Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180106
    Device Name
    ThinkQA
    Manufacturer
    Date Cleared
    2018-03-13

    (56 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ThinkQA is a radiation therapy dosimetry Quality Assurance (QA) device consisting of a software framework intended to contain a suite of modules to verify that radiation dose actually delivered to the patient is as intended.

    The Epibeam module contained in ThinkQA is intended to be used as follows:

    Epibeam is a standalone software tool independent of the linear accelerator, the TPS and the Record-and-Verify system. It is intended to assist in reducing the clinical risk in the delivery of radiotherapy treatments and does not alter the treatment delivery. It is to be used by a radiation oncology medical professional as a guide to provide pretreatment plan delivery verification.

    The software is to be used for the purposes of detecting errors in the delivery of radiation therapy prior to treatment, like corruption of the transferred plan data to the treatment unit, inappropriate multileaf collimator sequence or beam output malfunctioning. The software acquires data from the Electronic Portal Imaging Device (EPID) during a blank fraction dedicated to the pretreatment verification without the patient and subsequently processes it. The processed data is compared with data calculated by the Epibeam system. The comparison is derived on one hand, from the application of dose conversion to the EPID data and on the other hand, from the computation of a predicted dose image under ideal conditions of functioning. A gamma-index analysis is then performed according to the dose difference and distance-to-agreement criteria provided by the user.

    Epibeam is not a treatment planning system and cannot be used to generate radiotherapy treatment plans. It provides an independent means of checking the reliability of the dose delivery for each beam in reference to TPS data.

    Epibeam therefore provides an added level of treatment quality assurance, thus giving clinicians confidence especially when complex treatment techniques are employed (gantry-fixed and rotational intensity modulated radiation therapy).

    Epibeam is intended to support decision making in relation to the delivery of treatment plan to the patient with every clinical linear accelerators equipped with an EPID, but does not alter the existing Indications for Use of the treatment unit.

    Device Description

    ThinkQA is a modular software suite composed of the module Epibeam which is a quality assurance tool dedicated to Patient Specific QA for pretreatment verification of irradiation beams.

    The EPIbeam verification module integrated to the ThinkOA software platform is a Quality Assurance tool in external beam radiation therapy, used in combination with the electronic portal imaging device (EPID) and dedicated to the irradiation beam pre-treatment verifications, particularly for IMRT and VMAT techniques.

    EPIbeam principle is based on the comparison of two images expressed in terms of absolute dose: on the one hand, a RT Plan defined in the TPS is used for the acquisition of a real portal image (test image) with the EPID directly irradiated (without attenuating medium); on the same RT Plan is used to compute a theoretical portal image (reference image). Specific models and algorithms are applied to express both images in the same absolute dose terms.

    The dose images obtained from the same RT Plan, one by the conversion model of the acquired raw EPID images and the other by the prediction model, can be quantitatively compared through dose difference mappings or 2D gamma-index. Both models are based on dosimetric data provided from the TPS.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the ThinkQA Epibeam device, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of numerical acceptance criteria. Instead, it describes functional requirements and states that "acceptance criteria were met." The performance is generally framed as demonstrating substantial equivalence to the predicate device.

    Acceptance Criteria (Implied / Functional)Reported Device Performance
    Pretreatment check functionalityYes
    Independent software operationYes
    Ability to acquire pretreatment imagesYes
    Algorithm for computing predicted reference dose imageYes
    Algorithm for converting acquired portal image into dose imageYes
    Comparison of measured and reference dose images via gamma-index analysisYes
    Generation of reviewer reportsYes
    Inclusion of gamma agreement index per beamYes
    Inclusion of significant statistic gamma index values per beamYes
    Ability to view test and reference imagesYes
    Ability to view superimposed test/reference dose profilesYes
    Ability to view 2D gamma index distributionYes
    Patient control database integrationYes
    User-defined Alert Criteria for out-of-tolerance analysisYes
    Import Approved Plan data from Treatment Planning SystemYes
    Import Portal Images from pretreatment fractionYes
    Automatic offline analysisYes
    Support for multiple treatment techniques (Static, IMRT, VMAT)Yes
    Requirement for EPID panel Calibration for commissioningYes
    Use of TPS results for dose data referenceYes
    Demonstration of substantial equivalence to predicate device (K133572)Achieved through performance, functional, and algorithmic testing.
    Conformance to applicable technical design specificationMet
    Achievement of safety and effectivenessAchieved
    Meeting device requirements under normal conditions of useMet

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" with a particular sample size from real patient data. The validation seems to be based on "clinically representative conditions" and "test cases" rather than a specific patient cohort for a validation study.

    • Sample Size for Test Set: Not explicitly stated as a separate patient-based test set. The testing involved "unit, integration and system tests" and "validation of the system under clinically representative conditions."
    • Data Provenance: Not specified regarding country of origin or whether it was retrospective or prospective. Given the nature of a software release, it's likely synthetic or internally generated test cases reflecting various clinical scenarios, and potentially retrospective clinical data for "clinically representative conditions."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of external experts to establish a "ground truth" for a specific test set. The ground truth for the device's function appears to be established through:

    • Comparison of acquired EPID data with data calculated by the Epibeam system itself, based on TPS plans and prediction models.
    • The assumption that the TPS data and the device's prediction model represent the "ideal conditions of functioning" or "reference."

    4. Adjudication Method for the Test Set

    Not applicable/not mentioned. There's no indication of an adjudication method involving multiple human readers for establishing a ground truth or resolving discrepancies in a test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The document focuses on the standalone performance and substantial equivalence of the software tool.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    Yes, a standalone study was done. The entire premise of the "Performance Testing - Bench" section describes the testing of the ThinkQA software's functionalities and algorithms independently. The Epibeam module is described as a "standalone software tool independent of the linear accelerator, the TPS and the Record-and-Verify system." The performance testing demonstrates that the software itself "meets the requirements of the device."

    7. Type of Ground Truth Used

    The ground truth for the comparison performed by the Epibeam module is based on:

    • Predicted dose image: Calculated by the Epibeam system under ideal conditions, derived from the RT Plan defined in the Treatment Planning System (TPS).
    • Dose conversion of acquired EPID data: The software converts raw EPID images into dose terms.
    • TPS data: The models and algorithms used by Epibeam are based on dosimetric data provided by the TPS, which serves as a reference for the planned dose.

    Essentially, the "ground truth" for the device's internal comparison is the expected dose distribution as calculated by the validated Treatment Planning System and through the device's own prediction models.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" or its size. As a "software framework" and a "Quality Assurance tool," its development likely involved conventional software engineering practices, potentially including internal data for model development and calibration, but a specific "training set" like in deep learning models is not detailed.

    9. How the Ground Truth for the Training Set Was Established

    Since a "training set" is not explicitly mentioned, the method for establishing its ground truth is also not described. The device's foundational data relies on the principles of radiation dosimetry and verified TPS data.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1