Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161305
    Device Name
    Habib EUS RFA
    Manufacturer
    Date Cleared
    2016-07-07

    (58 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Habib EUS RFA 6700 is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radio frequency generator.

    Device Description

    The Habib EUS RFA 6700 is a catheter that is delivered during an EUS procedure, through 19 Gauge (G) endoscopic needles. The subject device is monopolar configuration and thus, must be used in conjunction with patient grounding pad. RF energy is produced by connecting the catheter to a compatible RF generator via accessory cable. When attached to a generator, RF current is emitted from the exposed portion of the electrode and this current translates into ion agitation within the surrounding tissue, which is converted by friction into heat and induces cellular death by means of coagulation necrosis.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Habib EUS RFA 6700 device, asserting its substantial equivalence to a predicate device (K150029). The submission outlines various performance, biocompatibility, and electrical tests conducted to demonstrate that the device meets required specifications.

    Here's the information extracted and organized as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Tests PerformedReported Device Performance
    Performance Bench TestsTrackability and Pushability TestMet required specifications for design verification
    Consistency of Heating Zone and Temperature TestMet required specifications for design verification
    Abrasion TestMet required specifications for design verification
    Mechanical Tests on the Electrical Assembly (Pull Test, Three Points Bending, Compression Test, Fatigue Resistance Test, Peel Resistance Test)Met required specifications for design verification
    Transport Simulation and Package IntegrityMet required specifications for design verification
    Accelerated AgingMet required specifications for design verification
    Seal Integrity (visual inspection)Met required specifications for design verification
    Bubble Test (package integrity)Met required specifications for design verification
    Peel Test (seal strength)Met required specifications for design verification
    Ambient PreconditioningMet required specifications for design verification
    Controlled ConditioningMet required specifications for design verification
    Distribution SimulationMet required specifications for design verification
    Design and Usability ValidationMet required specifications for design verification
    Functional Validation for Soft Tissue IndicationMet required specifications for design verification
    Label InspectionMet required specifications for design verification
    Usability of PouchMet required specifications for design verification
    Functional testing of catheter insertion in the endoscope needleMet required specifications for design verification
    Deflection TestMet required specifications for design verification
    Mechanical Tests on the Positioning GuideMet required specifications for design verification
    IPX2 TestMet required specifications for design verification
    Biocompatibility TestsISO 10993-1:2009 Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management processMet required specifications for biocompatibility tests
    ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: Tests for in vitro cytotoxicityMet required specifications for biocompatibility tests
    ISO 10993-7:2008 Biological Evaluation of Medical Devices Part 7: Ethylene oxide sterilization residualsMet required specifications for biocompatibility tests
    ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitizationMet required specifications for biocompatibility tests
    Electrical TestingIEC 60601-1:2006: Medical electrical equipment - part 1: General requirements for basic safety and essential performanceMet required specifications for electrical tests
    IEC 60601-2-2:2009 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessoriesMet required specifications for electrical tests
    IEC 60601-2-18:2009 Endoscopic equipmentMet required specifications for electrical tests

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for any of the performance tests. The tests appear to be laboratory-based bench tests and biocompatibility tests, rather than clinical studies with patient data. Therefore, details regarding country of origin or retrospective/prospective nature of patient data are not applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The evaluations described are primarily engineering and biocompatibility tests, not interpretations of clinical data requiring expert consensus or ground truth establishment in that context.

    4. Adjudication Method for the Test Set

    This information is not applicable as the tests are objective engineering and biological assessments, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This device is an electrosurgical tool, not an AI-powered diagnostic or assistive technology for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable. The device is a physical electrosurgical catheter and not a standalone algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests performed can be considered the established scientific and engineering standards outlined by the various ISO and IEC standards cited (e.g., ISO 10993 for biocompatibility, IEC 60601 for electrical safety). The performance bench tests would have had internal engineering specifications as their "ground truth."

    8. The Sample Size for the Training Set

    This information is not applicable as the device is not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as the device is not an AI algorithm that requires a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1