Search Results
Found 1 results
510(k) Data Aggregation
(59 days)
The CEDIA Heroin Metabolite (6-AM) Assay is a homogeneous enzyme immunoassay for the in vitro qualitative and/or semi-quantitative determination of the presence of heroin metabolite (6-AM) in human urine at a cut-off concentration of 10 ng/mL. The assay is intended to be used in laboratories and provides a rapid analytical screening procedure to detect 6-Acetylmorphine in human urine. The assay is designed for use with a number of clinical chemistry analyzers. This product is intended to be used by trained professionals only.
The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Liquid Chromatography/tandem mass spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/ mass spectrometry (GC/MS) or Liquid chromatography with tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method.
Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used. For In Vitro Diagnostic Use Only
The assay consists of buffers (1 and 2) and lyophilized reagents (1a and 2a). The components include mouse monoclonal antibodies to 6-Acetylmorphine, recombinant microbial "enzyme donor" - 6-Acetylmorphine conjugate, "enzyme acceptor", chlorophenol red ß-Dgalactopyranoside, stabilizers and preservatives.
Here's a summary of the acceptance criteria and study details for the CEDIA Heroin Metabolite (6-AM) Assay, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria (Implicitly Derived) | Reported Device Performance |
|---|---|---|
| Precision | All determinations at concentrations below the cutoff should be negative; all at concentrations above the cutoff should be positive. For cut-off concentration, readings can be mixed. | Qualitative Mode:- 0 to 7.5 ng/mL (below cutoff): 80/0 (Negative/Positive) determinations.- 10 ng/mL (at cutoff): 25/55 (Negative/Positive) determinations.- 12.5 to 20 ng/mL (above cutoff): 0/80 (Negative/Positive) determinations.Semi-Quantitative Mode:- 0 to 7.5 ng/mL (below cutoff): 80/0 (Negative/Positive) determinations.- 10 ng/mL (at cutoff): 46/34 (Negative/Positive) determinations.- 12.5 to 20 ng/mL (above cutoff): 0/80 (Negative/Positive) determinations. |
| Spike Recovery | Qualitative: No ± 2SD overlap between samples below, at, and above the cutoff. All replicates of spiked negative and positive controls should be accurately detected.Semi-quantitative: Spiked samples recover within 80-120% of nominal values. | Qualitative Mode: No ± 2SD overlap between 7.5 ng/mL, 10 ng/mL, and 12.5 ng/mL. All 20 replicates of 7.5 ng/mL were Negative, and all 20 replicates of 12.5 ng/mL were Positive, compared to 10 ng/mL.Semi-Quantitative Mode: Spiked samples recovered within 80-120% of nominal values (as seen in the Analytical Recovery table). |
| Analytical Recovery and Dilution Linearity | Semi-quantitative mode: Demonstrates ability for sample dilution and QC across the assay range, with acceptable percent recovery. | Average Recovery (%) ranged from 97.3% to 115.6% for target concentrations from 0 to 20 ng/mL. Range of Recovery (%) was also provided. |
| Method Comparison and Accuracy | High concordance (suggesting 100%) between the immunoassay and a confirmatory method (LC-MS/MS) for patient samples across various concentration ranges. | Qualitative & Semi-Quantitative Modes: Overall concordance between LC-MS/MS and CEDIA Heroin Metabolite (6-AM) Assay was 100% for the 103 samples tested. |
| Specificity (Cross-reactivity with Metabolites) | 6-Acetylmorphine shows 100% cross-reactivity and heroin shows appropriate cross-reactivity (e.g., in this case, 8.3%). | 6-Acetylmorphine: 100% cross-reactivity at 10 ng/mL.Heroin: 8.3% cross-reactivity at 120 ng/mL. |
| Specificity (Cross-reactivity with Structurally Related/Unrelated Compounds) | Structurally related or unrelated compounds should exhibit minimal cross-reactivity (e.g., <0.01% or as indicated by accurate detection of spiked controls). | Numerous compounds (e.g., 6-Acetylcodeine, Buprenorphine, Codeine, Fentanyl, Oxycodone, etc.) showed cross-reactivity <0.01%, or specific low cross-reactivity (e.g., Dextromethorphan (0.01%), Hydromorphone (0.05%), Levorphanol (0.07%), Morphine (0.07%), Nalorphine (0.1%), Normorphine (0.02%)). Controls in the presence of these compounds were detected accurately (Low Control as negative, High Control as positive), indicating minimal interference. |
| Interference (pH and Endogenous Substances) | Endogenous physiologic substances and varying pH levels should not interfere with the accurate detection of low and high controls. | Compounds like Acetone, Ascorbic acid, Creatinine, Ethanol, Glucose, Hemoglobin, Human serum albumin, Urea, and various pH levels (3.0-11.0) did not interfere with the assay, with both Low and High Controls detected accurately. |
| Interference (Specific Gravity) | Variations in specific gravity of urine samples should not interfere with accurate detection. | Drug-free urine samples across a specific gravity range of 1.002 to 1.029, spiked with Low Control and High Control levels, showed no interference. |
| Stability (Open Vial, Reagent On-Board, Reconstituted Reagent, Real-Time) | The device should demonstrate specified stability periods for open vial, on-board reagents, reconstituted reagents, and overall shelf-life. | Open Vial: 60 days (qualitative and semi-quantitative modes) - data from K173183.Reagent On-Board: 60 days (qualitative and semi-quantitative modes).Reconstituted Reagent: 60 days (qualitative and semi-quantitative modes).Real-Time Stability: 24 months (supported by 26 months of data from K173183). |
2. Sample Sizes and Data Provenance
- Test Set Sample Size:
- Precision Studies: 80 samples for both qualitative and semi-quantitative modes (two replicates per run, twice a day, for 20 days). These samples were spiked at various concentrations relative to the cutoff (0, 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20 ng/mL).
- Analytical Recovery and Dilution Linearity: Drug-free urine spiked to 20 ng/mL and diluted to generate 10 intermediate levels. Each sample was run in replicates of five.
- Method Comparison and Accuracy: 103 patient samples.
- Specificity (Cross-reactivity): Varies per compound, usually tested at one high concentration for structurally related/unrelated compounds, and two control levels (7.5 ng/mL and 12.5 ng/mL) for "structurally unrelated compounds spiked at the concentration listed below into Low Control and High Control".
- Interference (Endogenous & pH): Low Control (7.5 ng/mL) and High Control (12.5 ng/mL) samples, spiked with various interfering substances.
- Interference (Specific Gravity): Drug-free urine samples separated into 11 specific gravity levels, then each spiked at Low Control (7.5 ng/mL) or High Control (12.5 ng/mL).
- Data Provenance: The document does not explicitly state the country of origin. The studies are described as analytical performance studies conducted by the manufacturer (Microgenics Corporation, Fremont, CA, USA). The studies appear to be prospective for the current submission, though some stability data references previous 510(k) submission (K173183).
3. Number of Experts and Qualifications for Ground Truth
The document does not mention the use of experts to establish ground truth in the traditional sense (e.g., radiologists, clinicians reviewing images). For this type of in vitro diagnostic device (immunoassay), the "ground truth" for the test set is established by a highly accurate and precise reference method.
4. Adjudication Method for the Test Set
Not applicable. For this type of in vitro diagnostic, adjudication by human experts is not typically used for establishing ground truth. The comparison is made against a validated confirmatory laboratory method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This is an in vitro diagnostic assay, not an imaging device or AI-assisted diagnostic tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable.
6. Standalone Performance
Yes. The entire document describes the standalone performance of the CEDIA Heroin Metabolite (6-AM) Assay in both qualitative and semi-quantitative modes. The performance metrics (precision, accuracy, specificity, interference, stability) are all evaluated for the assay itself, without a human-in-the-loop component.
7. Type of Ground Truth Used
The primary ground truth for the test set (method comparison and accuracy, as well as for spiking concentrations in other studies) was established using Liquid Chromatography/tandem mass spectrometry (LC-MS/MS), and potentially Gas chromatography/mass spectrometry (GC/MS), as these are stated as the "preferred confirmatory methods."
8. Sample Size for the Training Set
Not applicable. This is an immunoassay, not a machine learning or AI-based device that would typically have a separate "training set" for model development. The reagents are chemical components that react based on established biochemical principles.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no "training set" in the context of an immunoassay. The development of the assay's reagents and methodologies would rely on chemical and biochemical principles and rigorous validation against known standards and reference methods.
Ask a specific question about this device
Page 1 of 1