Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172940
    Manufacturer
    Date Cleared
    2017-10-18

    (22 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.

    Device Description

    The Anchor Tissue Retrieval System™ by CONMED is a sterile, disposable retrieval pouch for use with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery. The Anchor Tissue Retrieval System™ by CONMED consists of a rip stop nylon pouch, two stainless steel arms attached to internal pusher rod connected to a handle at the distal end to the patient. Around the internal pusher rod is the introducer and handle. The introducer handle incorporates a spring loaded button and stop mechanism is built into the pusher rod. There is a braided drawstring which enters the introducer with the pusher rod and loops through the rip stop nylon pouch.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance CriteriaReported Device Performance
    Intended UseSterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.Meets criteria; same as predicate.
    Where UsedOperating roomMeets criteria; same as predicate.
    Prescription OnlyYesMeets criteria; same as predicate.
    DesignSpecimen bag and a delivery system consisting of an introducer shaft deployment handle.Meets criteria; same as predicate.
    MaterialsNylon, stainless steel, polycarbonateMeets criteria; same as predicate.
    Mechanism to deploy bagAutomatic with compression of pusher rodMeets criteria; same as predicate.
    Mechanism to separate bag from introducerPush buttonMeets criteria; same as predicate.
    PerformanceAdequately tested through bench testing methodologies; no FDA performance standards.Meets criteria; adequately tested, substantially equivalent to predicate.
    SterilizationEthylene Oxide per ISO 11135:2014Meets criteria; same as predicate.
    BiocompatibilityAccording to ISO 10993-1:2009Meets criteria; patient contacting materials compliant.
    Redeployment Capability (New Feature)Can be deployed more than once within a single patient procedure.Meets criteria; redeployment verification testing performed.
    Safety and Effectiveness (Overall)No new issues of safety or effectiveness identified, and redeployment has no impact on current device safety and effectiveness. Benefits outweigh residual risks.Meets criteria; concluded through risk management activities in accordance with ISO 14971.
    Compliance with StandardsDesign specifications and applicable sections of ISO 11607-1:2006, ISO 11135:2014, ISO 10993-7:2008.Meets criteria; compliance demonstrated through design verification testing.
    User Needs and Intended UseConforms to user needs and the intended use.Meets criteria; demonstrated through design validation testing.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" with a particular sample size from a patient population, nor does it mention data provenance in terms of country of origin or retrospective/prospective nature. The studies conducted were primarily targeted performance and design verification/validation testing.

    • Sample Size for performance testing: The document mentions "bench and simulated use testing" and "redeployment verification testing." It doesn't specify the exact number of devices or trials for these tests.
    • Data Provenance: The studies are "non-clinical bench and simulated use testing," implying laboratory-based testing rather than patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The testing described is primarily engineering and performance-based validation rather than clinical studies requiring expert ground truth for diagnostic accuracy.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. The testing described does not involve a process of adjudication by experts for a test set.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This device is a surgical tissue retrieval system, not an AI-powered diagnostic tool, so such a study would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a physical surgical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established through:

    • Design Specifications: The device is tested against its established design specifications.
    • Performance Standards (Industry/Regulatory): Compliance with relevant ISO standards (e.g., ISO 11607-1:2006 for packaging, ISO 11135:2014 for sterilization, ISO 10993-1:2009 & ISO 10993-7:2008 for biocompatibility).
    • User Needs: Demonstrated conformance to user needs and intended use from design validation.
    • Substantial Equivalence to Predicate: Performance is compared to a legally marketed predicate device (K091930) to demonstrate substantial equivalence.

    8. The Sample Size for the Training Set

    There is no mention of a "training set" in the context of this traditional medical device. This term is typically used for machine learning or AI models.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1