Search Results
Found 1 results
510(k) Data Aggregation
(204 days)
Hypocore™ is a single-use device intended to cool sterile saline, lactated ringers and other comparable low viscosity inert fluids during intravenous administration where clinically indicated for reduction of patient temperature.
The Hypocore device is an inline chiller for IV (Intravenous) solutions that uses an endothermic reaction for cooling. The IV fluids enter the device through a Luer-lock connector on a standard IV tube, is cooled inside the device and exits through another Luerlock connector on a longer section of IV tube. During expected use, the device inlet is connected to an IV solution bag via an IV administration set, and its outlet is connected to an intravenous access port. It is powered by an internal non-rechargeable battery. The device has a handle to turn the device on and off. The entire device is disposable, single use and is provided pre-sterilized.
The provided text describes the acceptance criteria and a series of performance tests for the Novocor Medical Systems, Inc. Hypocore device.
Here's an analysis of the requested information:
1. A table of acceptance criteria and the reported device performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| 10 Minute Standard Device Performance - Level of cooling, fast flow | To cool 1 liter of fluid to an average temperature of <11ºC | All samples met the average output temperature acceptance criteria (range = 6.9-7.8 °C) |
| 30 Minute Standard Device Performance - Level of cooling, slow flow | To cool 1 liter of fluid to an average temperature of <11ºC | All samples met the average output temperature acceptance criteria (range = 6.1-8.5 °C) |
| 30 Minute Standard Device Performance - Time to achieve required cooling level | Time required to achieve the stated temperature threshold is less than 2 minutes | All samples met the acceptance criteria (range = 46-60 seconds) |
| Elevated Environmental Temperature Test | To cool IV fluids to less than 11ºC when flowed at 48 ml/min in a 32°C environment | All samples met the acceptance criteria. The average temperature of the exiting fluid ranged from 8.8-10.5 ℃ inside a 32°C chamber. |
| Flow Rate Test - Comparative flow rate of standard IV set with and without the Hypocore device in line | Flow rate reduction be less than 25% of the standard flow rate | All samples met the acceptance criteria (flow rate reduction ranged from 10.35-15.87%) |
| Tilt Test | The temperature of the exit fluid be less than that of the initial fluid | All samples met the acceptance criteria. The average environmental temperature ranged from 22.5-22.8°C. The output temperatures during the test ranged from 8.6-9.2°C at +45 degrees and 9.1-10.1°C at -45 degrees. |
| Activation Handle Force Test | Activation handle must withstand 20.36 lbs | All samples withstood the applied force and no sample devices broke. |
| Strap Performance Test - Device attachment | Devices must stay attached to the limb | All samples met the acceptance criteria. |
| Strap Performance Test - Pressure on limb | Pressure of the strap must not exceed 200 mmHg | All samples met the acceptance criteria (pressures ranged from 50-80 mmHg). |
| Strap Performance Test - Force on outlet tubing upon drop | Force of the device be less than 52.9 N (force required to pull out a taped IV catheter) | All test iterations met the acceptance criteria (forces ranged from 14.0-27.0 N). |
| Biocompatibility (Cytotoxicity) - Fluid Path | Non-cytotoxic with scores of 0, 1, or 2 per ISO 10993-5 (grade less than 2) | The test article extract showed no evidence of causing cell lysis or toxicity. The test article extract met the requirements of the test, since the grade was less than a grade 2 (mild reactivity). |
| Biocompatibility (Cytotoxicity) - Skin Contact | Non-cytotoxic with scores of 0, 1, or 2 per ISO 10993-5 (grade less than 2) | The test article extract showed evidence of causing slight cell lysis or toxicity; however, the test article extract met the requirements of the test, since the grade was less than a grade 2. |
| Biocompatibility (Irritation) - Fluid Path | Non-irritant with scores of 0-1 comparison between control and test sample reaction (ISO 10993-10) | The test article met the requirements, since the difference between each test article extract overall mean score and corresponding control extract overall mean score was 0.0 for sodium chloride (SC) and sesame oil (SO) extracts. |
| Biocompatibility (Irritation) - Skin Contact | Non-irritant with scores of 0-1 comparison between control and test sample reaction (ISO 10993-10) | The test article met the requirements, since the difference between each test article extract overall mean score and corresponding control extract overall mean score was 0.0 and 0.1 for the SC and SO test article extracts, respectively. |
| Biocompatibility (Sensitization) - Fluid Path | Non-sensitizer or mild sensitizer with a score of 0 (non-sensitizer) or 1 (mild sensitizer), based on comparison between control and test sample (ISO 10993-10) | The test article extracts showed no evidence of causing delayed dermal contact sensitization in the guinea pig. The test article was not considered a sensitizer. |
| Biocompatibility (Sensitization) - Skin Contact | Non-sensitizer or mild sensitizer with a score of 0 (non-sensitizer) or 1 (mild sensitizer), based on comparison between control and test sample (ISO 10993-10) | The test article extracts showed no evidence of causing delayed dermal contact sensitization. (Result truncated in document, but implies acceptance criteria met for non-sensitization). |
| Biocompatibility (Pyrogenicity) | No single animal temperature rise of 0.5°C or more above its baseline temperature. | No single animal showed a temperature rise of 0.5°C or more above its baseline temperature. The total rise of the rabbits' temperature during 3 hours was 0.0°C. The test article was judged as non-pyrogenic. |
| Biocompatibility (Ethylene Oxide and Ethylene Chlorohydrin Testing) - EO and ECH residuals | EO residuals <= 4mg and ECH residuals <= 9mg. | EO and ECH levels were acceptable on Day 1. |
| Biocompatibility (Ethylene Oxide and Ethylene Chlorohydrin Testing) - Tolerable Contact Limits (TCL) | TCL for EO and ECH shall not exceed 10 μg/cm2 and 5 mg/cm2, respectively, or the device shall have negligible irritation as specified in ISO 10993-10. | The Hypocore device exhibited no irritation as demonstrated within biocompatibility irritation testing. |
| Biocompatibility (Partial Thromboplastin Time) | Test result clotting time > 50% of the negative control per ASTM F2382. | The plasma exposed to the test article had an overall average clotting time of 235.9 seconds, which represented 79.9% of the negative control. The test article was considered a minimal activator, and therefore, met the requirements of the test. |
| Biocompatibility (Acute Systemic Toxicity) | Reference ISO 10993-11 (Section 5.3) - no mortality or evidence of systemic toxicity. | There was no mortality or evidence of systemic toxicity from the extracts injected into mice. Each test article extract met the requirements of the study. |
| Biocompatibility (Hemocompatibility) | Reference ISO 10993-4 - criteria for non-hemolytic/slightly hemolytic. | The hemolytic index for the test article in direct contact with blood was 2.0%, and the hemolytic index for the test article extract was 0.8%. The test article in direct contact with blood was slightly hemolytic and the test article extract was non-hemolytic. |
| Sterilization | Suitable to pass the acceptance criteria of ISO 11135 for a 1x10-6 SAL. | Testing demonstrated that the Hypocore device is suitable to pass the acceptance criteria of ISO 11135 therefore demonstrating substantial equivalence for sterilization compared to the predicate device. |
| Medical Electrical Safety | Meets the acceptance criteria of IEC 60601-1 and IEC 60601-1-2. | Testing demonstrated that the Hypocore device meets the acceptance criteria of IEC 60601-1 and IEC 60601-1-2 therefore demonstrating substantial equivalence to the predicate device with regard to electrical safety. |
| Human Factors Design Validation | All users safely complete all use tasks successfully. | All users safely completed all use tasks successfully. |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the specific number of samples used for each test beyond indicating "All samples" met criteria. For some tests like "Strap Performance Test", it implies multiple tests were performed (e.g., three tests performed, "all samples met"). For the Biocompatibility tests, it refers to "test article extract" and "test article" but doesn't quantify the number of devices or extracts tested.
The data provenance (country of origin, retrospective/prospective) is not mentioned in the provided text. The tests appear to be laboratory-based performance and safety evaluations rather than clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this study. The study focuses on the physical and chemical performance, safety, and usability of a medical device, which are evaluated against predefined engineering specifications, international standards (ISO, IEC, ASTM, USP), and human factors guidance. There is no mention of expert consensus or interpretation required to establish a "ground truth" in the way it would be for diagnostic imaging or clinical assessment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable. Adjudication methods are typically used in studies involving human interpretation or clinical endpoints where agreement among observers is critical. This study involves objective measurements and adherence to engineering and safety standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The device (Hypocore) is a thermal regulating system, not an AI-assisted diagnostic or interpretive tool. Therefore, an MRMC study or assessment of AI's effect on human readers is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable. The Hypocore is a physical device, not an algorithm. Its "standalone" performance is assessed throughout the various performance and safety tests, which evaluate the device's ability to meet its specifications independently of human intervention during the measurement process. However, Human Factors Design Validation did evaluate user interaction.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the performance tests (cooling level, time to cool, flow rate, environmental conditions, tilt, handle force, strap performance), the "ground truth" or reference was predefined engineering specifications and benchmark values, often based on "known published literature regarding IV fluid cooling" or "standard of care" (e.g., refrigerated IV fluid bag temperature, strength limits for a human wrist).
For the biocompatibility tests, the "ground truth" or reference was international standards such as ISO 10993-1, ISO 10993-4, ISO 10993-5, ISO 10993-10, ISO 10993-11, ASTM F756, ASTM F2382, and USP General Chapter <151>.
For sterilization, it was ANSI/AAMI/ISO 11135.
For electrical safety, it was IEC 60601-1 and IEC 60601-1-2.
For Human Factors, it was FDA Draft Guidance "Applying Human Factors and Usability Engineering to Optimize Medical Device Design" and standards IEC 62366: 2007 + A1: 2014, IEC 60601-1-6: 2007 (3rd Ed.).
8. The sample size for the training set
This information is not applicable. This is a medical device performance and safety study, not an AI model development study that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set for an AI model.
Ask a specific question about this device
Page 1 of 1