510(k) Data Aggregation

    K Number
    K153490
    Manufacturer
    Date Cleared
    2016-05-13

    (161 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DD medical polymers are indicated for temporary (≤ 12 months) crowns, bridges and bite splints. Applications include both anterior and posterior structures.

    Device Description

    DD medical polymers, made from medical grade, are industrially polymerized, pre-colored or clear dental milling blanks designed for milled fabrication of temporary anterior and posterior crowns and bridges (tooth-colored variants) or bite splints (clear variant) on dental CAD/CAM systems.

    AI/ML Overview

    The document concerns the 510(k) premarket notification for "DD medical polymers" (DD tempMED and DD Bio Splint P) by Dental Direkt GmbH. The device is a resin for temporary crowns, bridges, and bite splints.

    Here's an analysis of the acceptance criteria and study data provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from international standards DIN EN ISO 20795-1 and ISO 10477, which pertain to denture base polymers and polymer-based crown and bridge materials, respectively.

    RequirementRequired Value (ISO 20795-1)Required Value (ISO 10477)Value Submission DeviceValue Predicate DevicePassed / Failed
    Flexural Strength [MPa]≥ 65≥ 50≥ 7590Passed
    Flexural Modulus [MPa]≥ 2000Not specified2800 (± 200)Not specifiedPassed
    Residual MMA Monomer [%]≤ 2.2Not specified0.41.4Passed
    Water Sorption [µg/mm³]≤ 32≤ 40≤ 2323Passed
    Solubility [µg/mm³]≤ 1.6≤ 7.50.20.0Passed

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes (number of specimens) used for each mechanical or chemical test. It only mentions that "Finished DD medical polymer products in all variants were tested."

    The data provenance is from non-clinical testing performed according to international standards (DIN EN ISO 20795-1 and ISO 10477) and biocompatibility standards (EN ISO 10993-1, -5). These tests were conducted by an accredited, GLP-certified testing laboratory. The country of origin of the data is not explicitly stated beyond the fact that Dental Direkt GmbH is based in Germany and the testing laboratory is accredited. The tests are non-clinical (laboratory-based) and are retrospective in the sense that they are conducted on finished products.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the provided document pertains to a medical device's physical and chemical properties and biocompatibility, not an AI/CADe/CADx system that would require expert-driven ground truth for performance evaluation.

    4. Adjudication Method for the Test Set

    This information is not applicable for the reasons stated in point 3. The "ground truth" here is defined by the requirements of the international standards themselves, measured objectively.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI Assistance

    This information is not applicable as the device is a dental material, not an AI-assisted diagnostic or treatment planning system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable as the device is a physical material, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is established by the specified limits set forth in the following international standards:

    • DIN EN ISO 20795-1:2013: Dentistry - Base polymers - Part 1: Denture base polymers
    • ISO 10477:2004: Dentistry Polymer-based Crown and Bridge Materials
    • EN ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (for biocompatibility)
    • EN ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (for biocompatibility)

    These standards define objective physical, chemical, and biological performance criteria.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a physical material, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated in point 8.