Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160352
    Date Cleared
    2016-10-21

    (256 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGEC® Spinal Bracing and Distraction System is intended for skeletally immature patients less than 10 years of age with severe progressive spinal abnormalities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) associated with or at risk of Thoracic Insufficiency Syndrome (TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.

    Device Description

    The MAGEC Spinal Bracing and Distraction System is comprised of a sterile single use spinal rod that can be surgiclly implanted using appropriate Stryker® Xia® fixation components (i.e. pedicle screws, hooks and/or connectors). The implanted MAGEC rod is used to brace the spine during growth to minimize the progression of scoliosis. The system includes a nonsterile hand held External Remote Controller (ERC or ERC 2) that is used periodically after implantation to non-invasively distract the implanted spinal rod. The titanium MAGEC rod includes an actuator portion that holds a small internal magnet. The actuator can be turned noninvasively by use of the ERC, which is electrically powered. The hand held non-invasive ERC is placed over the patient's spine and then manually activated, which causes the implanted magnet to rotate and distract the MAGEC rod. Once the physician determines that the implant has achieved its intended use and is no longer required, the implant is explanted. Additional accessories for the MAGEC System include the MAGEC Manual Distractor and the MAGEC Wand Magnet Locator. The MAGEC Manual Distractor is a sterile, single use device, which is used in the operating room to test the MAGEC rod prior to implantation. The MAGEC Manual Distractor used with the MAGEC System is made of Radel® and contains a rare-earth magnet. The MAGEC Wand Magnet Locator is a non-sterile device which is used during the distraction procedure to locate the magnet within the MAGEC rod. The ERC is placed over this location on the child's back.

    AI/ML Overview

    The provided text is a 510(k) summary for the MAGEC® Spinal Bracing and Distraction System, focusing on a labeling revision from MR Unsafe to MR Conditional. It does not contain information about acceptance criteria or a study proving the device meets those criteria, as typically understood for software or AI-enabled medical devices.

    Instead, the document details a comparison to a predicate device and non-clinical testing performed to establish MRI compatibility for the device itself. There is no mention of an algorithm or AI component in this specific submission.

    Therefore, many of the requested categories about acceptance criteria, AI performance, ground truth, and human reader studies are not applicable to the information contained in this document.

    However, I can extract the information relevant to the device's substantial equivalence and the testing related to its regulatory submission for the hardware component of the MAGEC System.

    Based on the provided document, here is the information that can be extracted:

    1. Table of Acceptance Criteria (for MRI Compatibility) and Reported Device Performance:

    Test DescriptionApplicable Test Standard / Performance MetricReported Performance
    Magnetically Induced Displacement ForceASTM F2052Evaluated for magnetically induced displacement force for MR Conditional labeling.
    Magnetically Induced TorqueASTM F2213Evaluated for magnetically induced torque for MR Conditional labeling.
    MR Image ArtifactASTM F2119Evaluated for MR image artifact for MR Conditional labeling.
    Radiofrequency Induced HeatingASTM F2182Evaluated for radiofrequency induced heating for MR Conditional labeling.
    Design VerificationNone (Custom setups were developed and utilized)Non-clinical testing indicated that, in accordance with relevant ASTM standards, MR Conditional labeling is appropriate.

    Explanation: The "acceptance criteria" here are implicitly defined by the ASTM standards for MRI compatibility and the FDA guidance document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment (12/11/2014)". The reported performance is that the testing indicated MR Conditional labeling is appropriate according to these standards and guidance.

    2. Sample size used for the test set and the data provenance:

    • Test Set (for MRI compatibility): "Test samples were provided in the final manufactured condition." The exact number of samples is not specified, but it refers to manufactured units of the MAGEC rod.
    • Data Provenance: Not explicitly stated for the MRI compatibility testing, but it would be from in-house or contracted lab testing of the physical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This testing is for the physical device's MRI compatibility, not an AI or diagnostic algorithm requiring expert ground truth.

    4. Adjudication method for the test set:

    • Not applicable. This testing is for the physical device's MRI compatibility, not an AI or diagnostic algorithm requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This submission concerns a physical medical device (spinal bracing system), not an AI or diagnostic algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This submission concerns a physical medical device (spinal bracing system), not an AI or diagnostic algorithm.

    7. The type of ground truth used:

    • For MRI Compatibility: The "ground truth" is defined by the physical properties of the device and its interaction with an MRI environment, as measured against the specified ASTM standards and FDA guidance.
    • For Clinical Efficacy (referenced from predicate): The clinical study endpoints for the predicate device included: Cobb angle correction, thoracic spine height increase, improvement in space available for lung (SAL), coronal and sagittal balance, reduction in subsequent surgical procedures, and weight gain. These are objective clinical measurements and patient outcomes.

    8. The sample size for the training set:

    • Not applicable. This submission is about a physical medical device, not an AI model.

    9. How the ground truth for the training set was established:

    • Not applicable. This submission is about a physical medical device, not an AI model.

    Additional Information from the document:

    • Clinical Performance Data (referenced from predicate): A retrospective clinical study was conducted for the predicate MAGEC System outside the United States. This study evaluated children who had either a primary or revision spinal bracing procedure using the MAGEC System. The study showed benefits similar to traditional growing rods without the need for regular surgical lengthenings.
    • The current submission (K160352) is identical to the predicate device (K150885) in terms of indications for use, principles of operation, and technological characteristics. The only purpose of this specific premarket notification is to revise the labeling from MR Unsafe to MR Conditional based on the non-clinical MRI compatibility testing.