Search Results
Found 1 results
510(k) Data Aggregation
(98 days)
Cardiovascular surgery - closure of the sternum following sternotomy.
The Sternal Cable System is an alternative over traditional monofilament sternal wire used for cardiovascular sternal closure following sternotomy. The system consists of multi-strand stainless steel and titanium cables and crimps, which are tensioned and secured around the bone using a tensioner/crimper instrument. Multiple “figure 8” constructs work as one unit to provide stabilization.
Cables are manufactured from titanium 6Al-4V ELI alloy (ASTM F136) or 316L stainless steel (ASTM F138); CP titanium (ASTM F67) or 316L stainless steel (ASTM F138) crimps are provided to match the corresponding cable materials. However, the different metals are never to be mixed. Non-implantable leaders are manufactured from 316L stainless steel (ASTM F138) or titanium 3Al/2.5V Alloy (ASTM B863) and needles are 420 or 470 stainless steel.
The implants are provided sterile for single-use but must never be re-sterilized; reusable instruments are supplied non-sterile and must be steam sterilized by the user prior to use in accordance with the instructions for use. Cases are supplied for sterilization and transport of the instruments.
The provided document is a 510(k) Summary for the Sternal Cable System and primarily focuses on demonstrating substantial equivalence to a predicate device, particularly for the addition of Magnetic Resonance (MR) conditional safety labeling. It is not a study that evaluates the performance of an AI medical device, but rather a conventional medical device.
Therefore, many of the requested criteria, such as "acceptance criteria for an AI medical device," "sample size for the test set," "number of experts," "adjudication method," "MRMC study," "standalone performance," and "ground truth establishment," are not applicable to this document as it does not describe an AI medical device or its performance study.
However, I can extract information relevant to the device's MR Conditional safety labeling, which is the core of this submission's "study" aspect:
1. A table of acceptance criteria and the reported device performance (for MR safety):
Acceptance Criteria (from referenced ASTM standards) | Reported Device Performance |
---|---|
Magnetically Induced Displacement Force: Sufficiently low so as not to pose an additional hazard. | The device meets the acceptance criteria (implied by "MR Safety Evaluation following standards listed below" and "Testing identified in this summary has all passed acceptance criteria established by the predicate device where applicable."). |
Magnetically Induced Torque: Sufficiently low so as not to pose an additional hazard. | The device meets the acceptance criteria. |
Radio-Frequency (RF) Induced Heating: Heating within acceptable limits (e.g., |
Ask a specific question about this device
Page 1 of 1