use in conjunction with endoscopic image processor (HDVS-S100A and HDVS-S100D) to provide images through the video monitor for observation, diagnosis, photography and treatment of the respiratory system such as trachea, bronchi, and lungs.
The proposed device, Broncho Videoscope System is consisting of a Single-use Broncho Videoscope and an Endoscopic Image Processor including the foot switch. The proposed device has been designed to be used for endoscopic diagnosis and therapies within the respiratory system such as trachea, bronchi and lungs.
The Single-use Broncho Videoscope is a single use device. The Single-use Broncho Videoscope has six models which are available in three kinds of outer diameter of insertion (Φ2.8mm, Φ4.2mm and Φ5.6mm), one working length (600mm) and two different material of the insertion (Nylon and PEEK). The single-use Broncho Videoscope is a single-channel endoscope. Only one working channel is in the distal end of the endoscope, and it bifurcates to two channels leading to the irrigation valve and suction section.
The Single-use Broncho Videoscope is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied in sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.
The Endoscopic Image Processor is a reusable device. The Endoscopic Image Processor has two models. The only difference between the two models is that the HDVS-S100A has Enhance function and the HDVS-S100D does not have the Enhance function.
Here's a breakdown of the requested information based on the provided text, outlining the acceptance criteria and study details for the Broncho Videoscope System:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria here are broad, based on compliance with various international standards and demonstrating equivalence to a predicate device. The performance is reported as meeting these standards and being comparable to or better than the predicate.
| Acceptance Criteria Category | Specific Acceptance Criteria (Standards/Tests) | Reported Device Performance |
|---|---|---|
| Biocompatibility | ISO 10993-7:2008 (Ethylene oxide sterilization residuals) | Complies; no negative impacts from materials |
| ISO 10993-5:2009 (In vitro cytotoxicity) | Complies; no negative impacts from materials | |
| ISO 10993-10:2010 (Irritation and skin sensitization) | Complies; no negative impacts from materials | |
| Intracutaneous (Biocompatibility test) | Complies; no negative impacts from materials | |
| Material/Packaging Sterility | ASTM F88/F88M-15 (Seal strength of flexible barrier materials) | Complies |
| ASTM F1929-15 (Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration) | Complies | |
| Endoscope Design/Optics | ISO 8600-1:2015 (General requirements for endoscopes) | Complies |
| ISO 8600-3:1997/Amd1:2003 (Field of view and direction of view) | Complies | |
| ISO 8600-4:2014 (Maximum width of insertion portion) | Complies | |
| Software | FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Moderate level of concern) | Software verification and validation testing conducted and documentation provided as recommended. |
| Electrical Safety/EMC | IEC 60601-1-2005+ (General requirements for basic safety and essential performance) | Complies |
| IEC 60601-1-2:2014 (Electromagnetic compatibility) | Complies | |
| IEC 60601-2-18:2009 (Particular requirements for endoscopic equipment) | Complies | |
| Shipping Performance | ASTM D4169-16 (Performance Testing of Shipping Containers and Systems) | Complies |
| Photobiological Safety | IEC 62471:2006 (Photobiological Safety of Lamps and Lamp Systems) | Complies |
| Risk Management | ISO 14971:2007 (Application of Risk Management to Medical Devices) | Complies |
| Bench Performance (Optical) | Photobiological safety test according to IEC 62471: 2006 | Passes |
| Color reproduction, Resolution, Depth of view, Geometric distortion, Image intensity uniformity, Image frame frequency and system delay testing | Compared with predicate device; results demonstrated equivalence or improvement (e.g., color reproduction better than predicate). | |
| Bench Performance (Physical/Functional) | Suction system function test | Performed and compared with predicate device. |
| Irrigation valve leakage test | Performed. | |
| Use-Life Testing | Optical performance comparison (un-aged vs. aged Single-use Broncho Videoscope) | Optical performance of aged endoscope similar to un-aged. |
| Use-life verification of image system (accelerated aging/running) | Performance of system doesn't reduce after accelerated aging/running; use-life of six years accepted. Optical performance of aged image processor similar to new. | |
| Substantial Equivalence (General) | Comparison to predicate device (Vathin Video Bronchoscope System, K191828) | Determined to be Substantially Equivalent (SE), with differences in expression of indications, anatomical sites, field of view, depth of field, outer diameter, deflection, work length, instrument channel, dimension, weight, input power, video output, white balance, and patient-contact materials addressed and deemed not to affect safety and effectiveness. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated as a number of devices or "cases" for human studies. The testing appears to be primarily bench/laboratory-based on the device components. For the use-life testing, "un-aged" and "aged" single-use bronchoscopes were compared, and "new" and "accelerated aged and running" image processors were compared, implying at least two samples for these comparisons.
- Data Provenance: The studies are non-clinical (bench performance, testing against standards). The country of origin for the data is not specified, but the applicant is "Scivita Medical Technology Co., Ltd." in "Suzhou, Jiangsu, China," suggesting the testing may have occurred there. The studies are prospective in nature, as they are conducted specifically to support this 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable for this submission. The ground truth for the non-clinical tests is established by compliance with international standards and direct objective measurements, not by expert consensus on clinical data.
4. Adjudication Method for the Test Set
Not applicable. This was a non-clinical submission relying on objective measurements against predefined standards and comparative bench testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission."
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, a standalone performance assessment was conducted for the device itself. The entire non-clinical testing section (bench performance, biocompatibility, electrical safety, etc.) is a standalone assessment of the device's technical specifications and performance against established standards and a predicate device. This is primarily a hardware device with associated software, not an AI algorithm in the typical sense that would necessitate a "human-in-the-loop" study for algorithm performance.
7. The Type of Ground Truth Used
The ground truth used for the non-clinical tests includes:
- International Standards: Compliance with various ISO, ASTM, and IEC standards (e.g., ISO 10993 for biocompatibility, IEC 60601 for electrical safety, ISO 8600 for endoscope performance).
- Predicate Device Performance: Direct comparison of objective measurements (e.g., optical parameters, suction function) against a legally marketed predicate device (Vathin Video Bronchoscope System K191828).
- Objective Measurements: Laboratory measurements of device specifications (e.g., field of view, depth of field, dimensions, weight, power).
8. The Sample Size for the Training Set
Not applicable. This is a medical device, not a machine learning or AI algorithm in the common sense that would require a "training set" of data for learning purposes. The "software" mentioned is for device control and image processing, not for diagnostic AI interpretation.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set."
Ask a specific question about this device
Page 1 of 1