Search Results
Found 1 results
510(k) Data Aggregation
(107 days)
The Tritanium® C Anterior Cervical Cage is indicated for use in cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level or two contiguous levels from the C2 to T1 disc.
DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have six weeks of non-operative therapy.
The Tritanium® C Anterior Cervical Cage System is to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and is to be implanted via an open, anterior approach.
The Tritanium® C Anterior Cervical Cage System is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical spine.
The Stryker Spine Tritanium® C Anterior Cervical Cage is a hollow, ring shaped titanium alloy (Ti-6Al-4V) interbody fusion cage intended for use in the cervical spine. The cage consists of an open window for bone graft containment and has serrations on the superior and inferior surfaces of the cage for fixation. The cage is offered in a variety of footprints, heights, and lordotic angles to adapt to varying patient anatomies. The Tritanium® C Anterior Cervical Cage consists of a unique configuration of both solid and porous structures that are simultaneously built using Laser Rapid Manufacturing (LRM) method applying Stryker's proprietary Tritanium® In-Growth Technology. The Tritanium® C Anterior Cervical Cage is provided sterile.
This document is a 510(k) summary for the Stryker Spine Tritanium® C Anterior Cervical Cage, an intervertebral body fusion device. It details the device's characteristics, intended use, and the non-clinical testing performed to establish substantial equivalence to predicate devices.
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria (Standard/Guidance Referenced) | Reported Device Performance |
|---|---|---|
| Mechanical Testing | Based on FDA's June 12, 2007 "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" and specific ASTM standards: | "demonstrated substantially equivalent performance to the identified predicate devices." Specific tests performed include: - Static Compression (per ASTM F2077) - Dynamic Compression (per ASTM F2077) - Static Compression Shear (per ASTM F2077) - Dynamic Compression Shear (per ASTM F2077) - Static Torsion (per ASTM F2077) - Dynamic Torsion (per ASTM F2077) - Expulsion (per ASTM F04-25-02-02 Draft) - Subsidence (per ASTM F2267) - Wear Debris Assessment - Impaction |
| Material Properties | FDA guidance documents: - "Guidance Document for testing orthopedic implants with modified metallic surfaces apposing bone or bone cement [April 28, 1994]" - "Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses [January 16, 2003]" - Draft guidance for "Technical considerations for Additive Manufactured Devices [May 10, 2016]" ASTM Standard: - ASTM 1472-08: Standard Specification for Wrought Titanium-6Aluminum-4Vanadium Alloy (UNS R56400) for Surgical Implant Applications | "established that the porous surface design of the cage meets at minimum the requirements outlined in the [referenced FDA guidance documents and ASTM standard]." |
| Material Mechanical Properties | ASTM standards: - ASTM F1147-05: Standard Test Method for Tension Testing of Calcium Phosphate and Metallic Coatings - ASTM F1044-05: Standard Test Method for Shear Testing of Calcium Phosphate and Metallic Coatings - ASTM F1160-05: Standard Test Method for Shear and Bending Fatigue Testing of Calcium Phosphate and Metallic Medical and Composite Calcium Phosphate/Metallic Coating - ASTM E8/E8M: Standard Test Methods for Tension Testing of Metallic Materials | Characterization of the Mechanical Properties of the Tritanium® C Anterior Cervical Cage material was performed in accordance with these standards. The conclusion states "results of the mechanical testing... demonstrated substantial equivalence." |
| Corrosion Susceptibility | ASTM F2129-15: Standard test method for conducting cyclic potentiodynamic polarization measurements to determine the corrosion susceptibility of small implant devices. | "Electromechanical performance evaluation of additive manufactured Ti-6Al-4V alloy was also performed per ASTM F2129-15." The overall conclusion implies satisfactory performance. |
| Pyrogenicity | ANSI/AAMI ST72:2011: For bacterial endotoxin testing (BET) to achieve an endotoxin limit of < 20 EU/Device. | "Bacterial endotoxin testing (BET) as specified in ANSI/AAMI ST72:2011 is used for pyrogenicity testing to achieve the Endotoxin limit of < 20EU/Device." The overall conclusion implies satisfactory performance. |
2. Sample Size Used for the Test Set and Data Provenance
This document describes non-clinical (bench) testing, not a clinical trial with a "test set" of patient data. Therefore, the concepts of human "sample size" and "data provenance" (country, retrospective/prospective) are not applicable in this context. The testing was performed on representative units of the medical device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This is a non-clinical device safety and performance evaluation. "Ground truth" in the context of human expert consensus is not applicable. The "ground truth" for this type of evaluation is defined by adherence to established engineering standards (ASTM) and FDA guidance documents. The expertise involved would be in manufacturing, materials science, and biomechanical engineering, typically internal or external testing labs, but not clinical "experts" establishing ground truth in the sense of medical diagnosis.
4. Adjudication Method for the Test Set
As this is non-clinical bench testing, an "adjudication method" involving human readers or experts in the clinical sense is not applicable. The results are objectively measured against predefined criteria from the standards and guidance documents.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study is relevant for imaging or diagnostic AI devices. This document is for an intervertebral body fusion device (an implant). Therefore, an MRMC comparative effectiveness study is not applicable. There is no AI component mentioned in the context of this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This device does not have an "algorithm only" or "human-in-the-loop" component in the sense of a diagnostic or AI-powered device. The device itself is an implant. Standalone performance refers to the device's mechanical and material properties as tested on a bench. All the performance data described (demonstrated substantially equivalent performance) represents the standalone performance of the physical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this non-clinical submission, the "ground truth" is based on:
- Engineering Standards: Specific ASTM standards (e.g., F2077, F2267, F1147, F1044, F1160, E8/E8M, F2129-15). These standards define test methodologies and often include performance benchmarks or criteria for material properties and mechanical performance.
- FDA Guidance Documents: Such as the "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" and guidance related to modified metallic surfaces and additive manufactured devices. These documents outline the expected performance characteristics and testing requirements for this type of device.
- Predicate Device Performance: The primary "ground truth" for substantial equivalence is the demonstrated performance of legally marketed predicate devices. The new device's performance is compared to these established benchmarks to show it is "substantially equivalent" and thus equally safe and effective.
8. The Sample Size for the Training Set
This document describes non-clinical testing for a physical implant, not an AI/machine learning device that would require a "training set." Therefore, the concept of a training set sample size is not applicable.
9. How the Ground Truth for the Training Set was Established
Since there is no "training set" for an AI algorithm in this context, the method for establishing its "ground truth" is not applicable.
Ask a specific question about this device
Page 1 of 1