Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This mask is intended to be worn to protect both the patient and healthcare worker of microorganisms, body fluids and particulate material. This mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use device provided non-sterile

    Device Description

    These 3 Layer Surgical Masks: Level 1, Level 2 or Level 3 (with Ear Loops or Ties) are 3 layer, flat-folded masks constructed of 100% nonwoven polypropylene materials. The masks are provided with Ear Loops (Polyester/Spandex) or Ties (Polypropylene). A malleable nose strip is placed within the binding for comfort and individualized fit. These masks meet the acceptance criteria for ASTM F 2100, Level 1, Level 2 or Level 3 performance standards and are provided in blue. These masks are single use and provided non-sterile.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for 3A Medical Products Co., Ltd.'s 3 Layer Surgical Masks, seeking substantial equivalence to a predicate device. It details the masks' performance characteristics against recognized consensus standards for surgical masks.

    1. A table of acceptance criteria and the reported device performance:

      Test MethodsAcceptance Criteria for Proposed Device LevelsReported Device Performance (Proposed Device)Predicate Device PerformanceComparison
      ASTM F2100 (Performance Standard)N/A (Categorizes into levels 1, 2, 3)Level 1, Level 2, Level 3Level 1, Level 2, Level 3Same
      ASTM F1862 (Fluid Resistance)Pass at 80 mmHg (Level 1)Pass at 80 mmHgPass at 80 mmHgSame
      Pass at 120 mmHg (Level 2)Pass at 120 mmHgPass at 120 mmHgSame
      Pass at 160 mmHg (Level 3)Pass at 160 mmHgPass at 160 mmHgSame
      ASTM F2299 (Particulate Filtration Efficiency)N/A (Measured for comparison against predicate)Pass at 98.86% (Level 1)Pass at 99.6%Similar
      Pass at 98.8% (Level 2)Pass at 99.6%Similar
      Pass at 98.6% (Level 3)Pass at 99.7%Similar
      ASTM F2101 (Bacteria Filtration Efficiency)N/A (Measured for comparison against predicate)Pass at 99.9% (Level 1)Pass at >98%Similar
      Pass at 99.9% (Level 2)Pass at >98%Similar
      Pass at 99.9% (Level 3)Pass at >99%Similar
      EN14683 (Proposed) / MIL-M36945C (Predicate) (Differential Pressure)N/A (Measured for comparison against predicate)Pass at 4.1 mm H2O/cm² (Level 1)Pass at 2.0mm H2O/cm²Similar
      Pass at 4.0 mm H2O/cm² (Level 2)Pass at 1.6mm H2O/cm²Similar
      Pass at 4.53 mm H2O/cm² (Level 3)Pass at 2.5mm H2O/cm²Similar
      16 CFR 1610 (Flammability)Class 1Class 1 (all levels)Class 1Same
      ISO 10993-5:2010 (Cytotoxicity)Non-toxic potential to L929 cellsNo potential toxicity to L929 cells (all levels)Non-cytotoxicSimilar
      ISO 10993-10:2010 (Irritation)No potential skin irritation on rabbitsNo potential skin irritation on rabbits (all levels)Non-irritatingSimilar
      ISO 10993-10:2010 (Sensitization)No potential skin sensitization on guinea pigsNo potential skin sensitization on guinea pigs (all levels)Non-sensitizingSimilar
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
      The document does not explicitly state the sample sizes used for each specific test (e.g., number of masks tested for fluid resistance). The data provenance is derived from tests conducted to demonstrate conformance to recognized consensus standards like ASTM and ISO. The manufacturer, 3A Medical Products Co., Ltd., is located in Liu An, Anhui Province, People's Republic of China, suggesting the testing was likely conducted in or for China on a prospective basis for market clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
      This document describes testing against performance standards for medical devices, not an algorithm that requires expert ground truth establishment. The "ground truth" here is defined by the objective, quantifiable metrics set forth in recognized international standards (e.g., ASTM, ISO). Therefore, no human experts for establishing ground truth are applicable in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable. Adjudication methods are typically used for subjective interpretations or diagnostic accuracy studies, not for objective physical and biological performance testing of a medical device like a surgical mask against established standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This is a submission for a physical medical device (surgical masks), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. This document pertains to a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      The "ground truth" for the performance of these surgical masks is established by the objective, quantitative results of standardized laboratory tests against recognized consensus standards (e.g., ASTM F2100, ASTM F1862, ASTM F2299, ASTM F2101, EN14683, 16 CFR 1610, ISO 10993-5, ISO 10993-10). These standards define specific methodologies and thresholds that a device must meet to achieve a certain performance level.

    8. The sample size for the training set:
      Not applicable. This document relates to a physical medical device and its performance testing, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:
      Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1