Search Results
Found 1 results
510(k) Data Aggregation
(181 days)
The Astute Medical NEPHROCHECK® Test System is intended to be used in conjunction with clinical evaluation in patients who currently have or have had within the past 24 hours acute cardiovascular and or respiratory compromise and are ICU patients as an aid in the risk assessment for moderate or severe acute kidney injury (AKI) within 12 hours of patient assessment. The NEPHROCHECK® Test System is intended to be used in patients 21 years of age or older.
The Astute NEPHROCHECK® Test System is comprised of the NEPHROCHECK® Test Kit, the ASTUTE140® Meter Kit, NEPHROCHECK® Liquid Controls Kit and the NEPHROCHECK® Calibration Verification (Cal Vers) Kit. The system is designed to be used by trained medical professionals in the central laboratory.
The NEPHROCHECK® Test Kit, includes the NEPHROCHECK® Test which is a single-use cartridge comprised of two immunoassays for the protein biomarkers, insulin-like growth factor-binding protein (IGFBP7) and tissue-inhibitor of metalloproteinases 2 (TIMP-2) on a membrane test strip enclosed in a plastic housing. The concentrations of the TIMP-2 and IGFBP-7 proteins are used to derive an AKIRISK® Score. The test procedure involves the operator applying a clinical urine sample mixed with labeled fluorescent conjugate to the NEPHROCHECK® Test cartridge, and then inserting the Test cartridge into the ASTUTE140® Meter for incubation, result calculation and result display. Internal positive and negative procedural controls in each NEPHROCHECK® Test cartridge monitor the function of each test cartridge. If the automatic check of these procedural controls shows that the control value results are not within pre-defined limits, the ASTUTE140® Meter will display an error message and the Test result will not be reported.
Included in the NEPHROCHECK® Test is test buffer and the NEPHROCHECK® Test Conjugate Vial which contains murine monoclonal and goat polyclonal antibodies against TIMP-2 and IGFBP-7, fluorescent dye, stabilizers and excipients. Each Kit has the materials necessary to perform 25 tests. Each NEPHROCHECK® Test Kit also contains a lot-specific radio-frequency identification (RFID) card containing lot and calibration information. The RFID card information must be loaded prior to using a new Test kit lot.
The AsTUTE140® Meter is a bench-top analyzer that converts the fluorescent signal from each of the two immunoassays (TIMP-2 and IGFBP7) contained within the NEPHROCHECK® Test cartridge into the AKIRISK® Score. The NEPHROCHECK® Test result is displayed on the Meter LCD screen in approximately 20 minutes from the addition of the specimen. Only the AKIRISK® Score appears on the Meter display. The ASTUTE140® Meter contains an internal printer that can print the AKIRISK® Score.
The NEPHROCHECK® Low Liquid Control and NEPHROCHECK® High Liquid Control are bi-level, lyophilized control materials prepared from human urine containing human TIMP-2 and human IGFBP7 proteins with protein stabilizers. TIMP-2 and IGFBP-7 proteins have been added to the urine to achieve specified target concentration levels. Each NEPHROCHECK® Liquid Controls Kit also contains a high and low RFID encoded with the Liquid Control Kit lot number, expiration date, and the expected range of concentration values based on + two standard deviations in the measurement of each protein biomarker (TIMP-2 and IGFBP-7) in each Liquid Control level. Each NEPHROCHECK® Liquid Control Kit Vial is intended for single use only.
The NEPHROCHECK® Calibration Verification Kit includes five levels of Iyophilized material prepared from human urine, containing TIMP-2 and human IGFBP-7 to achieve specified target concentration levels that evenly span the reportable range of the AKIRISK® Score. The expected concentrations and standard deviations of the individual biomarkers are printed on an enclosed Values Card.
This document describes a 510(k) premarket notification for a modified NEPHROCHECK® Test System. The submission claims substantial equivalence to a previously cleared device (K153165). Therefore, the "study that proves the device meets the acceptance criteria" in this context is primarily focused on demonstrating that the modifications do not negatively impact the device's performance characteristics or its intended use, rather than a de novo clinical trial establishing efficacy or accuracy.
Based on the provided text, here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with corresponding performance data in a detailed quantitative manner for the modified device's clinical performance. Instead, it states that the modified components "met its specifications in this testing." The acceptance criteria for this 510(k) are implicitly focused on ensuring the new modifications do not degrade the performance previously established for the predicate device.
Implicit Acceptance Criteria (for modifications) and Reported Performance:
| Acceptance Criteria (General) | Reported Device Performance (for modified components) |
|---|---|
| ASTUTE140® Electronic Quality Control (EQC) Device Shelf-Life and Ambient Shipping Performance | Met specifications. |
| ASTUTE140® Electronic Quality Control (EQC) Device - Verification of Performance | Met specifications. |
| NEPHROCHECK® Test Kit cartridge - Shelf-Life Extension Performance | Met specifications. |
| ASTUTE140® Meter Modified LCD Window Screen Compatibility with Standard Cleaning Process, Climate Testing, Electrostatic Discharge Testing Performance | Met specifications. |
| ASTUTE140® Meter Modified Optics Board Verification of Electrical, Optical, and Mechanical Performance; Noise Testing; Test Result Stability; Temperature Coefficient Reproducibility; Verification of Production; Climate Testing (Storage & Operating); Vibration and Drop Testing Performance | Met specifications. |
| No impact on previously established final release specifications | "The NEPHROCHECK® Test System final release specifications as cleared in K153165 are not impacted by these minor changes." and "None of these changes impact the indications, warnings or precautions for the device." |
Note: The detailed acceptance criteria (e.g., specific ranges for shelf-life, noise levels, etc.) are not provided in this summary document. The document primarily focuses on demonstrating that modifications did not negatively affect performance, relying on previously established performance for the predicate device.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document primarily describes engineering and stability testing for the modified components, as opposed to a clinical test set for diagnostic performance.
- Sample Size for Test Set: Not explicitly stated in numerical terms for each tested parameter (e.g., how many cartridges were tested for shelf-life, how many meters for climate testing etc.). The context suggests sufficient samples were used for engineering verification and validation testing, but specific numbers are not disclosed.
- Data Provenance: The nature of the studies (shelf-life, climate testing, electrical/optical performance, etc.) implies laboratory or manufacturing test data. No information on country of origin or whether it was retrospective/prospective is provided, as these are typically not human clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This section is not applicable to the type of studies described. The document does not describe a clinical study where human experts establish ground truth for diagnostic performance. The tests are engineering and stability evaluations.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods are typically used in clinical studies where multiple human readers' interpretations need to be reconciled to establish a "ground truth" for a diagnosis. The studies described are hardware/software verification and shelf-life testing.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The NEPHROCHECK® Test System measures biomarkers directly; it is not an AI-assisted diagnostic imaging device that human readers interpret. Therefore, an MRMC study is not relevant to this device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The NEPHROCHECK® Test System is an in vitro diagnostic device that produces a quantitative result (AKIRISK® Score) from a biological sample (urine) using immunoassays. The "algorithm" for this device is the formula used to derive the AKIRISK® Score from the concentrations of TIMP-2 and IGFBP-7.
The device itself, the ASTUTE140® Meter, performs the measurement, calculation, and display of the AKIRISK® Score. Therefore, its performance in generating the AKIRISK® Score from a sample is its "standalone" performance. The document states: "The NEPHROCHECK® Test result is displayed on the Meter LCD screen in approximately 20 minutes from the addition of the specimen. Only the AKIRISK® Score appears on the Meter display."
The initial 510(k) for the predicate device (K153165) would have established the clinical and analytical performance of this standalone functionality. This current submission verifies that the modifications to the device components (e.g., optics board, LCD screen) do not negatively impact this standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the modifications described in this 510(k) summary, the "ground truth" refers to engineering specifications and established analytical performance. For example:
- Shelf-life: Stability over time, confirming the device components maintain their specified performance characteristics (e.g., calibration, reagent integrity, mechanical function).
- Climate Testing: Performance within specified temperature and humidity ranges.
- Electrical/Optical Performance: Conformance to design specifications for signal detection, noise, and measurement accuracy.
- Cleaning Compatibility: Verification that the device withstands specified cleaning agents without degradation.
For the original clearance of the predicate device (K153165), the ground truth for establishing the clinical aid in risk assessment for AKI would have involved patient outcomes data (e.g., acute kidney injury diagnosis within a certain timeframe based on clinical criteria and/or biomarker levels) correlated with the AKIRISK® Score. However, this 510(k) for a modified device does not present such clinical ground truth data anew.
8. The sample size for the training set
This information is not provided and is likely not relevant to this specific 510(k) submission, as it focuses on modifications to an already cleared device. The original development of the NEPHROCHECK® Test System (predicate device K153165) would have involved extensive data for biomarker correlation and algorithm development, which would conceptually be its "training set," but those details are not in this document.
9. How the ground truth for the training set was established
As with point 8, this information pertains to the original development and clearance of the predicate device (K153165) and is not detailed in this document. For a diagnostic test like NEPHROCHECK®, the "ground truth" for its development would typically involve:
- Clinical Diagnosis of AKI: Based on standard clinical criteria (e.g., KDIGO guidelines, serum creatinine changes, urine output) established by clinicians.
- Correlation with Outcomes: Investigating the relationship between the biomarker levels/AKIRISK® Score and the subsequent development or severity of AKI in patient cohorts.
These studies would have been part of the clinical evidence submitted for the original 510(k) (K153165).
Ask a specific question about this device
Page 1 of 1